
(4) Communication Modes switches Figure 6-9 Figure 6-10(, )
A. From toSingle-device communication
Multi-device communication
Set the address of the meter(value of )
as (3).E. such as 255;
Set the meter to Multi-device communication
mode (select value of to P1-255 or P2-255)
as (3).D;
After setting, the address will show on the LCD.
B. From toMulti-device communication
Single-device communication
After setting, no address will show on the LCD.
Figure 6-10 Single-device communication mode
Set the meter to Signal-device communication
mode (select value of to P0-000) as (3).D;
Figure 6-9 Multi-device communication mode
E“SET GCF” Set the gas correction factor. See detailed operation in Figure 6-11.,
10
M
M
G“SET Addr” Set the address for,.Multi-device communication mode
H
Notes: Please remember the new password and placed it properly.
I“OFFSET” reset the offset of the meter.,See detailed operation in Figure 6-11.
J“CLEAr ACC” reset the flow accumulation reading to zero.,
“SET PAS” Set the password.,
C“UnITTyPE” select,instant flow units.
Press the display will show the instant flow units menu; if the value is “--n3--”, the instant,
flow unit is Nm /h; if the value is “--SL--”, the instant flow unit is SLPM.
3
Press to switch in two units.
After selection, press to confirm and exit.
D“InTErVAL” set response time.,
Press the display will show the response time (default value is 125 ms).,
Press to set as other response time, 250 ms, 500 ms, 1000 ms (1 s), 2000 ms (2 s), 4000 (4 s).
After selection, press to confirm and exit.
F“PrOTOCOL” select communication .,protocols.
Press the display will show the protocol menu; P0-000 means mode 0, it is Single-device,
communication mode; P1-xxx (between 001 and 255) means mode 1, Modbus mode , it is Multi-device
communication mode. P2-xxx (between 001 and 255) means mode 2, reserved, it is Multi-device
communication mode. (e.g, P1-153 means the meter is working in Multi-device communication mode,
protocol is Modbus, and the address is153.)
Press to switch in two communication modes.
After selection, press to confirm and exit.
See detailed operation in Figure 6-11.
Notes: The default address is 255.