ATAGO PAL-BX/RI User manual

Main scale : Brix
Secondary scale : Refractive Index
Brix represents the weight of sucrose in 100 grams of sucrose
solution as percentage by weight. When other dissolved solids
are present in the solution, Brix conversion may be applied.
Brix is a me
asure of the total dissolved solids in a solution and
indicates the combined concentration of all soluble substances,
such as sugar, salt, protein, and acids.
The displayed Refractive ndex value is not compensated for
temperature. The measured value is t
he Refractive ndex at the
displayed measurement temperature. Changes in temperature
directly affect the Refractive ndex of any sample (even for the
same sample at the same concentration level).
Automatic Temperature Compensation
The readings are corrected, based on the temperature of the prism, within the automatic temperature compensation range.
【Caution】
◇ Measurements may fluctuate with hot or cold samples. Wait for approximately 20 seconds to press the START button.
Measurements will stabilize once the instrument acclimates to the sample temperature.
Storage and Maintenance
Store the instrument in a dry place away from
direct sunlight. Exposure to humidity and heat
may damage the instrument.
Clean and dry the sample stage thoroughly,
following the “Cleaning” instructions.
Store the unit away from direct sunlight at a
stable temperature with
as little fluctuation as
possible.
Repair and Warranty
The instrument is warranted for one year from the date of purchase. This warranty is void if the instrument shows evidence of the
following. Send the included batteries as well if they are still in use.
•
Having been disassembled by unauthorized personnel
•
Damages to the prism and/or sample stage
•
Water damage or having been dropped
•
Having been misused and/or operated outside the environmental specifications
•
Leakage from batteries other than those included with the unit
Repair services are available for a fee after the warranty expires.s
Contact an ATAGO authorized service center for service and support.
Please have the serial number information ready when contacting a service center.
Specifications
Measurement range Brix : 0.0 to 93.0% (10.0 to 100℃)
Maximum number of
100
Refractive In ex (RI) : 1.3306 to 1.5284 (5.0 to 45.0℃)
ata history
Resolution Brix : 0.1% Temperature : 0.1℃
Output NFC Forum Type 4 Tag
Refractive In ex : 0.0001
ISO/IEC 14443 Type A
Accuracy Brix :±0.1% Temperature : ±1℃
Output category :
Refractive In ex :±0.0003 (Water at 20℃)
Date Time,Brix [%], RI, Temp [ egC]
Automatic temperature
Brix : 10 to 100℃
(e.g.) 2017/08/17 09:30:45, 12.3, 1.3425, 20.4
compensation range
Power supply Two (2) AAA alkaline batteries
Ambient temperature 10 to 40℃
Battery life Approx. 11,000 measurements
range
(when using alkaline batteries)
Sample volume At least 0.3mL
International IP65
Measurement time Approx. 3 secon s
Protection class
Backlight The backlight stays on for 30
Dimensions 55 (W) x 31 (D) x 109 (H)mm,
secon s after any button is presse .
an weight 100g (main unit only)
The product is in conformity with the requirements of the EMC Directive 2004/108/EC.
Patent Granted in Japan, United States, Germany, China and Taiwan.
Hea quarters: The Front Tower Shiba Koen,
23r Floor 2-6-3 Shiba-koen, Minato-ku,
Tokyo 105-0011, Japan
TEL: 81-3-3431-1943 overseas@atago.net
http://www.atago.net/
TEL: 1-425-637-2107
[email protected]
TEL: 91-22-2854-4915 / 4071-3232
customerservice@atago-in ia.com
TEL: 66-21948727-9 ,66-21171549
customerservice@atago-thailan .com
TEL: 55 16 3913-8400
[email protected]
TEL: 39 02 36557267
[email protected]
TEL: 86-20-38108256
[email protected]
TEL: 7-812-777-96-96
[email protected]
TEL: 234-707-558-1552
[email protected]
TEL: 7-727-257-08-95
[email protected]
1803K Printed in Japan
Digital Hand-held “Pocket” Dual scale (BX/RI) Refractometer
