
MXN488 iss01 ed A page 14
14
IT IS MOST IMPORTANT THAT THE LOAD SENSING PROBE IS ALWAYS
PLACED IN THE LOAD .OR STERILIZING WILL NOT TAKE PLACE!
USE OF "LOAD SIMULATORS" WITH LOAD SENSED STERILIZE TIMING
These may be used in front-loading models ; If it is possible to restrict the LOAD Sensed Timing to
one type of load, then a Simulated Load can be arranged , to mimic the behaviour of the real load, but
with the advantage that the LOAD SENSING PROBE can be left fixed in place in the simulator .
LOAD SIMULATOR SENSING IN PLACE OF "TRUE" LOAD SENSING
The basic Load Simulator comprises a short Stainless shelf or basket that fits at the rear of the
chamber on FRONT LOADING models. This supports a sample bottle or container (as used in the
load) which is filled generally with Liquid Paraffin, Glycol, Silicone oil, or any other non-volatile liquid
with a similar specific heat to the load material. The Load Sensing Probe is positioned in the container
to sense the temperature of the simulator, and give an approximation to the load temperature.
NOTE;- The LOAD SIMULATOR is an additional optional item and should not be confused with the
DUMMY LOAD which has a different function and. This is fitted as standard to ensure safest operation
of the COOLING LOCK for all likely load types and affects the Temperature Sensor that operates the
PRESET Cooling Lock.
STERILIZING TIMING "DROPBACK" WITH LOAD SENSING SELECTED
Refer to "Dropback" on page 13 The Dropback Function operates if the LOAD TEMPERATURE drops
2 C below Sterilizing Temp., once sterilizing has begun and resets the Sterilizing timer.
6. COOLING SYSTEM:
The Cooling System is supplied in two versions- Basic Cooling and Deluge Cooling.
Basic cooling uses circulated cold water in a chamber coil or jacket. Deluge Cooling circulates the
chamber water through a pump and heat-exchanger and sprays it over the load. Cooling operates with
Air Ballasting.
7. COOLING LOCKS
The Cooling Locks prevent the STERILIZER opening under unsafe conditions.
There are two cooling lock sensors.
PRESET COOLING LOCK
This is a thermostatic device with Dial & scale.
On VERTICAL models it is mounted on the side of the cabinet behind a cover panel,
On HORIZONTAL models it is at the rear behind a cover plate secured by two screws, or on some
models behind a drop-down top section of the rear panel. It is calibrated in Deg.C(+/- 5C), and
measures the temperature at the outside of the chamber wall.
The Preset cooling lock responds to the temperature of the chaamber wall which acts as a thermal
mass. & cools in a similar way to the real load . The temperature in the chamber wall and real loads will
not be the same during cooling, due to differing thermal masses and Thermal inertia.
PROGRAMMABLE COOLING LOCK.
When LOAD SENSING option is NOT selected. The Release Temperature threshold is set within the
Program and is compared with the CHAMBER TEMPERATURE SENSOR. When the CHAMBER
temperature is above this threshold the lock is active (engaged) and Completion is inhibited stopping
the door from unlocking.. As the CHAMBER sensor detects the temperature within the steam space it
will always Heat or Cool faster than the load or the "Chamber Wall”.
LOAD SENSED COOLING LOCK ( LOAD SENSING OPTION fitted & selected )
In this case the Programmable Cooling Lock Temperature set for the Program is compared with the
Temperature measured by the LOAD SENSING PROBE. When the sensed Load temperature is
above this threshold the Cooling-lock is active (engaged) and Completion is inhibited , so the door
cannot be unlocked. As the LOAD SENSING PROBE detects the temperature within the LOAD itself ,
it will always represent the actual load temperature.
DOUBLE COOLING LOCKS
The PRESET cooling lock , (sensing the Chamber Wall ) is always operative , so the temperature of
both of ACTUAL LOAD and DUMMY LOAD must be below the set thresholds to allow the cycle to
complete.
PDF created with FinePrint pdfFactory Pro trial version http://www.fineprint.com