Amico Go Lift Portable 450 User manual

Installation and Operation Instructions
www.amico.com
Portable450 Portable450

2Amico Mobility Solutions Corporation
Contents
Introduction 3-4
Symbols used in this Manual 4
Markings 4
Safety Instructions 5
Technical Specifications for Amico GoLift Portable 6
Anatomy of the GoLift Portable 7
Basics in Transferring a Patient 8-9
Installing the Portable trolley in the Amico PLP Track 9-10
Installing the Endstop on the Amico Track 10
Using the Reacher Bar 11
Connecting the Hand Control to the GoLift Portable 11
Operating the GoLift Portable 12
Charging the GoLift Portable 12
LED and Buzzer Functions 13
Emergency Stop 13
Emergency Lowering 14
Manual Emergency Lowering 14
Overspeed Cam 14
Cleaning and Disinfection 15
Troubleshooting 15
Inspection and Maintenance 16
GoLift Portable Accessories 17
GoTrack Systems and Support Structure 18
Environmental Conditions 19
Electro-Magnetic Compliance Data for Amico GoLift Portable 20-23
Warranty 24

www.amico.com 3
Introduction
APPLICATION
The primary purpose of the GoLift Portable system is to safely lift and transfer a patient within an institutional setting
or at home. The GoLift Portable allows these transfers with as little effort as possible for the caregiver, while providing
dignity to the patient. A ceiling lift is simple and safe for the caregiver as well as the patient. The lift can be easily
moved by a caregiver by along the track from one location to another.
DESCRIPTION
The Amico GoLift Portable is an ideal ceiling lift system designed for routine transfers of patients. The most compact
portable lift in its class, the Amico GoLift Portable is designed to be aesthetically pleasing to both the caregiver and
the patients. In an effort to address infection control requirements, we gave careful consideration to the smooth
edges and rounded corners of our ceiling lift, carry Bar and hand control. When you look under the cover of this
compact lift, you will find an impressive set of all metal gears and state-of-the-art battery technology that will allow
the caregiver to safely, and effortlessly, transfer a patient weighing up to 450 lbs on a single lift.
The GoLift Portable weight capacity is 450 lbs. This weight limit must not be exceeded.
The Amico GoLift Portable is a fixed lift and can be installed in a variety of track profiles (the trolley can be customized
to fit these track profiles) and it also works seamlessly with the Amico GoLift Portable Pendant (PLP) System.
Contents of Packaging:
1. Amico GoLift Portable
2. Hand Control
3. Charger
4. Owner’s Manual
5. Portable Trolley
Upon receipt of the packages, verify it against the packing
slip to ensure the shipment is complete and inspect the equipment for
possible damage. If there is any damage, DO NOT USE the equipment
and notify the carrier immediately to file a claim. Provide complete
information concerning damage claims or shipping errors to Amico
Mobility Solutions Corporation. Include all equipment identification
numbers along with a description of the damaged parts.

4Amico Mobility Solutions Corporation
SYMBOLS USED IN THIS MANUAL
Symbol Reference Title
!
ISO 7000-0434A Caution risk of danger
TUV Certified by TUV
IEC 60417-5840 Type BF Applied Part
CE Certification of Conformity
ISO 7010-M002 Refer to instruction manual/booklet
Amico Mobility Emergency Lowering
N/A N/A
!
WARNING: This symbol is intended to alert the user of hazard or unsafe practices, which could result in
serious bodily harm.
MARKINGS
The GoLift Portable 450 are designed to comply with the following Standards:
Standard(s): CAN/CSA-C22.2 No. 60601-1:08 Medical Electrical Equipment – Part 1: General requirements for
basic safety and essential performance. ISO10535:2006 Hoists for the transfer of disables persons
– Requirements and Test Methods. CAN/CSA-C22.2 60601-1-11:2-15 home use environments.
Product: Amico GoLift Portable
Brand Name: GoLift
Models: 450 lbs
Introduction
AJW Technology Consulting GmbH
Königsallee 106
40215 Dusseldorf
Germany
7.338
Amico Mobility Solutions Corporation
Email: [email protected]
1. 8 7 7. 4 6 2 . 6 4 2 6
M
A
D
E
I
N
N
O
R
T
H
A
M
E
R
I
C
A
!
Charger: Wall Industries 36V 1.11A 40W
Duty Cycle: 1 Min ON / 9 Min OFF
Serial #:
Mfg. Date:
GLP450-9999
450 lbs (205 kg)
7.315
A A
B B
4
4
3
3
2
2
1
1
Front
Back
0.8341”
1.6256”
IP23