PAL-BX/RI
Cat.No.3852
3852-E08
Instruction Manual
Parts
Contents
Main unit.........1
Instruction Manual (this book) .....1 Calibration Report .. 1
 batteries ...........2
AAA alkaline batteries are included. Remove the tape from the battery compartment before inserting the batteries.
ATAGO instruments are rigorously inspected to ensure each unit meets the highest standards of quality assurance.
Introduction
Thank you for purchasing the instrument. Carefully rea an follow all instructions. Keep this manual for future reference.
Safety Instructions
Rea an follow all safety instructions before operating the instrument. Failure to comply with the following instructions may result in personal injury or
property amage.
◇ Ensure safety when han ling hazar ous materials. Observe precautionary measures an use protective equipment, Be aware of the hazar s of such
chemicals an emergency response gui elines.
◇ ATAGO may not be hel liable for any injury or amage arising in connection with han ling of hazar ous materials uring the use of the instrument.
◇ Do not rop the instrument or subject it to strong physical shock.
◇ Do not attempt to repair, mo ify, or isassemble the instrument.
◇ Carefully rea this manual to have basic knowle ge of the function of each component.
◇ ATAGO is not liable for any loss an amage cause by the measurement an use of this instrument.
◇ Some aci s may corro e the glass prism an /or metal sample stage, which may cause erroneous measurements.
◇ Do not use metal tools, such as a spoon, as they may scratch the prism, resulting in erroneous measurements.
◇ Do not use water above 50℃ to rinse the instrument.
◇ Only use the specifie battery type. Observe proper polarities, properly aligning the ano es an catho es.
◇ Store the instrument away from irect sunlight/heat sources an excessive amounts of ust/ ebris.
◇ Do not expose the instrument to a rapi change in ambient temperature.
◇ Do not subject the instrument to strong vibration.
◇ Do not subject the instrument to extreme col temperature.
◇ Do not place the instrument un er anything heavy.
◇ Loosen the battery compartment cover for air transportation.
International Protection Classification IP65>
◇ The instrument is water-resistant, not waterproof, an shoul not be submerge .
Chemical Resistance of Body Case>
◇ The bo y case is ma e of PBT resin. Do not expose it to water vapor. Some solvents may compromise the structural integrity of the instrument.
Press to take
measurements and
hold down to turn off
the display.
ST RT button
(Power botton)
Place and remove
batteries from here.
Battery
compartment
ZERO button
Press to perform
zero-setting.
ST RT button
&
ZERO button
Press to set date, time,
and delete data history.
The warranty period
extension method
1 year
2 years
The warranty period will
be extended from year
to 2 years when you
register customer
information.
ATAGO Logger NFC
can also be downloaded at the same
time.
Trouble scanning the code? Access
this link
http://www.atago.net/ur/
QR co e
Lanyard hole
Sample stage
Apply water and
samples on the glass
prism located in the
center of the sample
stage.
Image is for explanation purposes only. It may be
ifferent than the actual pro uct purchase .
LCD
Measurement results,
prism temperature,
remaining battery
charge, etc., are
displayed.
The displayed value
is an example.
Zero-setting and Measurement
Zero-setting
【Caution】
◇ Zero-set the instrument at the beginning of each day before use as well as after replacing the batteries.
◇ Let water on the prism acclimate to the temperature of the instrument before zero-setting.
◇ When “AAA” is displayed, wipe the prism clean, apply water, and press the ZERO button again.
3.
Measurement : not Brix 0.0%
Press the ZERO
button while the
water is on the
prism.
4.
Zero-setting is
complete when “000”
