ATAGO 3850 User manual

Refractive Index
The displayed Refractive Index value is not compensated for temperature. The measured value is the Refractive Index at the
displayed measurement temperature. Changes in temperature directly affect the Refractive Index of any sample (even for the
same sample at the same concentration level).
Storage and Maintenance
Store the instrument in a dry place away from
direct sunlight. Exposure to humidity and heat
may damage the instrument.
Clean and dry the sample stage thoroughly
following the “Cleaning” instructions.
Store the unit away from direct sunlight at a
stable temperature with as little fluctuation as
possible.
Repair and Warranty
The instrument is warranted for one year from the date of purchase. This warranty is void if the instrument shows evidence of the
following. Send the included batteries as well if they are still in use.
•
Having been disassembled by unauthorized personnel
•
Damages to the prism and/or sample stage
•
Water damage or having been dropped
•
Having been misused and/or operated outside the environmental specifications
•
Leakage from batteries other than those included with the unit
Repair services are available for a fee after the warranty expires.
Contact an ATAGO authorized service center for service and support.
Please have the serial number information ready when contacting a service center.
Specifications
Measurement range Refractive index 1.3306 to 1.5284
Maximum number of 100
Temperature 5.0 to 45.0
data history
Resolution Refractive index 0.0001
Output NFC Forum Type 4 Tag
Temperature 0.1
ISO/IEC 14443 Type A
Accuracy Refractive index ±0.0003 (water at 20 )
Output category : Date Time, RI, Temp [degC]
Temperature ±1
(e.g.) 2017/08/17 09:30:45, 1.3425, 20.4
Automatic temperature
10 to 40
Power supply Two (2) AAA alkaline batteries
compensation range
Battery life Approx. 11,000 measurements
Sample volume At least 0.3mL
(when using alkaline batteries)
Measurement time Approx. 3 seconds
International IP65
Backlight The backlight stays on for 30 seconds
Protection class
after any button is pressed.
Dimensions and weight
55 (W) x 31 (D) x 109 (H)mm, 100g (main unit only)
The product is in conformity with the requirements of the EMC Directive 2004/108/EC.
Patent Granted in Japan United States Germany China and Taiwan.
Headquarters: The Front Tower Shiba Koen,
23rd Floor 2-6-3 Shiba-koen, Minato-ku,
Tokyo 105-0011, Japan
TEL: 81-3-3431-1943 overseas@atago.net
http://www.atago.net/
TEL: 1-425-637-2107
[email protected]
TEL: 91-22-2854-4915 / 4071-3232
[email protected]
TEL: 66-21948727-9 ,66-21171549
[email protected]
TEL: 55 16 3913-8400
[email protected]
TEL: 39 02 36557267
[email protected]
TEL: 86-20-38108256
[email protected]
TEL: 7-812-777-96-96
[email protected]
TEL: 234-707-558-1552
[email protected]
TEL: 7-727-257-08-95
[email protected]
1803K Printed in Japan
Digital Hand-held “Pocket” Refractive index Refractometer
PAL-RI
Cat.No.3850
3850-E12
Instruction Manual
Parts
Contents
Main unit.........1
Instruction Manual (this book) .....1 Calibration Report .. 1
 batteries ...........2
AAA alkaline batteries are included. Remove the tape from the battery compartment before inserting the batteries.
ATAGO instruments are rigorously inspected to ensure each unit meets the highest standards of quality assurance.
Introduction
Thank you for purchasing the instrument. Carefully read and follow all instructions. Keep this manual for future reference.
Safety Instructions
Read and follow all safety instructions before operating the instrument. Failure to comply with the following instructions may result in personal injury or
property damage.
◇ Ensure safety when handling hazardous materials. Observe precautionary measures and use protective equipment, Be aware of the hazards of such
chemicals and emergency response guidelines.
◇ ATAGO may not be held liable for any injury or damage arising in connection with handling of hazardous materials during the use of the instrument.
◇ Do not drop the instrument or subject it to strong physical shock.
◇ Do not attempt to repair, modify, or disassemble the instrument.
◇ Carefully read this manual to have basic knowledge of the function of each component.
◇ ATAGO is not liable for any loss and damage caused by the measurement and use of this instrument.
◇ Some acids may corrode the glass prism and/or metal sample stage, which may cause erroneous measurements.
◇ Do not use metal tools, such as a spoon, as they may scratch the prism, resulting in erroneous measurements.
◇ Do not use water above 50 to rinse the instrument.
◇ Only use the specified battery type. Observe proper polarities, properly aligning the anodes and cathodes.
◇ Store the instrument away from direct sunlight/heat sources and excessive amounts of dust/debris.
◇ Do not expose the instrument to a rapid change in ambient temperature.
◇ Do not subject the instrument to strong vibration.
◇ Do not subject the instrument to extreme cold temperature.
◇ Do not place the instrument under anything heavy.
◇ Loosen the battery compartment cover for air transportation.
<International Protection Classification IP65>
◇ The instrument is water-resistant, not waterproof, and should not be submerged.
