ARMANO DPG 2600 User manual

www.armano-messtechnik.com
Operating Instructions
Digital Pressure Gauge DPG 2600
B78
01/22
1. Information on This Operating Instruction
• The manual is aimed at specialists and semi-skilled
personnel.
• Please read the instructions carefully before carry-
ing out any operation and keep the specified order.
• Thoroughly read and understand the information in
chapter 2 “Safety Instructions”.
If you have any problems or questions, please contact
your supplier or contact us directly at:
ARMANO Messtechnik GmbH
Location Beierfeld
Am Gewerbepark 9 •08344 Grünhain-Beierfeld
Tel.: +49 3774 58 – 0 •Fax: +49 3774 58 – 545
mail@armano-beierfeld.com
Location Wesel
Manometerstraße 5 •46487 Wesel-Ginderich
Tel.: +49 2803 9130 – 0 •Fax: +49 2803 1035
1. Information on This Operating Instruction 1
1.1 Pictographs Used 2
1.2 Exclusion of Liability 2
2. Safety Instructions 2
3. Device Description 4
3.1 Intended Use 4
4. Technical Data 4
5. Mounting 5
6. Commissioning 5
7. Operation 5
8. Maintenance/Cleaning,
Storage and Transport 7
9. Return 7
10. Dismounting and Disposal 8
11. Warranty Conditions 8
12. CE Conformity 8
13. Declaration of Conformity 9
Contents

B78 • 01/22 • GB • p. 2 of 12 • www.armano-messtechnik.com
Operating Instructions
Digital Pressure Gauge DPG 2600
1.1 Pictographs Used
In this manual, pictographs are used as hazard warnings.
Particular information, instructions and restrictions
designed for the prevention of personal or substantial
property damage:
WARNING! Is used to warn you against an
imminent danger that may result in personal
injury or death.
IMPORTANT! Is used to warn you against
a possibly hazardous situation that may re-
sult in personal, property or environmental
damage.
CAUTION! Is used to draw your attention to
important recommendations to be observed.
Disregarding them may result in property
damage.
Passages in the text containing explana-
tions, information or advice are highlight-
ed with this pictograph.
The following symbol highlights
actions you have to conduct
or
instructions that have to be strictly observed.
1.2 Exclusion of Liability
We accept no liability for any damage or malfunction
resulting from incorrect installation, inappropriate use
of the device or failure to follow the instructions in this
manual.
2. Safety Instructions
Please read this operating instruction thoroughly be-
fore operating the digital pressure gauge.
Disregarding the containing warnings, especially the
safety instructions, may result in danger for people,
the environment, and the device and the system it is
connected to.
The digital pressure gauge corresponds with the state
of engineering at the time of printing. This concerns
the accuracy, the operating mode and the safe op-
eration of the device. In order to guarantee that the
device operates safely, the operator must act compe-
tently and be conscious of safety issues.
The ARMANO Messtechnik GmbH provides support
for the use of its products either personally or via rele-
vant literature. The customer verifies that our product
is fit for purpose based on our technical information.
The customer performs customer and application
specific tests to ensure that the product is suitable
for the intended use. With this verification, all hazards
and risks are transferred to our customers. Our war-
ranty expires in case of inappropriate use.
Qualified personnel:
• The personnel that is charged for the installation,
operation and maintenance of the instrument must
hold a relevant qualification. This can be based on
training or relevant tuition. The personnel must be
aware of this manual and have access to it at all
times.
• The electrical connection shall be carried out by a
fully qualified electrician only.

