ATI Model A15/82 FluorideMonitorwith Auto-Calibration
19
O&M Manual
Rev-G (6/10)
9. SET AF.:LO (ORHI) Thisroutine allowsthe function ofsetpoint Atobe defined aseitherloworhigh.
When LOisselected,relayAactslikealowalarm,withthe relayactivating when the measured fluoride
isbelowthe setpoint. When HIisselected,the relayactslikeahigh alarm,activating when fluoride is
abovethe setpoint. Setpointsareprogrammed inanotherroutine described laterinthismanual.
10. SET BF.:LO (ORHI) Thisroutine allowsthe function of setpointBtobe defined aseitherloworhigh
similartorelayAdescribed in7above. Notethatrelay Bcannotbeused as an alarm relay when the
automaticcalibration function isused. WhentheAutoCalisactivated, relay Bisusedfor
initiating thecalibration sequence.
11. ALSET A:OFF Aseparatealarmrelayisprovided inthe monitorwhichisindependent ofrelaysAand
B.It ismarked C/Don the back ofthe instrument. Thisrelaycan be used toactuatean emergency
alarmshouldthe fluoride concentration remainoutside ofthe Asetpointforsomefixed period oftime.
The defaultforthisfunction isoff.If you wanttousethisfunction, usethe UPkeytoselect ONand then
press ENTER.The displaywill change to“TIME SETA:60 M”.Usethe UPorDOWNkeytoselect the
timeperiod afterwhichyou wantthe C/Drelaytoactuate.Forinstance,ifyou havealreadydefined
relayAasalowalarmwithasetpointof0.50 ppmand wanttoactivatean additionalalarmiffluoride
remainsbelow0.5formorethan 30 minutes,press the UPkeyuntil the displayindicates30 Mand then
press ENTER.Thisfunction can be extremelyusefulforapplicationswherethe Arelayisbeing used for
control,sinceitprovidesan alarmifthe controlfunction doesnotmaintainthe desired value forsome
period oftime.
12. ALSETB:OFF Thisroutine issimilarto8above,butwithrespecttorelayB.It associatesthe alarm
relaywiththe operation ofrelayBasdescribed above.
13. ALCALIBR.:OFF Thisroutine allowsthe alarmrelay(relayC)tobe used toindicatefailureofthe
automaticcalibration sequence. Set thisfunction toONifyou wishtousethe Crelayforthispurpose.
14. ALRELAY: ACT(ORDEA) Thisroutine allowsthe C/Drelaytobe defined fornormalorfail-safe
operation.SelectACT(activated)fornormaloperation.Thisreferstothe factthatthe relaycoil isnot
normallyenergized and energizeson alarmcondition.SelectDEA (deactivated)forfail-safeoperation.
Thiscan be usefulifthisrelayistobe used forapowerfailureindication.Notethatthe Cand NC relay
indicationson the wiring diagramsrefertothe contact designationsinfail-safeoperating mode. Ifnormal
operation isselected, the NC contactsbecomeNO contacts.
15. ALTYPE: CONT orFLASHThisroutine givesthe option, when alarmisactivated, toholdthe relaycoil
continuouslyofhavethe coil flash(cyclebetween energized and de-energized).The sub menu when AL
TYPE isset toFLASH, allowsforthe selection ofslow, med, orfast flashing.
16. CLEAN:DISABLEDThe automaticcleaning featureisnotimplemented inthe fluoride monitor.This
function shouldnot be changed.
17. CLEANING T:15.0”Not used. Do not adjust.
18. HOLDING T:3.O’ Not used. Do not adjust.
19. CALIBR.:AUTO Thisroutine allowsyou tospecifywhetherthe automaticcalibration featurewill be run
automaticallybased on the internaltimer,manuallyatthe option ofthe operator,ordisabled.The
defaultisAUTO. Ifyou wish,you can usethe UPorDOWNkeystosetthe calibration function to
ManualorDisabled.