blinks
twice and then stops.
Measurement : Brix 0.0%
1.
Wipe the prism clean and
apply about 0.3mL of tap water.
2.
Press the START
button, and after “---”
appears, the
measurement and the
prism temperature are
displayed.
The instrument is zero-set
correctly.
Proceed to measurement.
Wipe the prism clean with
tissue paper.
Measurement
【Caution】
◇ Do not use metal tools to apply samples on the prism as they may scratch the prism.
◇ nitial measurements may fluctuate with hot or cold samples. Wait for the instrument to acclimate to the sample temperature,
approximately 20 seconds, to press the START button. Alternatively, press the START button multiple times until
measurements become stable.
◇ Do not splash water above 50℃. The plastic may warp, which may compromise the water resistance.
•When measuring hot samples, place only the necessary amount and do not let it overflow from the sample stage well.
•When hot water is necessary to clean off hardened samples, use water-soaked gauze around the prism area and keep hot
water away from the body case.
◇ The displayed temperature is that of the prism and may not necessarily match the temperature of the sample.
1.
Wipe the prism clean and
apply about 0.3mL of the sample.
2.
Press the START
button
3.
After “---” appears, the measurement
and the prism temperature are displayed.
LCD Auto Shut-off>
The instrument will turn itself off after 2 minutes of inactivity. To manually turn it off, hold down the START
button for more than 2 seconds.
For oily/fatty samples>
Try stirring the sample on the prism while measuring to improve the repeatability of oily/fatty samples.
Cleaning
【Caution】
◇ Do not scratch the prism.
◇ The instrument is water-resistant, not waterproof, and should not be submerged.
1. Wipe off the sample.
2. Clean the prism and sample stage using a mild soap and thoroughly rinse with water.
3. Dry the area with tissues thoroughly.
Error Messages
The following messages alert the user when an operation has failed.
The battery is low.
The prism temperature is below the temperature
range.
The ZERO button was pressed with nothing or
something other than water on the prism.
The prism temperature is above the temperature
range.
The START button was pressed with nothing or an
insufficient amount of sample on the prism.
The sample measured outside the measurement
range.
The instrument is faulty.
(Replace the batteries. Contact ATAGO if this
error persists.)
Too much light is entering the prism, and the
instrument cannot measure accurately.
(Shade the sample stage with your hand and take a
measurement again.)
Replacing the Batteries
【Caution】
◇
Fasten the battery compartment cover tightly to prevent water ingress or poor connection, which will
cause erroneous measurements. Push the cover in firmly and turn.
◇ When the O-ring on the cover is dirty or damaged, the water resistance may be compromised.
(Figure 1)
◇ When the battery icon indicates the low power level (
), replace both batteries with a brand new
set of AAA alkaline batteries (1.5V).
◇ Static images may occasionally appear on LCD. Such retained pixel charges do not indicate a faulty
display, consume the battery power, or affect the instrument’s performance in any way.
◇ Check the expiration dates on batteries before purchase.
◇ Zero-set the instrument after the batteries are replaced.
1.
nsert a coin in the
groove on the battery
compartment cover. Turn
the coin counterclockwise
to remove the cover.
2.
nsert batteries,
observing the correct
polarity.
3.
Align the cover and push
it down.
4.
Close the battery
compartment cover by
pushing the cover in with a
coin in the groove and turning
it clockwise until it stops.
*Turning excessively may cause
malfunction.
Approx.
2mm
Approx. 0.3mL of sample
Groove
Approx. 0.3mL of water
Approx.
2mm
Figure1
O-ring
Mild
soap

