ATAGO PAL-98S User manual

Condiment Concentration Percentage
The Condiment Concentration Percentage represents the total concentration of all soluble solids within a water solution.
Soluble solids include elements such as sugar, salts, minerals and proteins. When measuring a pure sucrose in water
solution, the measured value displayed directly represents the actual concentration percentage.
When a solution contains multiple substances other than or including sugar, a conversion chart may be needed to
determine the percent concentration of the sample for the element it is being measured for.
Automatic Temperature Compensation
The readings are corrected, based on the temperature of the prism, within the automatic temperature compensation range.
【Caution】
◇ Measurements may fluctuate with hot or cold samples. Wait for approximately 20 seconds to press the START button.
Measurements will stabilize once the instrument acclimates to the sample temperature.
Storage and Maintenance
Store the instrument in a dry place
away from direct sunlight. Exposure to
humidity and heat may damage the
instrument.
Clean and dry the sample stage thoroughly,
following the “Cleaning” instructions.
Store the unit away from direct sunlight at a
stable temperature with as little fluctuation as
possible.
Repair and Warranty
The instrument is warranted for one year from the date of purchase. This warranty is void if the instrument shows
evidence of the following. Send the included batteries as well if they are still in use.
•
Having been disassembled by unauthorized personnel
•Damages to the prism and/or sample stage
•
Water damage or having been dropped
•
Having been misused and/or operated outside the environmental
specifications
•Leakage from batteries other than those included with the unit
Repair services are available for a fee after the warranty expires.
Contact an ATAGO authorized service center for service and support.
Please have the serial number information ready when contacting a service center.
Specifications
Measurement range Condiment Concentration 0.0 to 70.0%
10.0 to 100˚C
Maximum number of
data history
100
Resolution Condiment Concentration 0.1%
Temperature 0.1˚C
Output NFC Forum Type 4 Tag
ISO/IEC 14443 Type A
A
ccurac
y
Condiment Concentration ±0.2%
Temperature ±1˚C
Output category
Date Time, Condiment [%] ,Temp [degC]
A
utomatic temperature 10 to 100˚C (e.g.) 2017/08/17 09:30:45, 12.3, 20.4
compensation range Power supply Two (2) AAA alkaline batteries
A
mbient temperature 10 to 40˚C Battery life
A
pprox. 11,000 measurements
range (when using alkaline batteries)
Sample volume
A
t least 0.3mL International Protection
class
IP65
Measurement time
A
pprox. 3 seconds Dimensions and weight 55 (W) x 31 (D) x 109 (H)mm,
100g (main unit only)
Backlight The backlight stays on for 30 seconds
after any button is pressed.
The product is in conformity with the requirements of the EMC Directive 2004/108/EC.
Patent Granted in Japan, United States, Germany, China and Taiwan.
Headquarters: The Front Tower Shiba Koen,
23rd Floor 2-6-3 Shiba-koen, Minato-ku,
Tokyo 105-0011, Japan
TEL: 81-3-3431-1943
overseas@atago.net
http://www.atago.net/
TEL: 1-425-637-2107
customerservice@atago-usa.com
TEL: 91-22-2854-4915 / 4071-3232
customerservice@atago-india.com
TEL: 66-21948727-9 ,66-21171549
customerservice@atago-thailand.com
TEL: 55 16 3913-8400
customerservice@atago-brasil.com
TEL: 39 02 36557267
customerservice@atago-italia.com
TEL: 86-20-38108256
info@atago-china.com
TEL: 7-812-777-96-96
[email protected]
TEL: 234-707-558-1552
atagonigeria@atago.net
TEL: 7-727-257-08-95
info@atago-kazakhstan.com
2107K Printed in Japan
4498-E12
Digital Hand-held “Pocket” Condiment Refractometer
Instruction Manual
PAL-98S Cat.No. 4498
Parts
Contents
Main unit..... 1 Instruction Manual (this book) .1 Calibration Report 1 AAA batteries....... 2
AAA alkaline batteries are included. Remove the tape from the battery compartment before inserting the batteries.
ATAGO instruments are rigorously inspected to ensure each unit meets the highest standards of quality assurance.
Introduction
Thank you for purchasing the instrument. Carefully read and follow all instructions. Keep this manual for future reference.
Safety Instructions
Read and follow all safety instructions before operating the instrument. Failure to comply with the following instructions may result in personal
injury or property damage.
◇ Ensure safety when handling hazardous materials. Observe precautionary measures and use protective equipment, Be aware of the
hazards of such chemicals and emergency response guidelines.
◇ ATAGO may not be held liable for any injury or damage arising in connection with handling of hazardous materials during the use of the
instrument.
◇ Do not drop the instrument or subject it to strong physical shock.
◇ Do not attempt to repair, modify, or disassemble the instrument.
◇ Carefully read this manual to have basic knowledge of the function of each component.
◇ ATAGO is not liable for any loss and damage caused by the measurement and use of this instrument.
◇ Some acids may corrode the glass prism and/or metal sample stage, which may cause erroneous measurements.
◇ Do not use metal tools, such as a spoon, as they may scratch the prism, resulting in erroneous measurements.
◇ Do not use water above 50˚C to rinse the instrument.
◇ Only use the specified battery type. Observe proper polarities, properly aligning the anodes and cathodes.
◇ Store the instrument away from direct sunlight/heat sources and excessive amounts of dust/debris.
◇ Do not expose the instrument to a rapid change in ambient temperature.
◇ Do not subject the instrument to strong vibration.
