Anest Iwata Cliq DV-350 User manual

DV-350 Aspirator
Manual
Please read this manual carefully before using this device. Keep this manual for future reference.
Cliq_DV350_Manual (NUS_NCE_20201126)
3510387D

Table of Content
Symbol Definitions .........................................................................................
Important Safety Information ....................................................................
Product Introduction .....................................................................................
Part List ...............................................................................................................
Accessories / Replacement Parts ...............................................................
Operation ...........................................................................................................
Power and Indicator Lights .........................................................................
Charging the Battery ....................................................................................
Battery Operation ...........................................................................................
Battery Replacement .....................................................................................
Cleaning .............................................................................................................
Maintenance & Repair ...................................................................................
Warranty .............................................................................................................
Troubleshooting ..............................................................................................
Transportation, Storage, and Disposal ....................................................
Guidance and Manufacturer’s Declaration ............................................
P. 3
P. 4
P. 5
P. 6
P. 7
P. 8
P. 9
P. 9
P. 9
P. 10
P. 11
P. 11
P. 11
P. 12
P. 13
P. 14 - 15
2/16
Copyright © 2020 ANEST IWATA SPARMAX Co., Ltd. All rights reserved. Cliq_DV350_Manual (NUS_NCE_20201126)
3510387D

Symbol Definitions
3/16
Off
On
Manufacturer
IP22
Solid particle protection > 12.5 mm. Vertically falling drops shall have no
harmful effects when the enclosure is tilted at an angle up to 15° on either
side of the vertical.
Temperature limits.
Humidity limits.
Serial number.
SN
One time use gloves required.
Disposal of Electrical & Electronic Equipment (WEEE): This device contains
electrical and/or electronic equipment that must be recycled and should
be handed over to a suitable collection point for proper disposal.
Class II equipment
Manufacture date
Protection against electrical shock; type of applied part, body floating.
Caution: Consult instructions for warnings and safety precautions.
Refer to instruction manual.
MR Unsafe. This device must not be used in an MRI environment.
Copyright © 2020 ANEST IWATA SPARMAX Co., Ltd. All rights reserved. Cliq_DV350_Manual (NUS_NCE_20201126)
3510387D

