Projecta DT910 User manual

DIGITAL
P/No. DT910
CLAMP METER
UP TO 400 AMPS

2
WARNING
To avoid possible electric shock or personal injury, follow these guidelines:
• Do not use the meter if it is damaged. Before you use the meter, inspect the case.
Pay particular attention to the insulation surrounding the connectors.
• Inspect the test leads for damaged insulation or exposed metal. Check the test leads
for continuity. Replace damaged test leads before you use the meter.
• Do not use the meter if it operates abnormally. Protection may be impaired. When in
doubt, have the meter serviced.
• Do not operate the meter around explosive gas, vapour, or dust.
• Do not apply more than the rated voltage, as marked on the meter, between terminals
or between any terminal and earth ground.
• Before use, verify the meter’s operation by measuring a known voltage.
• When servicing the meter, use only specified replacement parts.
• Use with caution when working above 30V ac rms, 42V peak, or 60 dc. Such voltages
pose a shock hazard.
• When using the probes, keep your fingers behind the finger guards on the probes.
• Connect the common test lead before you connect the live test lead. When you
disconnect test leads, disconnect the live test lead first.
• Remove the test leads from the meter before you open the battery door.
• Do not operate the meter with the battery door or portions of the cover removed
or loosened.
• To avoid false readings, which could lead to possible electric shock or personal injury,
replace the batteries as soon as the low battery indicator (“ ”) appears.
• CAT II – Measurement Category II is for measurements performed on circuits
directly connected to low voltage installation. (Examples are measurements on
household appliances, portable tools and similar equipment.) Do not use the meter
for measurements within Measurement Categories III and IV.

INTRODUCTION
This manual provides all safety information, operation instruction, specifications and
maintenance for the meter, which is compact, handheld, and battery operated.
This instrument performs AC/DC voltage, AC/DC Current (clamp), Resistance, Audible
Continuity, Diode, Frequency measurements and Non-Contact ACV Detecting(NCV).
It is an 3 3/4 digit (4000 counts) auto ranging DMM with DC & AC clamp meter function.
CAUTION
To avoid damage to the meter, don’t apply input which exceeds the limit shown below:
Function Terminals Input Limits
DCV/ACV VΩ& COM 600V DC or 600V rms AC
ΩVΩ& COM 250V DC or rms AC
Hz VΩ& COM 250V DC or rms AC
GENERAL CHARACTERISTICS
Display: 4000 counts, updates 2/3 sec.
Polarity Indication: “-” displayed automatically
Over-range Indication: “OL” displayed
Low Battery Indication: “ ” displayed
Operation Temperature: 0°C to 40°C
Storage Temperature: -10°C to 50°C
Battery Type: (AAA) 1.5V×2PCS
Dimension(H×W×D): 186×75×28mm
Weight: Approx. 180g
Accessories: Operator’s Manual, Battery, Test Leads
PANEL DESCRIPTION
1. Transformer jaw
2. Clamp trigger
3. Function button
4. LCD display
5. COM Input Terminal
6. NCV LED Indication
7. Function rotated switch
8. Hold button
9. REL(Zero) button
10. VΩInput Terminal
1
2
3
4
5
6
7
8
9
10
3

4
SPECIFICATIONS
AC VOLTAGE (Auto ranging)
Range Resolution Accuracy
400mV 0.1mV ±(1.5% of rdg + 5dgts)
4V 1mV
40V 10mV ±(1.2% of rdg + 3dgts)
400V 100mV
600V 1V
Input Impedance: more than 10MΩ
Frequency Range: 40 to 200Hz
DC VOLTAGE (Auto ranging)
Range Resolution Accuracy
400mV 0.1mV ±(1.2% of rdg + 5dgts)
4V 1mV
40V 10mV ±(1.0% of rdg + 2dgts)
400V 100mV
600V 1V
Input Impedance: more than 10MΩ
RESISTANCE (Auto Ranging)
Range Resolution Accuracy
400Ω0.1Ω
4KΩ1Ω
40KΩ10Ω±(1.2% of rdg + 3dgts)
400KΩ100Ω
4MΩ1KΩ
40MΩ10KΩ±(2.5% of rdg + 3dgts)
AC CURRENT (Auto ranging)
Range Resolution Accuracy
AUTO RANGE 4A 1mA ±(2.5% of rdg + 15dgts)
40A 10mA
AUTO RANGE 200A 100mA ±(2.5% of rdg + 5dgts)
400A 1A
Measuring voltage drop: 200mV; Frequency Range: 40 to 400Hz.

DC CURRENT (Auto ranging)
Range Resolution Accuracy
AUTO RANGE 4A 1mA ±(3.0% of rdg + 5dgts)
40A 10mA
AUTO RANGE 200A 100mA ±(2.5% of rdg + 7dgts)
400A 1A
FREQUENCY (Auto ranging)
Range Resolution Accuracy
10 ~ 1M Hz 1 Hz ±(1.0% of rdg + 2dgts)
1M ~ 10M Hz 1A ±(1.5% of rdg + 4dgts)
Input Voltage Limit: Max.250V(<1M Hz) Max. 10V(>1M Hz).
Non-Contact AC VOLTAGE Detection(NCV)
Special LED Lighting is indicating once AC>100V.
CONTROL BUTTON DESCRIPTION
HOLD BUTTON
When this button is pushed, LCD will show the last reading, and “H” symbol will appear
till pushed again. Data holding will be cancelled automatically when the function switch
is rotated.
FUNC BUTTON
Push this button to select ACA/DCA, ACV/DCV, measuring function when the
function switch is set at A,V, or position.
REL/ZERO BUTTON
When this button is pushed, LCD will show the relative reading or Zero reading till
pushed again.
OPERATION INSTRUCTION
AC/DC VOLTAGE MEASUREMENT
Connect the BLACK test lead to the COM jack and the RED to the VΩjack.
Set the function switch at V position, connect test leads across the source or load
under measurement.
Push “FUNC” button to select AC or DC Mode, and then the symbol ~ or
is shown on LCD.
Read LCD display. The polarity of RED test lead will be indicated when making
a DC measurement.
5

