13. Press the RBW soft key and use the
Up/down arrow key to select 10 kHz.
Press ENTER to set the resolution band-
width to 10 kHz.
14. Press the VBW soft key and use the
Up/down arrow key to select 3 kHz. Press
ENTER to set the video bandwidth to 3
kHz.
15. Press the MARKER key, then the M1 soft
key.
16. Select the EDIT soft key and use the
Up/down arrow key to center the marker
on the waveform. Verify that the marker
frequency is 1000 MHz, ±2 kHz.
5. PHASE NOISE VERIFICATION
a. Equipment Required:
·Anritsu 68047C Synthesized Signal
Source, with options 11 and 15A
b. Procedure:
1. Connect the output of the source to the
MS2711 RF Input.
2. Connect the external power supply
(Anritsu part number 40-115) to the
MS2711.
3. Press and hold the ESCAPE/CLEAR key,
then press the ON/OFF key to turn on the
Spectrum Analyzer. (This sets the instru-
ment to the factory preset state.)
NOTE
Before continuing, allow a 30-min-
ute warm up for the internal cir-
cuitry to stabilize.
4. Set the 68047C output to 1000 MHz,
with an RF output level of –30 dBm.
5. Press the BW/SWEEP key.
6. Press the RBW soft key and use the
Up/down arrow key to select 10 kHz.
Press ENTER to set the resolution band-
width to 10 kHz.
7. Press the VBW soft key and use the
Up/down arrow key to select 3 kHz. Press
ENTER to set the video bandwidth to 3
kHz.
8. On the MS2711, press the FREQ/SPAN
key and the CENTER soft key.
9. Enter 1000 and press the ENTER key to
set the center frequency to 1000 MHz.
10. Press the SPAN soft key and enter 0.100.
Press the ENTER key to set the span to
0.100 MHz.
11. Press the AMPLITUDE key.
12. Press the REF LEVEL soft key and enter
–27. Press ENTER to set the reference
level to –27 dBm.
13. Press the MARKER key, then the M1 soft
key.
14. Press EDIT and enter 1000. Press ENTER
to set the M1 marker frequency to 1000
MHz.
15. Press the BACK soft key and the M2 soft
key.
16. Press EDIT and enter 1000.030. Press
ENTER to set the M2 marker frequency to
1000.030 MHz (30 kHz higher than the
center frequency).
17. Press the DELTA (M2-M1) soft key.
18. Press the RUN/HOLD key and read and
record the amplitude of the signal at the
M1 30 kHz offset.
19. Press the RUN/HOLD key to read and re-
cord five values, then calculate the aver-
age of the five recorded values.
20. Add –40 dB to the average value and ver-
ify that the result is £–74 dBc/Hz.
(For example: –35 dBc measured +
(-40 dB) = –75 dBc/Hz.)
6. MEASUREMENT ACCURACY
Measurement accuracy involves testing the MS2711
over three frequencies at four power levels. Table 1
provides a guide to the frequencies, power levels and
reference levels required for each measurement, and
can be used to record the readings.
a. Equipment Required:
·Anritsu 68047C Synthesized Signal
Source, with options 11 and 15A
MS2711 HHSA MM 3