Asap CUBE Dual User manual

ENDOSCOPE REPRECESSOR
CUBE Dual Manual
ASAP Co., Ltd.

Table of Contents
1. Product Specification ···················································································· 3
2. Precautions ······································································································· 4
3. Name & Description of Each Unit·································································5
4. Using Method
1) Preparation Items Before Use ·················································· 9
2) Using Method & Operation Sequence ········································· 9
3) Method of Storage & Management After Use ························· 12
5. Customer Service(A/S) ················································································ 13
6. Marking ········································································································· 13
7. Product Warranty ······················································································ 14

3
1. Product Specification
N
o
Classification
Contents
Remarks
1
Cleaning & Rinsing
Method
·Exterior Surface: Cleaning by rotational
spray of washing water
·Channel Interior: Cleaning using washing
water
·Buttons: Cleaning by rotational spray of
washing water
2
Disinfection
Method
·Exterior Surface: Soaking in disinfection
solution
·Channel Interior: Soaking in & spay
disinfection solution
·Buttons: Soaking in disinfection solution
3
Program Mode
As configured as automatic mode in 1 to 4
modes and manual mod in 5 to 9 modes,
realize the convenience of automatic mode
and the efficiency of manual mode
4
Operation Unit
Button
As configured with (Leakage Test, Alcohol
Selection, Disinfection Solution Supply,
Disinfection Solution Drain, Cleaning Tank
Drain) buttons, may use easily with
pressing a button once
5
Number of
cleaning
endoscopes
2
6
Cleaning & Rinsing
Time
0 to 5 Min (Set in unit of 1 second)
7
Disinfection Time
0 to 60 Min (Set in unit of 1 minute)
8
Disinfection
Solution
Temperature
15℃to 32℃Heating (Set in unit of 1 ℃)
Option
9
Alcohol Supply
Time
5 Sec
1
0
Air Supply Time
0 to 30, 30, 60 Sec (Set in unit of 1
second)

4
1
1
Leakage
Inspection
Pressure
0.3 bar
1
2
Disinfection
Solution
Recommend using of the approved
disinfection solution from the Ministry of
Food & Drug Safety (The disinfection time
is based on the use criteria of the
disinfection solution manufacturer)
1
3
Alcohol Tank
Capacity
2ℓ
1
4
Disinfection Tank
Capacity
19ℓ
1
5
Cleaning Tank
Capacity
14ℓ
1
6
Water Supply
Requirement
Pressure: 1~ 5bar
2. Precautions
1. Do not Use the product except a skilled person.
2. Do not touch with wet hands.
3. Do not apply excessive force, such as holding the wire in case of removing the
power cord.
4. Be careful of the following in case of installing the product:
1) Install in a well-ventilated place that there is no moisture and exposure to
direct sunlight.
2) Install away from heat generating objects.
3) Install on a hard and flat floor.
4) Connect the power cord exactly to AC 220V (60Hz) single-phase ground outlet.
5. Before using the product, the water-resistant cap of soft endoscope must be fastened.
6. Measure sequence in case of product malfunction or failure during the use of
product
1) Stop using the product and turn off the power switch.
2) Close the water supply valve.
3) Unplug the power cord.
4) Dry the moisture in case of water leakage.
5) Mark the failure and failure symptoms outside the product for other users and
technicians to see it.
6) Contact an expert engineer and take the measure.

5
※CAUTION: Do not touch the product arbitrarily.
7. Be careful of the following after using the product:
1) Turn off the power switch.
2) The water supply valve must be closed.
8. When reusing a product that has not been used for a long time, use it after checking if
the product is operating properly before use.
3. Name & Function Description of Each Unit
1. Front Side
No
Name
Description
①
Cleaning Tank Cover
Cover that can be opened and closed during cleaning
②
Alcohol Gauge
Display the remaining amount of alcohol in the alcohol tank
③
Disinfection Gauge
Display the remaining amount of disinfection solution in the
disinfection tank
④
Carrying Caster
Carrying caster for product movement

6
⑤
Operation Unit
Control the operation of product
⑥
Power Switch
Switch to turn on and off the product’s power
⑦
Disinfection Solution
Manual Water Drain
Outlet
Drain manually the disinfection solution by opening the
valve
2. Cleaning Tank
No
Name
Description
①
Cleaning Tank Water Level
Sensor
Sensor to detect the level of water or disinfection solution
within cleaning tank
②
Water Supply Pipe
Supply water to the cleaning tank
③
Cleaning Tank Nozzle
Cleaning the endoscopic surface by water spray from
four nozzles
④
Alcohol Inlet
Inlet to supply the alcohol into the alcohol tank
⑤
Channel Connector
Connector to be connected with the channels of
endoscope
⑥
Disinfection Solution
Supply Pipe
Supply disinfection solution to the cleaning tank