<Che ical Resistance of Body Case>
◇ The body case is made of PBT resin. Do not expose it to water vapor. Some solvents may compromise the structural integrity of the instrument.
Press to take
measurements and
hold down to turn off
the display.
ST RT button
(Power button)
Place and remove
batteries from here.
Battery
compartment
ZERO button
Press to perform
zero-setting.
ST RT button &
ZERO button
Press to set date time
and delete data history.
The warranty period
extension method
1 year
2 years
The warranty period will
be extended from year
to 2 years when you
register customer
information.
ATAGO Logger NFC
can also be downloaded at the same
time.
Trouble scanning the code? Access
this link
http://www.atago.net/ur/
QR code
Lanyard hole
Sample stage
Apply water and
samples on the glass
prism located in the
center of the sample
stage.
Image is for explanation purposes only. It may be
different than the actual product purchased.
LCD
Measurement results
prism temperature
remaining battery
charge etc. are
displayed.
The displayed value
is an example.
Zero-setting and Measure ent
Zero-setting
【Caution】
◇ Zero-set the instrument at the beginning of each day before use as well as after replacing the batteries.
◇ Let water on the prism acclimate to the temperature of the instrument before zero-setting.
◇ When “AAA” is displayed wipe the prism clean apply water and press the ZERO button again.
4.
Zero-setting is
complete when “000” blinks
twice and then stops.
1.
Wipe the prism clean and
apply about 0.3mL of tap water.
2.
Press the START
button and after “---”
appears the measurement
and the prism temperature
are displayed.
3.
Press the ZERO button
while the water is on the
prism.
Proceed to measurement.
Wipe the prism clean with
tissue paper.
Measurement
【Caution】
◇ Do not use metal tools to apply samples on the prism as they may scratch the prism.
◇ Initial measurements may fluctuate with hot or cold samples. Wait for the instrument to acclimate to the sample temperature
approximately 20 seconds to press the START button. Alternatively press the START button multiple times until
measurements become stable.
◇ Do not splash water above 50℃. The plastic may warp which may compromise the water resistance.
•When measuring hot samples place only the necessary amount and do not let it overflow from the sample stage well.
•When hot water is necessary to clean off hardened samples use water-soaked gauze around the prism area and keep hot
water away from the body case.
◇ The displayed temperature is that of the prism and may not necessarily match the temperature of the sample.
1.
Wipe the prism clean and
apply about 0.3mL of the sample.
2.
Press the START
button
3.
After “---”
appears the
measurement and the prism
temperature are displayed.
<LCD Auto Shut-off>
The instrument will turn itself off after 2 minutes of inactivity. To manually turn it off hold down the
START button for more than 2 seconds.
<For oily/fatty sa ples>
Try stirring the sample on the prism while measuring to improve the repeatability of oily/fatty samples.
Cleaning
【Caution】
◇ Do not scratch the prism.
◇ The instrument is water-resistant not waterproof and should not be submerged.
1. Wipe off the sample.
2. Clean the prism and sample stage using a mild soap and thoroughly rinse with water.
3. Dry the area with tissues thoroughly.
Error Messages
The following messages alert the user when an operation has failed.
The battery is low.
The prism temperature is below the temperature
range.
The ZERO button was pressed with nothing or
something other than water on the prism.
The prism temperature is above the temperature
range.
The START button was pressed with nothing or an
insufficient amount of sample on the prism.
The sample measured outside the measurement
range.
The instrument is faulty.
(Replace the batteries. Contact ATAGO if this
error persists.)
Too much light is entering the prism and the
instrument cannot measure accurately.
(Shade the sample stage with your hand and take a
measurement again.)
Replacing the Batteries
【Caution】
◇
Fasten the battery compartment cover tightly to prevent water ingress or poor connection which will
cause erroneous measurements. Push the cover in firmly and turn.
◇ When the O-ring on the cover is dirty or damaged the water resistance may be compromised.
(Figure 1)
◇ When the battery icon indicates the low power level (
) replace both batteries with a brand new
set of AAA alkaline batteries (1.5V).
◇ Static images may occasionally appear on LCD. Such retained pixel charges do not indicate a faulty
display consume the battery power or affect the instrument’s performance in any way.
◇ Check the expiration dates on batteries before purchase.
◇ Zero-set the instrument after the batteries are replaced.
1.
Insert a coin in the
groove on the battery
compartment cover. Turn
the coin counterclockwise
to remove the cover.
2.
Insert batteries
observing the correct
polarity.
3.
Align the cover and push
it down.
4.
Close the battery
compartment cover by
pushing the cover in with a
coin in the groove and turning
it clockwise until it stops.
*Turning excessively may cause
malfunction.
Approx. 0.3mL of water
Approx.
2mm
Groove
Approx.
2mm
Approx. 0.3mL of sample
Figure1
O-ring
Mild
soap