www.armano-messtechnik.com • p. 3 of 12 • GB • 01/22 • B78
Operating Instructions
Digital Pressure Gauge DPG 2600
General safety instructions:
• In all work, the existing national regulations for ac-
cident prevention and safety at the workplace must
be complied with. Any internal regulations of the
operator must also be complied with, even if these
are not mentioned in this manual.
• Please ensure that the process is unpressurised
before installing or removing the device. Other-
wise, there is a risk that hot, corrosive, toxic or ex-
plosive substances leak.
IMPORTANT! Risk of burns, chemical
burns, poisoning or explosion!
IMPORTANT! Risk of injury or material
damage due to overpressure!
• Exceeding the maximum permissible overload val-
ues may lead to material failure of the digital pres-
sure gauge. This may also cause serious damage
to health.
• Please check if the instrument model is suitable for
your application before ordering and installation.
• Degree of protection according to DIN EN 60 529:
Ensure that the ambient conditions at the installa-
tion location do not exceed the requirements of the
specified degree of protection (chapter 4 “Tech-
nical Data”).
• Use the instrument in its perfect technical condition
only. Damaged or defective instruments need to be
checked immediately and replaced if necessary.
• Only use appropriate tools for mounting, connecting
and dismounting the instrument.
• Nameplates or other information on the device shall
neither be removed nor obliterated, since other-
wise any warranty and manufacturer responsibility
expires.
• In order to ensure measurement accuracy and du-
rability of the instrument and to avoid damage, the
limit values indicated in the technical data have to
be observed.
• In case of visible damage or malfunctions, the in-
strument must be put out of operation immediately.
Special safety instructions:
Warnings, which are specifically relevant to individual
operating procedures or activities, are to be found at
the beginning of the relevant sections of this operating
instruction.
• The operating pressures of the digital pressure
gauge have to be within the specifications of the
device.
• Ensure that construction type and materials of the
digital pressure gauge are resistant regarding ap-
plication conditions and medium.
• Minimise external mechanical influences, such as
oscillations, vibrations and shocks, by an appropri-
ate installation.
• Reduce the influence of vapour, abrasive / aggres-
sive media, dust and soot or others by selecting a
suitable installation location.
• Avoid direct sunlight and immediate vicinity to hot
objects as far as possible.
• Avoid strong electromagnetic fields.
• Modifications or other technical changes of the in-
strument by the customer are not permitted. Other-
wise, you will lose your warranty.
Please observe the following warnings for safe
operations:
• Ensure that the battery terminals are in the cor-
rect position.
• Do never short-circuit the batteries.
• Do never cause the batteries to overheat.
• Do not use batteries that show signs of damage.
• Do never deeply discharge the batteries.
• Do never attempt to open the batteries.
• Dispose of the batteries properly ( chapter 10
“Dismounting and Disposal”).

B78 • 01/22 • GB • p. 4 of 12 • www.armano-messtechnik.com
Operating Instructions
Digital Pressure Gauge DPG 2600
3. Device Description
The digital pressure gauge DPG 2600 has been de-
veloped for the calibration of pressure transmitters,
pressure gauges and pressure switches.
The DPG 2600 is available in two versions:
• all-rounder DPG 2600 A
with a measurement uncertainty of ≤±0.05 % FS
• all-rounder+ DPG 2600 A+
with a measurement uncertainty of ≤±0.02 % FS
With the latest microprocessor technology, the
DPG 2600 is the most advanced portable digital pres-
sure gauge in terms of accuracy. Fully compensated
in the operating temperature range, the polynomial
error compensation ensures accurate measurements
at each point of every pressure range.
It is suitable for calibrating and testing pressure
measuring instruments such as analogue pressure
gauges, digital pressure gauges, pressure transmit-
ters, pressure switches.
The DPG 2600, combined with one of ARMANO's
hand test pumps, is the perfect solution for on-site
calibrations. In calibration laboratories, the DPG 2600
can be combined with one of our DPC 3800 or DPC
4800 series high-precision pressure controllers.
Nameplate and sticker:
The nameplate is located on the device. It contains
the most important technical data and information.
Figure 1: nameplate
Scope of delivery:
• Digital pressure gauge DPG 2600
• Calibration certificate
• Operating instruction
• Mains adapter 230 V AC – 9 V DC at 500 mA
• 4 NiMH batteries
3.1 Intended Use
The technical data specified in the current data sheet
are binding and must be adhered to. If you do not
have the data sheet, please request it or download
it from our website (www.armano-messtechnik.com).
Applications that are not explicitly listed as according
to regulations, are improper to intended purpose!
The operational safety of the device supplied is only
guaranteed by intended use. The specified limit values
(chapter 4 “Technical Data”) must not be exceeded.
4. Technical Data
Process connection
• material
• connection thread
stainless steel 316L
1⁄2" BSP
Case stainless steel 1.4301
Pressure units mbar, bar, Pa, hPa, kPa, MPa,
kg/cm2, kg/m2, mmHg, cmHg,
mHg, mmH2O, mH2O, mH2O,
torr, atm, psi, inHg, inH2O, ftH2O
Measurement
accuracy
DPG 2600 A with ≤±0.05 % FS
DPG 2600 A+ with ≤±0.02 % FS
Pressure ranges overpressure, compound
pressure or absolute pressure
from −200 / +200 mbar up to
0 – 1000 bar
Storage temperature −20 °C to +70 °C
(−4 °F to +158 °F)
humidity < 95 %
non-condensing
Supply either via 4 NiMH batteries or
via the included mains adapt-
er; the batteries are charged
during mains operation

