Baker Hughes Masoneilan 4411 Series User manual

Masoneilan™
4411 Series
Electropneumatic Transducer
Instruction Manual
Baker Hughes Data Classication : Public

b | Baker Hughes Copyright 2020 Baker Hughes Company. All rights reserved.
THESE INSTRUCTIONS PROVIDE THE CUSTOMER/OPERATOR WITH IMPORTANT PROJECT-
SPECIFIC REFERENCE INFORMATION IN ADDITION TO THE CUSTOMER/OPERATOR’S
NORMAL OPERATION AND MAINTENANCE PROCEDURES. SINCE OPERATION AND
MAINTENANCE PHILOSOPHIES VARY, BAKER HUGHES COMPANY (AND ITS SUBSIDIARIES
AND AFFILIATES) DOES NOT ATTEMPT TO DICTATE SPECIFIC PROCEDURES, BUT TO
PROVIDE BASIC LIMITATIONS AND REQUIREMENTS CREATED BY THE TYPE OF EQUIPMENT
PROVIDED.
THESE INSTRUCTIONS ASSUME THAT OPERATORS ALREADY HAVE A GENERAL
UNDERSTANDING OF THE REQUIREMENTS FOR SAFE OPERATION OF MECHANICAL AND
ELECTRICAL EQUIPMENT IN POTENTIALLY HAZARDOUS ENVIRONMENTS. THEREFORE,
THESE INSTRUCTIONS SHOULD BE INTERPRETED AND APPLIED IN CONJUNCTION WITH
THE SAFETY RULES AND REGULATIONS APPLICABLE AT THE SITE AND THE PARTICULAR
REQUIREMENTS FOR OPERATION OF OTHER EQUIPMENT AT THE SITE.
THESE INSTRUCTIONS DO NOT PURPORT TO COVER ALL DETAILS OR VARIATIONS IN
EQUIPMENT NOR TO PROVIDE FOR EVERY POSSIBLE CONTINGENCY TO BE MET IN
CONNECTION WITH INSTALLATION, OPERATION OR MAINTENANCE. SHOULD FURTHER
INFORMATION BE DESIRED OR SHOULD PARTICULAR PROBLEMS ARISE WHICH ARE NOT
COVERED SUFFICIENTLY FOR THE CUSTOMER/OPERATOR’S PURPOSES THE MATTER
SHOULD BE REFERRED TO BAKER HUGHES.
THE RIGHTS, OBLIGATIONS AND LIABILITIES OF BAKER HUGHES AND THE CUSTOMER/
OPERATOR ARE STRICTLY LIMITED TO THOSE EXPRESSLY PROVIDED IN THE CONTRACT
RELATING TO THE SUPPLY OF THE EQUIPMENT. NO ADDITIONAL REPRESENTATIONS OR
WARRANTIES BY BAKER HUGHES REGARDING THE EQUIPMENT OR ITS USE ARE GIVEN
OR IMPLIED BY THE ISSUE OF THESE INSTRUCTIONS.
THESE INSTRUCTIONS ARE FURNISHED TO THE CUSTOMER/OPERATOR SOLELY TO
ASSIST IN THE INSTALLATION, TESTING, OPERATION, AND/OR MAINTENANCE OF THE
EQUIPMENT DESCRIBED. THIS DOCUMENT SHALL NOT BE REPRODUCED IN WHOLE OR IN
PART WITHOUT THE WRITTEN APPROVAL OF BAKER HUGHES.

Masoneilan Model 4411 Electropneumatic Transducer Instruction Manual | c
Copyright 2020 Baker Hughes Company. All rights reserved.
Table Of Contents
Safety Warning ...............................................................................................................................1
Brief User Guide .............................................................................................................................2
General Description .......................................................................................................................2
Instrument Mounting......................................................................................................................3
Captured Bleed Version �������������������������������������������������������������������������������������������������������������3
Pneumatic Installation ...................................................................................................................3
Electrical Installation......................................................................................................................4
Wiring and Cable Entry ��������������������������������������������������������������������������������������������������������������4
Conduit Entry �����������������������������������������������������������������������������������������������������������������������������4
Calibration ����������������������������������������������������������������������������������������������������������������������������������4
Tight Shut-O Adjustment Instructions .......................................................................................5
Simple Functional Checks.............................................................................................................5
Technical Specications................................................................................................................6
Performance....................................................................................................................................6
Potentially Explosive Atmospheres — Markings ........................................................................7
Instructions Specic to Hazardous Area Installations ...............................................................8
Instructions — Special Condition for Safe Use...........................................................................8
FM Control Drawing .......................................................................................................................9
CSA Control Drawing ...................................................................................................................10

