7. Spectrum Analyzer Resolution Bandwidth Accuracy Test
The following test can be used to verify the resolution bandwidth accuracy of the Spectrum Master at dif-
ferent frequencies.
a. Equipment Required:
·Anritsu MG3692A Synthesized Signal Source with options 2A, 4 and 15, or equivalent
·Anritsu 34RSN50 50 ohm adapter or equivalent
·Anritsu 15NNF50-1.5C RF Coaxial Cable or equivalent
·10 MHz Reference Standard
·BNC male to BNC male coaxial cable
b. Procedure:
1. Connect the 10 MHz reference source to the Anritsu MG3692A Synthesized Signal Source.
2. Connect the output of the Anritsu MG3692A Synthesized Signal Source to the Spectrum Master
RF Input.
3. On the Spectrum Master, press and hold the ESCAPE/CLEAR key, then press the On/Off key to
turn on the Spectrum Master. (This sets the instrument to the factory preset state.)
NOTE: Before continuing, allow a 30-minute warm up for the internal circuitry to stabilize.
4. Set the MG3692A output to 1 GHz, with an RF output level of –30 dBm.
5. Press the AMPLITUDE key and the Ref Level soft key.
6. Enter –27 and press the ENTER key to set the Reference Level to –27 dBm.
7. Press the Scale soft key and enter 3, then press ENTER.
8. Press the FREQ/SPAN key and the Center soft key.
9. Enter 1and press the GHz soft key to set the center frequency to 1 GHz.
NOTE: The Measured Occupied BW value for each RBW setting can be recorded in Table 1, page 8.
1 MHz RBW Test
10. Press the Span soft key, enter 1.5 and press the MHz soft key to set the span to 1.5 MHz.
11. Press the BW/SWEEP key and the Bandwidth soft key.
12. Press the RBW Manual soft key and use the Up/Down arrow key to select 1 MHz. Press ENTER to
set the resolution bandwidth to 1 MHz.
13. Press the VBW Manual soft key and use the Up/Down arrow key to select 3 kHz and press ENTER to
set the video bandwidth to 3 kHz and select BACK.
14. Press the MEAS key, the OBW soft key, the dBc soft key, and then the ENTER key.
15. Press the MEAS key, then the MARKER key. Press the M2 soft key and then the Delta (M2–M1) soft
key. The displayed D2 frequency should be ± 5% of the RBW.
300 kHz RBW Test
16. Press the FREQ/SPAN key.
17. Press the Span soft key, enter 450 and press the kHz soft key to set the span to 450 kHz.
18. Press the BW/SWEEP key and the Bandwidth soft key.
19. Press the RBW Manual soft key and use the Up/Down arrow key to select 300 kHz. Press ENTER to
set the resolution bandwidth to 300 kHz and select BACK.
20. Press the MEAS key and then the MARKER key.
21. Press the M2 soft key. The displayed D2 frequency should be ± 5% of the RBW.
6 Spectrum Master MM