7
⑦
Cleaning Tank Water Drain
Outlet (Including Filter)
Drain of water or disinfection solution from cleaning Tank
(filter to filter out foreign substances)
⑧
Leakage Connection
Connector
Connect water leakage inspection plug in case of water
leakage inspection
3. Operation Unit
No
Name
Function
①
LCD Display Unit
Display the current status of product
·COUNT: Display the number of disinfection solution
use
·WASH: Display the cleaning time
·AIR: Display air supply time inside endoscope
channel
·DISIN: Display the disinfection time
·RINS: Display the rinsing time
·SEATE: Display total used time
②
Warning Indicator
Indicator that can notice if something is wrong during
operation
·DOOR: Turn on in case that the cleaning tank cover is
open
·DISN: Turn on in case that there is no disinfection
solution in disinfection tank
·DRAIN: Turn on in case that there is fluid in the
cleaning tank
③
Progress Indicator
Display the progress phase of program
·W: Turn on at the cleaning phase
·D: Turn on at the disinfection phase
·S: Turn on at the rinsing phase

8
④
Selection Button
Program Mode Selection Button
(1~4:Automatic Mode, 5~9:Manual Mode)
⑤
Selection Display Unit
Display the program mode selected at ④
⑥
TIME
CONTROL
UP Button
TIME CONTROL Sequence
1) Move to desired process by pressing UP Button or
DOWN Button, and
2) Push ENTER Button,
3) Set the desired time by pressing UP Button or DOWN
Button, then
4) Push ENTER Button again to save it
DOWN
Button
ENTER
Button
⑦
Leakage Test Button
Execute the leakage test by supplying air inside the
endoscope and water outside
⑧
Alcohol Selection
Button
If the button is selected during automatic mode, supply
alcohol to the endoscopic inside for 5 seconds.
⑨
Disinfecti
on
Solution
Discharge
Button
Drain used disinfection solution to the outside
Supply
Button
Supply new disinfection solution to the disinfection tank
⑩
Drain Button
Drain the water and disinfection Solution in the cleaning
tank
⑪
Start/Stop Button
Start: Push at the start of program mode (1 to 8 modes)
Pause: If pressing once softly, it is paused
Stop: If pressing more than 3 seconds in use, the
executed order is cancelled.
4. Rear Side

9
No
Name
Description
①
Short Circuit
Breaker
Safety device shutting off power automatically in case of
occurring leakage current or excessive current
②
Power Socket
Connect the power cord to supply the power
(Protect from excessive current due to built-in fuse)
③
Hinge
Hinge that opens and closes the cover
④
Water Drain
Outlet
Connect with hose between water drain outlet and drain pine of
building
⑤
Water Supply
Inlet
Connect with hose between water supply inlet and water supply
pipe of building
5. Using Method
1. Preparation Items Before Use
1) Check if there is any damage to the product in case of warehousing.
2) Installation Location
①Install in a well-ventilated place that there is no moisture and exposure to
direct sunlight.
②Install away from heat generating objects.
③Install on a hard and flat floor.
3) Product Installation
①Connect th water supply hose to faucet.
②Connect the end of drain hose to the water drain outlet.

10
③Connect the power plug to the AC 220V (60Hz) single-phase
ground outlet.
4) Be sure to read the manual before using the product, then use it after learning
how to use.
5) In case of using alcohol, supply an appropriate amount of alcohol into the inlet
of alcohol.
6) Supply appropriate amount of the disinfection solution tank.
(For more information, refer to “9) Other Using Method”on page 12)
2. Using Method & Operation Sequence
< Operation Unit >
《※Reference: For a quick review, refer to only underlined letters. 》
1) Turn on the power switch.
2) The inspection completed endoscope must be fastened by a water-resistant cap, roll,
and put it into the cleaning tank.
3) Set the process (cleaning, disinfection, air, rinsing) time in the operation Unit.
(1) Press UP or DOWN button of the operation unit to move the cursor to the
desired process (WASH, AIR, DISIN, RINS).
Disinfection Solution
Used Times
LCD
Display Unit
Warning
Indicator
Progress
Indicator
Selection
Display Unit
Used Time