www.amico.com 5
Safety Instructions
!
READ THESE INSTRUCTIONS CAREFULLY OR SERIOUS INJURY MAY OCCUR
• The Amico GoLift Portable must be installed only by personnel authorized by Amico Mobility Solutions Corporation.
• Do not use this equipment prior to understanding the contents of this manual.
• Contents of this manual are subject to change without prior notice to users. Keep for future reference.
• Never place the Amico GoLift Portable, track/PLP and sling(s) in control of a person who has not been properly trained
in the use and care of this equipment.
• The Amico GoLift Portable and associated Track/PLP and sling(s) are for transferring patients only. Never use the GoLift
Portable for any other purpose.
• Amico Mobility’s Warranty is void if unauthorized personnel perform service on the Amico GoLift Portable system.
• In facilities where more than one caregiver is be responsible for using the Amico GoLift Portable and associated track
and slings, it is important that all caregivers be trained in the proper use of this equipment. A training program should
be established by the facility to familiarize new caregivers with this equipment.
• Do not expose the Amico GoLift Portable directly to water. Warranty does not cover any misuse or abuse of the Amico
GoLift Portable.
• The Amico GoLift Portable should be inspected and maintained on a regular basis to keep it operating safely and
correctly. Refer to page 16 of this manual.
• Any accessories used with the Amico GoLift Portable including the track/PLP and sling(s) should be checked to ensure
that they are in good working order. Check for signs of wear or fraying prior to use. Report any unusual wear or damage
immediately.
• Amico Mobility will not be responsible for any damage caused by misuse, neglect or purposeful destruction of the lift
and its associated components. Do not attempt to modify/alter the Amico GoLift Portable.
• Do not in any circumstance exceed the maximum allowable load of this lift. Refer to the “Technical Specifications”
section of this manual and/or the labels on the lift.
• The installation of the lift, track and sling are certified to a maximum load. Do not exceed the maximum rated load of
any of the components.
• There is a risk of explosion if the lift is used in the presence of flammable anesthetics.
• The Amico GoLift Portable should be decommissioned/disposed of after the recommended service lift in accordance
with local law regulations.
• If the battery case cracks and the contents of the battery contact your skin or clothing rinse immediately with warm
water.
• If the contents of the battery contact your eyes, rinse immediately with plenty of water and seek medical attention
• If the contents of the battery can cause respiratory irritation. Provide fresh air and seek medical attention.
!
There are no known contraindications associated with the use of the Amico GoLift Portable and its accessories,
provided they are used per our recommendations and guidelines.
However, for any independent uses of the Amico GoLift Portable, it is extremely important that the patient is able to receive
assistance, during the transfer in the event of an equipment failure. This assistance can be provided in the form of; a nearby
qualified caregiver, a phone or other communication device.

6Amico Mobility Solutions Corporation
Technical Specifications for Amico GoLift Portable
M
A
D
E
I
N
N
O
R
T
H
A
M
E
R
I
C
A
Maximum Lifting Speed:
No load: 1 inch/second
150 lbs/68 kg: 0.6 inches/seconds
450 lbs/204 kg: 0.5 inch/seconds
Unit Weight:
6 lbs
Safe Working Load (SWL):
450 lbs
Dimensions (see diagram):
Length: 19.4" (049.27 cm)"
Height: 5.2" (13.20 cm)
Depth of Lift: 3.5" (8.89 cm)
Strap Length: 74" (187.96 cm)
Number of Lifts per Charge (Duty: 10/90):
89 with 150 lbs (25% of strap at midrange)
52 with 450 lbs (25% of strap at midrange)
Charging time: 2 - 4 Hours
Lift Case:
Flame retardant ABS
IP27 hand control - 1m water resistance
IP23 Lift - protected from fingers and water
spray less 60 degrees from vertical
Batteries:
High Capacity, Long Lasting NiMH
1 x 24V (1.6Ah)
Battery Charger:
100-240 VAC 40 Watt
Hand Control:
Wireless
Safety:
Emergency Stop: Yes
Emergency Lowering Device: Independent Electrical
Emergency Manual Lowering
Upper Limit Detection
Lower Limit Detection
Slack Tape Sensor
Free Fall Brake (Over Speed Governor)
Low Battery and Dead Battery indication
Soft Start and Stop
Overload Protection
Approvals:
CAN/CSA-C22.2 No 60601-1:08
CAN/CSA-Z10535.1:15
Expected Service Life:
22,500 cycles or 10 years. It is recommended that the lift
be replaced after this time.
The expected life of slings, batteries fuses, straps and cord
is dependent upon proper maintenance.
19.4"
18.2"
5.2"
81"

