3 Points Calibration
Make sure that you have selected 3 points calibration in the setup menu.
3.1 Repeat steps 1.1 through 1.3 above. When the first calibration point is completed, the display will show "pH 4.01/CAL 2", the tester prompts
you to continue with second point calibration.
3.2 Rinse the electrode with deionized water and place into the pH 4.01 buffer solution. Stir the tester gently.
3.3 Press the Enter key to begin the calibration. When the reading has stabilized, the display will show the pH slope and "pH 10.01/CAL 3"
(or pH 9.18/CAL 3).
3.4 Rinse the electrode with deionized water and place into the pH 10.01 (or 9.18) buffer solution. Stir the tester gently. When the reading
has stabilized, the display automatically shows the pH slope and "End". Calibration is completed.
•During the calibration, if the display shows indicating that the measured mV value for the current calibration point deviates by
more than 60 mV ( approximately ± 1 pH ) from the theoretical value of the pH buffer. The calibration will not be accepted. Please check the
electrode and ensure the buffer solutions are fresh and uncontaminated.
•To exit the calibration without saving calibrated values, press the Meas key.
Temperature Calibration
The PHscan30 tester has a built-in temperature sensor that used for automatic temperature compensation (ATC). Durning the measurement, if
the measured temperature reading differs from that of an accurate thermometer, the tester needs to be calibrated.
1. Press and hold the Cal key to enter the setup menu.
2. Press the Enter key until the display shows /or /.
3. Press the Cal key, the tester enters the temperature calibration mode.
4. Place the electrode into a solution with a known accurate temperature and wait for measurement is stable.
5. Press the Cal key to modify the temperature value.
6. Press the Enter key to save and press the Meas key to return to the pH measurement mode.
Measurement
Rinse the electrode with deionized water. Place the electrode into the sample solution and stir gently. Note that the end of the electrode must be
completely immersed into the solution. Wait for the measurement to stabilize and record the reading.
•During the measurement, never wipe the pH-sensitive glass membrane as this will cause static interference, blot dry with a lint-free tissue
to remove waterdrops on electrode.
•If the display shows "----" indicating the measurement exceeds the range, remove the tester from the sample immediately.
•If the option is enabled in the setup menu, the tester will automatically lock a measurement endpoint and show icon.
Press the Meas key to resume measuring.