11
< Process Description >
(2) Press the Enter button.
(3) If the cursor blinks, press UP or DOWN button to set the desired time.
①The cleaning process, rinsing process time can be set in unit of 1 second
from 0 second to 5 minutes
②The disinfection process can be set in unit of 1 minute from 0 min to 1 hour
③The air process can be set in unit of 1 second 0 sec to 30 sec.
(4) Press the Enter button to save it.
4) Perform the water leak test of endoscope.
(1) The purpose of the scope leak test is to protect the endoscope by checking
the damage of the endoscope.
(2) Scope Leak Test
①Endoscopy must be fastened with a water-resistant cap.
(If there is moisture, foreign substances inside and outside the water-
resistant cap, remove them and use it)
②Connect the leak inspection plug of this product to the water-resistant cap.
③Press the leakage test button of operation unit,
④Then, execute the leakage test by pressing the start button of operation unit.
⑤The water supply stops when the cleaning tank starts to fill up and the
water is filled to the normal level. At this time, the user checks visually for air
bubbles in the endoscope.
⑥If air bubbles occur continuously, do not clean and disinfect, but take the
endoscope out of the cleaning tank to be able to dry, and contact an expert
technician to take appropriate measure.
⑦If there is no problem (leakage from endoscope) in the above, cleaning can
be performed.
5) Connect the channel connection to each channel of the endoscope. (Refer to
below picture)
Cleaning
Process
Disinfection
Process
Rinsing
Process
Air Process (Supply air into endoscope channel)

12
< Olympus Endoscopic Channel Connection Example >
6) Press the selection button to select the desired program mode.
<<Program Mode>>
<No.1 to 4 Mode: Automatic Mode>
No.1 Mode (=A-Auto Mode): Cleaning-Air-Disinfection-Air-Deposition Cleaning
(Drain + Air)- Alcohol (Supply if selected)-Air
No.2 Mode (=B-Auto Mode): Cleaning-Air-Disinfection-Air-Rinsing-Alcohol
(Supply if selected)-Air
No.3 Mode (=Simple Cleaning Mode): Disinfection-Air-Deposition Cleaning (Drain
+ Air)-Alcohol (Supply if selected)-Air
No.4 Mode (=Ultra Simple Cleaning Mode): Disinfection-Air-Rinsing-Alcohol (If
selected)-Air
- In Automatic Mode, press the alcohol selection button of operation unit to select
alcohol process, then supply alcohol to endoscopic channels for 5 seconds -
<No.5 to 9Mode: Manual Mode>
No.5 Mode: Cleaning (=Perform the cleaning process only for the set time)
No.6 Mode: Disinfection (=Perform the disinfection process only for the set time)
No.7 Mode: Air (=Perform the cleaning process only for the set time)
No.8 Mode: Alcohol (=Supply alcohol to endoscopic channel for 5 seconds)
No.9 Mode(Option): Heater (=Perform the heating of the disinfection solution until
the set temperature. However, the disinfection solution shall
be filled with 10ℓor more within the disinfection tank)
Set the desired temperature according to the previous 3) Time
Setting Method.
7) Press the Start Button to execute it.
8) Terminated with beep sound.
9) Other Using Method
-Disinfection Solution Exchange Method

13
(1) Disinfection Solution Drain
①Press the disinfection solution drain button.
②Press the Start button.
(2) Disinfection Solution Supply
①Press disinfection solution supply button.
②Press the Start button.
3. Method of Storage & Management After Use
1) Management Method
(1) Turn off after use.
(2) After use, keep the cleaning tank cover open to be dried.
(3) Keep the product clean not to make any trouble for the next use.
- In case that the cleaning tank is dirty, wipe it with a neutral detergent.
- Be careful not to contact to water for the interior and operation unit of the
product in case of cleaning.
(4) Wipe the product with a dry cloth to keep dry always.
(5) Be sure to lock the water supply valve.
2) Storage method in case of not in use for a long time
(1) Store the power cord by removing it.
(2) Empty the disinfection solution in the disinfection tank and the alcohol in the
alcohol tank.
(3) For the place of storage, be careful of the following:
- Store at the place away from water.
- No effect from dust, salt and etc.
- Without slope, vibration, impact (in case of transportation) and etc.
- Do not store in a place of chemicals or where gas or others is generated.
- Temperature: -20°C to +60°C, humidity: less than 95%.
6. Customer Service(A/S)
- Measure sequence in case of product malfunction or failure
①Stop using the product and turn off the power switch.
②Close the water supply valve.
③Unplug the power cord.
④Dry the moisture in case of water leakage.
⑤Mark appropriately on the outside of product for other users and technicians can
see it.
⑥Contact an expert engineer and take the measure.
- In case of failure, do not repair by yourself, but request to a sales shops or
manufacturer.