◇ Do not subject the instrument to extreme cold temperature.
◇ Do not place the instrument under anything heavy.
◇ Loosen the battery compartment cover for air transportation.
<International Protection Classification IP65>
◇ The instrument is water-resistant, not waterproof, and should not be submerged.
<Chemical Resistance of Body Case>
◇ The body case is made of resin. Do not expose it to water vapor. Some solvents may compromise the structural integrity of the instrument.
Press to take
measurements and
hold down to turn
off the display.
START button
(POWER botton)
Place and remove
batteries from here.
Battery
compartment
ZERO button
Press to perform
zero-setting.
START button &
ZERO button
Press to set date, time,
and delete data history.
Lanyard hole
Sample stage
A
pply water and
samples on the glass
prism located in the
center of the sample
stage.
Image is for explanation purposes only. It ma
y
be
different than the actual product purchased.
LCD
Measurement
results, prism
temperature,
remaining battery
charge, etc., are
displayed.
The displayed value
is an example.
The warranty period will
be extended from 1 year
to 2 years when you
register customer
information.
ATAGO Logger NFC
can also be downloaded at the same
time.
Trouble scanning the code? Access
this link
https://www.atago.net/ur/
index.php?l=en
QR code
The warranty period
extension method
1 year 2 years
Zero-setting and Measurement
Zero-setting
【Caution】
◇ Zero-set the instrument at the beginning of each day before use as well as after replacing the batteries.
◇ Let water on the prism acclimate to the temperature of the instrument before zero-setting.
◇ When “AAA” is displayed, wipe the prism clean, apply water, and press the ZERO button again.
1.Wipe the prism clean and
apply about 0.3mL of tap
water.
2.Press the START
button, and after “---”
appears, the
measurement and the
prism temperature are
displayed.
3. 4. Zero-setting is
complete when “000”
blinks twice and then
stops.
Measurement : not 0.0%
Press the ZERO
button while the
water is on the
prism.
Measurement : 0.0% Proceed to measurement.
Wipe the prism clean with
tissue paper.
The instrument is zero-set
correctly.
Measurement
【Caution】
◇ Do not use metal tools to apply samples on the prism as they may scratch the prism.
◇ Initial measurements may fluctuate with hot or cold samples. Wait for the instrument to acclimate to the sample
temperature, approximately 20 seconds, to press the START button. Alternatively, press the START button multiple
times until measurements become stable.
◇ Do not splash water above 50˚C. The plastic may warp, which may compromise the water resistance.
•When measuring hot samples, place only the necessary amount and do not let it overflow from the sample stage well.
•When hot water is necessary to clean off hardened samples, use water-soaked gauze around the prism area and
keep hot water away from the body case.
◇ The displayed temperature is that of the prism and may not necessarily match the temperature of the sample.
1.Wipe the prism clean and
apply about 0.3mL of the
sample.
2.Press the START
button
3.After “---” appears, the
measurement and the
prism temperature are
displayed.
<LCD Auto Shut-off>
The instrument will turn itself off after 2 minutes of inactivity. To manually turn it off, hold down the START button for more
than 2 seconds.
<For oily/fatty samples>
Try stirring the sample on the prism while measuring to improve the repeatability of oily/fatty
samples.
Cleaning
【Caution】
◇ Do not scratch the prism.
◇ The instrument is water-resistant, not waterproof, and should not be submerged.
1. Wipe off the sample.
2. Clean the prism and sample stage using a mild soap and thoroughly rinse with water.
3. Dry the area with tissues thoroughly.
Error Messages
The following messages alert the user when an operation has failed.
The battery is low. The prism temperature is below the
temperature range.
The ZERO button was pressed with nothing or
something other than water on the prism.
The prism temperature is above the
temperature range.
The START button was pressed with nothing or
an insufficient amount of sample on the prism.
The instrument is faulty.
(Replace the batteries. Contact ATAGO if this
error persists.)
The sample measured outside the
measurement range.
Too much light is entering the prism, and the
instrument cannot measure accurately.
(Shade the sample stage with your hand and
take a measurement again.)
Replacing the Batteries
【Caution】
◇ Fasten the battery compartment cover tightly to prevent water ingress or poor connection,
which will cause erroneous measurements. Push the cover in firmly and turn.
◇ When the O-ring on the cover is dirty or damaged, the water resistance may be compromised.
(Figure 1)
◇ When the battery icon indicates the low power level ( ), replace both batteries with a
brand new set of AAA alkaline batteries (1.5V).
◇ Static images may occasionally appear on LCD. Such retained pixel charges do not indicate a
faulty display, consume the battery power, or affect the instrument’s performance in any way.
◇ Check the expiration dates on batteries before purchase.
◇ Zero-set the instrument after the batteries are replaced.
1.Insert a coin in the
groove on the battery
compartment cover. Turn
the coin counterclockwise
to remove the cover.
2.Insert batteries,
observing the correct
polarity.
3.Align the cover and
push it down.
4.Close the battery
compartment cover by
pushing the cover in with a
coin in the groove and
turning it clockwise until it
stops.
*Turning excessively may cause
malfunction.
Groove
Approx.
2mm
Approx. 0.3mL of sample
Approx. 0.3mL of water
Approx.
2mm
Figure 1
O-ring
Mild
soap

