American Standard 605XRCT User manual

Selectronic Remote Control
Operating Manual
Software Version 1.6
M965370 REV. 1.6 (June 2019)

2
OPERATION: The remote programmer has 4 keys:
“Enter” or validate
“Cancel”
“Scroll UP”
“Scroll Down”
The ‘Down’ Key is also used to turn “ON” the remote programmer. After
not being used for 2 minutes, the programmer automatically turns itself off.
DISPLAY
Mini-SUB Port
Transmission Area

3
When using the programmer, the sensor will begin blinking for approximately 5 seconds after the
“ENTER” key is pressed and the display shows “TRANSMISSION OK”.
While it is blinking move your finger to within 1 inch of the sensor (nearly covering it). When the
light stops blinking and stays on steady, remove your finger and, with the palm of your hand facing
the sensor, move your hand to the distance to be set. When the sensor starts to blink again, move
you hand away from the detection area. The flashing will stop and the sensor will now be
programmed with the new setting.
The Programmer will now display "BACKUP FUNCTION". You can program the next sensor(s) by
pointing the programmer at the sensor and pressing the “ENTER” key. The programmer will
remember the programmed settings for approximately 3.5 minutes before turning off.
NOTE: See “Product Settings & Specifications Chart”
Setting The Detection Distance
***ONLY DONE AFTER SELECTING ALL THE MODE SETTINGS***

4
ON
MENU 1
CLEANING
MENU 2
DISINFECTION
Press to activate the menus
Press to turn on the programmer
(Version VI.6 will display)
DISINFECTION
DURATION :1-120 min
Press up or down to adjust time
Align the programmer with the sensor and press
to transmit the signal to sensor to activate the
DISINFECTION cycle for the selected period of time
Align the programmer with the sensor and press to
transmit the signal to sensor to activate the
DIAGNOSTIC feature.
NOTE:
Use the up and down
arrow keys to scroll through the current settings
Press to go to MENU 2
Press to go to MENU 3
Press to go to MENU 4
To MENU 4
Start
Cleaning
Disinfection
Diagnostics
TIP: Cleaning – Use this mode to disable the sensor for
1 minute. This is sometimes used by the cleaning crew to keep
the faucet “off” while cleaning the fitting and or fixture.
TIP: Diagnostics – Use this mode to determine the settings of
the faucet or flush valve.
NOTE: this feature will only work with products
manufactured starting in 2008.
TIP: Disinfection –This is typically used in healthcare facilities
that conduct periodic high temperature or high anti-bacterial
chemical purging of the water lines. The maintenance person
should activate this mode and stay present to avoid any
users from exposure to the purged water. This ʻmodeʼcan
be activated while in any function (hand-wash faucet, urinal
flush valve, etc.). The duration (run time) can be programmed
from 1 minute to 120 minutes in 1 minute intervals.
MENU 3
DIAGNOSTIC

To
5
MENU 8.2
ENGLISH
MENU 9
BACKLIGHT
From MENU 3
Press up or down to select
language – French, English,
Spanish, Italian, or German
BACKLIGHT
Press up or down to select
display backlight intensity
Press to set selected language
Press to set selected display
intensity for easy reading in
different lighting conditions.
NOTE: Not all features are
available with this version
From this menu, scroll down to
the specific mode you want to program:
Parameter
Language
Backlight
• Hand-Wash
• Metering
• On/OFF
•Urinal
• LF Urinal + PF **(NOT USED)**
• Flushometer
• Kitchen **(NOT USED)**
• Dual Flush
From this menu, scroll down to
the specific mode you want to program:
• Hand-Wash
• Metering
• On/OFF
• Kitchen **(NOT USED)**
From this menu, scroll down to
the specific mode you want to program:
• Hand-Wash
• Metering
From this menu, scroll down to
the specific mode you want
to program:
• Urinal
• LF Urinal + PF **(NOT USED)**
• Flushometer
• Dual Flush
MENU 5
PARTIAL COUNT
MENU 6
TOTAL COUNT
MENU 7
BATTERY VOLTAGE
SENSOR
2016 VERSION
SENSOR
2008 VERSION
SENSOR
2003 VERSION
NEXTGEN
SENSOR
2018 FAUCET
SENSOR
2018 URI. & WC
TIP: Sensor Version – Select 2016 for all Selectronic
products manufactured from 2008 through 2015.
**If unsure, use the 2016 setting. If repeated transmission
errors occur, then try the 2008.
NOTE: Not all features are
available with this version
NOTE: Not all features are
available with this version
Align the programmer
with the sensor and press
to transmit the program
to the sensor.
Align the programmer
with the sensor and press
to transmit the program
to the sensor.
Align the programmer
with the sensor and press
to transmit the program
to the sensor.
MENU 8
LANGUAGE
MENU 4
PARAMETER

