Apel BR-5000N User manual

______________________________________________________________________________________
OPERATION MANUAL
BILIRUBIN METER
BR – 5000N
IMPORTANT!
READ THIS OPERATION MANUAL!
*Prior to operation, please read this manual
carefully and retain it for future reference
Failure to read, understand and follow the
instructions in this manual may result in
damage to the unit, injury to operating
personnel and/or poor equipment
performance
All rights are reserved This material may not be reproduced or copied, in whole or part, without the written permission of APEL CO , LTD
BR5N-1-OMA-103a

BR5N-1-OMA-103a 1
TABLE OF CONTENTS
Section 1 - Introduction
1.1 Introduction --------------------------------------------------- 2
1.2 Features ------------------------------------------------------ 2
1.3 Important Safet Instruction ------------------------------ 4
Section 2 – Contents of the unit
2.1 Description of the unit -------------------------------------- 6
2.2 Control panel Ke s, Displa and Indicator -------------- 7
Section 3 – Operation procedure
3.1 Preparation ---------------------------------------------------- 8
3.2 Taking blood sample ----------------------------------------- 8
3.3 Measurement of the specimen ------------------------------ 10
3.4 Setting the name or ID number of the test ------------------ 12
3.5 Other functions ----------------------------------------------- 12
3.6 How to use Serum gauge ------------------------------------ 15
Section – Maintenance and inspection
4.1 Adjustment with standard solution ------------------------ 16
4.2 Indication of alarms -------------------------------------------- 17
4.3 Inspection abnormalities ------------------------------------- 17
4.4 Light bulb replacement -------------------------------------- 18
Section 5 – Specifications
5.1 Specifications -------------------------------------------------- 19

BR5N-1-OMA-103a 2
Section 1 - Introduction
1.1 Introduction
Congratulations on purchasing an APEL product. It’s an important and ver smart
decision that’s sure to reward ou for man ears to come. To ensure that ou take
full advantage of our Bilirubin meter, BR-5000N’s offered features and warrant
benefits, be sure to carefull read and full comprehend this Operation Manual.
It is designed to acquaint ou with the features and help ou effectivel operate the
new instrument in safet . We sincerel hope ou will enjo man benefits of being
associated with our product which represents both qualit and customer satisfaction.
We are proud of the advanced engineering and qualit construction of each instrument
we build. If ou have an questions, please contact our supplier and refer to the
model number and seven-digit serial number located at the back of our unit, when
inquiring.
The BR-5000N Bilirubin meter is a digital, direct reading measuring equipment
developed with the purpose of enabling the measurement of total bilirubin in serum
of neonate while compensating for hemol sis, turbidit , and other components of
obstruction. With the BR-5000N, an alarm is displa ed automaticall whenever
there is a high-densit specimen or a high level of obstruction that falls outside the
tolerance.
1.2 Features
1. Handles hemol sis and turbidit easil
Careful selection of the dual wavelength color filters and the combination of
electronic technolog minimize the interference of hemol sis and turbidit .
2. Eas operation
Microprocessor enables the Auto Calibration and eas operation. And after the
equipment is calibrated with a standard solution, ou can measure specimen
without using the standard solution after that.
3. Eas set of the sample tube
A serum gauge is provided on the surface of BR-5000N, which indicates the
proper sample volume to be inspected. Onl simple action is required to set the
capillar tube into the holder.
4. Alarm lamp informs user of abnormalities
Alarm lamp signals either abnormal value or situations such as burned out of
the bulb, bilirubin concentration of 30mg/ ㎗or more, or high hemol sis of
250mg/㎗ or more.

BR5N-1-OMA-103a 3
5. The print out b the printer
The measurement data can be printed out b the option printer.
(The measurement data are memorized up to 100 samples in maximum while
the power switch in being turned on.)
The printer connector is based on Centronics standard.
6. The equipment can be connected to computer b serial interface.
The measurement data can be sent to computer through RS-232C interface.
7. mg/㎗- and μmol/ℓ-indication
One of mg/㎗- and μmol/ℓ-indication can be selected b the control panel.
8. Eas to operate b on touch ke .
Normal measurement can be performed b touching start button after inserting
the capillar tube to the capillar tube holder.
9. Compact design
BR-5000N has a compact design that does not occup much space.
10. Flexible power source
Power requirement is flexible, from 100 to 240VAC.

