AquaMetrix SHARK TX User manual

SHA K
TX
& SHA K
TXP
MULTI-PA AMETE
T ANSMITTE USE ’S MANUAL
AquaMetrix Inc.
1245 Maple ill Ct., Unit 7
Newmarket, ON
Canada, L3Y 9E8
Tel: (800) 742-1413
(905) 954-0841
Fax: (905) 954-0415
www.aquametrix.com
N116-35 Rev 3 A57-24

Subject Page No.
Introduction 1
Section 1 - Specifications 2
Section 2 - Installation 3 - 5
Section 3 - Electrical Connections & Setup 6 - 11
Differential Probe connection & setup 7
Combination Probe connection & setup 8
Conductivity Cell (Contacting style) connection & setup 9
Paddle Wheel Flow Sensor connection & setup 10
4-20mA Isolated Output 11
Section 4 - Using the Transmitter in pH Mode 12
LCD Menu 13 - 28
Section 5 - Using the Transmitter in ORP Mode 29
LCD Menu 30 - 44
Section 6 - Using the Transmitter in Conductivity Mode 45
LCD Menu 46 - 62
Section 7 - Using the Transmitter in Flow Mode 63
LCD Menu 64 - 78
Appendix A - Probe Configuration Table 79
Return Policy & Warranty Plan 80
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Table of Contents

MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Introduction
Introduction Page 1
The SHARKTX & SHARKTXP multi-parameter transmitter is a microprocessor based
transmitter capable of measuring one of the following parameters, pH, ORP, conductivi-
ty or flow.
When shipped from the factory, the Shark is not set to measure any one parameter.
When the Shark is powered up for the first time, it will display the meter selection
screen where the meter type must be selected. (refer to section 4.5 Meter Selection)
This meter selection screen will only be displayed when the Shark is powered up for the
first time.
After the user selects a meter type the Shark will remain set to that meter type until it is
changed with the meter selection menu function in the Utilities menu.
To return the Shark to its factory settings, the user must re-select the current meter type
from the meter selection menu function. This will override all set-points and return all
settings back to the factory settings.
The Shark Users menu has been divided into five main categories
- Calibration, used to calibrate the Shark with the selected sensor
- Utilities, used to manually control or override the outputs.
- Setup, used to configure the Sharks many options
- Diagnostics, used to troubleshoot problems with the Shark or sensor
- Outputs, used to configure the Shark’s 4 to 20mA output.
SHARKTX is packaged in a rugged NEMA 4X polycarbonate enclosure with a universal
mounting kit for surface, panel and pipe-mount applications. This enclosure is perfect for
stand-alone or panel-mount operation.
SHARKTXP enclosure is also polycarbonate with a NEMA 4X front panel. Panel mount,
and DIN rail mounting hardware are supplied.