4/16
Please read this manual carefully before using this device. Keep this manual for future reference.
Important Safety Information
Important information is highlighted by these terms:
WARNING: Danger for patient or operating staff.
CAUTION:Information for preventing damage to the product.
ATTENTION:Important operating instructions.
WARNING
• This device is intended for use by trained and certified healthcare professionals such as doctors, nurses, and therapists, or by those under their
guidance. Use only after consulting with your doctor and receiving proper training from a trained healthcare professional.
• Using this device for any purpose other than the stated intended use may result in injury to the patient or operator.
• For patients for whom a breakdown of the device can lead to a critical situation, a replacement or back-up device must always be readily available.
• To ensure proper hygiene, collection container, filter, suction tubing, and intermediate tubing must only be used by the same patient. Do not use
these accessories across multiple patients. Do not attempt to sterilize these accessories for use across multiple patients. Failure to abide by these
warnings may result in infection or cross contamination between patients.
• The suction tube must never come into direct contact with the suction area.
• Fluid collected in the collection container must be disposed of according to local regulations, institution policy (as applicable), and the instructions
of a certified healthcare professional.
• Do not use this device with accessories which have not been approved by the distributor or manufacturer of the product. Use of unapproved
accessories may cause harm to the patient and will void the warranty coverage for this device.
• Suction catheters are not included with this device and must be purchased separately, in accordance with the instructions of a trained
and certified physician or healthcare professional. Only use disposable, single-use, sterile suction catheters compliant with local and/or national laws,
regulations, and directives. Failure to do so could cause harm to the patient.
• To avoid the risk of infection, always use sterile single-use suction catheters. After use, immediately dispose of the suction catheter in accordance
with local regulations, institution policy (as applicable), and the instructions of a certified healthcare professional. Do not attempt to re-use or
re-sterilize suction catheters.
• During suction, do not exceed the maximum vacuum pressure recommended by your healthcare professional or physician.
• Ensure that the battery is fully charged before operating the device without access to mains.
• Never open, disassemble, or modify the device, or any accessories included with the device.
• Device is MR Unsafe. This device must not be used in an MRI environment. Further the device shall not be placed or operated in the following
environments or near the following devices: CT, diathermy machine, RFID transmitters, electromagnetic security systems (such as metal detectors).
To reduce the risk of burns, electrocution, fire or injury to persons:
• Do not store the device where it can fall or be pulled into a tub, sink, or other water source.
• Never place the device in water or other liquids. If it falls in water, do not reach for it. Unplug immediately.
• Never leave the device unattended when switched on.
• The device must stand upright during use.
• Do not use this device to suction explosive, flammable, or corrosive liquids.
• Keep the power cord away from hot surfaces.
• Keep the plug and ON/OFF switch away from moisture.
• Never use the device (1) while bathing, showering, or close to a water source (2) when tired or not fully alert (3) at high room temperatures
(4) in an environment where there is a risk of explosion.
• Close supervision is required when this device is used for, with, or near children or infants. This device should not be operated by children.
• Do not operate where aerosol products are being used or where oxygen is being administered in a closed environment such as an oxygen reservoir.
• Never operate this product if: (1) It has a damaged power cord or plug. (2) It is not working properly. (3) It has been dropped or damaged. (4) It has
been dropped into water. Send the product to an authorized service center for examination and repair.
CAUTION
• Before plugging in the device, check that local mains conforms to the specified voltage requirements listed on the AC adapter.
• Unless charging battery, unplug the device when not in use. Never pull the plug out of the mains socket by pulling on the power cable.
• This machine is an oil-less suction pump and does not require any lubrication. Do not add any lubricants to this machine.
• This aspirator is not suitable for prolonged continuous use.
• Before first use, battery must be charged continuously for at least 8 hours. Failure to do so may shorten the battery's lifetime.
• Fully charge the battery at least once every month. Failure to do so may shorten the battery’s lifetime.
ATTENTION
• Prior to first use, carefully unpack the unit and check for transportation damage. In case of damage, do not use.
• The device creates a negative pressure (vacuum) that draws fluids through a disposable suction catheter that is connected by suction tubing to a
collection container. The suctioned fluids are then trapped in the collection container for proper disposal.
• This aspirator is suitable for use at home, in hospitals or clinics, and in nursing homes.
• Before using any approved third party accessories with this device, check to make sure they are compatible - that all fittings are tight. Loose fittings
may degrade the performance of this device and may prevent the device from operating at its optimal level.
• Battery will gradually discharge when not in use; this is normal behaviour.
Copyright © 2020 ANEST IWATA SPARMAX Co., Ltd. All rights reserved. Cliq_DV350_Manual (NUS_NCE_20201126)
3510387D

Product Introduction
Device Description
The DV-350 aspirator is a portable AC/DC high-vacuum / high-flow suction pump. The DV-350 can be powered from a standard wall outlet or from a
rechargeable battery pack that provides up to 40 minutes of operational lifetime. The DV-350 creates a negative pressure (vacuum) that draws fluids
through tubing and into a collection container where the fluids are trapped for proper disposal. The device is comprised of a maintenance-free pump
unit, an AC power adapter with power cord, an on/off switch, a pressure relief valve and pressure adjustment knob, a pressure gauge, an inline filter,
intermediate tubing, collection canister, suction tubing, and a rechargeable battery pack. Disposable suction catheters are not packaged with this
device and must be purchased separately. Use only sterile single-use suction catheters that are compliant with local and/or national laws, regulations,
and directives.
This device must only be used on the order and instruction of a physician. If practiced outside of the hospital setting, the care giver must be trained
by a certified healthcare professional and the training must be recorded and documented.
Intended Use / Indications for Use
The DV-350 aspirator provides a portable, battery-powered medical vacuum source. The DV-350 can also be powered from a standard wall outlet with
the use of an AC power adapter. It is intended for use in the homecare or hospital environment.
Service Life
The expected service life of the device is 4 years. The respective service life of each accessory is listed on page 7.
5 / 16
Product Specification
Input Voltage: 12 VDC
Power consumption: 36 W
Max working air flow: ~20 l/min
Vacuum: 0 - 620 mmHg
Dimensions: L34 x W16.5 x H19 cm
Weight: 3.5 kg / 7.7 lbs
Collection container capacity: 800 ml
Power cord length: 180 cm
Operating temperature: 5 - 40 ˚C
Operating relative humidity: 15 - 93%
Operating atmospheric pressure: 70 - 106 kPa (10.2 - 15.4 psi)
Storage and transport temperature : -20 - 50 ˚C
Storage and transport humidity: 15 - 93%
Storage atmospheric pressure: 50 - 106 kPa (7.3 - 15.4 psi)
Storage and transport temperature : -25 - 70 ˚C
Storage and transport humidity: 0 - 93%
Storage and transport temperature : -20 - 50 ˚C
Storage and transport humidity: 0 - 93%
Switching Power Specification
Product Manufacturer: Fuhua Electronic Co., Ltd.
Model: UE48-120300SPA1
Input: 100-240 VAC
Output: 12 VDC, 3.0A
Battery Specification
Battery type: NiMH
Voltage: 12V
Capacity: 2000 mAh
Copyright © 2020 ANEST IWATA SPARMAX Co., Ltd. All rights reserved. Cliq_DV350_Manual (NUS_NCE_20201126)
3510387D