BR5N-1-OMA-103a 4
1.3 Important Safety Instruction
When utilizing our BR-5000N, alwa s take basic safet precautions and warnings, in
order to avoid damage to our unit, propert and/or personal injur , which ma result
from misusage, exposure to fire or electric shock. Important warnings and safet
precautions are as follows:
1. To ensure accurac of the measurement, alwa s warm-up the instrument for at
least 15 minutes, prior to operation.
2. Never cover the unit when it is turned-on. Doing so will prevent the electrical
components from being cooled properl and cause damage.
3. Unplug the BR-5000N from the wall outlet prior to cleaning. Do not use organic
liquid solutions such as alcohol or aerosol cleaners. Use a soft dr cloth for
cleaning.
4. Do not directl expose our unit to water or liquid. Avoid spills, which could
damage electronic components.
5. Alwa s use appropriate capillar tube in specification, when obtaining bilirubin’s
data or using standard solution.
6. Place this product on stable cart, stand, or table.
7. If stored in non-room temperature setting, allow the unit to reach room
temperature before turning on. Do not place this instrument in an environment
where there is severe temperature variation.
8. Place the unit awa from dust, excessive moisture, or corrosive chemicals.
Cover the instrument when not in use to protect electrical components from dust.
9. Do not dismantle or modif the instrument. Damage or performance problems
resulting from disassembl or alteration of the unit, will not be covered under the
warrant . Contact our supplier when maintenance, service or repair work are
required.
10. When instrument is not in use, make sure the power switch has been turned-off
and unplug the power cord from the electrical outlet. Do not pull the cord, but
hold the receptacle of the plug and pull it.
11. Onl use light bulb, capillar tubes and power cord that come with this instrument.
Such replacement parts are available through our supplier.
12. Do not overload wall outlets with extension cords as this can result in the risk of
fire or electric shock. Attaching multiple loads to the same wall outlet is
dangerous and ma cause improper operation of BR-5000N.

BR5N-1-OMA-103a 5
13. Do not allow an thing to rest on the power cord. Do not locate this product
where the power cord ma trip an person or object.
14. To keep the accurac of this instrument, calibration should be done periodicall
and after a period of long storage prior to operation.
15. Make sure that the GND code is grounded properl .
16. Unplug this product from the outlet and refer to our supplier under the following
conditions:
* When the power cord or plug is damaged and/or fra ed.
* If the unit has been directl exposed to water or liquid.
* If the instrument displa s unusual change in performance.
* If the instrument has been abused or dropped.
* If the unit does not function normall b following the operating instructions.
(Adjust onl those controls that are presented and mentioned in the
instruction.)
*IT IS OWNER’S RESPONSIBILITY TO MAKE SURE THAT THE SPECIFIED
SAFETY AND OPERATING INSTRUCTIONS ARE FOLLOWED.
SAVE THESE INSTRUCTIONS FOR FUTURE REFERENCE.

BR5N-1-OMA-103a 6
Section 2 - Contents of the unit
2.1 Description of the unit
Unpacking BR-5000N:
Open the carton from the top and carefull unpack the BR-5000N. Do not discard
the packing materials. Save all the packing material in the carton, in case the unit
ma need to be repacked and returned to our supplier for maintenance, repair, or
service work. In the event of missing parts, defects, damages or other problems,
immediatel contact our supplier. Make sure that our carton contains the
following items:
Main Unit Quantit
BR-5000N ------------------------------------------ 1
Standard Accessories
AC power cord --------------------------------- 1
Capillar tube (heparinized, red line) ------- 1 (pk of 100)
Capillar tube (plain, blue line) -------------- 1 (pk of 100)
Putt (tube sealant) ---------------------------- 1
Bilirubin standard ----------------------------- 1
Fuse (1A) --------------------------------------- 2
Vin l dust cover ------------------------------- 1
Operation Manual ----------------------------- 1
*Note: All standard accessories are replaceable through our supplier at
an additional cost.