14
Never proceed disassemble, repair or modify except the medical
equipment manufacturer
(A/S is not available for the product that is repaired and modifications except
those who involved)
May cause personal injury by ignition or abnormal operation.
7. Marking
•Product Name: Soft Endoscope Reprocessor
•Model Name: CUBE Dual
•Protective Type On Rating: Class 1 Equipment
•Power Voltage: AC 220[V], 60[Hz]
•Power Consumption: 220VA (1KAV in case of heater option use)
•Safety Device: Output Current Limit Hugh & Short Circuit Breaker
•Dimension: 480㎜(W)×620㎜(D)×960㎜(H)
•Weight: 43[㎏]
1. Product Name: Soft Endoscope Reprocessor
Type: CUBE Dual
2. Company Name of Manufacturer: ASAP Co., Ltd.
Address: Room No. 302, 37, Cheonho-daero 221-gil,
Gangdong-gu, Seoul, Republic of Korea
3. Manufacturing License: No. 4001
Product Authorization: No. 15-4039
Purpose of Use: Equipment of disinfection by circulating disinfection
solution within soft endoscope pipe.
4. Manufacturing No:
Manufacturing Date: 201 . .
Packaging Unit / Weight: 1 SET / 43KG
5. Performance & Using Method: Refer to the attached document
6. Precautions or Warning Mark in Use: Refer to the attached document
7. Storage Method: Store at room temperature
8. Electrical Rating, Protective Type & Degree
AC 220V / 60Hz / 1KAV / Class 1 Equipment
*The product is a medical equipment

15
P r o d u c t Wa r r a n t y
Product Name
Soft Endoscope Reprocessor
Model Name
CUBE Dual
Warranty
Period
1 Year
Manufacturing No
Purchasing
Date
(D/M/Y)
Product
Authorization No
15-4039 호
Customer
Purchasing Place
Name
TEL
Address
※When purchasing a product, you have to get the confirmation from the purchasing place for
future free warranty.
※Other Items:
______________________________________________________________________________________________________________
<Warranty Regulation>
The warranty period of the equipment main body is 1 year after purchasing and accessories
are excluded.
In case that failure occurs, please request the repair of product by attaching this warranty
certificate to the sales shop, which purchased the equipment, or manufacturer.
Pay attention that repair fees will also be asked even within the warranty period as followings:
- Failure due to careless handling
- Failure caused by product modification or repair not by A/S point.
- Failure and damage caused by force majeure due to fire, earthquake, and flood disasters.
- In case that the cause of failure is due to other matters except the product.
- In case that the content is not written on the warranty certificate
※The above regulation is applied for domestic use.
※The regulation does not restrict customer rights.
※The shipping and other expenses incurred in repairing the goods is supposed to bear by
the customer.

-1-
ENDOSCOPE REPROCESSOR
[CUBE LITE Manual]
Using Method

-2-
1. Preparation Items Before Use
1) Check if there is any damage to the product in case of warehousing.
2) Installation Location
①Install in a well-ventilated place that there is no moisture and
exposure to direct sunlight.
②Install away from heat generating objects.
③Install on a hard and flat floor.
3) Product Installation
①Connect with water supply hose between main body and faucet.
②Connect the end of drain hose to the water drain outlet.
③Connect the power cord exactly to AC 220V (60Hz) single-phase
ground outlet.
4) Be sure to read the manual before using the product, then use it after learning how to
use.
5) In case of using alcohol, supply an appropriate amount of alcohol into the inlet
of alcohol.
6) Supply appropriate amount of the disinfection solution tank.
(For more information, refer to “10) Other Using Method of 2. Using Method &
Operation Sequence”)
2. Using Method & Operation Sequence
< Operation Unit >
1) Turn on the power switch.
2) The inspection completed endoscope must be fastened by a water-resistant cap, roll,
and put it into the cleaning tank
3) Perform the water leak test of soft endoscope.
Disinfection Solution
Used Times
LCD
Display Unit
Warning
Indicator
Progress
Indicator
Selection
Display Unit
Used Time