Part List
1. Intermediate tubing
2. Inline filter
3. Vacuum port
4. Patient port
5. Collection container holder
6. Suction tubing
7. Collection container
8. Collection container coaster
9. Power switch
10. Vacuum gauge
11. Inlet
12. Power cord
13. AC adapter
14. Vacuum regulator knob
15. Handle
16. Power and indicator lights
17. Battery cover
18. Power inlet
19. Rechargeable battery
1
8
6
7
9
10
12 *
16
17
18
19
11 15
14
2
3
4
5
* Illustration of plug is for reference only. Plug type may vary depending on your region.
6/16
13
Copyright © 2020 ANEST IWATA SPARMAX Co., Ltd. All rights reserved. Cliq_DV350_Manual (NUS_NCE_20201126)
3510387D

Rechargeable battery AA*10*2000mA, NiMH
7/16
Accessories / Replacement Parts
The following accessories and replacement parts are available for purchase. Please address all inquiries to our distributors or authorized service
centers in your region.
Item Part No. Specification
Collection container
Collection container coaster
Intermediate tubing
(silicone)
Inline filter
Power cord
ø 150 x 185 (H) mm / 800 ml
ø 108 x 10 (H) mm, silicone
7 (ID) x 13 (OD) x 250 (L) mm
1/4" hose barb x 1/4" hose barb
3304025C
3402019A
3406073A
3406074A
3710006A
Use only with the same patient. Do not use across multiple patients.
WARNING:Suction catheters are not provided with this device and must be purchased separately, in accordance with the instructions of a trained
healthcare professional or physician:
• Only use disposable, single-use, sterile suction catheters.
• Only use the type and size of suction catheters that are recommended by your healthcare professional.
• Only use suction catheters that are compliant with local and/or national laws, regulations, and directives. Failure to do so, may cause harm to the
patient.
• Disposable suction catheters are single-use only; do not attempt to re-use or re-sterilize.
WARNING:If third party suction tubing is used, the third party suction tubing must also be compliant with local and national laws, regulations and
directives. Failure to do so may cause harm to the patient.
Must be replaced if damaged. One
month expected service life.
Must be replaced if damaged. One
month expected service life.
Must be replaced if damaged.
Must be replaced if damaged.
Replace if damaged, and/or if there is a noticeable
decline in battery lifetime.
Must be replaced if damaged or has been
in contact with fluid. One month expected
service life.
Suction tubing
(PVC)
7 (ID) x 10.4 (OD) x 1800 (L) mm3318007A Must be replaced if damaged. One
month expected service life.
AC adapter Input:
Output:
3709015A Must be replaced if damaged.
12 VDC, 3.0A
100-240 VAC, 50-60 Hz
Copyright © 2020 ANEST IWATA SPARMAX Co., Ltd. All rights reserved. Cliq_DV350_Manual (NUS_NCE_20201126)
3510387D
Type A plug
Type C plug
Type G plug
Type I plug
length: 1.8 m
3703057A
3703053B
3703060A
3703077A