BR5N-1-OMA-103a 7
2.2 Control panel Keys, Display and Indicator
Fig.1
11
1
1. LCD Displa 7. Power switch
2. Pilot lamp 8. Fuse holder
3. Control panel 9. AC Power socket
4. Serum gauge 10. Printer Interface Connector(D-sub25p,female)
5. Capillar tube holder 11. RS-232C Interface Connector(D-sub9p,male)
6. Serial number label
123
4
56
789 10 11
FUSE FUSE
1
4 5 6
78
9
DEL
START
SET
CAL
PRINT
No
A~Z
23 0
ZERO
μmol/dl
mg/dl
EXIT
( ) ( )
Control Panel

BR5N-1-OMA-103a 8
Section 3 – Operation Procedure
3.1 Preparation
1. Make sure that the power cord is plugged to the electric wall outlet and turn on
the Power switch. Then, ou can see the pilot lamp is lit and
the number of seconds are counted down on the LCD displa .
2. LCD Displa will show “Read ” as the right figure 15 minutes
later, and Control panel can be used.
* Press “Start” ke , if ou want to quit the count down.
3. Measurement of BLANK liquid
(1) Fill distilled water into a capillar tube (blue line) and seal one end with putt
and get rid of the bubbles.
(2) After wiping off the outside of the tube well, set it in the capillar tube holder.
( Fig. 1 - ⑤)
(3) Press the “ZERO” ke (for more than 2 seconds)
and confirm that the displa reads 00.0.
(4) ”Measure End” shows the end of the procedure.
If the LCD shows “Measure Error”. Press “ZERO”
again after getting rid of the bubbles and/or stains
of the capillar tubes.
4. When the printer is used, confirm the printer cable is connected properl and turn
on the switch of the printer. And turn on the “Printer Use” of “SET-4”.(Refer to the
section 3.5.2)
3.2 Taking blood sample
1. When taking blood sample from neonates, warm the tip of the big toe, the heel
of the foot, or wherever blood is to be drawn to get the blood flowing well.
*Taking the blood sample within 2 hours after suckling or taking meal ma
result high turbidit .
2. Disinfect the area that has been rubbed or warmed with alcohol then puncture
using a blood-taking needle. (Fig. 2-1)
3. After puncturing, wipe awa the first drop of blood, then take the following
droplets into a capillar tube (red line) that is provided with the BR-5000N.
(Fig. 2-2)
4. Seal one end of the capillar tube with putt when half or more of the capillar
tube has been filled with blood. (Fig. 2-3)
5. Put the capillar tube that has been prepared using steps 1 - 4 above, into a
high-speed hematocrit centrifuge, facing the end of the capillar tube that is
not sealed with putt toward the center of the centrifuge, as shown in Figure 3.
6. After the capillar tube is set in the manner described above, rotate 12,000
rpm over 3 minutes.
7. After the treatment is concluded, the blood in the capillar tube has been
Warming Up 900
Warming Up 900 Warming Up 900
Warming Up 900
TEST Ready
TEST Ready TEST Ready
TEST Ready
[Auto ZERO]
[Auto ZERO] [Auto ZERO]
[Auto ZERO]
Measure Start
Measure Start Measure Start
Measure Start
[Auto ZERO]
[Auto ZERO] [Auto ZERO]
[Auto ZERO]
Measure End
Measure End Measure End
Measure End

BR5N-1-OMA-103a 9
centrifugall separated into serum and blood cells, as shown in Figure 4.
(Make sure that there is at least 15mm of serum part. A serum gauge is
provided on the surface of BR-5000N)
Fig. 2-1 Fig. 2-2 Fig. 2-3
Fig. 3
Fig. 4-1 Fig. 4-2
遠心器中心
パテ
毛細管
パテ
血球部分
血清部分
パテ
血球 血清
スリット
毛細管及びホルダー断面図
Top of the big toe, or heel of foot
Capillary tube
Capillary tube
Blood
Blood
Capillary
tube
Putty
Putt
Capillar
tube
Center
Serum
Serum
Blood
cell
Putt
Slit
Blood cell
Putt
Cross section of Capillar tube
and the holder