6
Press up or down to adjust time
Press up or down to adjust time
To MENU 4.2
Press to set
Hand-Wash
From MENU 4
Press up or down to adjust distance
Align the programmer with the sensor and
press to transmit the program to the sensor
Press up or down to turn ON
the Auto Rise feature and
adjust frequency
Setting the sensor to the Hand-Wash faucet mode:
- Set the "Comfort Timeout"
- Set the "Safety Timeout"
- Set "Auto-Rinse" frequency (if the frequency is set to 0,
“auto-rinsing” (auto-purging) is Off)
- Set the "Duration" (run time) of the auto-rinsing process.
TIP: The “Auto. RINSE” (Auto Purge) allows the faucet to
automatically turn ON after a set time period of non-use,
preventing the water from becoming stagnant.
Default Setting: ‘OFF’
• This time period can be set to activate from 1 to 72 hrs (of non-use)
in 1 hr increments.
• The duration of run time can be set from 1 sec. to 240 sec.
MENU 4.1
HAND-WASHER
COMF. TIMEOUT
= 1 Sec.
SAFETY TIMEOUT
= 55 Sec.
Auto. RINSE
= OFF.
Auto. RINSE
= 10 Sec.
SENSOR DISTANCE
= 16 cm (DEFAULT)
TIP: The “COMF. TIMEOUT” is the (Comfort Timeout) determines
how long the water will run after the users’ hands are removed from
the detection zone. This prevents the water from quickly turning
ON & OFF as the user’s hands move in & out of the detection zone.
Recommended setting: 1.5 sec
•
Adjustable from 0 to 10 sec in 0.5 sec increments
TIP: The “SAFETY Timeout” is sometimes referred to as the ʻVandal
Timerʼ. This should be set ON to prevent someone from intentionally
blocking the eye (chewing gum, tape, etc) to keep the faucet running
and have the sink overflow.
Recommended setting: 55 sec
• Adjustable from 5 to 120 sec in 5 sec increments
NOTE: This is the standard ʻinstant on / instant offʼ
faucet mode.
TIP: The "SENSOR DIST." (Sensor Distance) is the maximum
distance that the sensor can detect a user. (Recommended setting
for Hand-Wash faucet is 16 cm)

7
From MENU 4.1
To MENU 4.3
Metering -- Faucet
(RUN TIME) Setting the sensor to the Metering faucet mode:
- Set the "Comfort Timeout" (run time)
-Set"Auto-Rinse" frequency (if the frequency is set to 0,
‘auto-rinsing” (auto-purging) is Off)
-Setthe"Duration" (run time) of the auto-rinsing process.
Align the programmer with the sensor
and press to transmit the program
to the sensor.
Press up or down to adjust distance.
Press up or down to adjust time
Press up or down to adjust
metering run time
Press up or down to turn ON
turn ON
the Auto Rise
Auto Rise
feature and
adjust frequency
TIP: In this Metering Mode, the “COMF. TIMEOUT” (Comfort Timeout)
is the RUN TIME - how long the water will run once activated by
the user.
•Run Times: 1, 2, 3.25, 7.5, 10, 15, 30, 60, 120, or 240 seconds
TIP: The “Auto. RINSE” (Auto Purge) allows the faucet to
automatically turn ON after a set time period of non-use,
preventing the water from becoming stagnant.
Default Setting: ‘OFF’
• This time period can be set to activate from 1 to 72 hrs (of non-use)
in 1 hr increments.
• The duration of run time can be set from 1 sec. to 240 sec.
MENU 4.2
METERING
COMF. TIMEOUT
= 15 Sec.
Auto. RINSE
= OFF.
Auto. RINSE
= 10 Sec.
SENSOR DISTANCE
= 16 cm (DEFAULT)
TIP: The "SENSOR DIST." (Sensor Distance) is the maximum
distance that the sensor can detect a user. (Recommended setting
for Metering faucet is 16 cm)