www.armano-messtechnik.com • p. 5 of 12 • GB • 01/22 • B78
Operating Instructions
Digital Pressure Gauge DPG 2600
5. Mounting
Prior to mounting, please check the following aspects:
• Are the goods undamaged and complete?
• Do the goods match the shipping documents?
• Is the instrument suitable for the case of application?
• Is the maximum possible process pressure within
the measuring range of the device to be installed?
• Does the process connection comply with the
requirements?
• Pay attention to adequate protection against weather!
• Avoid direct sunlight.
• Avoid proximity to heat sources.
• Note the degree of protection according to
DIN EN 60 529 (chapter 4 “Technical Data”).
• Operation and control shall only be carried out by
authorised personnel.
• Take appropriate precautions to protect the device
from damage!
IMPORTANT! Mounting and dismounting
of the device shall only be carried out in an
unpressurised state!
Connection
Please use two spanners (¼" BSP F or ½" BSP F)
when connecting the pressure port of the device to
avoid damage to the sensor and / or the device itself.
The DPG 2600 can be operated without batteries us-
ing only the mains adapter.
Four NiMH batteries (size AAA) are installed as
standard power supply, which can also be recharged
via the mains adapter.
6. Commissioning
Before commissioning, check for proper installation
and for any visible defects of the device.
Commissioning shall only be carried out by qualified
personnel, who read and understood this manual!
The device has to be operated within the specifica-
tions only! (Please refer to the technical data in the
data sheet.)
7. Operation
ARMANO Messtechnik GmbH
CHARGE
ZERO UNIT
+
MENU
–
ENTER
Figure 7.1: membrane keypad
The DPG 2600 has five buttons
Top right:
on/off button
Four control buttons below the display:
MENU
–
The MENU button for selecting and exiting
menu functions and for reducing the dis-
play contrast when pressed together with
the ZERO button.
ZERO The ZERO button for zeroing the meas-
urements in the overpressure mode, for
adjusting the contrast and for scrolling to
the left in the menu mode.
UNIT
+
The UNIT button for changing the unit.
When pressed together with the ZERO
button, it increases the display contrast.
ENTER With the Enter button, the backlight can be
switched on and off.
The buttons ZERO UNIT
+
and ENTER are also used for
navigation in the menu functions.