BR5N-1-OMA-103a 10
3.3
Measurement of the specimen
3.3.1 Normal measurement [ No. 0 ]
Normal measurement is used onl for temporar measurement. The data is not
stored.
1. Take the capillar tube containing distilled water out of the capillar tube holder,
insert the specimen tube that has been through centrifugal treatment. The serum
part must fall within the same height as the part of the slit of cell holder.(Fig.4-2)
*Make use of the serum gauge on the suface of the instrument to inset the capillar
tube correctl . Refer to section 3.6.)
2. Make sure the cell holder is inserted completel to
the equipment.
3. Select No. 0 and press “START” ke .
4. The end of the measurement.
* When an errors occur, the buzzer sounds 3 times
and the displa value is turned on and off.
(1) Abnormalit of the concentration of Bilirubin: [ *BIL.high
*BIL.high *BIL.high
*BIL.high ]
(2) Abnormalit of the concentration of
Hemoglobin and/or turbidit : [ *Hem.high
*Hem.high *Hem.high
*Hem.high ]
5. Displa Unit - mg/㎗ or μmol/ℓcan be changed b the ke s
”mg/㎗” or “μmol/ℓ” respectivel .
3.3.2 Measurement with ID_No. [ No. 1 ~
~~
~ 25 ]
In this measurement, ID_No. or name of individual bab s can be stored and can be
controled with them. Measured data are stored up to 100 until the power switch is
turned off.
For example, if the samples of 10 bab s are measured, 10 data per bab can be
stored. Or, 5 data per bab for 5bab s and 15 data per bab for 5 bab s can be
stored.
* Once the power switch is turned off, whole data are cleared.
1. Press “No.” ke , and
set the ID Number. (1 ~ 25)
2. Take the capillar tube containing distilled water
out of the capillar tube holder, insert the specimen tube
that has been through centrifugal treatment. (Fig. 4-2)
3. Press “START” ke
4. The end of the measurement.
When an errors occur, the buzzer sounds 3 times and
the displa value is turned on and off.
(3) Abnormalit of the concentration of Bilirubin: [ *BIL.high
*BIL.high *BIL.high
*BIL.high ]
(4) Abnormalit of the concentration of
Hemoglobin and/or turbidit : [ *Hem.high
*Hem.high *Hem.high
*Hem.high ]
N
NN
No. 0 TEST
o. 0 TEST o. 0 TEST
o. 0 TEST
Ready
Ready Ready
Ready
TEST 16.5mg/dl
TEST 16.5mg/dlTEST 16.5mg/dl
TEST 16.5mg/dl
End
End End
End
No.12 ID
No.12 IDNo.12 ID
No.12 ID-
--
-
K012
K012 K012
K012
SN.001 Ready
SN.001 Ready SN.001 Ready
SN.001 Ready
No.12
No.12 No.12
No.12
SN.001 Start
SN.001 Start SN.001 Start
SN.001 Start
No.12 13.5mg/dlS
No.12 13.5mg/dlSNo.12 13.5mg/dlS
No.12 13.5mg/dlS
N.001 End
N.001 End N.001 End
N.001 End