This manual suits for next models

1

Other ATAGO Measuring Instrument manuals

ATAGO PAL-HIKARi 5 User manual

ATAGO

ATAGO PAL-HIKARi 5 User manual

ATAGO PAL-1 BLT/ i User manual

ATAGO

ATAGO PAL-1 BLT/ i User manual

ATAGO PAL-121S User manual

ATAGO

ATAGO PAL-121S User manual

ATAGO 4439 User manual

ATAGO

ATAGO 4439 User manual

ATAGO MASTER Series User manual

ATAGO

ATAGO MASTER Series User manual

ATAGO PAL-31S User manual

ATAGO

ATAGO PAL-31S User manual

ATAGO PAL-36S User manual

ATAGO

ATAGO PAL-36S User manual

ATAGO PAL-91S User manual

ATAGO

ATAGO PAL-91S User manual

ATAGO PAL-43S User manual

ATAGO

ATAGO PAL-43S User manual

ATAGO RX-5000 i-Pius User manual

ATAGO

ATAGO RX-5000 i-Pius User manual

ATAGO PAL-HIKARi 2 User manual

ATAGO

ATAGO PAL-HIKARi 2 User manual

ATAGO PAL-Easy ACID9 User manual

ATAGO

ATAGO PAL-Easy ACID9 User manual

ATAGO PAL-67S User manual

ATAGO

ATAGO PAL-67S User manual

ATAGO PAL-COFFEE(BX/TDS) User manual

ATAGO

ATAGO PAL-COFFEE(BX/TDS) User manual

ATAGO PAL-04S User manual

ATAGO

ATAGO PAL-04S User manual

ATAGO PAL-87S User manual

ATAGO

ATAGO PAL-87S User manual

ATAGO PAL-102S User manual

ATAGO

ATAGO PAL-102S User manual

ATAGO PAL-92S User manual

ATAGO

ATAGO PAL-92S User manual

ATAGO PAL-03CS User manual

ATAGO

ATAGO PAL-03CS User manual

ATAGO PAL-Fish Fillets User manual

ATAGO

ATAGO PAL-Fish Fillets User manual

ATAGO PAL-89S User manual

ATAGO

ATAGO PAL-89S User manual

ATAGO 5468 User manual

ATAGO

ATAGO 5468 User manual

ATAGO PAL-Grape Must KMW User manual

ATAGO

ATAGO PAL-Grape Must KMW User manual

ATAGO PR-50HO User manual

ATAGO

ATAGO PR-50HO User manual

Popular Measuring Instrument manuals by other brands

Dräger Polytron 7500 Instructions for use

Dräger

Dräger Polytron 7500 Instructions for use

horiba LAQUA F-71G instruction manual

horiba

horiba LAQUA F-71G instruction manual

Mark-10 EK3-100 quick start guide

Mark-10

Mark-10 EK3-100 quick start guide

MRC PVB-820 Operation manual

MRC

MRC PVB-820 Operation manual

Kreg Precision KMS7101 Instructional manual

Kreg

Kreg Precision KMS7101 Instructional manual

Ametek JOFRA CTC-140 A user manual

Ametek

Ametek JOFRA CTC-140 A user manual

Kipp & Zonen SMP series instruction manual

Kipp & Zonen

Kipp & Zonen SMP series instruction manual

Perel EVM125 quick start guide

Perel

Perel EVM125 quick start guide

Nexus 1500+ quick start guide

Nexus

Nexus 1500+ quick start guide

Zeiss GOM Scan 1 User manual hardware

Zeiss

Zeiss GOM Scan 1 User manual hardware

Tektronix 1L40 instruction manual

Tektronix

Tektronix 1L40 instruction manual

Vacom BARION atm II manual

Vacom

Vacom BARION atm II manual

SOCTESTER SOC AERO operating manual

SOCTESTER

SOCTESTER SOC AERO operating manual

SenseAir aSENSE mIII user manual

SenseAir

SenseAir aSENSE mIII user manual

Extech Instruments EX622 user guide

Extech Instruments

Extech Instruments EX622 user guide

Tegam R1L-B instruction manual

Tegam

Tegam R1L-B instruction manual

Sinmag Europe SE42 manual

Sinmag Europe

Sinmag Europe SE42 manual

Keithley SourceMeter 2400 user manual

Keithley

Keithley SourceMeter 2400 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.