www.amico.com 7
Anatomy of the GoLift Portable
Emergency lowering/stop feature
Unit controls
Quick magnetic hand
control connection
Wireless hand control
Durable all metal gears
Lifts up to
450 lbs!
GoLift Portable weighs
just 6 lbs

8Amico Mobility Solutions Corporation
Basics in Transferring a Patient
Lifting Sling:
A lifting sling with four to six straps designed for mounting on hooks
should be used with the Amico Portable 450 lift. Place the straps on the
hooks and make sure the loops are not in the latch so the straps do not
unintentionally fall off.
Amico Mobility shall not be liable for faults or accidents due to
incorrect use of the lifting sling, or for reasons of inadequate attention
on the part of the caregiver or patient.
Working with the Amico GoLift Portable:
The Amico GoLift Portable moves freely in the track system and does
not have any special requirements for space or power in connection
with moving. Attention can be fully focused on the user’s functional level and the caregivers technique.
To use the GoLift Portable correctly, the patient should only be lifted to the extent that she/he is clear of the surface
and should be moved at this height.
Attaching the Lifting Sling:
Place the straps from the lifting sling on the hooks of the GoLift Portable Start with the uppermost set of straps (from
the back) and then take the lowest set of straps (from the legs).
!
Be careful when attaching the lifting sling on the hooks. Check that the straps have been completely through
the opening and into place in the Carry Bar hooks. When pressing the up button to lift the patient, check
again that all straps remain correctly placed.
Lifting to and from a seated position:
• When lifting a patient from e.g. a wheelchair, move the Amico
GoLift Portable towards the patient to be lifted.
• The lifting hooks should be at the same height as the patient’s
chest and should not be moved further in over the user to
approximately mid-though position.
• Place the lifting hooks parallel to the patient’s shoulders.
• Place the lifting sling behind the user’s back between the back of
the chair and the user’s back.
• The center band of the lifting sling should follow the user’s spine.
Lead the leg straps along the outer sides of the patient’s shins and
beneath the thighs between the hollow of the knees and the hip
joints. Cross the leg straps in front of the user.
• All four lifting straps are now ready to be attached. The lifting sling
can now be mounted on lifting the hooks.

www.amico.com 9
Lifting to and from lying position in bed:
• Bring the GoLift Portable over the center of the patient to be lifted.
• Place the hooks parallel to the patient’s shoulders.
• Turn the patient onto his or her side. The sling should be placed so that the top of the sling is at the same height
as the top of the user’s head. Now position the sling over the user so that the center band follows the user’s spine.
Turn the user onto his or her back and pull out the remaining part of the lifting sling. Place the leg straps beneath
the user’s thighs and cross them. All four lifting straps are now ready to be attached and the lifting sling can now
be mounted on the hooks. It is an advantage to elevate the head of the bed so that the patient is sitting up.
• Only persons who have received competent instruction regarding the use of the lifting equipment and fitting of
slings should use the Amico GoLift Portable.
!
Important: Plan the move and avoid leaving the patient in the sling unattended. Before lifting, check that the
patient is completely free of his/her surroundings. The patient’s head, arms, hands and feet must not be in
danger or becoming trapped. Be careful with any tubes and wires that are attached to the user. Check that
the hand control and hand control cable is free of hanger, patient and other object before the lift is activated
up or down moved.
Never leave the lift unattended in an area such as a Psychiatric Ward with the patient.
Installing the Portable Trolley in the Amico PLP Track
1. Remove both the end caps on the PLP Arm.
2. Make sure the portable trolley is inserted correctly in the PLP track.
Basics in Transferring a Patient