BR5N-1-OMA-103a 11
5. The sample No.( SN.XXX
SN.XXX SN.XXX
SN.XXX) shown on the LCD displa is automaticall
incremented(+1) for the next measurement.
6. When the total measurement numbers reach 100, and one more measurement is
tried to be executed, the alarm buzzer sounds and
the LCD displa shows “
* Memory Full *
* Memory Full ** Memory Full *
* Memory Full *
”
as shown right figure.
Press “SET” or “PRINT” to print out the whole data stored in the instrument.
If ou want to skip the print, Press “EXIT”.
After the print or when the EXIT is pressed, LCD displa
shows “Clear all data?” Press “SET” for clearning all data,
or press “EXIT” to skip the procedure.
* An measurement can not be done unless stored cata are not cleared, when
*Memory Full*
is shown.
3.3.3 Print-out of measured data
The printer can be connected to BR-5000N and the measured data can be
printed out.
Connect the instrument and the printer with the printer cable, and turn on the
switch of the printer.
Turn on “PRINTER USE” of “SET 4” which is described in the section 3.5.2
“The setting of the printer”.
1. Auto print at each measurement.
Turn on the “Auto printer (SET 5)”. (Refer to 3.5.3)
2. Manual operation of printing right after the measurement.
Press “PRINT” after a data is measured.
3. Printing of whole data with designated ID_No.
Set the ID_No. and Press “EXIT” ke then press
“PRINT” ke .
4. Printing of the whole data.
When the memor area is full stored, the whole stored data are printed.
Refer to 3.3.2 “Memor Full”
3.3. Data communication function
Displa and collection of data on the Computer can be done through
the RS-232C cable from the instrument.
Turn on “COMM(RS-232C), SET 6” as described in the section 3.5.4,
if the communication is used.
Adjust the baud rate and the protocol to the computer.
Data format for output.
2
22
2
1
11
1,
, ,
, S
SS
S
A
AA
A
M
MM
M
P
PP
P
–
0
00
0
0
00
0
5
55
5,
, ,
, 0
00
0
0
00
0
1
11
1
,
, ,
, 2
22
2
1
11
1
.
..
.
3
33
3 0D 0A
ID_No. ID_Code, Name Sample No. Data CR LF
* Memory Full
* Memory Full * Memory Full
* Memory Full
*
* *
*
Print all data ?
Print all data ?Print all data ?
Print all data ?
* Memory Full *
* Memory Full * * Memory Full *
* Memory Full *
Clear all data ?
Clear all data ?Clear all data ?
Clear all data ?
No.12
No.12No.12
No.12
[Ready]
[Ready] [Ready]
[Ready]
[Print] No.12
[Print] No.12 [Print] No.12
[Print] No.12

BR5N-1-OMA-103a 12
3.
Setting the ID number
1. Press “No” ke , and set the ID number (1 ~ 25)
2. Press “SET” ke .
Then, ID No. can be set.
3. Set ID Number with the following ke s.
[A
AA
A-
--
-Z
ZZ
Z] : Alphabet/Numeral Change ke
(The displa shows “NUM” or “ALP” according to the condition.)
[▲],[▼] : Select ke for alphabets and numerals.
[
],[
] :
Move cursor at inputting alphabets and numerals.
[
0
00
0
]
~
[
9
99
9
] : Inputting numerals.
[
],[
➪
] :
Move cursor.
[DEL
DELDEL
DEL] :
Delete the letter on the cursor
[SET
SETSET
SET] :
Fix and memorize the characters shown on the LCD displa .
[EXIT
EXITEXIT
EXIT]
:
Quit the storage and end the procedure.
Up to 8 letters can be used for the ID numbers.
The basic parameters are memorized even after the power switch is turned off.
3.5
Other functions
Following parameters can be set for BR-5000N.
Confirm the present state of parameter before measurement.
SET 0 : Standard CONC 10.0 ~ 30.0 mg/㎗
SET 4 : Printer Use 1:on, 2:off
SET 5 : Auto-Print 1:on, 2:off
SET 6 : Comm(RS-232C) 1:on, 2:off
SET 7 : Baud rate 1:9600, 2:19200, 3:34800 bps
SET 8 : RS-232C mode d:Data bit 7/8, s:Stop bit 1/2
p:Parit bit odd/even/none
3.5.1 The setting of standard solution
1. Press “SET”, and then “0” ke .
The standard concentration value at present can be seen.
2. Set CONC value with the following ke s.
『0』~『9』 : Input numerals.
Decimal point is fixed, and the effective range is
from 9.0 to 30.0.
[DEL] ke clears the displa ed numbers to “0.0”.
3. Press “SET” ke .
The datum is fixed and memorized.
Displa comes back to “Read ” without memorizing the data when “EXIT”
ke is pressed.
No. 5 ID
No. 5 IDNo. 5 ID
No. 5 ID-
--
-
E005
E005 E005
E005
SN.001 Ready
SN.001 Ready SN.001 Ready
SN.001 Ready
No. 5 [A
No. 5 [ANo. 5 [A
No. 5 [A-
--
-
Z]
Z]Z]
Z]
3F
3F 3F
3F
-
--
-
12
1212
12
-
--
-
N
NN
N
5
55
5
No. 5 [A
No. 5 [ANo. 5 [A
No. 5 [A-
--
-
Z]
Z]Z]
Z]
_ID
_ID _ID
_ID
-
--
-
E005
E005 E005
E005
Set
Set Set
Set-
--
-
0
0 0
0
STD
STD STD
STD-
--
-
CONC 20.0
CONC 20.0CONC 20.0
CONC 20.0
1 9 .
1 9 . 1 9 .
1 9 .
8
88
8