8/16
Operation
Operator must wear disposable single-use gloves when operating and handling unit.
Prior to each use, inspect collection container and tubing for cracks or leaks, and replace all damaged accessories immediately. Do not use device if
accessories are damaged.
Place the machine on a flat and stable surface. WARNING:Inspect AC adapter and power cord. If either are damaged, do not use.
Place the collection container securely into the collection container holder and into the coaster cut-out.
Connect the filter into the inlet, ensuring that the“FLUID SIDE” is facing upwards. Connect one end of the intermediate tubing to the filter on the
“FLUID SIDE”. Connect the other end of the intermediate tubing to the“vacuum port” on collection container lid.
Connect one end of suction tubing to the “patient port” on the collection container lid.
Check all connections and make sure they are secure and tight. A poor connection could result in an air leak that diminishes pump performance.
Turn the power switch on.
Adjust the unit to its desired suction level before beginning patient suction.
a. Create a vacuum by completely blocking the open end of the suction tubing.
b. Adjust the vacuum regulator knob until the vacuum gauge displays the desired vacuum level.
c. As long as the open end of the suction tubing remains completely blocked, the vacuum level should remain steady.
Expert recommendations for vacuum settings during airway suctioning:
(AARC and Respiratory Care Journal, September 2004, Vol. 49, No 9, p1080)
Adult: 100-150 mmHg
Child: 80-120 mmHg
Infant: 60-100 mmHg
Once unit is set to the desired vacuum pressure, attach a disposable suction catheter to the open end of the suction tubing.
Proceed with suctioning as instructed by your healthcare professional. Remove suction catheter from patient immediately after suctioning.
After each use, turn off the power switch. Unplug the unit if not charging battery. Clean the unit and accessories as instructed in the “Cleaning”
section found on page 11 of this manual.
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
CAUTION
• Make sure the power switch is turned off before inserting or removing the plug.
• The unit should not be left running for over 1 hour continuously. If the machine has been operating continuously for 1 hour, please turn off the
power for at least 30 minutes before turning back on.
• Keep the machine and relevant accessories dry. Avoid direct sun exposure during storage.
WARNING:Set vacuum pressure in accordance with instructions provided by your healthcare professional. Never exceed their maximum
pressure guidelines. Failure to follow the instructions and guidance of your healthcare professional could cause bodily injury.
WARNING:Disposable medical suction catheters are not provided with this unit and must be purchased separately in accordance with the
instructions of a trained healthcare professional or physician. Only use suction catheters that are compliant with local and/or national
laws, regulations, and directives.
WARNING:Never operate the unit without the filter.
Connect to suction catheter
Connect suction tubing to “patient port” on
collection container lid.
Connect intermediate tubing to filter and
“vacuum port” on collection container lid.
“FLUID SIDE”
facing upwards.
Disposable suction catheter
(not included with DV-350)
WARNING:The suction tube must never come into direct contact with the suction area.
Copyright © 2020 ANEST IWATA SPARMAX Co., Ltd. All rights reserved. Cliq_DV350_Manual (NUS_NCE_20201126)
3510387D

Battery Operation
WARNING:To prevent interruption of suction, plug the unit into a working power source as soon as the red indicator light turns on. Operating while
the red indicator light is on will result in gradual performance decline and eventual shut-off of the unit.
Power and Indicator Lights
• Solid Blue Light: The unit is currently plugged into a working power source.
• Solid Green Light: Optimal battery level
• Solid Red Light: Battery requires charging. Charge battery as soon as possible.
• Flashing Red Light: Battery charging.
The unit is “ON”.
The unit is “OFF”.
Battery is charging.
The unit is “OFF”.
Battery is full.
The unit is “ON”and running on battery power,
but battery levels are low. Recharge battery
as soon as possible.
The unit is “ON”and running on battery
power.
The unit is “OFF”and not charging.
• When switched “ON”and connected to a working power source, the unit does not charge. To charge the unit, plug the unit into a working power
source when the unit is“OFF”.
• A fully charged battery will provide up to ~40 minutes of continuous operation.
• When the red indicator light turns on, the battery is low and will only provide ~5 minutes more of operation. Plug the unit into a working power
source as soon as possible.
Charging the Battery
WARNING:Ensure that the battery is fully charged before operating the device without access to mains.
CAUTION:If the unit is not in use for a prolonged period, make sure to charge the battery to full at least once every month.
• Before running the unit on battery for the first time, allow it to charge uninterrupted for 8 hours.
• To charge the unit, plug it into an AC power source. The blue indicator light confirms that the unit is plugged in. A flashing red light indicates that
the battery is charging.
• When the green light appears, the battery is charged and ready for use.
Unit plugged in Unit not plugged in
9/16
Copyright © 2020 ANEST IWATA SPARMAX Co., Ltd. All rights reserved. Cliq_DV350_Manual (NUS_NCE_20201126)
3510387D