B78 • 01/22 • GB • p. 6 of 12 • www.armano-messtechnik.com
Operating Instructions
Digital Pressure Gauge DPG 2600
The DPG 2600 has the following menu functions:
DECIMALS increase or decrease the number of
the indicated decimal places.
FILTER enables the following function:
OFF/ON activate or deactivate measured
value filtering.
FILTER RESET
resets the measured value filter.
FILTER WEIGHT
sets the weighting of the filter
between 10…100 %.
MAX-MIN RESET
resets the minimum and maximum
values.
OFFSET tares the indication and the
currently displayed measured value.
FREEZE freezes the indication at the current
measured value.
USE BAROMETRIC
(with optional barometric reference
only)
when ON, the measurement mode
changes to Absolut.
NOTE: The Gshown in the display
indicates the sensor type and
not the measuring mode (when
absolute measuring mode is active,
“g--> a” is displayed.
LOG SAMPLE
TIME sets the sample time of the
measured value logging.
LOG TIME configures the time period of the
measured value logging.
LOG SLOT software slot (CSV file).
LOG switches logging on and off. A
flashing symbol is displayed on the
screen during logging. Up to 8 logs
in CSV format can be downloaded
via the RS-232 port (optional RS-232
data cable and software).
BAUD RATE sets the baud rate of the serial
communication.
available values are OFF-300-600-
1200-2400-4800-9600-19200-
38400-57600-115200
SERIAL ADDR currently not in use
(please keep set to 1).
AUTO BKL-OFF when activated (ON), the display
backlight turns off after some time
to extend the battery life.
AUTO PWR-OFF when activated (ON), the device
switches off automatically after
some time to extend the battery life.
Pressure units
mbar, bar, Pa, hPa, kPa, MPa, kg/cm2, kg/m2,
mmHg, cmHg, mHg, mmH2O, cmH2O, mH2O, torr,
atm, psi, inHg, inH2O, ftH2O
(note: Hg @ 0 °C; H2O @ 4 °C; 1 atm = 101325 Pa)
Battery life
When the batteries are low, a warning appears on
the display. Connect the device to the power supply.
During charging, the battery status indicator remains
at 100 %. The end of charging is reached when the
yellow charging LED goes out.
Interfaces
RS-232 interface
On the back of the DPG 2600 in the form of an M12
connector plug.
An RS-232 data cable and an RS-232 USB converter
are available upon request.
Power supply
On the back of the DPG 2600.
9 V DC – 600 mA
Polarity of the coaxial power connector:

www.armano-messtechnik.com • p. 7 of 12 • GB • 01/22 • B78
Operating Instructions
Digital Pressure Gauge DPG 2600
8. Maintenance / Cleaning,
Storage and Transport
CAUTION! Material damage and loss
of warranty!
Any modifications or interventions in the
device, made by the customer, might dam-
age important parts or components. Such
intervention leads to the loss of any warran-
ty and manufacturer’s responsibility!
ÆNever modify the device or perform any
repairs yourself.
Maintenance:
In principle, the device is maintenance-free. If neces-
sary, the housing of the device can be cleaned with a
damp cloth and a non-aggressive cleaning solution
when switched off.
Cleaning:
Depending on the medium, deposits or contamination
may occur. If such a contamination of the medium is
known, the operator has to specify appropriate clean-
ing intervals.
CAUTION! Incorrect cleaning may cause
irreparable damage. Therefore, never use
any sharp objects or compressed air for
cleaning.
IMPORTANT! Improper transport can de-
stroy the device and cause considerable
personal and property damage.
Please inspect the transport packaging and
the delivered items immediately upon their
receipt to determine their integrity, com-
pleteness and conformity with the delivery
documents.
The permissible ambient conditions for
storage and transport can be found in the
data sheet.
Storage:
• If possible, store the instrument in its original pack-
aging.
• If possible, remove the packaging not until installa-
tion of the device.
• Store the instruments in a dry place, not exposed
to direct sunlight.
• The storage temperature of the instruments should
not fall below or exceed the permissible tempera-
ture limitations specified in the data sheet.
Transport:
Electronic components!
The device is equipped with sensitive elec-
tronic components and has to be handled
with due care.
Remove batteries!
If the device is not in use for extended peri-
ods, the batteries should be removed from
the device to preclude any leaking damage.
• Use the original packaging or comparable packag-
ing for transport.
• Avoid impacts or strong vibrations.
• Protect the device against moisture.
9. Return
Prior to any return, whether for recalibration, decalci-
fication, for modification or for repair, the instrument
has to be cleaned thoroughly and packaged careful-
ly. Please enclose a notice of return with a detailed
description of the faults when returning a defective
device. If your instrument came into contact with
harmful substances, a declaration of contamination is
required additionally. A corresponding template can be
found on our website www.armano-messtechnik.com.
If you send in your device without a declaration of con-
tamination and our service department has doubts
regarding the medium used, then the repair will only
be started once a corresponding declaration has been
submitted.
If the device came into contact with
harmful substances, appropriate precau-
tions are to be taken during cleaning!