1 | Baker Hughes Copyright 2020 Baker Hughes Company. All rights reserved.
Safety Information
Important - Please read before installation
These instructions contain DANGER, WARNING, and CAUTION
labels, where necessary, to alert you to safety related or other
important information. Read the instructions carefully before
installing and maintaining your control valve. DANGER and
WARNING hazards are related to personal injury. CAUTION
hazards involve equipment or property damage. Operation
of damaged equipment can, under certain operational
conditions, result in degraded process system performance
that can lead to injury or death. Total compliance with all
DANGER, WARNING, and CAUTION notices is required for safe
operation�
This is the safety alert symbol. It alerts you to potential personal
injury hazards. Obey all safety messages that follow this symbol
to avoid possible injury or death.
Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided,
could result in serious injury.
Indicates a potentially hazardous situation which, if not avoided,
could result in minor or moderate injury.
When used without the safety alert symbol, indicates a
potentially hazardous situation which, if not avoided, could
result in property damage.
Note: Indicates important facts and conditions.
About this Manual
• The information in this manual is subject to change without
prior notice�
• The information contained in this manual, in whole or part,
shall not be transcribed or copied without Baker Hughes’s
written permission.
• Please report any errors or questions about the information in
this manual to your local supplier.
• These instructions are written specically for the 4411 IP
Transducer, and do not apply for other instruments or IP
transducers outside of this product line.
Warranty
Items sold by Baker Hughes are warranted to be free from
defects in materials and workmanship for a period of one
year from the date of shipment provided said items are used
according to Baker Hughes recommended usages� Baker
Hughes reserves the right to discontinue manufacture of any
product or change product materials, design or specifications
without notice. This instruction manual applies to the Masoneilan
4411 IP Transducer.
Note: Prior to installation:
• The 4411 IP Transducer must be installed, put into service
and maintained by qualied and competent professionals who
have undergone suitable training.
• All surrounding pipe lines must be thoroughly ushed to ensure
all entrained debris has been removed from the system.
• Under certain operating conditions, the use of damaged
equipment could cause a degradation of the performance of
the system which may lead to personal injury or death.
• Changes to specications, structure, and components used
may not lead to the revision of this manual, unless such
changes aect the function and performance of the product.
IMPORTANT: SAFETY WARNING
Please read these instructions carefully BEFORE this instrument
is installed or maintained�
These converters are intended for use in industrial compressed
air systems only. Ensure that adequate pressure relief provision
is installed if application of system supply pressure could cause
downstream equipment to malfunction. Installation should
be in accordance with local and national compressed air and
instrumentation codes�
Products certified for use in explosion-proof or intrinsically safe
installations MUST:
a) Be installed in accordance with local and national codes for
hazardous area installations, and in accordance with this
manual�
b) Only be used in situations which comply with the certification
conditions stated in this handbook.
c) Only be maintained by qualified personnel with adequate
training on hazardous area implementation�
Before using these products with fluids other than air or for non-
industrial applications consult Baker Hughes�

Masoneilan Model 4411 Electropneumatic Transducer Instruction Manual | 2
Copyright 2020 Baker Hughes Company. All rights reserved.
Brief User Guide
This is a quick guide to connecting the instrument for the conven-
ience of personnel who are familiar with this type of product.
More comprehensive instructions are contained later in this
manual:
1. Connect a clean air supply of about 2.0 bar (30 psig for the
3-15 psig output model, or 2.4 bar (35 psig) for the 6-30 psig
output model) to the IN port (1/4" NPT).
2. Connect a pressure gauge or actuator to the OUT port.
DO NOT use PTFE tape or similar to seal the
ports. Use a minimum of soft setting anaerobic
compound, e.g. Loctite Hydraulic Seal 542.
3. Unscrew the lid to obtain access to the terminals.
The cable will need to be sealed by an approved
gland for explosion-proof installation.
The second un-used conduit entry must be
sealed with an explosion-proof blanking plug.
4. Connect a 4-20mA current source, either through the Jack-
Plug source or Terminal connections. Ensure that it can
supply at least 6.5V at 20mA. Check for correct polarity.
DO NOT use a voltage source; it will irreversibly dam-
age the converter.
5. Switch on the air supply and current source and allow a
few seconds stabilisation time. Adjust span and zero via the
trimpots if necessary.
If the instrument fails to operate refer to the Simple Functional
Checks�
General Description
These converters are precision electronic pressure controllers
designed for continuous process control applications. The units
operate a closed loop system incorporating a feedback sensor
ensuring long-term stability and high accuracy performance, with
‘fail-safe’ operation, i.e. the output falls to a low pressure upon
failure of the loop current.
These instruments combine substantial flow capacity, with the
ability to give precise control into closed volume ‘dead end’
applications�
The instrument has been designed to withstand the rigours of
industrial environments; such as shock, vibration and positional
effects. As with any other precision instrument dropping the
unit onto a hard surface may result in shifts in calibration or
possibly permanent damage. The unit may be mounted directly
upon a control valve. A mounting bracket is provided for surface
mounting, or the instrument may be fixed directly to rigid
pipework.
The electronics are enclosed within a watertight housing, which
is suitable for outdoor use if a suitable cable gland is used. The
lid should be firmly tightened, the use of a tool for this purpose is
not recommended�
Figure 1
JACK SOCKET: TO CALIBRATE UNIT ON
INSTALLATION USING A 4-20MA INPUT
SIGNAL. WHEN USING JACK PLUG A SIGNAL
DOES NOT HAVE TO BE SUPPLIED VIA THE
TERMINAL BLOCK. THE INPUT SIGNAL CAN
BE MONITORED THROUGH THE JACK PLUG
SOCKET WHEN UNIT IS IN OPERATION.
1/2"NPT or M20
Conduit Entry
3.5mm Jack Socket
Alternative Power
Supply & Monitor
Span
Zero
Tight Shut O
4-20mA Input Terminals
1/2"NPT or M20
Conduit Entry
Lid Locking
External Earth
1/8" Gauge Port
External Earth
Air Filter;
DO NOT REMOVE WHEN PRESSURIZED.
DO NOT USE WITH FILTER REMOVED.
Air Filter. BS2000 with 98% eciency
@ 2µM with water and oil absorption.
Internal Earth

3 | Baker Hughes Copyright 2020 Baker Hughes Company. All rights reserved.
To meet the EMC specifications, screened cable should be
used for installation. The cable screen should be connected to
the internal earth bonding point of the I/P. An earth strap should
also be connected from the external earth bonding point of
the I/P to a common earth point. The cable screen should not
be connected at the signal source when used in Intrinsically
Safe (IS) environments. The I/P has been thoroughly tested in
accordance with:
Electromagnetic compatibility (EMC)
Generic Standards – Immunity for industrial environments
Electromagnetic compatibility (EMC)
Generic Emissions Standard – Industrial environment
Please consult wiring diagram for further information. The results
obtained from these tests show that with the screen arrangement
shown in the installation diagram, the typical sustained output
pressure shift is <+/-5% full scale. Consult supplier for further
details�
Instrument Mounting
The converter can be mounted directly onto a valve, onto any
suitable flat surface, or onto 2” (50mm) pipe. Dimensional details:
Figure 2
These converters will operate in any position, with slight re-
calibration adjustments, normal operation is in the upright
position, i.e. with the screwed lid horizontal. The unit is weather
resistant when the lid is removed for connection or calibration,
but should not be left uncovered during operation or for extended
periods of time. They are unaffected by shock and vibration, but
severe vibration environments are best avoided.
The instruments are protected against environmental effects, to
a level of Type 4X (IP66). Sensible positioning should be used if
the local environment is severe.
Temperature compensation is designed into these converters
which will operate over a range from –40° to +85°C (-40° to
+185°F). Maximum life is obtained if extreme temperature cycling
is avoided. In extreme direct sunlight, sun-shading is advised.
The pneumatic exhaust and the bleed are routed through the
baffles. The baffles must not be blocked with paint and/or plastic
sheeting etc�
Captured Bleed Version
The baffles and the baffle cover are fitted to maintain the adequate
Ingress Protection (IP) rating of the instrument. Removal of these
protective parts will invalidate the IP and the NEMA 4X rating of
the instrument�
For captured bleed installation, remove the baffle cover and
unscrew both the baffles, fit 1/8" NPT fittings to two separate
lengths of pipes, to exhaust and bleed gas to a safe area
installation�
Vented gas from the enclosure through bleed port must not
be re-circulated and introduced back into the process stream.
Exhaust port installation must not have any restriction to create
backpressure to cause instrument to malfunction. Instrument
installation for captured bleed version must be under the
installer’s discretion.
Note: Exhaust Port is nearer the conduit entry (LH side)
Enclosure Bleed Port is on the RH side. (See Figure 3.)
Figure 3
Pneumatic Installation
These instruments are recommended for use with clean, dry, oil
free instrument grade air to BS.6739 or ISA-7.0.01-1996.
Dew Point: At least 10°C (18°F) below (-40°C) minimum
anticipated ambient temperature.
Dust: Filtered to below 50 microns.
Oil Content: Not to exceed 1ppm mass.
The instrument is factory calibrated with a supply pressure of 30
psig (2 bar) ±10%. Operation is possible at any pressure between
18 and 150 psig (1.3 to 10 bar), though recalibration may be
necessary towards these limits to maintain specified accuracy.
The inlet and outlet ports are threaded 1/4" NPT female and
suitable fittings should be used. For most installations 1/4" (6mm)
pipe will be adequate. If a large actuator, high flow rates or long
pipework is necessary then a larger diameter should be used.
Plastic tubing, e.g. Nylon is preferable where circumstances
permit, since it is normally very clean internally. In all cases,
purge the supply pipework before connection to the converter.
Two gauge ports are provided to facilitate direct mounting of
a pressure gauge. To use one of these ports remove the plug
(using a 1/4" or 3/16" Hexagon Key) in a de-pressurized state,
and connect the gauge. The ports are threaded 1/4" NPT at the
back of the unit and 1/8" NPT at the front.
Exhaust Port Enclosure Bleed Port