10 Amico Mobility Solutions Corporation
Installing the GoLift Portable in the Amico PLP Track
3. By using an adjustable wrench, remove the end stopper on the
PLP track by loosening the bolts.
4. Fasten the end stops back on to the PLP track. Ensure there is enough clearance room to place the end cap on the
PLP arm and place the end cap back on the arm.
5. Fasten all end stops tightly using an adjustable wrench.
!
NOTE: If you are installing the Amico GoLift Portable in an existing track system you must ensure that the max
load of the track system is equal or higher than the max load of the Amico GoLift Portable.
6. To disconnect the lift from the trolley, open up the hook in the carabiner to release.
Installing the Endstop on the Amico Track
Amico Track Endstop
Use a 3/16 Allen key to
secure the endstop to
the track.

www.amico.com 11
1. With the lift resting on a bed, cart, table or
other stationary object, use the down button
on the lift or hand control to release enough
of the strap to the point that will allow you to
connect the carabiner to the trolley without
having to pick up the lift from its position.
2. Open the carabiner latch by pressing the
colored latch towards the back of the
carabiner. This "opens" the carabiner so
that it can be attached to the latch of the
Reacher Arm. Once the carabiner is hooked
onto the Reacher latch, release the latch of
the carabiner so that it will be closed, and
therefore locked in place.
3. Raise the Reacher Arm up to the trolley. Make
sure there is enough slack on the lift strap to
accomplish this step otherwise it will not be
possible to hook on.
4. Slide the top hook of the Reacher Arm
through the trolley. Make sure there is enough
slack on the lift strap to accomplish this step
otherwise it will not be possible to hook on.
5. The lift has now been securely hooked onto
the trolley. The Reacher arm remains in
position until the lift is removed.
Using the Reacher Bar
Open carabiner to attach
in slot of the Reacher bar.
Raise the Reacher bar to
the trolley.
Slide the top hook
through the trolley.
Lift is securely hooked
onto the trolley. The
Reacher arm remains in
position.
Closed carabiner attached to latch of the
Reacher bar.
Connecting the Hand Control to the GoLift Portable
Capacitive touch hand control – Allows users with limited dexterity to
effortlessly operate the unit.
Magnetic Hand control connect for quick attachment to the GoLift Portable.
!
Do not place the hand control within 5" of a pacemaker.
Patients with pacemakers must follow the instructions provided by
their doctors.

12 Amico Mobility Solutions Corporation
The Amico GoLift Portable is switched on automatically when a button on the
hand control is pressed. The Amico GoLift Portable switches off automatically
after approximately three minutes without activation.
To replace the battery in the hand control, use a Phillips screwdriver to remove
the cover.
To pair & unpair the hand control to the portable lift:
1. Press and hold both the up and down button simultaneously on the hand control until the LED flashes Red and
Green (5 seconds), this will put the hand control in paring mode and unpair from the current lift. The hand control
will search for 10 seconds before going to sleep. If the lift is not found and paired during this time, wake the hand
control up again by pressing any button. The hand control will wake up still in pairing mode.
NOTE: The hand control holds the memory, if the lift is unpaired before the hand control, as soon as the hand control
is awake, it will override the lift and they will remain paired.
2. Press and hold both the buttons on the portable lift for 5 seconds, until the lift blinks Green. This indicates that the
lift is ready to be paired to a new hand control.
Charging the GoLift Portable
The charger plugs in into the port on the GoLift Portable.
The batteries should be charged on a regular basis. It is recommended that the lift be left on charge when not in
operation, and at the end of each day. This will maximize the life cycle of the batteries.
The GoLift Portable may remain connected to the charger indefinitely since the charger has a built-in regulator,
eliminating the danger of overcharging.
As a general rule, it is recommended that the lift be raised to a height that will not interfere with anything or
anyone while the lift is not in use.
!
Do not drain the batteries excessively. This will dramatically reduce the lifespan of the batteries.
If the slow beep sounds, be sure to recharge the battery as soon as possible.
!
If the GoLift Portable is not returned to the charger the batteries will be drained and damaged.
Operating the GoLift Portable
DOWN button
UP button

