6. Front Panel to Millimeter-Wave Module Connections
QSG-18 PN: 10410-00771 Rev. B VectorStar ME7838E4 QSG
Inverting a Module
If necessary, a module can be turned over in the bracket to change the elevation of the W1 connector. To turn
the module over:
1. Remove the six knurled-head thumbscrews from the module.
2. Turn the module over.
3. Install the cable assembly, then install the module into the bracket.
4. Install the six thumbscrews.
Millimeter-Wave Module Operating Environment
The following notes should be observed when operating the 3743E and 3744E-xx mm-Wave Modules:
•Thermal heat sinking similar to the supplied mounting brackets of the Millimeter-Wave Module should
be considered in custom mounting applications.
•Each 3743E Module consumes a maximum of 12 watts.
•Each 3744E-EE and 3744E-EW Module consumes a maximum of 12 watts.
•Each 3744E-Rx Module consumes a maximum of 7 watts.
•The primary heat sink paths for the module is the two side mounting surfaces of the mounting bracket.
•With the attached cable mounting brackets, the case temperature rise is approximately 15 °C to 20 °C
above ambient.
•A three (3) year warranty is valid for all versions of 3744E and 3743E modules, with or without
mounting brackets attached.
3
75685-1a
or
75685-3b,c
TEST – SSMC Connector (3743E, 3744E-EE, and 3744E-EW modules)
• Tighten using a 4 mm (5/32 in) torque end wrench set to less than 0.17 N·m (1.5 lbf·in).
• Recommended is Anritsu 01-529-Rtorque wrench.
4
REF –SSMC Connector (3743E, 3744E-EE, and 3744E-EW modules)
• Tighten using a 4 mm (5/32 in) torque end wrench set to less than 0.17 N·m (1.5 lbf·in).
• Recommended is Anritsu 01-529-R torque wrench.
5Power/Signal Bi-Lobe™ Connector (3743E, 3744E-EE, 3744E-EW, 3744E-Rx)
6
LO – K Connector (3743E, 3744E-EE, and 3744E-EW modules)
• Tighten using an 8 mm (5/16 in) torque end wrench set to 0.9 N·m (8 lbf·in).
• Recommended is Anritsu 01-201.
7 Module Power and I/O Cable Restraint
8 806-209-Rd
SRC – V Connector (3743E module only)
• Tighten using an 8 mm (5/16 in) torque end wrench set to 0.9 N·m (8 lbf·in).
• Recommended is Anritsu 01-201.
9 75685-1a
RF – V Connector (3743E, 3744E-EE, and 3744E-EW modules)
• Tighten using an 8 mm (5/16 in) torque end wrench set to 0.9 N·m (8 lbf·in).
• Recommended is Anritsu 01-201.
10 N/A Factory Calibrated Port Assignment Label
11 N/A Module Serial Number Label
a. 3743E, 3744E-EE, and 3744E-EW modules use cable assembly 75685-1.
b. The REF cable is not used in the 75685-3 cable assembly.
c. 3744E-Rx module uses cable assembly 75685-3.
d. The 806-209-R Coaxial Cable is used only with the 3743E modules.
Table 5. Millimeter-Wave Module Connections (2 of 2)
Index Cable P/N Description