BR5N-1-OMA-103a 13
3.5.2
The setting of Printer
Choose whether Printer Interface is active or not.
. 1. Press “SET”, and then “4” ke .
2. Choose the condition with [1]/[2] ke s.
On : Printer effective
Off : Printer ineffective
3. Press “SET” ke .
The condition is fixed and memorized.
Displa comes back to “Read ” without memorizing the data.
when “EXIT” ke is pressed.
3.5.3 The setting of Auto-Print
When “Auto-Print is ON, the measured data are printed automaticall on the printer
after each measurement. When it is OFF, data can be printed with the “PRINT” ke .
1. Press “SET”, and then “5” ke .
2. Choose the condition with [1]/[2] ke s.
On : serial output effective
Off: serial output ineffective
3. Press “SET” ke .
The condition is fixed and memorized.
Displa comes back to “Read ” without memorizing the data
when “EXIT” ke is pressed.
* Even when the “Auto print” is off, the measured data can be printed out after the
measurement with [PRINT] ke .
3.5. The setting of RS-
RS-RS-
RS-232
232232
232C
CC
C output
Choose whether RS-232C Interface is active or not.
1. Press “SET”, and then “6” ke .
2. Choose the condition with [1]/[2] ke s.
on : Serial output effective
off : Serial output ineffective
3. Press “SET” ke .
The condition is fixed and memorized.
Displa comes back to “Read ” without memorizing the data
when “EXIT” ke is pressed.
Printer Use
Printer Use Printer Use
Printer Use
1
11
1
:on 2:off
:on 2:off :on 2:off
:on 2:off
Set
Set Set
Set-
--
-4
44
4
Set
Set Set
Set-
--
-5
55
5
Auto Print
Auto Print Auto Print
Auto Print
1:on
1:on 1:on
1:on
2
22
2
:off
:off :off
:off
Set
Set Set
Set-
--
-6
66
6
Comm (RS
Comm (RS Comm (RS
Comm (RS-
--
-
232C)
232C) 232C)
232C)
1
11
1
:on 2:off
:on 2:off :on 2:off
:on 2:off

BR5N-1-OMA-103a 14
3.5.5 The setting of the Baud rate for the RS-232C
Baud rate of RS-232C output can be set.
1. Press “SET”. and then “7” ke .
2. Choose the condition with [1]/[2]/[3] ke s.
1
11
1:9600bps 2
22
2:19200bps 3
33
3:38400bps
3. Press “SET ke .
The condition is fixed and memorized.
Displa comes back to “Read ” without memorizing the data
when “EXIT” ke is pressed.
3.5.6 The setting of the other RS-232C protocols
Protocols for RS-232C output can be set.
1. Press “SET”, and then “8” ke .
2. Choose the condition with [
],[
] ke s.
d
dd
d:Data_bit s
ss
s:Stop_bit p
pp
p:Parit
3. Press “SET”, and then “8” ke .
Data bit : [
7
](7bit) / [
8
](8bit)
Stop bit : [
1
](1bit) / [
2
](2bit)
Parit : [
1
](ODD) / [
2
](EVEN) / [
](NONE)
4. Press “SET ke .
The condition is fixed and memorized.
Displa comes back to “Read ” without memorizing the data
when “EXIT” ke is pressed.
3.5.7 Deletion of data
You can delete the stored data for the designated ID_No..
And “SN._” comes back to “1”.
1. Press “No.” ke .
e.g. in case of ID_No. 5, press [NO] then, [5].
2. Press “DEL” ke .
3. Press “SET ke .
The condition is fixed and memorized.
Displa comes back to “Read ” without memorizing the data
when “EXIT” ke is pressed.
No.5 ID
No.5 IDNo.5 ID
No.5 ID-
--
-
A105
A105 A105
A105
SN.20 Ready
SN.20 Ready SN.20 Ready
SN.20 Ready
No.5 ID
No.5 IDNo.5 ID
No.5 ID-
--
-
A105
A105 A105
A105
Delete data ?
Delete data ? Delete data ?
Delete data ?
Baud rate
Baud rate Baud rate
Baud rate
1
11
1
:9600bps
:9600bps :9600bps
:9600bps
Set
Set Set
Set-
--
-7
77
7
Set
Set Set
Set-
--
-8
88
8
RS
RS RS
RS-
--
-
232C mode
232C mode 232C mode
232C mode
d:
d: d:
d:
8
88
8
s:1 p:none
s:1 p:none s:1 p:none
s:1 p:none