This manual suits for next models

1

Other ATAGO Measuring Instrument manuals

ATAGO PAL-HIKARi 2 User manual

ATAGO

ATAGO PAL-HIKARi 2 User manual

ATAGO PAL-RI User manual

ATAGO

ATAGO PAL-RI User manual

ATAGO PAL-10S User manual

ATAGO

ATAGO PAL-10S User manual

ATAGO PAL-Easy ACID15 User manual

ATAGO

ATAGO PAL-Easy ACID15 User manual

ATAGO PAL-Fish Fillets User manual

ATAGO

ATAGO PAL-Fish Fillets User manual

ATAGO PALETTE Series User manual

ATAGO

ATAGO PALETTE Series User manual

ATAGO RePo-1 User manual

ATAGO

ATAGO RePo-1 User manual

ATAGO PAL-15S User manual

ATAGO

ATAGO PAL-15S User manual

ATAGO PAL- HIKARi 51 User manual

ATAGO

ATAGO PAL- HIKARi 51 User manual

ATAGO PAL-12S User manual

ATAGO

ATAGO PAL-12S User manual

ATAGO PAL-Easy ACID16 User manual

ATAGO

ATAGO PAL-Easy ACID16 User manual

ATAGO PAL-121S User manual

ATAGO

ATAGO PAL-121S User manual

ATAGO 4440 User manual

ATAGO

ATAGO 4440 User manual

ATAGO RX-7000a Manual

ATAGO

ATAGO RX-7000a Manual

ATAGO Pen - Urine S.G User manual

ATAGO

ATAGO Pen - Urine S.G User manual

ATAGO PAL-2 User manual

ATAGO

ATAGO PAL-2 User manual

ATAGO PAL-102S User manual

ATAGO

ATAGO PAL-102S User manual

ATAGO PAL-79S User manual

ATAGO

ATAGO PAL-79S User manual

ATAGO PAL-84S User manual

ATAGO

ATAGO PAL-84S User manual

ATAGO RX-5000i User manual

ATAGO

ATAGO RX-5000i User manual

ATAGO PAL-23S User manual

ATAGO

ATAGO PAL-23S User manual

ATAGO PAL-OWE2 User manual

ATAGO

ATAGO PAL-OWE2 User manual

ATAGO PAL-26S User manual

ATAGO

ATAGO PAL-26S User manual

ATAGO PAL-3 User manual

ATAGO

ATAGO PAL-3 User manual

Popular Measuring Instrument manuals by other brands

Bender ISOMETER isoNAV685-D-B manual

Bender

Bender ISOMETER isoNAV685-D-B manual

EA Technology UltraTEV Plus2 operating manual

EA Technology

EA Technology UltraTEV Plus2 operating manual

Olson Technology CTG-2 System reference manual

Olson Technology

Olson Technology CTG-2 System reference manual

Amprobe SM70 Operaing instructions

Amprobe

Amprobe SM70 Operaing instructions

Dakota Digital Odyssey II Series quick start guide

Dakota Digital

Dakota Digital Odyssey II Series quick start guide

Bushnell 1K owner's guide

Bushnell

Bushnell 1K owner's guide

WOBIT ADT42 user manual

WOBIT

WOBIT ADT42 user manual

Berger & Schroter LW800 PRO operating instructions

Berger & Schroter

Berger & Schroter LW800 PRO operating instructions

forlong PM835 user manual

forlong

forlong PM835 user manual

Pioneer SGY-PM90 quick guide

Pioneer

Pioneer SGY-PM90 quick guide

Milwaukee MC510 user manual

Milwaukee

Milwaukee MC510 user manual

DANIEL 1500 Series user manual

DANIEL

DANIEL 1500 Series user manual

TPS Aqua-pH manual

TPS

TPS Aqua-pH manual

Benning IT 130 operating manual

Benning

Benning IT 130 operating manual

GESTRA NRG 16-38 S Original Installation Instructions

GESTRA

GESTRA NRG 16-38 S Original Installation Instructions

PCB Piezotronics 138A10 Installation and operating manual

PCB Piezotronics

PCB Piezotronics 138A10 Installation and operating manual

Ozito ODSM-130D instruction manual

Ozito

Ozito ODSM-130D instruction manual

BRUEL & KJAER 2107 Instructions and applications

BRUEL & KJAER

BRUEL & KJAER 2107 Instructions and applications

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.