MΩ/cm30 to 19.99 0.01
uS/cm3
0 to 2.000 0.01
0 to 20.00 0.1
0 to 200.0 0.1
0 to 5000 1.0
mS/cm30 to 20.00 10
0 to 200.0 50
Section I - Specifications Page 2
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Section I - Specifications
pH ORP Conductivity Flow
Display 2 x 16 alpha-numeric LCD display
Power Requirements 4 to 20mA, Loop Powered, 16 to 32 VDC
Flow: 0 to 9999 with
selectable
flow rate units
pH: 0.01 to 14.00 ORP: -1999 to +1999mV
Volume: 0 - 999 with Auto Range
Measuring Range (Dependent on sensor)
Flow rate units: Gallons (GP), Cubic
Temp: 0 to 100°C or Temp: 0 to 100°C or
Feet (CF), Liters (LP), Cubic Meters
32° to +212°F 32° to +212°F (
CM), custom by entering factor
related to Gallons
Time units: Seconds (S), Minutes (M)
Temp: 0 to 100°C or 32° to +212°F
Hours (H)
Automatic or Manual
Temperature Automatic or Manual Not required User selectable temperature Not required
Compensation 0 - 100°C (32° to +212°F)
compensation slope 0.0 to 10.0%/°C.
0 to 100°C (32° to +212°F)
Temperature Unit °C or °F Not required
Temperature Sensor User selectable: 300ΩNTC Thermistor, 3000ΩNTC Thermistor or Pt. 1000 RTD Not required
Auto-Calibration Manual Calibration Dry Calibration
Calibration Modes Manual Calibration Temperature Calibration Sample Calibration K factor Input
Temperature Calibration Temperature Calibration
Ambient Conditions Temperature: -20°C to +60°C or -4°F to +140°F Humidity: 0 to 90% RH (non-condensing)
Sensor to Transmitter Differential Sensor: 3000 ft 300 ft 2000 ft
Distance Combination Sensor: 10 ft
Analog Output 4 to 20mA
Isolated Output, Range expand 0 to 100% of full scale (min segment 10% of full scale), max. load 800Ω
Memory Back-up All user settings are retained indefinitely in memory (EEPROM)
SHARKTX Enclosure: NEMA 4X, 1/4 DIN, polycarbonate enclosure with two 1/2”conduit holes
Mechanical SHARKTXP Enclosure: NEMA 4X front panel, 1/4 DIN, polycarbonate
SHARKTX Mounting: Universal Mounting kit for surface, pipe and panel mount included
SHARKTXP Mounting: Panel and DIN rail mount included
Sensor Input Probe: -600 to +600mV Probe: -1999 to +1999mV Cell: 0 to 9999ΩPaddle: 0 to 2000Hz
Temp. Sensor: 0 to 9999ΩTemp. Sensor: 0 to 9999ΩTemp. Sensor: 0 to 9999Ω
Invalid Entries Invalid entries cannot be stored
Manual Test Mode Process value can be simulated with arrow keys to verify correct setup of output
Output Hold 4 to 20mA output is placed on hold when the transmitter is in Menu mode
Recall data from last
Recall data from last calibration, calibration mode, calibration, calibration buffer
Calibration Data 1st & 2nd accepted buffer value and probe mV output, accepted value, and cell Recall store K factor.
calibration temperature, calibration slope, and probe resistance, calibration
efficiency temperature
Auto Return User selectable auto return if the transmitter is left in menu mode for more than 10 min.
Display Damping User can select rate at which the transmitter updates display. Enables display damping of unstable process
Net Weight SHARKTX: 0.71 lbs (0.32 kg)
SHARKTXP: 0.25 lbs (0.12 kg)
Approvals ULC (pending)

MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Section 2 - Installation
Section 2 - Installation Page 3
2.1 Unpacking
Save the shipping carton and packing material in case the instrument needs to be stored or
returned. Inspect the instrument and packing material for shipping damage and report any
problems immediately.
2.2 Location
Locate the transmitter close to the sensor. The list below gives typical maximum distances for
various sensors. Refer to the sensor specifications for exact information.
• Aquametrix Differential PH Probe 3000 ft (914 meters)
• Aquametrix Combination PH Probe 10 ft (3 meters)
• Aquametrix Conductivity Probe 300 ft (91 meters)
• Aquametrix Flow sensor 2000 ft (610 meters)
2.3 Mounting
Fig 2.1 SHARKTX Transmitter dimensions
Dwg# N106-128
This manual suits for next models
1
Table of contents
Popular Transmitter manuals by other brands

Geo
Geo Web Pack quick start guide

Inovonics
Inovonics EchoStream EN1210W installation instructions

IKONNIK
IKONNIK KA-6 quick start guide

Rohde & Schwarz
Rohde & Schwarz SR8000 Series System manual

Audio Technica
Audio Technica UniPak ATW-T93 Installation and operation

NIVELCO
NIVELCO EasyTREK SCA-300 Series Programming manual