www.amico.com 13
LED and Buzzer Functions
Charging Battery Main Lift LED Main Lift Buzzer
Yes Full Green N/A
No Full Green N/A
Yes Not full Orange N/A
No Low Flashing Orange N/A
No Very low Solid Red 1 beep every 3 seconds
No Max Current State Flashing Red Continuous beep until max
current state excited
Yes Charging Solid Orange N/A
Yes Top-up Solid Orange for 5 seconds when charger plugged in then
solid Green N/A
Yes Battery over temp/
Sensor error Flashing red 1 beep every 5 seconds
Yes No temp sensor detected Flashing red 3 times per second then 1 second off 1 beep every 5 seconds
Function/Error Main Lift LED Main Lift Buzzer
Ready to Repair Flashing Green 4 times N/A
Emergency down Solid Red 2 Beeps per second
Connection Function/Error Hand Control LED Hand Control Buzzer
N/A Unpair/Search Flashing green then Red 3 times N/A
N/A Low battery Solid Red N/A
Connected Show main lift LED status Show main lift LED status N/A
Not connected Not paired Slow flashing Green N/A
Not connected Paired with main lift Fast flashing Green N/A
Maintenance Alarm
At 1001 lifts, the LED will flash Green
To reset, press Up button, then Down button, then Up button, then Down button and then hold both buttons for 10 seconds.
If successful, the LED turns Red, Green and Orange. The buzzer beeps three times.
Please note, that a reset can be performed at any time, not just after the 1000th lift.
Emergency Stop
The GoLift Portable also has an Emergency Shut-off feature that allows the operator
to shut-off the power to the lift completely in the event of an emergency. By
pressing once on the RED emergency tab, located on the top of the lift, the lift will
immediately stop and all its functions will be disabled. The ON indicator light will
turn off, and the Emergency Shut-off button located inside the lift will pop out.
After an emergency, the lift must be inspected prior to restoring to use. In order to
restore power to the lift, the tab must be gently pulled.
!
Do not press the red tab forcefully. For assistance after an emergency,

14 Amico Mobility Solutions Corporation
Emergency Lowering
In the event that the DOWN button on the hand control does not function, or in power failure situations, the patient
may be lowered by pressing and holding down the RED emergency stop button located on the top of the lift.
Continue to press down until the patient is safely lowered to the desired position. The lift will beep as you continue
to press down on the red button and will continue beeping until the button is released after the desired lowering has
been achieved.
NOTE: The emergency lowering button does not provide a raising function. The failure of any of the lowering devices
should be reported to Amico Mobility.
!
In an event of emergency when normal lowering system of the lift malfunctions and the “emergency
lowering”function is used, the lift must be reset by a qualified lift technician before re-use.
Manual Emergency Lowering
The manual emergency and raising should only be used if the emergency lowering does not work. A proper safety
ladder or stool may be required to remove the plug from the cover. Remove the round plug from the lift cover and use
3/16" or 2.5 mm Allen key to rotate the motor in the up or down direction.
!
Do not attempt to use the lift while using manual lowering.
Remove the round plug from the lift cover and use 3/16" or 2.5 mm Allen key to rotate the motor in the up or down
direction.
Overspeed Cam
The Overspeed Cam brake is made of a metal bar fixed to the drum. In case of gear or motor breakage, the centrifugal
force created will block the bar against the frame.
Emergency RED button
Continue to press down until the
patient is safely lowered to the
desired position.

www.amico.com 15
Cleaning and Disinfection
The exterior of the GoLift Portable should only be cleaned, disinfected using the recommended cleaning agents
shown below. Damp a cloth with the cleaning agent and wipe down the entire exterior of the lift. Other chemicals
and/or liquids not listed should not be used to clean and disinfect this lift.
!
Take great care to ensure that no liquids get inside the Amico GoLift Portable. The lift is not drip proof or
water tight. Failure to protect the lift from liquids may result in damage to the lift and may cause personal
injury.
Recommended cleaning agents:
• Virox Accel TB
• Virox 5
• Dispatch Hospital Cleaner disinfectant towels with Bleach
• Clorox Healthcare Professional Disinfecting Bleach Wipes
• Sani-Cloth super germicidal disposable wipes
• Virocidin-X
Clean the inside of the tracks every 4 months using a damp cloth to ensure a better trolling surface for the trolley.
Troubleshooting
Should problems arise with the use of the Amico GoLift Portable, review the following chart. Find the fault and
complete the recommended solution. If the fault is not found and or/the solution does not correct the problem
contact Amico Mobility.
Fault Recommended Solution
The GoLift Portable emits a slow beep while in use. The
lift only goes down but not in the upwards direction. The batteries are low and the lift should be charged.
The GoLift Portable cannot lift. If the load is in excess of the safe working load the GoLift Portable will
not work due to over-current protection.
The batteries are always dead after a few charges Replace the batteries as they may be at the end of their service life
The GoLift Portable does not operate when you press
the buttons on the hand control. Check that the emergency cord is not pulled.
The lift does not go up There may be a twist in the lift strap.

