Pump Head Operation
Pump Head Installation:
The pump head is mounted on drive before the pump leaves the factory.
Follow below procedure for changing another pump head.
1.Loose the three M3 screws which connect the pump head and the pump
head bracket. Dismantle the pump head slightly.
2.Put the new pump head in the old pump head location. Pay attention to
the inlet and outlet locations of the pump head. Use three M3 screws to
connect the pump head and the pump head bracket.
Turn off the power supply before changing the pump head.
Pump Heads and Flow rates
85.7~857.1
171.4~1714.3
257.1~2571.4
MS(MG)204
MS(MG)209
MS(MG)213
Acceptable
Pump Heads
Flow Rates(mL/min)
0 00(rpm)30 30~
Part Number
05015010001
05015010002
05015010003
MG MS()Strong Magnet
05015010011
05015010012
05015010013
Pump Head Setting
Press the to highlight Pump Head line, press again to enter pump
head selection interface, turn to select the pump head, press the
for confirmation and return to previous menu, or press not to store
the selection and return to previous menu.
Knob Knob
Knob Knob
Exit Key
Flow Mode Operation
Enter Flow Mode
Press and turn the to highlight control mode line. Press to enter
next interface, turn to select control mode, press for confirmation.
Press not to store the selection and return to previous menu.
Knob Knob
Knob Knob
Exit Key
Flow Control: Turn the to adjust the flow rates.Knob
Flow Calibration
Press and turn to highlight calibration line. Press to enter next
interface. Turn to select . Press to highlight test time
line. Press again to highlight the value of time. Turn to select
test time (its range is from 30 to 1800 seconds), press for
confirmation. Press not to store the selection and return to
previous menu.
Press , the LCD displays running time and actual volume. Running
time counts down. Turn to input actual test volume. Press for
confirmation. The calibration operation can be repeated.
Knob Knob
Knob Flowrate Knob
Knob Knob
Knob
Exit Key
Run Key
Knob Knob
5 6
Micro Gear Pump Drive Operating ManualMicro Gear Pump Drive Operating Manual
If the actual volume is known in advance, enter the calibration interface and
input the actual volume directly.