Masoneilan Model 4411 Electropneumatic Transducer Instruction Manual | 4
Copyright 2020 Baker Hughes Company. All rights reserved.
Under no circumstances should PTFE tape be used
for sealing the ttings as this tends to shred small
particles which may nd their way into the instrument
causing malfunctions.
The use of a soft setting anaerobic hydraulic seal is recommended,
(e.g. Loctite Hydraulic Seal 542). Follow the manufacturers
recommendations�
Do not use an excessive amount as this will not set
and could nd its way into the instrument.
If the air supply is not of adequate quality, the device performance
can be affected. Adequate quality can normally be achieved by
the use of air filter regulators.
Electrical Installation
The electrical connections should be made as shown in the
dimensional drawing Figure 1. The instrument is protected
against reverse polarity to –100mA, no operation is possible in
this condition�
The Model 4411 approximates a constant voltage load of 6.5
volts across the loop terminals, therefore it is essential that the
loop controller be capable of providing a constant current in the
range 4-20mA with an output voltage of at least 6.5 volts.
Voltage output controllers (e.g. variable voltage power
supplies) are entirely unsuitable for the Model 4411
and could severely damage the electronic circuits.
Important Note
The control electronics of the Model 4411 incorporate
precision electronics. The calibration of the unit may be
affected by very high voltage spikes. Consequently, in
environments where static electricity may be present ESD
precautions should be used.
Wiring And Cable Entry
These instruments must be installed in accordance
with local and national codes of practice, especially for
hazardous area installations. The instruments are fully
isolated from ground and therefore grounding is un-
necessary for functional purposes. However, ground-
ing may be necessary to conform to installation codes.
Note: It is strongly recommended that shielded cable or a
grounded conduit be used to achieve maximum RFI immunity,
if the installation has any risk of electromagnetic interference.
Conduit Entry
The instrument has twin conduit entry threaded 1/2" NPT. For
explosion proof installation, a sealed conduit gland conforming to
explosion-proof specifications must be used. A ground terminal
is provided both internally and externally and should be used if
ground continuity is essential.
The unit is supplied with plastic blanking plugs, which
must be removed before operation/calibration and re-
placed with a blanking plug conforming to the hazard-
ous area certication rating applicable.
Calibration
The instruments are designed for continuous operation without
the necessity for routine overhaul, with continuous monitoring,
adequate precautions and replacement of the filter no longer
than every 5 years.
The most common source of failure for pneumatic instrumentation
has been found to be inadequate air quality, allowing
contaminants to block internal orifices. Air filtering is included
within the instrument but cannot cope with sustained poor air
quality, which may ultimately lead to failure.
The recommendations in the Pneumatic Installation section
should be rigorously observed.
Note: These instruments are factory calibrated at a supply
pressure of 30psig (2 bar).
The instrument cover must be unscrewed to obtain access to
the trimpots�
Do not remove the instrument cover in a potentially
explosive atmosphere when the instrument is powered.
An accurate current source of 4-20mA and pressure gauge are
required. These should be of good quality with an accuracy of
0.1% or better. The current source should be checked to ensure
that it provides at least 6.5V at 20mA output compliance.
• Connect the instrument as described in the installation
section or the test-jack section below.
• Remove the instrument cover to gain access to the trimpots
and jack-socket.
• Set the current to 4.00mA – the instrument outlet should be
3.00+/-0.05psig (0.200+/-0.003 bar). Adjust the Zero trimpot
if necessary.
• For 6-30 psig output, the output pressure should be set to
6.00 ±.05 psig (0.400 ± .003 bar). Adjust trimpot as necessary.
• Set the current source to 20.00mA – the instrument outlet
should be 15.00 +/-0.05psig (1.00+/-0.003 bar). Adjust the
Span trimpot if necessary.
• For 6-30 psig output, the output pressure should be set to
30.00 ±.05 psig (2.00 ± .003 bar). Adjust trimpot as necessary.
If either Span or Zero controls are adjusted it may be necessary
to repeat the above steps until both ends are within the calibration
limits�
Alternatively the Jack Plug can be connected to calibrate and
test the unit. The Jack Plug can be either set-up for Monitoring or
Calibration/Operation set-up.