B78 • 01/22 • GB • p. 8 of 12 • www.armano-messtechnik.com
Operating Instructions
Digital Pressure Gauge DPG 2600
10. Dismounting and Disposal
WARNING! Risk of injury!
Never remove the device from a system in
operation.
Make sure that the system is switched off
professionally.
Before dismounting:
Check before dismounting, whether the system
• is switched off,
• is in a safe and currentless state,
• is unpressurised and cooled down.
Dismounting:
Pay attention to potentially leaking media. Take ap-
propriate precautions to collect them.
Disposal:
In compliance with the directives 2011/65/EU (RoHS)
and 2012/19/EU (WEEE), the device must be disposed
of separately as electrical and electronic waste. Please
regard legal regulations of the country of distribution.
NO DOMESTIC WASTE!
The instrument comprises various materials.
It shall not be disposed of together with do-
mestic waste.
ÆBring the device to your local recycling plant
or
Æsend the device back to your supplier or to the
ARMANO Messtechnik GmbH.
Disposal of used batteries:
• Cover the poles with tape during storage and dis-
posal to avoid short circuits.
• Dispose of used batteries properly in commercial
collection boxes or at municipal collection points.
WARNING! Depending on the medium
used, residues on the device may cause
a risk for the user and the environment.
Therefore, take appropriate precautions
and dispose of the device properly.
11. Warranty Conditions
The warranty conditions are subject to the statutory
warranty period of 24 months, valid from the date of
delivery.
Any warranty claims are excluded in case of improper
use, modification of or damage to the device. Dam-
aged membranes are not accepted as warranty claim.
Furthermore, defects resulting from normal wear are
not subject to warranty services.
12. CE Conformity
The CE marking of the instruments certifies
the conformity with prevailing EU directives
for placing products on the market within
the European Union. The following direc-
tives apply:
2014/30/EU (EMC)
2014/68/EU (PED)
The corresponding declaration of conform-
ity is enclosed or available upon request.