Battery Replacement
WARNING
• It is normal for the battery pack to get warm during use. Please wait for at least 30 minutes after use before changing battery.
• Before replacing battery pack, please make sure the unit is turned off and unplugged. All indicator lights should be off.
The DV-350 includes a single battery pack. Additional and/or replacement battery packs can be purchased separately.
To replace the battery pack:
1. Open the battery cover next to the indicator lights, by sliding the cover to the right.
2. Slide the existing battery out of the slot.
3. Disconnect the battery by squeezing the connector.
4. Insert the new battery into the connector until you hear a click to ensure the connection is secure. Slide the battery into the battery slot and
re-attach the battery cover.
5. Fully charge battery before operating the device without access to mains.
10/16
1. 2.
3. 4.
Copyright © 2020 ANEST IWATA SPARMAX Co., Ltd. All rights reserved. Cliq_DV350_Manual (NUS_NCE_20201126)
3510387D

11/16
Cleaning
Maintenance & Repair
• Replace the filter if the fibrous material in the filter is wet or in the case of a collection container overflow. With the vacuum regulator fully closed
cand the tubing disconnected from the “Fluid Side” of the filter (filter open to atmosphere), a vacuum reading in excess of 50 mmHg on the gauge
indicates the filter should be replaced. The filter must be replaced in the event fluids have been in contact with the filter, such as in a collection
container overflow.
• Do not disassemble the unit; there are no user-serviceable components inside the device. If repair or servicing is required, please send unit to an
authorized service center. Disassembly or tampering of the device will damage the device and void the warranty.
Warranty Length
• DV-350 is warranted to be free from defective workmanship and materials for two years from the date of purchase. Within the warranty period, any
defective part(s) will be repaired or replaced at manufacturer’s option. Manufacturer is not liable, and warranty is voided, for any damage or
consequential damage resulting from incorrect operation of the unit, inappropriate usage of the unit, disassembly of the unit, or use by
unauthorized or untrained persons. This warranty does not cover the collection container, filter, intermediate tubing, or suction tubing.
• Rechargeable battery has a 90 day limited manufacturer's warranty.
WARNING
• Prior to cleaning, make sure machine is turned off and unplugged.
• To ensure proper hygiene, collection container, filter, suction tubing, and intermediate tubing must only be used by the same patient.
Do not use these accessories across multiple patients. Do not attempt to sterilize these accessories for use across multiple patients. Failure to abide
by these warnings may result in infection or cross contamination between patients.
• Any parts of the device or accessories that come into contact with bodily secretions must be cleaned after each suction operation. All single-use
accessories, such as the suction catheter, must be disposed of according to local regulations, institution policy (as applicable) and the instructions
of a certified healthcare professional.
• Wear disposable gloves during cleaning procedures.
Cleaning the surface of the unit
• Wipe machine clean with a damp cloth.
• Do not clean machine with powder-type cleaning agents or soap.
• Do not clean machine in water.
WARNING:If the inside of the unit comes into direct contact with liquids, the aspirator must be sent to an authorized service center for examination
and repair.
ATTENTION: Slight discoloration of plastic housing or the stickers on the housing may occur over time. This will not impact the performance or
function of the device.
Cleaning reusable accessories
• Clean the collection container after each use. Empty contents and wash all components using a mixture of 3 parts white vinegar 1 part warm tap
water.
• Clean intermediate and suction tubing after each use. Disconnect tubing from machine and clean by rinsing inside and outside of tubing with hot
tap water. Air dry.
• Clean the collection container coaster after each use. Rinse with hot tap water and air dry.
WARNING:Fluid collected in the collection container must be disposed of according to local regulations, institution policy (as applicable), and the
instructions of a certified healthcare professional.
CAUTION:Do not immerse filter in liquid. Doing so will damage the filter. If the fibrous material inside the filter is wet, replace immediately.
Copyright © 2020 ANEST IWATA SPARMAX Co., Ltd. All rights reserved. Cliq_DV350_Manual (NUS_NCE_20201126)
3510387D