5 | Baker Hughes Copyright 2020 Baker Hughes Company. All rights reserved.
Calibration with the Test Jack:
Connect pin 1 of the Jack Plug to the positive (+) lead of the
current calibrator and pin 3 of the Jack Plug to the negative (-)
lead of the current calibrator and then insert the Jack Plug into the
Test Jack. The current calibrator is now the input signal source.
Calibrate as stated above. Removing the Jack Plug will return
operation of the unit back over to the original current source.
Figure 4
Monitoring with the Test Jack:
Connect pin 1 of the Jack Plug to the negative (-) lead of the
ammeter and pin 2 of the Jack Plug to the positive (+) lead of
the ammeter. Insert the Jack Plug into the Test Jack and use the
ammeter to monitor the input current loop�
Tight Shut-Off Adjustment
Instructions
Figure 5
The Tight Shut-Off potentiometer can be adjusted to set the zero
point at which the instrument ‘shuts off’.
For Example:
Tight shut-off is deactivated in the factory.
Figure 6
Setting of shut-off point:
1. With the instrument under test connected to the required
supply pressure and load ports, apply a demand current
signal of that value at which shut-off is required.
2. If the output pressure is >start-up pressure (~0psi), adjust
potentiometer slowly clockwise until the output pressure
falls to zero. Do not turn the pot beyond this point. The shut-
off point is now set.
Shut-off point check:
Increase the input signal by approximately 0.5mA (e.g. to 3.8mA)
the output pressure will rise above 0psi. Reset the input signal to
the required shut off point (e.g. 3.3mA) and the output pressure
will fall to zero.
Simple Functional Checks
Apply a 4-20mA signal and an air supply of 30psig and observe
the output on a pressure gauge. It should control smoothly. The
Test Jack can also be connected as a monitor measure.
Problem Possible Causes Suggested
Action
Maximum
output
not available
Supply pressure
too low
Check and adjust
supply
Calibration error Recalibrate
Air leak in
instrument
Locate leak and
repair
Excess outlet flow Check with
specification
Minimum output
too high
Calibration error Replace
Recalibrate
Delay on
start up
A delay of a few
seconds is normal
None
Erratic
operation at low
pressure
Signal currents
below 3.5mA are
insufficient for
normal operation
Increase current
May need
recalibration
Tight shut-off set
No output
available
Tight shut off
potentiometer
adjusted
Turn until a
pressure output
is observed
on gauge or
pressure sensor
Erratic at all
pressures
Controller cannot
provide 6.5V
continuously
Reduce loop
resistance or
change controller
Contamination Check I/P filter
status