BR5N-1-OMA-103a 15
3.6
How to use the Serum gauge.
Capillar tube containing the sample can be inserted correctl without pulling
out the capillar tube holder, if the serum gauge is used as follows.
1. Put the capillar tube treated
centrifugall on the serum gauge.
The border line between blood corpuscles
and serum must be left of the line “A”
as shown in the figure.
2. Hold the capillar tube with the thumb
touching the right side of the gauge.
3. Lift up the capillar tube as it is and
insert it into the capillar holder
until the thumb touchs the top of
the capillar holder
4. Then the slit of the capillar tube holder
must be within the range of the serum
part of the capillar tube.
.
5. The capillar tube can be inserted to
the inside of the instrument because the
holder has no stopper.
In case the capillar tube is broken,
the fragment of it can drop inside without
remaining inside of the capillar holder.
A
Capillar tube
Serum gauge
Capillar tube
holder
Front view of
the instrument

BR5N-1-OMA-103a 16
Section – Maintenance and inspection
.1
Adjustment with standard solution
The standard value that is set in the equipment will fluctuate corresponding to
deterioration of the light bulb, so to keep the accurac of the instrument,
confirm the value ever 1 month (assuming 2 hours of usage/da ) or
after the light bulb is replaced or the instrument is not used for a long interval.
*Standard solution to be used for the adjustment must be diluted precisel .
Otherwise, measured value could have error.
.1.1 The procedure of adjustment
1. Turn on the power switch and wait until the LCD displa shows “Read ”.
2. Adjust “ZERO” according to the procedure of 3.1, 3-6 .
3. Prepare standard solution with "Hi Level Check-BIL", manufactured b
International Reagents Corp., Japan or an substitute standard solution.
[Prepare original standard solution]
Remove cap and stopper of Hi Level Check-BIL. Reconstitute each vial with 3mℓ
distilled or deionized water of low bacterial count. Replace the stopper. Leave it as it
is for 5 minutes, then mix well b swirling. A high concentration standard is
obtained.
[To use with BR-5000N]
Empt the standard solution into a larger container. Add 6mℓof distilled or
deionized water and dissolve the standard solution again. Assa value × 1/3
concentration standard is obtained. This is the specified value to be used with
BR-5000N.
* The assa value of Hi Level Check-BIL is specified per each production Lot.
Confirm the value in note enclosed with Hi Level Check-BIL, e.g. assa value for
Lot No. 7324 is 54.0mg/㎗. Specified value is 54.0 ×1/3 = 18.0mg/㎗.
4. Set the standard vbalue of the standard solution according to the section 3.5.1.
5. Fill standard solution into a capillar tube (blue line)
and set it in the capillar holder, then press the “CAL” ke
not less than 2 second.
6. The adjustment will end if the displa shows “Measure End”.
If “Error BBK = x.xxx” is shown, confirm the standard
solution and the procedure of calibration, and calibrate it again.
* Measure the standard solution as the normal sample,
then confirm if the measured value falls within ±0.5 of
the calibrated value.
[Auto CAL]
[Auto CAL] [Auto CAL]
[Auto CAL]
Measure Start
Measure Start Measure Start
Measure Start
[Auto CAL]
[Auto CAL] [Auto CAL]
[Auto CAL]
Measure End
Measure End Measure End
Measure End
[Auto CAL] 20.0
[Auto CAL] 20.0[Auto CAL] 20.0
[Auto CAL] 20.0
Error BBK=0.000
Error BBK=0.000Error BBK=0.000
Error BBK=0.000

