3 | A n y l o a d D D - K C 1 M a n u a l V ( 1 6 0 3 )
4.2.2 Load cell output sensitivity:
1~2m V/V
4.3 Weighing
When power-on, with nothing on the platform, press button once, the instrument
will then display “0 ”. Place an object on the platform. After the “kg”, “lb” or “pcs”
character stops flashing, which means that the data is stable, you can begin to read it.
4.4 Animal Weighing/Hold (On by default)
When the scale is on, press and hold key until the display "0FF" or "ON", then release.
This will turn off/on animal weighing/hold. When the ‘running person’ symbol above the
units symbol flashes, the current mode is animal weighing/hold. In this state, once the
weighing is stable, it will automatically hold for about 10 seconds.
4.5 Calibration Methods
4.5.1 Random Single-Point Calibration
a) At weighing mode, press and hold button until the window displays“------”;
b) Press or buttonuntil display shows “CAL” ;
c) Press button to enter the calibration mode;
d) After entering the calibration mode, the window will display“Load”and“0.0kg”
alternately,which means that the weight“0.0kg”should be loaded on the platform.
e) After making sure that there is nothing on the platform, press button to
confirm so that the scale begins“0.0kg”calibration. In the process of
“0.0kg”calibration, “0.0kg” will flicker rapidly on the display window;keep the
platform calm, stable, and without other interference factors which will influence the
weighing performance. If the weighing remains stable for longer than 2s, the scale will
finish “0.0kg”calibration automatically, it will then begin the next calibration point;
otherwise, “0.0kg” will keep flickering to wait for the scale to the stabilize.
f) Suppose that the next calibration point should be“xx.xkg”,the window will display
“Load”and“xx.xkg” alternately,which means that the weight“xx.xkg”(WEIGHT)