6
RESISTANCE MEASUREMENT
1. Connect the BLACK test lead to the COM jack and the RED to the VΩjack.
(NOTE: The polarity of the RED lead is positive “+”)
2. Set the function switch at Ωposition, connect test leads across the resistance
under measurement.
NOTE:
1. For resistance above 2MΩ, the meter may take a few seconds to stabilize reading.
This is normal for high resistance measuring.
2. When the input is not connected, i.e. at open circuit, the figure “OL” will be displayed
under over-range condition.
3. When check in circuit resistance, be sure the circuit under test, has all power removed
and all capacitors are fully discharged.
AC/DC CURRENT MEASURE
1. Set Function/Range Switch to the 400A range. If the display indicates one or more
leading zeros. Shift to 40A range to improve the resolution of the measurement.
2. Push “FUNC” button to select AC or DC Mode, and then the symbol ~ or
is shown on LCD.
3. Press the “REL(ZERO)” button to zero the meter.
4. Press the trigger to open the transformer jaws. And clamp one conductor only it is
impossible to make measurements when two or three conductors are clamped at the
same time.
5. Display reading is showing the current value.
CONTINUITY/DIODE TEST
1. Connect the BLACK test lead to the COM jack and the RED to the VΩjack.
2. Set the function switch at position and push the button “FUNC” to select
continuity or diode test mode.
3. In continuity test, if the circuit resistance under test is lower than 50Ω, built-in buzzer
will sound.
4. If diode test mode is selected, connect the RED and BLACK test leads to anode and
cathodes of the diode under test separately. The forward voltage drop of this diode in
V will be displayed.

FREQUENCY TEST
1. Connect the BLACK test lead to the COM jack and the RED test lead to the V/ΩHz jack.
2. Put the range selector at “Hz” position, and connect the test leads across the
resistance under measurement.
3. Take the reading from LCD.
Note:
When test the high Frequency >1MHz, the input Voltage must be less than 10V for getting
the accurate value.
NON-CONTACT AC VOLTAGE DETECTION (NCV)
1. Set the function range switch at the NCV position. Then the NCV indicating LED light
is flashing.
2. Take the clamp jaw close to the AC signal.
3. Once the ACV is above 100V, the NCV indicating LED flashes quickly.
NOTE:
Don’t try to contact the testing place with this tester. No display on LCD at this range.
BATTERY REPLACEMENT
If the sign “ ” appears on the display, it indicates battery should be replaced.
Remove screws and open the back case, replace the exhausted batteries with
new ones (AAA 1.5V *2pcs or equivalent).
7

Distributed by
AUSTRALIA
Brown & Watson International Pty Ltd
Knoxfield, Victoria 3180
Telephone (03) 9730 6000
Facsimile (03) 9730 6050
National Toll Free 1800 113 443
NEW ZEALAND
Narva New Zealand Ltd
22–24 Olive Road
PO Box 12556 Penrose
Auckland, New Zealand
Telephone (09) 525 4575
Facsimile (09) 579 1192 IS263
Issue 1 26.08.13
WARRANTY STATEMENT
Brown & Watson International Pty Ltd (“BWI”) of 1500 Ferntree Gully Road, Knoxfield,
Vic., telephone (03) 9730 6000, fax (03) 9730 6050, warrants that all products described in
its current catalogue will under normal use and service be free of failures in material and
workmanship for a period of one (1) year from the date of the original purchase by the
customer as marked on the invoice. This warranty does not cover ordinary wear and tear,
abuse, alteration of products or damage caused by the purchaser.
To make a warranty claim the consumer must deliver the product at their cost to the
original place of purchase or to any other place which may be nominated by either
BWI or the retailer from where the product was bought in order that the warranty
assessment may be performed. The consumer must also deliver the original invoice
evidencing the date and place of purchase together with an explanation in writing as
to the nature of the claim.
In the event that the claim is determined to be for a minor failure of the product then BWI
reserves the right to repair or replace it at its discretion. In the event that a major failure is
determined the consumer will be entitled to a replacement or a refund as well
as compensation for any other reasonably foreseeable loss or damage.
This warranty is in addition to any other rights or remedies that the consumer may have
under State or Federal legislation.
IMPORTANT NOTE
Our goods come with guarantees that cannot be excluded under the Australian Consumer
Law. You are entitled to a replacement or refund for a major failure and compensation for
any other reasonably foreseeable loss or damage. You are also entitled to have the goods
repaired or replaced if the goods fail to be of acceptable quality and the failure does not
amount to a major failure.
Table of contents
Other Projecta Measuring Instrument manuals