16 Amico Mobility Solutions Corporation
Inspection and Maintenance
Prior to using the Amico GoLift Portable, the inspections should be conducted per the following schedule:
Item Before Use Every Month Annual
GoLift Portable •
Ensure that end stops are installed •
Inspect strap for wear or fraying •
Ensure Batteries are charged •
Inspect hooks for damage or sharp edges •
Inspect the wheels in the portable trolley. Replace if damaged. •
Sling
Check all straps for wear or fraying •
Inspect the sling for any damage in the fabric •
Ensure there are no loose threads in the stitching •
Maintenance by a certified technician
Check the strap and replace only if frayed or damaged •
Inspect the gearbox for any unusual noises •
Inspect the hook components for any signs of damage •
Inspect the gears for any broken or worn teeth •
** Verify the overspeed cam is operating freely •
Check emergency stop tab •
** Check emergency lowering device •
*Annual load test with SWL (maximum safe working load) •
Ensure the end stops are installed •
* In accordance to the ISO 10535 Standard“Hoists for the transfer of disabled persons – Requirements and test
methods”an inspection should be performed on the GoLift Portable at least once a year. This inspection should be
performed by a qualified technician and should include a working load test of one (1) lifting cycle with the maximum
load.
** These two functions must be checked by a qualified technical to ensure the essential performance of the GoLift
Portable.
!
Do not operate the GoLift Portable until any issues discovered during the inspection have been addressed by
a certified technician.

www.amico.com 17
GoLift Portable Accessories
Slings Material S M L XL
BasicGo
LongGo
ActiveGo
ComfortGo
HygieneGo
CustomGo
TrackGo
LimbGo
TurnGo
GoBasic
With Head Support Polyester SLG-GO-BASIC-WV-HS-S SLG-GO-BASIC-WV-HS-M SLG-GO-BASIC-WV-HS-L SLG-GO-BASIC-WV-HS-XL
GoBasic
With Head Support Mesh SLG-GO-BASIC-MSH-HS-S SLG-GO-BASIC-MSH-HS-M SLG-GO-BASIC-MSH-HS-L SLG-GO-BASIC-MSH-HS-XL
GoBasic
No Head Support Polyester SLG-GO-BASIC-WV-S SLG-GO-BASIC-WV-M SLG-GO-BASIC-WV-L SLG-GO-BASIC-WV-XL
GoBasic
No Head Support Mesh SLG-GO-BASIC-MSH-S SLG-GO-BASIC-MSH-M SLG-GO-BASIC-MSH-L SLG-GO-BASIC-MSH-XL
BasicGo
LongGo
ActiveGo
ComfortGo
HygieneGo
CustomGo
TrackGo
LimbGo
TurnGo
GoLong Polyester SLG-GO-LONG-WV-42X78 SLG-GO-LONG-WV-56X78
GoLong Mesh SLG-GO-LONG-MSH-42X78 SLG-GO-LONG-MSH-56X78
BasicGo
LongGo
ActiveGo
ComfortGo
HygieneGo
CustomGo
TrackGo
LimbGo
TurnGo
GoHygiene Polyester SLG-GO-HYGIENE-WV-S SLG-GO-HYGIENE-WV-M SLG-GO-HYGIENE-WV-L SLG-GO-HYGIENE-WV-XL
BasicGo
LongGo
ActiveGo
ComfortGo
HygieneGo
CustomGo
TrackGo
LimbGo
TurnGo
GoComfort Polyester SLG-GO-COMFORT-WV-S SLG-GO-COMFORT-WV-M SLG-GO-COMFORT-WV-L SLG-GO-COMFORT-WV-XL
Guidelines For Choosing Your Sling Size
S M L XL
Height 31.4 - 34.5" / 798 - 876 mm 33.4 - 36.2" / 848 - 820 mm 35.4 - 38.2" / 899 - 970 mm 37.4 - 41.3" / 950 - 1049 mm
Width 14.2 - 15.8" / 361 - 401 mm 15.4 - 17" / 391 - 432 mm 16.5 - 17.7" / 419 - 450 mm 19 - 20.5" / 483 - 521 mm
Washing Instructions
176°F
80°C
Normal washing at
indicated temperature
Do not bleach Tumble dry low Do not iron Do not dry clean