-3-
①The purpose of the scope leak test
It has a purpose to protect endoscopy by detecting damage of endoscopy in
advance. If air bubbles appear continuously during endoscopy, do not proceed
cleaning and disinfection, and Immediately stop using it and contact an expert
technician to take an appropriate measure. If it is used in violation of this, it can
cause fatal damage to the endoscope.
②Scope Leak Test
A) For soft endoscope, be sure to close water-resistant cap on the electrical
connection portion. In case of closing the water-resistant cap, if there is moisture,
foreign substances inside the water-resistant cap, remove them and use it.
B) Connect the leak inspection connector of this product to the water-
resistant cap (Venting connector) (If water or foreign substances are on the
part that connects the Leakage tester's connector cap, remove and connect
it.)
C) Press the leakage test button of operation unit,
D) Then, execute the leakage test by pressing the start button of operation
unit.
E) The water supply stops when the cleaning tank starts to fill up and the
water is filled to the normal level. The user checks visually for air bubbles in the
endoscope. (At this time, turn the knob on the endoscopic handle to the left, right,
up and down, then press the switch parts once to check if air bubbles come out)
F) If there is no problem (leakage from endoscope) in the above, cleaning can
be performed.
G) If air bubbles occur continuously, do not clean and disinfect, but take the
endoscope out of the cleaning tank to be able to dry, and contact an
endoscope company to take appropriate measure.
(Request the endoscope repair.)
4) Connect the channel connection to each channel of the endoscope. (Refer to
below picture)
- Connect the channel plug to the air/water and suction with the endoscope that
the leakage test is completed. And connect the cleaning biopsy valve to the
instrument channel port.

-4-
< Olympus Endoscopic Channel Connection Example >
5) Set the process (cleaning, disinfection, air, rinsing) time in the operation Unit.
(1) Press UP or DOWN button at the time control of the operation unit to move
the cursor to the desired process (WASH, AIR, DISIN, RINS).
< Process Description >
(2) Press the Enter button.
(3) If the cursor blinks, press UP or DOWN button to set the desired time
①The cleaning process, rinsing process time can be set in unit of 1 second
from 0 second to 5 minutes
②The disinfection process can be set in unit of 1 minute from 0 min to 60
Min.
③The air process can be set in unit of 1 second 0 sec to 30 sec.
(4) Press the Enter button to save it.
6) Press the selection button to select the desired program mode
<<Program Mode>>
Cleaning
Process
Disinfection
Process
Rinsing
Process
Air Process (Supply air into endoscope channel)

-5-
< No.1 to 4 Mode: Automatic Mode>
No.1 Mode (=A-Auto Mode): Cleaning-Air-Disinfection-Air-Deposition Cleaning
(Drain + Air)-Alcohol (Supply if selected)-Air
No.2 Mode (=B-Auto Mode): Cleaning-Air-Disinfection-Air-Rinsing-Alcohol
(Supply if selected)-Air
No.3 Mode (=Simple Cleaning Mode): Disinfection-Air-Deposition Cleaning
(Drain + Air)-Alcohol (Supply if selected)-
Air
No.4 Mode (=Ultra Simple Cleaning Mode): Disinfection-Air-Rinsing-Alcohol (If
selected)-Air
* In Automatic Mode, press the alcohol selection button of operation unit to
select alcohol process, then supply alcohol to endoscopic channels for 5
seconds.
<No.5 to 8 Mode: Manual Mode>
No.5 Mode: Cleaning (=Perform the cleaning process only for the set time)
No.6 Mode: Disinfection (=Perform the disinfection process only for the set time
No.7 Mode: Air (=Perform the cleaning process only for the set time)
8번Mode: Alcohol (=Supply alcohol to endoscopic channel for 5 seconds)
7) Press the Start Button to execute it.
8) Terminated with beep sound.
9) Turn off the power switch.
10) Other Using Method
-Disinfection Solution Exchange Method
(1) Disinfection Solution Drain
①Press the disinfection solution drain button.
②Press the Start button.
(2) Disinfection Solution Supply
①Press disinfection solution supply button.
②Press the Start button.
3. Method of Storage & Management After Use
1) Management Method
(1) Turn off after use.
(2) After use, keep the cleaning tank cover open to be dried.
(3) Keep the product clean not to make any trouble for the next use.
- In case that the cleaning tank is dirty, wipe it with a neutral detergent.
- Do not spray water on the interior and operation unit of the product in
case of cleaning.
(4) Wipe the product with a dry cloth to keep dry always
(5) Be sure to lock the water supply valve as much as possible on holidays.
(6) Measure sequence in case of product malfunction or failure
①Stop using the product and turn off the power switch.
This manual suits for next models
2
Table of contents
Popular Laboratory Equipment manuals by other brands

Agilent Technologies
Agilent Technologies PlateLoc Quick reference guide

Velp Scientifica
Velp Scientifica HSC instruction manual

Gamry Instruments
Gamry Instruments Reference 30k Booster Operator's manual

ELITechGroup
ELITechGroup AEROSPRAY 7122 Applications manual

Labnet
Labnet BioPette PLUS instruction manual

Chemglass
Chemglass OptiTherm CG-1992 Quick setup guide