Align the programmer with the sensor and
press to transmit the program to the sensor
Press up or down to adjust distance
(16 cm recommended)
Press up or down for short range or long
range (DO NOT USE)
Press up or down to adjust time
Press up or down to adjust
metering run time
Press up or down to turn ON
turn ON
the Auto Rise
Auto Rise
feature and
adjust frequency
8
From MENU 4.2
To MENU 4.4
Metering with On/Off -- Faucet
Setting the sensor to the Metering faucet mode:
-Setthe"Comfort Timeout" (run time)
-Set"Auto-Rinse" frequency (if the frequency is set to 0,
‘auto-rinsing” (auto-purging) is Off)
-Setthe"Duration" (run time) of the auto-rinsing process.
TIP: In this “Metering with On/Off Mode”, the “COMF. TIMEOUT” is the
RUN TIME - how long the water will run once activated by the user.
The detection distance is set very short, so the user must start & stop
the water by moving hands close to the sensor. - This is suited for
gooseneck type fittings where the movement of the user may be
outside the normal detection zone.
• Run Times: 3.25, 7.5, 10, 15, 30, 60, 120, or 240 seconds
TIP: The “Auto. RINSE” (Auto Purge) allows the faucet to
automatically turn ON after a set time period of non-use, maintaining
the trap seal & preventing the water from becoming stagnant.
Default Setting: ‘OFF’
• This time period can be set to activate from 1 to 72 hrs (of non-use)
in 1 hr increments.
• The duration of run time can be set from 1 sec. to 240 sec.
MENU 4.3
ON/OFF
COMF. TIMEOUT
= 15 Sec.
Auto. RINSE
= OFF.
Auto. RINSE
= 10 Sec.
SENSOR DISTANCE
= 5 cm
SENSOR DISTANCE
SHORT RANGE
NOTE: This is actually uses the ‘Metering’ mode, but with the
added feature of being able to start and stop the water flow
before the metering time runs out.
TIP: The "SENSOR DIST." (Sensor Distance) is the maximum
distance that the sensor can detect a user.
(Recommended setting 5 cm)