18 Amico Mobility Solutions Corporation
GoTrack Systems and Support Structure
Description Part Number Length x Width x Height Product Weight Material
Regular TRK-REG-96 96 x 2.63 x 2.36" / 2440 x 67 x 60 mm 8.3 lbs / 3.77 kg White powder coated extruded aluminum
Regular TRK-REG-120 120 x 2.63 x 2.36" / 3050 x 67 x 60 mm 8.3 lbs / 3.77 kg White powder coated extruded aluminum
Regular TRK-REG-240 240 x 2.63 x 2.36" / 6100 x 67 x 60 mm 8.3 lbs / 3.77 kg White powder coated extruded aluminum
Double TRK-DBL-120 120 x 2.63 x 3.55" / 3050 x 67 x 90 mm 10.4 lbs / 4.72 kg White powder coated extruded aluminum
Double TRK-DBL-150 150 x 2.63 x 3.55" / 3800 x 67 x 90 mm 10.4 lbs / 4.72 kg White powder coated extruded aluminum
Double TRK-DBL-240 240 x 2.63 x 3.55" / 6100 x 67 x 90 mm 10.4 lbs / 4.72 kg White powder coated extruded aluminum
Triple TRK-TRP-120 120 x 2.63 x 5.75" / 3050 x 67 x 146 mm 12.4 lbs / 5.65 kg White powder coated extruded aluminum
Triple TRK-TRP-240 240 x 2.63 x 5.75" / 6100 x 67 x 146 mm 12.4 lbs / 5.65 kg White powder coated extruded aluminum
Custom Custom lengths in any track type available upon request
GoTrack Configurations
We offer three different track profiles, as well as
modular and custom configurations that come in:
• Straight Track
• Curved Systems
• Full Room Coverage (X/Y or H Systems)
• Optional Infection Control Track
• Optional recessed, flush mount track
systems
All GoTrack Systems feature:
• Low weight and high capacity
• Fast and flexible installation
• Hidden End Stops for safety
• Anywhere charge
Approved by:
PROJECT MANAGER : ___________________
ENGINEERING : __________________________
PRODUCTION : __________________________
REV.
DESCRIPTION
BY
DATE
5
4
3
1
55 East Wilmot St.
Richmond Hill, Ontario
L4B 1A4, CANADA
Tel: (905)764-0800
Fax: (905)764-0862
DESCRIPTION:
PART NO:
SHEET:
OF
DRAWING NO:
11
PROJECTION
PRIVATE AND CONFIDENTIAL
THIS PRINT IS PROPERTY OF AMICO CLINICAL AND IS LOANED IN CONFIDENCE SUBJECT TO RETURN UPON
REQUEST AND WITH THE UNDERSTANDING THAT NO COPIES ARE TO BE MADE WITHOUT THE CONSENT OF
AMICO CLINICAL.
ALL RIGHTS TO DESIGN OR INVENTION ARE RESERVED
DATE:
DATE:
CHECKED BY:
DRAWN BY:
NTS
SCALE:
2
A
B
C
D
FINISH:
MATERIAL:
YYMMDD-XXX
UNLESS SPECIFIED:
ALL DIMENSIONS IN INCHES
SURFACE FINISH 63 uin
BREAK ALL EDGES 0.005 - 0.010
REMOVE SHARP EDGES AND BURRS
ANGLE TOLERANCES
0.25
CONCENTRICITY OF 0.005"
STRAIGHTNESS & FLATNESS
PARALLELISM
DECIMAL TOLERANCES: X.XX
.05
X.XXX
.005
X.XXXX
.0005
Regular
Approved by:
PROJECT MANAGER : ___________________
ENGINEERING : __________________________
PRODUCTION : __________________________
REV.
DESCRIPTION
BY
DATE
5
4
3
1
55 East Wilmot St.
Richmond Hill, Ontario
L4B 1A4, CANADA
Tel: (905)764-0800
Fax: (905)764-0862
DESCRIPTION:
PART NO:
SHEET:
OF
DRAWING NO:
11
PROJECTION
PRIVATE AND CONFIDENTIAL
THIS PRINT IS PROPERTY OF AMICO CLINICAL AND IS LOANED IN CONFIDENCE SUBJECT TO RETURN UPON
REQUEST AND WITH THE UNDERSTANDING THAT NO COPIES ARE TO BE MADE WITHOUT THE CONSENT OF
AMICO CLINICAL.
ALL RIGHTS TO DESIGN OR INVENTION ARE RESERVED
DATE:
DATE:
CHECKED BY:
DRAWN BY:
NTS
SCALE:
2
A
B
C
D
FINISH:
MATERIAL:
YYMMDD-XXX
UNLESS SPECIFIED:
ALL DIMENSIONS IN INCHES
SURFACE FINISH 63 uin
BREAK ALL EDGES 0.005 - 0.010
REMOVE SHARP EDGES AND BURRS
ANGLE TOLERANCES
0.25
CONCENTRICITY OF 0.005"
STRAIGHTNESS & FLATNESS
PARALLELISM
DECIMAL TOLERANCES: X.XX
.05
X.XXX
.005
X.XXXX
.0005
Double
Approved by:
PROJECT MANAGER : ___________________
ENGINEERING : __________________________
PRODUCTION : __________________________
REV.
DESCRIPTION
BY
DATE
5
4
3
1
55 East Wilmot St.
Richmond Hill, Ontario
L4B 1A4, CANADA
Tel: (905)764-0800
Fax: (905)764-0862
DESCRIPTION:
PART NO:
SHEET:
OF
DRAWING NO:
11
PROJECTION
PRIVATE AND CONFIDENTIAL
THIS PRINT IS PROPERTY OF AMICO CLINICAL AND IS LOANED IN CONFIDENCE SUBJECT TO RETURN UPON
REQUEST AND WITH THE UNDERSTANDING THAT NO COPIES ARE TO BE MADE WITHOUT THE CONSENT OF
AMICO CLINICAL.
ALL RIGHTS TO DESIGN OR INVENTION ARE RESERVED
DATE:
DATE:
CHECKED BY:
DRAWN BY:
NTS
SCALE:
2
A
B
C
D
FINISH:
MATERIAL:
YYMMDD-XXX
UNLESS SPECIFIED:
ALL DIMENSIONS IN INCHES
SURFACE FINISH 63 uin
BREAK ALL EDGES 0.005 - 0.010
REMOVE SHARP EDGES AND BURRS
ANGLE TOLERANCES
0.25
CONCENTRICITY OF 0.005"
STRAIGHTNESS & FLATNESS
PARALLELISM
DECIMAL TOLERANCES: X.XX
.05
X.XXX
.005
X.XXXX
.0005
Triple