This manual suits for next models

1

Other ATAGO Measuring Instrument manuals

ATAGO PRM-TANK100 (FER) User manual

ATAGO

ATAGO PRM-TANK100 (FER) User manual

ATAGO PAL-24S User manual

ATAGO

ATAGO PAL-24S User manual

ATAGO RX-5000 i-Pius User manual

ATAGO

ATAGO RX-5000 i-Pius User manual

ATAGO DR-M2 User manual

ATAGO

ATAGO DR-M2 User manual

ATAGO PAL-Easy ACID101 User manual

ATAGO

ATAGO PAL-Easy ACID101 User manual

ATAGO AP-300 User manual

ATAGO

ATAGO AP-300 User manual

ATAGO PAL-RI User manual

ATAGO

ATAGO PAL-RI User manual

ATAGO MASTER-M Series User manual

ATAGO

ATAGO MASTER-M Series User manual

ATAGO PAL-04S User manual

ATAGO

ATAGO PAL-04S User manual

ATAGO PAL-80S User manual

ATAGO

ATAGO PAL-80S User manual

ATAGO PAL-Easy ACID9 User manual

ATAGO

ATAGO PAL-Easy ACID9 User manual

ATAGO PAL-Easy ACID181 User manual

ATAGO

ATAGO PAL-Easy ACID181 User manual

ATAGO PAL-32S User manual

ATAGO

ATAGO PAL-32S User manual

ATAGO PALETTE Series User manual

ATAGO

ATAGO PALETTE Series User manual

ATAGO PAL-64S User manual

ATAGO

ATAGO PAL-64S User manual

ATAGO PAL-Easy ACID121 User manual

ATAGO

ATAGO PAL-Easy ACID121 User manual

ATAGO PAL-131S User manual

ATAGO

ATAGO PAL-131S User manual

ATAGO PAL-61S User manual

ATAGO

ATAGO PAL-61S User manual

ATAGO 4437 User manual

ATAGO

ATAGO 4437 User manual

ATAGO PAL-84S User manual

ATAGO

ATAGO PAL-84S User manual

ATAGO PAL-55S User manual

ATAGO

ATAGO PAL-55S User manual

ATAGO PAL-Easy ACID16 User manual

ATAGO

ATAGO PAL-Easy ACID16 User manual

ATAGO 4553 User manual

ATAGO

ATAGO 4553 User manual

ATAGO PAL-26S User manual

ATAGO

ATAGO PAL-26S User manual

Popular Measuring Instrument manuals by other brands

EKO LS-100 instruction manual

EKO

EKO LS-100 instruction manual

PCB Piezotronics ICP TLD356A34 Installation and operating manual

PCB Piezotronics

PCB Piezotronics ICP TLD356A34 Installation and operating manual

LaserLiner ThermoVisualizer manual

LaserLiner

LaserLiner ThermoVisualizer manual

BTI Disfix 80 Original instructions

BTI

BTI Disfix 80 Original instructions

Agilent Technologies 8590 Series Service guide

Agilent Technologies

Agilent Technologies 8590 Series Service guide

Extech Instruments ExStik EC500 user manual

Extech Instruments

Extech Instruments ExStik EC500 user manual

Elcometer 204 operating instructions

Elcometer

Elcometer 204 operating instructions

IFM Electronic Efector 300 SI0556 operating instructions

IFM Electronic

IFM Electronic Efector 300 SI0556 operating instructions

TDE Instruments Digalox DPM72-PP instruction manual

TDE Instruments

TDE Instruments Digalox DPM72-PP instruction manual

Transmille 8700 Operation manual

Transmille

Transmille 8700 Operation manual

Precision Digital Corporation PD6801-0K1-M20 instruction manual

Precision Digital Corporation

Precision Digital Corporation PD6801-0K1-M20 instruction manual

Xylem WTW photoLab 7600 UV-VIS operating manual

Xylem

Xylem WTW photoLab 7600 UV-VIS operating manual

natus Embla Dx Series User and service manual

natus

natus Embla Dx Series User and service manual

ALLEN FACE & COMPANY D-Meter instruction manual

ALLEN FACE & COMPANY

ALLEN FACE & COMPANY D-Meter instruction manual

PRO SHOT AS2 MAGNUM Operation guide

PRO SHOT

PRO SHOT AS2 MAGNUM Operation guide

Kirchner und Tochter KLA Assembly and operating instructions

Kirchner und Tochter

Kirchner und Tochter KLA Assembly and operating instructions

Tascam dr-44wl Reference manual

Tascam

Tascam dr-44wl Reference manual

Balmac 205ce Operation manual

Balmac

Balmac 205ce Operation manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.