www.armano-messtechnik.com • p. 9 of 12 • GB • 01/22 • B78
Operating Instructions
Digital Pressure Gauge DPG 2600
13. Declaration of Conformity
EU Declaration of Conformity
EU-Konformitätserklärung
www.armano-messtechnik.de
abgegeben durch/by
Grünhain-Beierfeld, 2021-08-30
Bernd Vetter
Geschäftsführender Gesellschafter/Managing Director
Diese Erklärung wird verantwortlich für den Hersteller:
This declaration is issued under the sole responsibility of the manufacturer:
ARMANO Messtechnik GmbH
ARMANO Messtechnik GmbH
Standort Beierfeld
Am Gewerbepark 9
08344 Grünhain-Beierfeld
Tel.: +49 3774 58 – 0
Fax: +49 3774 58 – 545
Standort Wesel
Manometerstraße 5
46487 Wesel-Ginderich
Tel.: +49 2803 9130 – 0
Fax: +49 2803 1035
Für die nachfolgend bezeichneten Erzeugnisse
Digital-Manometer Typ DPG 2600
gem. Datenblatt 9668
wird hiermit bestätigt, dass sie den wesentlichen Schutzanforderun-
gen entsprechen, die in den nachfolgend bezeichneten Richtlinien
festgelegt sind:
RICHTLINIE 2014/30/EU DES EUROPÄISCHEN PARLAMENTS
UND DES RATES vom 26. Februar 2014
zur Angleichung der Rechtsvorschriften der Mitgliedstaaten über die
elektromagnetische Verträglichkeit – kurz: EMV-Richtlinie
RICHTLINIE 2014/68/EU DES EUROPÄISCHEN PARLAMENTS
UND DES RATES VOM 15. Mai 2014
zur Angleichung der Rechtsvorschriften der Mitgliedstaaten über
Druckgeräte – kurz: Druckgeräterichtlinie
RICHTLINIE 2011/65/EU DES EUROPÄISCHEN PARLAMENTS
UND DES RATES vom 8. Juni 2011
zur Beschränkung der Verwendung bestimmter gefährlicher Stoffe
in Elektro- und Elektronikgeräten – kurz: RoHS-Richtlinie
Die Geräte entsprechen folgenden technischen Vorschriften:
We hereby declare for the following named goods
Digital Pressure Gauge Model DPG 2600
according to data sheet 9668
that they meet the essential protective requirements, which have
been fixed in the following directives:
DIRECTIVE 2014/30/EU OF THE EUROPEAN PARLIAMENT AND
THE COUNCIL from February 26, 2014
on the approximation of the laws of the Member States relating to
the electromagnetic compatibility – short: EMC Directive
DIRECTIVE 2014/68/EU OF THE EUROPEAN PARLIAMENT AND
THE COUNCIL from May 15, 2014
on the approximation of the laws of the Member States relating to
pressure equipment – short: Pressure Equipment Directive
DIRECTIVE 2011/65/EU OF THE EUROPEAN PARLIAMENT AND
THE COUNCIL from June 8, 2011
on the restriction of the use of certain hazardous substances in
electrical and electronic equipment – short: RoHS Directive
The devices comply with following technical specifications:
149 EU-Konformitätserklärung DPG 2600 Ausg. 08/21
Norm:
Standard:
Richtlinienbezug
Reference to directive
DIN EN 55011:2018-05 Industrielle, wissenschaftliche und medizinische Geräte - Funkstörungen -
Grenzwerte und Messverfahren
Industrial, scientific and medical equipment - Radio-frequency disturbance characteristics -
Limits and methods of measurement
DIN EN 61000-4-2:2009-12 Elektromagnetische Verträglichkeit (EMV): Prüf- und Messverfahren -
Prüfung der Störfestigkeit gegen die Entladung statischer Elektrizität
Electromagnetic compatibility (EMC): Testing and measurement techniques -
Electrostatic discharge immunity test
DIN EN 61000-4-3:2011-04 Elektromagnetische Verträglichkeit (EMV): Prüf- und Messverfahren -
Prüfung der Störfestigkeit gegen hochfrequente elektromagnetische Felder
Electromagnetic compatibility (EMC): Testing and measurement techniques -
Radiated, radio-frequency, electromagnetic field immunity test
DIN EN 61000-4-8:2010-11 Elektromagnetische Verträglichkeit (EMV): Prüf- und Messverfahren -
Prüfung der Störfestigkeit gegen Magnetfelder mit energietechnischen Frequenzen
Electromagnetic compatibility (EMC): Testing and measurement techniques -
Power frequency magnetic field immunity test
DIN EN 61326-1:2013-07 Elektrische Mess-, Steuer-, Regel- und Laborgeräte -
EMV-Anforderungen: Allgemeine Anforderungen
Electrical equipment for measurement, control and laboratory use -
EMC requirements: General requirements

B78 • 01/22 • GB • p. 10 of 12 • www.armano-messtechnik.com
Operating Instructions
Digital Pressure Gauge DPG 2600

www.armano-messtechnik.com • p. 11 of 12 • GB • 01/22 • B78
Operating Instructions
Digital Pressure Gauge DPG 2600

www.armano-messtechnik.com
Operating Instructions
Digital Pressure Gauge DPG 2600
© 2022 ARMANO Messtechnik GmbH · Technical changes, replacement of materials and printing errors excepted!
ARMANO Messtechnik GmbH
Location Beierfeld
Am Gewerbepark 9
08344 Grünhain-Beierfeld
Tel.: +49 3774 58 – 0
Fax: +49 3774 58 – 545
mail@armano-beierfeld.com
Location Wesel
Manometerstraße 5
46487 Wesel-Ginderich
Tel.: +49 2803 9130 – 0
Fax: +49 2803 1035