BR5N-1-OMA-103a 17
.2
Indication of alarms
When an alarm is given as the buzzer sounds 3 times and displa value turned on and
off, it is because of one or more of following reasons.
1. The high concentration of the specimen.
(Bilirubin that exceeds a measurement value of 30mg/㎗.)
2. High concentration of hemol sis or turbidit .
(Specimen with hemol sis or turbidit exceeding 250mg/㎗.)
* With case 1. and 2. above, accuratel dilute the specimen with an equal amount of
distilled water, then measure. Then, double the measurement value obtained.
3. The outside of the capillar tube being dirt , or bubbles exist inside.
4. The placement of the serum part being incorrect.
5. The light bulb being deteriorated, or light bulb cables being cut.
6. The placement of the light bulb being incorrect.
.3 Inspection abnormalities
S mptoms Possible cause Corrective action
1)The unit is not plugged in.
Plug in the power cord into
electrical wall outlet and the unit
securel .
LCD displa and/or LED does
not glow when the power
switch is turned on.
2)DC power supl has some
troubles.
3)Certain cables inside are not
securel connected.
4)Main Circuit board, LED
and/or LCD have some
troubles.
Consult our supplier.
The ke s of the operation
panel do not work.
1)Ke switch has some
troubles.
2)Main Circuit board has some
troubles.
Consult our supplier.
1)The stain of the capillar
tube.
Replace the capillar tube.
2)Deterioration of capillar
tube or snapping of lamp
cable.
3)Lamp is not mounted
correctl .
Replace the lamp.
Confirm the mounting of the
lamp.
LCD displa can not be
adjusted to “0.0”.
4)The circuit board has some
troubles.
Consult our supplier.
1)The serum portion of the
sample is not correctl
inserted.
Check it with the serum gauge.
2)Scars and/or stains on the
capillar tube.
Replace the capillar tube.
3)Incorrect calibration.
Calibrate the instrument with the
new standard solution again.
Measured value is not
dipla ed correctl .
4)Circuit board has some
troubles.
Consult our supplier.

BR5N-1-OMA-103a 18
. Light bulb replacement
Caution : Prior to replacing the light bulb, be sure to turn off the power switch, unplug
the power cord from the wall outlet, and make sure that the bulb is not hot.
(1)
Remove a decorative screw and the cover from the bottom of the main unit.
(fig5)
(2)
Remove a decorative screw which fixed the lamp holder.(Fig. 6)
(3)
Disconnect the lamp connector.(fig7)
(4)
Turn the light bulb counterclockwise toward the bulb, and remove it. Do not
touch the bulb b our fingers directl , but hold it with a cloth.(fig8)
(5)
Replace the bulb and reverse the procedure.
(6)
Connect the lamp connectors.
(7)
Fix the lamp holder with the decorative screw.
(8)
Put the lamp cover and fix it with the decorative screw.
Never fail to adjust the instrument with the standard solution.
(Refer to 4.1 Adjustment with standard solution)
Fig5
Decorative scre
w
Cover
Lamp connect
or
Fig7
Fig8
化粧
ビス
Fig6
Decorative screw

BR5N-1-OMA-103a 19
Section 5 – Specifications
5.1 Specifications
Range of Measurement : 0 ~30mg/㎗(Total bilirubin)
: 0 ~ 513.09 μmol/ℓ
Measurement accurac : ±5%
Correcting Hemol sis : 0 ~250mg/㎗ Hbcv
Measuring method : Dual wavelength differential s stem (461nm,551nm)
Sample volume : 50 ~ 60 ㎕
Sample container : Hematocrit capillar tube (Heparinized)
Light source : 5V, 0.75Atungsten lump
Detector : Silicon photocell
Displa : LCD Displa 16 rows × 2 lines
Power requirements : 100~240VAC 50/60Hz, 10VA
Dimensions : 280mm(W)×230mm(D)×115mm(H)
Weight : Net 3.4kg (Main bod )
Operating temperature range : +10°C ~ +40°C
Storage temperature range : 0°C ~ +55℃
Humidit : 30% ~ 75%
Dew condensation must be avoided.
Pollution Degree :
2
( IEC 1010-1, IEC 664 )
Overvoltage Categor : Ⅱ ( IEC 1010-1, IEC 664)
Fuse ( In Power Unit ) : T 3.15A (IEC 127)
Table of contents