Masoneilan Model 4411 Electropneumatic Transducer Instruction Manual | 6
Copyright 2020 Baker Hughes Company. All rights reserved.
Technical Specifications
Functional
Input: 4-20mA (0-100%)
Output: 3-15 psi or 3-30 psi
Minimum output greater than 0.2psi
Supply Pressure: 150psi maximum, minimum 3psi above max required output pressure
Supply Pressure Effect: Span over full supply pressure range <0.1%
Medium: Standard instrument quality air to 50 micron
Air Consumption: <2.5l/min (0.09 scfm) at 50% signal
Operating Temperature: -40°C to 85°C (-40° F to +185° F)
Relative Humidity: 0 to 100% Relative Humidity
Output Capacity: >300l/min (12scfm) delivery and exhaust at 100% signal
Maximum Terminal Voltage: Maximum 6.5Volts
Minimum Operating Current: Less than 3.5mA
Current Reversal Protection: No effect within normal 4-20mA range protected to 100mA continuous.
Insulation: Electrical circuits are isolated from housing. Tested to 725V DC, 100MΩ
Performance
(At 20°C (68°F), 30psi supply, 3-15psi range, for a typical instrument except otherwise stated);
Instrument Accuracy: mean <0.1%
Independent Linearity: mean <±0.05% of span
Hysteresis, Resolution & Deadband: mean <=0.05% of span
Accuracy Rating: <±0.75% span (including factory calibration error)
Above figures are in accordance with ANSI/ISA 51.1:1993; N.B. Instrument accuracy excludes calibration errors
Temperature Effect: Span and Zero
Mean temperature coefficient over full operating range –40 to +85°C less than 0.035%
span/degC
Vibration: <3% of span; 4mm peak-peak 5-15Hz, 2g sine 15-150Hz
Mounting Position: Integral bracket allows for mounting in any orientation
Long Term Stability: Span and Zero
Typically better than 0.25% span per year
Electromagnetic Compatibility: Compliant with EC requirements:
BS EN61000-6-4 - Generic Emissions Standard – Industrial environment
BS EN61000-6-2 - Generic Standards – Immunity for industrial environments
To meet the EMC specifications, screened cable should be used for installation. The
cable screen should be connected to the internal earth bonding point of the I/P. An
earth strap should also be connected from the external earth bonding point of the I/P to
a common earth point. The cable screen should not be connected at the signal source
when used in IS environments.
Controls: Span and Zero and tight shut-off trimpots with 10% adjustment
Physical
Air Supply and Output Connections: 1/4" NPT supply and output; 1/8" NPT exhaust baffles to allow for captured bleed
Electrical: 1/2" NPT (M20 option); 2 internal screw terminals for 2.5mm2 cable
Weight: 2.07Kg
Materials
Aluminium and zinc diecasting with nitrile diaphragms, epoxy powder coat painting as standard. Weatherproof to Type 4X (IP66)
[mounted upright]�

7 | Baker Hughes Copyright 2020 Baker Hughes Company. All rights reserved.
Potentially Explosive Atmospheres - Markings:
Certification
Agency
Flame Proof /
Explosion Proof Intrinsically Safe Non-Incendive Others
SIRA ATEX
Approved
To EN60079
Sira 01ATEX1223
Ex d IIC T4 Gb
(Ta = -20° to +40°C)
Ex d IIB+H2 T5 Gb
(Ta = -20° to +80°C)
Ex d IIB+H2 T6 Gb
(Ta = -20° to +65°C)
Ex t IIIC T95°C Db
(Ta = -20° to +85°C)
Umax = 30Vdc
II
2GD
2GD
Sira 01ATEX2224X
Ex ia IIC T4 Ga
Ex ia IIIC Da T95°C
(Ta = -40° to +85°C)
Ui = 30Vdc
Ii = 110mA
Pi = 0.84W
Ci = 6nF
Li = 100µH
II
2GD
1GD
Factory Mutual Class I, Division 1,
Group BCD.
T6 Ta = 75°C,
T5 Ta = 85°C.
Class I, II & III,
Division 1,
Group ABCDEFG.
T4 Ta = 85°C.
Vmax = 30Vdc
Imax = 110mA
Ci = 0.006µF
Li = 100µH
Install as per Control
drawing 2001-082.
Class I, Division 2,
Group ABCD.
T6 Ta = 75°C,
T5 Ta = 85°C.
Dust Ingress Protection:
Class II & III, Division 1,
Group EFG.
T6 Ta = 75°C,
T5 Ta = 85°C.
Suitable for:
Class II & III, Division 2,
Group FG,
T6 Ta = 75°C,
T5 Ta = 85°C
CSA
Class I, Division 1,
Group BCD.
Class II,
Group EFG;
Class III;
Ta = 85°C; T5
Ta = 65°C; T6
Class I, Division 1,
Group ABCD.
Class II,
Group EFG;
Class III;
Ta = 85°C; T4
Vmax = 30Vdc
Imax = 100mA
Pmax = 0.75W
Ci = 10.5nF
Li = 100µH
(30Vdc max, 300 Ohms).
Install as per Control
drawing 2001-083.
Class I, Division 2,
Group ABCD.
Class II, Division 2,
Group EFG;
Class III;
Ta = 85°C; T5
Ta = 75°C; T6
Ii = 24mA
Ci = 6nF
Li = 100µH
Note: Because of standard updates, product changes or improvements, certification level may have changed and the manual
may have not been updated yet.
Please refer to device label to check updated certification level.
POTENTIAL ELECTROSTATIC CHARGING, IMPACT & FRICTION HAZARD. This shall be taken in consideration when
apparatus is installed in Division 1 locations or locations that specically require Equipment Protection Levels Ga
and Da.
For additional warnings that may apply for safe use and installation, refer to the Instructions – Special Conditions for
Safe Use section of this manual as identied on ATEX certicates numbers with an ‘X’ sux.

Masoneilan Model 4411 Electropneumatic Transducer Instruction Manual | 8
Copyright 2020 Baker Hughes Company. All rights reserved.
Instructions Specific to
Hazardous Area Installations
(Reference European ATEX Directive 2014/34/EU, Annex II,
1.0.6)
The following instructions apply to equipment covered by
certificate numbers SIRA 01ATEX2224X & SIRA 01ATEX1223
1. The Model 4411 may be installed as intrinsically safe or
flameproof:
• Zone 0 or 20 installations require the equipment to be
installed as intrinsically safe via suitable associated
apparatus�
• Zone 1 or 21 installations require the equipment to be
installed as intrinsically safe or flameproof; if installed
as flameproof, associated apparatus is not required,
but there are more onerous requirements for cable
entry – refer to EN60079-0.
• Zone 2 or 22 installations may be intrinsically safe or
flameproof.
• It is recommended that the installer indicates on the
equipment which certification code applies.
2. The equipment should not be used outside the stated
ambient temperature range.
3. The equipment has not been assessed as a safety-related
device (as referred to by Directive 2014/34/EU Annex II,
clause 1.5).
4. Installation and maintenance of this equipment shall be
carried out by suitably-trained personnel in accordance
with the applicable codes of practice (EN 60079-14 and EN
60079-17 within Europe).
5. Repair of this equipment shall be carried out by the
manufacturer or in accordance with the applicable code of
practice (IEC 60079-19).
6. When installing the equipment as flameproof cable gland
(M20 parallel thread) or a conduit (1/2" NPT tapered thread),
the installer should verify the thread form and ensure that
the appropriate matching thread is used� The pneumatic
connections are not flameproof entries.
7. The certification of this equipment relies on the following
materials used in its construction:
Enclosure: aluminium alloy and zinc alloy
O-ring: nitrile rubber
Encapsulant: polyurethane
If the equipment is likely to come into contact with aggressive
substances, then it is the responsibility of the user to take suitable
precautions that prevent it from being adversely affected, thus
ensuring that the type of protection is not compromised.
• Aggressive substances: e.g. acidic liquids or gases that
may attack metals, or solvents that may affect polymeric
materials�
• Suitable precautions: e.g. regular checks as part of routine
inspections or establishing from the material’s data sheet
that it is resistant to specific chemicals.
• Not to be used with pure oxygen or oxygen enriched media
as the process fluid.
Consult the supplier for approval certificates.
8. Only approved cable glands are to be used which must
maintain an Ingress Protection of IP66.
9. Unused entry port hole must be blanked off using suitably
approved blanking plug conforming to the hazardous area
certification rating applicable.
Instructions – Special
Condition for Safe Use
For Certificate Number Sira 01ATEX1223 – (Ex d) None.
For Certificate Number Sira 01ATEX2224X – (Ex ia)
1. For operation reasons, it may be necessary to connect the
screen of the cable to the enclosure of the equipment in
the hazardous area. If this is the case, then the equipment
shall be installed in accordance with EN 60078-14:2014
(specifically clause 16.2.2.3 and 16.2.4), typically using a
galvanically-isolated interface with no other connection to
Earth apart from via the enclosure of the equipment.
2. The enclosure is made from light metal which could cause
ignition due to impact and friction. This shall be taken into
consideration when the apparatus is installed in location that
specifically requires equipment protection level Ga.
3. Under certain extreme circumstances, the non-metallic
parts incorporated in the enclosure of this equipment may
generate an ignition-capable level of electrostatic charge.
Therefore, when it is used for applications that specially
require equipment protection level Ga, the equipment shall
not be installed in a location where the external conditions
are conductive to the build-up of electrostatic charge on
such surfaces. Additionally, the equipment shall only be
cleaned with a damp cloth.
4. The option for the air vented from the enclosure to be re-
circulated and introduced back into the process stream shall
not be used when the equipment is installed in locations that
specifically require equipment protection level Ga for group
II category 1G or 1D for group III category 1D equipment.