9
From MENU 4
Press up or
down to adjust
the required
user detection
activation time
Press up or down
to adjust how long
to wait after the user
leaves to begin
the flush
Press up
or down
to adjust
frequency
Press up
or down
to turn ON
turn ON
the Heavy
Heavy
Flush Count
Flush Count
feature
and adjust
frequency
Press up or down
to adjust the A
Auto
Rinse
run time
Press up
or down
to adjust
the distance
To MENU 4.5
LF URINAL = PF
(DO NOT USE)
Scroll down to turn
the Auto Rinse
feature OFF
Align the programmer
with the sensor and
press to transmit the
program to the sensor
Urinal -- Urinal Flush Valve
TIP: The "DETEC. INTERVAL" (Detection Interval) is how long
a user must be in the detection zone to arm the flush valve. This
eliminates false flushing due to people walking past the urinals.
Recommended setting is 5 sec.
TIP: The "STADIUM MODE" feature automatically
switches to flush every 2 minutes during periods of
heavy usage (2 sec between users) and return to normal
when there are more than 2 sec. between users.
TIP: The "RINSE START INT." (Rinse Start Interval) is how long
after the user leaves the detection zone before the valve will flush.
Recommended setting is 2 sec.
TIP: The "Rinse"is the Flush Time (Flush Duration).
See Flush Valve Settings Chart
TIP: The "Auto. RINSE "(Sanitary Flush) allows the flush valve to automatically flush,
after a set time period of non-use, to maintain the trap seal. This time period can be set
to activate from OFF to 72 hrs (of non-use) in 1 hr increments. You can set the duration
of flush time from 1 sec. to 6 sec. – never more than the fixture can drain before
overflowing! (Recommended setting is OFF unless prolonged non-use is
expected – then 2 s should maintain the trap seal). Default Setting: ʻOffʼ
TIP: A large amount of water is delivered after a pre-programmed
number of flushes (or time) to prevent the drain from clogging.
The number of flushes can be set to activate from 10 to 100
flushes (in 10 flush increments) before the large flush occurs.
Default setting: OFF. The duration of the flush time can be
adjusted from 1 s to 15 s in 1 sec. intervals.
Press up
or down
to turn ON
or OFF
Press up or down
to adjust the Heavy
Heavy
Rinse
run time
TIP: The "SENSOR DIST." (Sensor Distance)
is the maximum distance that the sensor can
detect a user. (Recommended setting for
Urinal Flush Valve is 40 cm)
Press up
or down
to adjust
Flush Time
MENU 4.4
URINAL
DETEC INTERVAL
= 5 Sec
*RINSE START
INT = 2 Sec
HEAVY FLUSH
COUNT = OFF
Auto RINSE
= OFF
Auto RINSE
= 1 - 72 h
Auto RINSE
= 3 Sec.
HEAVY FLUSH
DURATION = 10 s
*RINSE
= 3 Sec
*SENSOR
DISTANCE = 40 cm
STADIUM MODE
= INACTIVE
* REFER TO PAGE 12 FOR
RECOMMENDED SETTINGS

Press up or
down to turn
turn
ON
ON
the Heavy
Heavy
Flush Count
Flush Count
feature
and adjust
frequency
Press up or
down to adjust
user detection
activation time
Press up or down to
adjust how long after
the user leaves to
begin the flush
Press up or down
to turn ON
turn ON
the
Auto Rinse
Auto Rinse
feature and
adjust frequency
Press up or down
to adjust the Auto
Auto
Rinse
Rinse
run time
Press up or down
to adjust the distance
10
From MENU 4
Flushometer – Toilet Flush Valve
NOT
USED
Press up
or down
to adjust
Flush Time
MENU 4.6
FLUSHOMETER
MENU 4.7
KITCHEN
Auto RINSE
= OFF
Auto RINSE
= 1 - 72 h
Auto RINSE
= 3 Sec.
HEAVY FLUSH
COUNT = OFF
HEAVY FLUSH
DURATION = 10 s
DETEC INTERVAL
= 5 Sec
*RINSE START
INT = 2 Sec
*RINSE
= 3 Sec
*SENSOR DISTANCE
= 60 cm
* REFER TO PAGE 12 FOR
RECOMMENDED SETTINGS
Align the programmer with the sensor
and press to transmit the program,
to the sensor
TIP: The "DETEC. INTERVAL" (Detection Interval) is how long
a user must be in the detection zone to arm the flush valve. This
eliminates false flushing due to people walking past the urinals.
Recommended setting is 5 sec.
TIP: The "RINSE START INT." (Rinse Start Interval) is how long
after the user leaves the detection zone before the valve will flush.
Recommended setting is 2 sec.
TIP: A large amount of water is delivered after a pre-programmed
number of flushes (or time) to prevent the drain from clogging.
The number of flushes can be set to activate from 10 to 100
flushes (in 10 flush increments) before the large flush occurs.
Default setting: OFF. The duration of the flush time can be
adjusted from 1 s to 15 s in 1 sec. intervals.
TIP: The "Rinse"is the Flush Time (Flush Duration).
See Flush Valve Settings Chart
TIP: The "Auto. RINSE "(Sanitary Flush) allows the flush valve to automatically flush,
after a set time period of non-use, to maintain the trap seal. This time period can be set
to activate from OFF to 72 hrs (of non-use) in 1 hr increments. You can set the duration
of flush time from 1 sec. to 6 sec. – never more than the fixture can drain before
overflowing! (Recommended setting is OFF unless prolonged non-use is
expected – then 2 s should maintain the trap seal). Default Setting: ʻOffʼ
TIP: The "SENSOR DIST."
(Sensor Distance) is the
maximum distance that the
sensor can detect a user.
(Recommended setting for
Toilet Flush Valve is 60 cm)