www.amico.com 19
Environmental Conditions
Operation
Minimum Maximum
Temperature +10°C +40°C
Relative atmospheric humidity 30% 75%
Air pressure 700 hPa 1060 hPa
Transport/Storage
Minimum Maximum
Temperature -10°C +50°C
Relative atmospheric humidity 20% 90%
Air pressure 700 hPa 1060 hPa
NOTE: It is recommended that the lift we brought back to an area kept at room temperature between uses.
For example after transport in a vehicle do not leave the lift in the trunk or backseat.
Also, try to keep the lift from staying in rooms with a metallic roof where temperatures may increase the allowable
operating temperature.
References on the Package
RH
20% - 90%
P
700hPa - 1060hPa
-10 C°
+50 C°
Temperature range
for transport
and storage
Atmospheric
humidity
for transport
and storage
Air pressure
for transport
and storage

20 Amico Mobility Solutions Corporation
Electro-Magnetic Compliance Data for Amico GoLift Portable
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The [EQUIPMENT or SYSTEM] is intended for use in the electromagnetic environment specified below.
The customer or the user of the [EQUIPMENT or SYSTEM] should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11 Group 1
The [EQUIPMENT or SYSTEM] uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11 Class A
The [EQUIPMENT or SYSTEM] is suitable for use in all establishments
other than domestic establishments and those directly connected to
the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2 Class A
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in
improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify
that they are operating normally.
Use of accessories and cables other than those specified by Amico Mobility, could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment, and result in improper operation.
Portable RF communications equipment should be no closer than 30 cm to any part of the Amico GoLift Portable 450.
Otherwise, degradation of the performance of this equipment could result.
Table of contents
Other Amico Lifting System manuals
Popular Lifting System manuals by other brands

KAPTARLIFT
KAPTARLIFT MCHW user manual

Posi-Plus
Posi-Plus Line Runner 700-40 Operator's manual

Stannah
Stannah PICCOLO User handbook

Challenger Lifts
Challenger Lifts 12000 Operation installation maintenance manual

HADEF
HADEF 19/90-EX Installation, Operating and Maintenance Instruction

Yale
Yale MPB040-E manual