9 | Baker Hughes Copyright 2020 Baker Hughes Company. All rights reserved.

Masoneilan Model 4411 Electropneumatic Transducer Instruction Manual | 10
Copyright 2020 Baker Hughes Company. All rights reserved.

11 | Baker Hughes Copyright 2020 Baker Hughes Company. All rights reserved.
Notes

Masoneilan Model 4411 Electropneumatic Transducer Instruction Manual | 12
Copyright 2020 Baker Hughes Company. All rights reserved.
Notes

Tech Field Support & Warranty:
Phone: +1-866-827-5378
valvesupport@bakerhughes.com
bakerhughes.com
Copyright 2020 Baker Hughes Company. All rights reserved. Baker Hughes provides this information on an
“as is” basis for general information purposes. Baker Hughes does not make any representation as to the
accuracy or completeness of the information and makes no warranties of any kind, specic, implied or oral,
to the fullest extent permissible by law, including those of merchantability and tness for a particular purpose
or use. Baker Hughes hereby disclaims any and all liability for any direct, indirect, consequential or special
damages, claims for lost prots, or third party claims arising from the use of the information, whether a claim
is asserted in contract, tort, or otherwise. Baker Hughes reserves the right to make changes in specications
and features shown herein, or discontinue the product described at any time without notice or obligation.
Contact your Baker Hughes representative for the most current information. The Baker Hughes logo and
Masoneilan are trademarks of Baker Hughes Company. Other company names and product names used in
this document are the registered trademarks or trademarks of their respective owners.
BHMN-4411-19084E-0720 07/2020
[Ref: YM14002 / Ctrl Ref: 2011-037c]
Direct Sales Office Locations
Find the nearest local Channel Partner in your area:
valves.bakerhughes.com/contact-us
valves.bakerhughes.com
Australia
Brisbane
Phone: +61-7-3001-4319
Perth
Phone: +61-8-6595-7018
Melbourne
Phone: +61-3-8807-6002
Brazil
Phone: +55-19-2104-6900
China
Phone: +86-10-5738-8888
France
Courbevoie
Phone: +33-1-4904-9000
India
Mumbai
Phone: +91-22-8354790
New Delhi
Phone: +91-11-2-6164175
Italy
Phone: +39-081-7892-111
Japan
Tokyo
Phone: +81-03-6871-9008
Korea
Phone: +82-2-2274-0748
Malaysia
Phone: +60-3-2161-03228
Mexico
Phone: +52-55-3640-5060
Russia
Veliky Novgorod
Phone: +7-8162-55-7898
Moscow
Phone: +7-495-585-1276
Saudi Arabia
Phone: +966-3-341-0278
Singapore
Phone: +65-6861-6100
South Africa
Phone: +27-11-452-1550
South & Central
America and the Caribbean
Phone: +55-12-2134-1201
Spain
Phone: +34-935-877-605
United Arab Emirates
Phone: +971-4-8991-777
United Kingdom
Phone: +44-7919-382-156
United States
Houston, Texas
Phone: +1-713-966-3600
This manual suits for next models
1
Table of contents
Other Baker Hughes Transducer manuals