Press up or
down to turn
turn
ON
ON
the Heavy
Heavy
Flush Count
Flush Count
feature
and adjust
frequency
Press up or down
to adjust duration
Press up or
down to adjust
how long to
wait after the
user leaves to
begin the flush
Press up or down
to turn ON
turn ON
the
Auto Rinse
Auto Rinse
feature and
adjust frequency
Press up or down
to adjust the distance
Align the programmer with the sensor
and press to transmit the program,
with the
with the
SENSOR DIST. = 60 cm
SENSOR DIST. = 60 cm
TIP: The "SENSOR DIST."
(Sensor Distance) is the
maximum distance that the
sensor can detect a user.
(Recommended setting for
Toilet Flush Valve is 60 cm)
Press up or
down to adjust
user detection
activation time
TIP: The "Auto. RINSE" (Sanitary Flush) allows the flush valve to automatically flush,
after a set time period of non-use, to maintain the trap seal . This time period can be set
to activate from OFF to 72 hrs (of non-use) in 1 hr increments. You can set the duration
of flush time from 1 sec. to 10 sec. – never more than the fixture can drain before
overflowing! (Recommended setting is OFF unless prolonged non-use is
expected –then 2 s should maintain the trap seal). Default Setting: ʻOffʼ
11
From MENU 4
Dual Flush – Toilet Flush Valve
TIP: The "Short Rinsing" is the is the flush time used when the
user stays in the detection zone for less than 1 minute. – This
should be the setting for the lowest of the 2 flush volumes listed for
this model flush valve. – See Flush Valve Settings Chart
TIP: The "Rinse" is the Long Flush Time (Flush Duration)
if the user stays in the detection zone for longer than 1 minute.
See the Flush Valve Settings Chart
TIP: The "DETEC. INTERVAL" (Detection Interval) is how long a
user must be in the detection zone to arm the flush valve. This
eliminates false flushing due to people walking past the urinals.
Recommended setting is 5 sec.
TIP: The "RINSE START INT." (Rinse Start Interval) is how long
after the user leaves the detection zone before the valve will flush.
Recommended setting is 2 sec.
MENU 4.8
Dual FlushToilet
*SHORT RINSING
= 1.5 Sec
Press up
or down
to adjust
Flush Time
Auto RINSE
= OFF
Auto RINSE
= 1 - 72 h
Auto RINSE
= 3 Sec.
HEAVY FLUSH
COUNT = OFF
HEAVY FLUSH
DURATION = 10 s
DETEC INTERVAL
= 5 Sec
*RINSE START INT
= 2 Sec
*RINSE
= 3 Sec
*SENSOR DISTANCE
= 60 cm
* REFER TO PAGE 12 FOR
RECOMMENDED SETTINGS

4
OFF - 72 hr
(in 1 hour increments)
OFF - 72 hr
(in 1 hour increments)
OFF - 72 hr
(in 1 hour increments)
OFF - 72 hr
(in 1 hour increments)
OFF - 72 hr
(in 1 hour increments)
OFF - 72 hr
(in 1 hour increments)
OFF - 72 hr
(in 1 hour increments)
OFF - 72 hr
(in 1 hour increments)
OFF - 72 hr
(in 1 hour increments)
OFF - 72 hr
(in 1 hour increments)
Other American Standard Remote Control manuals