12/16
If you run into problems with the product, refer to the possible causes below to perform simple troubleshooting. If the problem cannot be solved,
send the product to an authorized service center for examination and repair.
Troubleshooting
Problem Possible cause Remedy
Unit does not start
Battery troubleshooting
Reduced performance
No suction
Unit is not plugged into mains
properly.
The fibrous material inside the filter is
wet, stained, or damaged.
Mains not supplying power.
The fibrous material inside the filter is
wet, stained, or damaged.
Wrong voltage and current levels.
Check that the mains is live by trying another device.
Replace the filter.
Replace the filter.
Check if any tubing is blocked. If so, replace tubing.
Clean or replace suction tubing.
Check that the mains provides the voltage and current
levels required by this device.
Unit over heated.
Tubing is blocked.
Suction tubing is blocked.
Make sure that all tubing is securely connected and that there
are no leaks.
Check filter connection. Make sure fiter is tightly
connected to inlet.
Replace collection container.
Empty the collection container.
Tubing is leaking air.
Collection container is damaged.
Collection container overflowed.
Connection between filter and inlet is
loose.
Please contact qualified service personnel for repair.
Check plug connection. Make sure plug is properly
inserted into mains.
The fibrous material inside the filter is
wet, stained, or damaged.
Replace the filter.
Problem Possible causeIndicator light Remedy
Unit does not start
Reduced performance
Battery is empty.All lights OFF
Red light ON
All lights OFF
Charge the battery.
Charge the battery.Battery is almost empty.
Unit will not charge
Battery is broken. Replace the battery.
All lights OFF Battery is not properly connected. Make sure the battery is inserted into the
battery slot, and is tightly connected to the
connector.
All lights OFF Unit is not plugged into mains
properly.
Check plug connection. Make sure
plug is properly inserted into mains.
Copyright © 2020 ANEST IWATA SPARMAX Co., Ltd. All rights reserved. Cliq_DV350_Manual (NUS_NCE_20201126)
3510387D

13/16
Transportation, Storage, and Disposal
Transportation
• General transportation of the unit should correspond to the conditions outlined in the 'Product Introduction' section of this manual.
• In the event that the DV-350 Aspirator needs to be sent to an authorized service center for inspection and repair, do not package the collection
container, suction tubing, filter, intermediate tubing, collection container coaster, or any single-use accessories (such as a suction catheter) in the
shipment.
Storage
Storage of the DV-350 should correspond to the conditions outlined in the "Product Introduction" section of this manual.
Disposal
• When the product’s life cycle is at an end, all components of the unit must be disposed of in accordance with local regulations.
• Take care that the unit is clean and that the various materials are carefully separated for disposal.
• The European Union (EU) directives 2012/19/EU, known as the WEEE (Waste Electrical and Electronic Equipment) Directive, and 2011/65/EU, referred
to as the RoHS (Restrictions on Hazardous Substances) Directive prohibits the disposal of this device in the domestic waste.
• Outside the EU: Follow disposal regulations in the country of use.
Copyright © 2020 ANEST IWATA SPARMAX Co., Ltd. All rights reserved. Cliq_DV350_Manual (NUS_NCE_20201126)
3510387D

