4 2014-07-30 127139 13
4.9.1Running an Auto step calibration................................... 72
4.9.2Auto Step test results..................................................... 75
4.10Sensor Setup menu .................................................................... 77
4.10.1Setting the additional stability time (A version).............. 77
4.10.2Setting the parameters for TRUE – reference sensor (B
and C versions only) ...................................................... 78
4.10.3Setting the parameters for DLC– dynamic load
compensation (B and C versions only).......................... 80
4.10.4Setting the parameters for SUT– Sensor under test (B
versions only)................................................................. 80
4.10.5Viewing the Reference and DLC data (B and C versions
only) ............................................................................... 83
4.11Calibrator Setup menu................................................................ 85
4.11.1Setting the temperature parameters .............................. 85
4.11.2Setting the temperature resolution................................. 87
4.11.3Setting the sound, volume and operating mode ............ 88
4.11.4Setting calibration interval.............................................. 89
4.11.5Changing the date and time........................................... 89
4.11.6Saving a setup ............................................................... 90
4.11.7Loading a setup ............................................................. 91
4.11.8Resetting the instrument setup to factory defaults ........ 92
4.11.9Network Configuration (for service use only)................. 92
4.12Selecting the stirrer speed (RTC-158/250 A/B/C only)............... 93
4.13Information Screen ..................................................................... 94
4.14About the calibrator..................................................................... 95
4.15Simulation or training .................................................................. 95
5.0After use..................................................................................97
5.1Storing and transporting the calibrator ....................................... 97
5.1.1Transporting the dry-block calibrator ............................. 99
5.1.2Transporting the liquid bath calibrator ......................... 100
5.2Emptying the well (liquid baths only) ........................................ 100
6.0Replacing the main fuses....................................................102
6.1Error messages ........................................................................ 103
6.2Returning the calibrator for service........................................... 105
7.0Maintenance..........................................................................108
7.1Cleaning.................................................................................... 108
7.2Adjusting and calibrating the instrument................................... 110
7.3Maintenance of STS-reference sensor..................................... 111
7.4Maintenance of DLC-sensor..................................................... 113
8.0Technical specifications......................................................114