GUIDANCE AND MANUFACTURER’S DECLARATION
• All Medical Electrical Equipment requires special precautions regarding Electromagnetic Compatibility (EMC) and should only be installed and
put into service according to the EMC information provided in the accompanying section.
• Portable and Mobile RF Communications Equipment could affect Medical Electrical Equipment.
• The Medical Electrical Equipment should not be stacked with, or close to other equipment. If this is necessary, ensure that the equipment is
observed as having normal operation in the configuration of its intended use.
14/16
Recommended separation distances between portable and mobile RF communications equipment and the
ME equipment
The DV-350 suction pump is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or
the user of the DV-350 suction pump can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the DV-350 suction pump recommended below, according to the maximum output power
of the communications equipment.
150 kHz to 80 MHz
d=
1
5,3
V
P
80 MHz to 800 MHz
d=
1
5,3
E
P
800 MHz to 2.5 GHz
d=
1
7
E
P
Rated maximum output
power of transmitter
(W)
Separation distance according to frequency of transmitter
(m)
0.01
0.1
1
10
100
0.12
0.37
1.17
3.7
11.67
0.12
0.37
1.17
3.7
11.67
0.23
0.74
2.33
7.37
23.33
WARNING: Use of this device with electrical accessories such as extension cords or plug adapters may result in non-compliance and may negatively
affect EMC performance.
Copyright © 2020 ANEST IWATA SPARMAX Co., Ltd. All rights reserved. Cliq_DV350_Manual (NUS_NCE_20201126)
3510387D

15/16
Declaration – electromagnetic immunity
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Emissions test Compliance Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Electrostatic
discharge (ESD)
IEC 61000-4-2
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
3 V
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any part of
the EQUIPMENT or SYSTEM including cables, than
the recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter. Interference may occur in the
vicinity of equipment marked with the following
symbol.
The DV-350 suction pump is intended for use in the electromagnetic environment specified below. The customer or the user of the DV-350 suction
pump should assure that it is used in such an environment. NOTE: The DV-350 suction pump is contraindicated from placement and use in the
following environments: MR environments, CT environments, in proximty to diathermy therapy machines, near RFID transmitters and
electromagnetic security systems (such as metal detectors).
Declaration – electromagnetic emissions
The DV-350 suction pump is intended for use in the electromagnetic environment specified below. The customer or the user of the
DV-350 suction pump should assure that it is used in such an environment.
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
If the user of the EQUIPMENT or SYSTEM requires
continued operation during power mains
interruptions, it is recommended that the
EQUIPMENT or SYSTEM be powered from an
uninterruptible power supply or a battery.
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV for power supply lines
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
< 5 % UT
(> 95 % dip in UT) for 0.5 cycle
40 % UT
(60 % dip in UT) for 5 cycles
70 % UT
(30 % dip in UT) for 25 cycles
< 5 % UT
(> 95 % dip in UT) for 5 sec
< 5 % UT
(> 95 % dip in UT) for 0.5 cycle
40 % UT
(60 % dip in UT) for 5 cycles
70 % UT
(30 % dip in UT) for 25 cycles
< 5 % UT
(> 95 % dip in UT) for 5 sec
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.
CE emissions
CISPR11
Group 1
RE emissions
CISPR11
Class B
Class A
Complies
The DV-350 suction pump uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby electronic equipment.
The DV-350 suction pump is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
Copyright © 2020 ANEST IWATA SPARMAX Co., Ltd. All rights reserved. Cliq_DV350_Manual (NUS_NCE_20201126)
3510387D

Authorized Service Center Contact Infomation:
ANEST IWATA SAPRMAX Co., Ltd
No.121, Sec. 3, Zhongshan Rd.
Dacun, Changhua 51542 Taiwan, R. O. C.
www.cliqmedical.com
Tel: +886.2.2345.1868
4B-03 Taipei World Trade Center
4F, No. 5, Sec. 5, Hsinyi Rd.
Taipei, Taiwan 11011
Copyright © 2020 ANEST IWATA SPARMAX Co., Ltd. All rights reserved. Cliq_DV350_Manual (NUS_NCE_20201126)
3510387D
Table of contents
Popular Medical Equipment manuals by other brands

Polymer Technology Systems
Polymer Technology Systems CardioChek user guide

Labex
Labex INSPIRATION ELECTROLARYNX user manual

Spectranetics
Spectranetics Stellarex Instructions for use

Acare
Acare AH-M1 instruction manual

Hitachi
Hitachi UST-5417 instruction manual

Vancare
Vancare GoLift Assembly, installation and operation instructions