BC Group International BC Biomedical DA-2006-VL User manual

DEFIBRILLATOR ANALYZER
VARIABLE LOAD
DA-2006-VL
USER MANUAL


i
WARNINGS, CAUTIONS, NOTICES ............................................................................ ii
DESCRIPTION ............................................................................................................. 1
LAYOUT ....................................................................................................................... 2
INSTRUCTIONS……………………………………………………………………………... 3
MANUAL REVISIONS .................................................................................................. 4
LIMITED WARRANTY ................................................................................................. 4
SPECIFICATIONS ....................................................................................................... 5
NOTES ......................................................................................................................... 6
BC BIOMEDICAL
DA-2006-VL
TABLE OF CONTENTS

ii
WARNING - USERS
The DA-2006-VL is for use by
skilled technical personnel only.
WARNING - USE
The DA-2006-VL is intended for testing only and
they should never be used in diagnostics,
treatment or any other capacity where they would
come in contact with a patient.
WARNING - CONNECTIONS
All connections to patients must be removed
before connecting the DUT to the DA-2006-VL. A
serious hazard may occur if the patient is
connected when testing with the Analyzer. Do
not connect any leads from the patient directly to
the unit or DUT.
WARNING - LIQUIDS
Do not submerge or spill liquids on the DA-2006-
VL. Do not operate the unit if it may have been
exposed to fluid.
WARNING - MODIFICATIONS
The DA-2006-VL is intended for use within the
published specifications. Any application
beyond these specifications or any unauthorized
user modifications may result in hazards or
improper operation.
CAUTION - CLEANING
Do not immerse. The DA-2006-VL should be
cleaned by wiping gently with a damp, lint-free
cloth.
A mild detergent can be used if desired.

iii
CAUTION - ENVIRONMENT
Exposure to environmental conditions outside
the specifications can adversely affect the
performance and accuracy of the DA-2006-VL. If
the unit is outside the Operating Specifications,
allow it to acclimate to specified conditions for at
least 30 minutes before attempting to operate it.
CAUTION - AC ADAPTOR
Unplug the AC Adaptor before
cleaning the surface of the DA-2006 Series
Analyzers.
CAUTION - SERVICE
The DA-2006-VL is intended to be serviced only
by authorized service personnel.
Troubleshooting and service procedures should
only be performed by qualified technical
personnel.
CAUTION - INSPECTION
The DA-2006-VL should be
inspected before each use for wear and should
be serviced if any parts are in question.
NOTICE – INDICATIONS FOR USE
The DA-2006-VL is used to determine that
defibrillators are performing within their
performance specifications by providing multiple
loads of 25, 50, 75, 100, 125, 150, 175 and 200
Ohms. The DA-2006-VL is used in conjunction
with the DA-2006/DA-2006P.

iv
NOTICE – SYMBOLS
Symbol
Description
Caution
(Consult Manual for Further Information)
Per European Council Directive 2002/95/EC,
do not dispose of this product as unsorted
municipal waste.
Hazardous Voltage
CAT I
IEC Measurement Category I – CAT I
equipment designed to protect against
transients in equipment on circuits not directly
connected to MAINS. Under no circumstances
should the terminals of the Load be connected
to any MAINS voltage
CAUTION - INSPECTION
The DA-2006-VL should be inspected before each
use for obvious signs of abuse or wear. The DA-
2006-VL should not be used and should be
serviced if any parts are in question.
NOTICE – ABBREVIATIONS
C
Celsius
°
degree(s)
F
Fahrenheit
Hz
hertz
IEC
International Electrotechnical Commission
Lbs
Pounds
LED
Light Emitting Diode
mm
millimeter(s)
PSID
Pounds per Square Inch Differential
sec
seconds
USA
United States of America

v
NOTICE – CE
The DA-2006-VL bears the mark
Based on the following testing standards:
ELECTROMAGNETIC COMPATIBILITY DIRECTIVE
EN 61326-1:2006
According to the Guide for the EMC Directive 2004/108/EC, section 1.1.4 the DA-
2006-VL is inherently benign equipment and does not require EMC testing.
LOW VOLTAGE DIRECTIVE
EC – Directive 73/23/EC
EN 61010-1:2001
“Safety requirements for electrical equipment for measurement, control, and
laboratory use – General requirements”
This equipment has been type tested and compliance was demonstrated
to the above standard to the extent applicable.

vi
DA-2006-VL User Manual Copyright © 2014
www.bcgroupintl.com Made in the USA
09/14 Rev 01
NOTICE – CONTACT INFORMATION
BC BIOMEDICAL
BC GROUP INTERNATIONAL, INC.
3081 ELM POINT INDUSTRIAL DRIVE
ST. CHARLES, MO 63301
USA
1-800-242-8428
1-314-638-3800
www.bcgroupintl.com
sales@bcgroupintl.com
NOTICE – DISCLAIMER
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO
MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT
ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE
DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST
POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS
BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE
ACCURATE. HOWEVER, NO RESPONSIBILITY IS ASSUMED FOR
INACCURACIES.
NOTICE – DISCLAIMER
BC GROUP INTERNATIONAL, INC. WILL NOT BE RESPONSIBLE
FOR ANY INJURIES SUSTAINED DUE TO UNAUTHORIZED
EQUIPMENT MODIFICATIONS OR APPLICATION OF EQUIPMENT
OUTSIDE OF THE PUBLISHED INTENDED USE AND
SPECIFICATIONS.

vii
This Page Intentionally Left Blank

1
The Model DA-2006-VL is an adjustable load box used in conjunction with the DA-2006
Series Analyzers in the testing of defibrillators. It allows for analysis of defibrillators over a
25-200 ohm range with 25 ohm steps.
The following are highlights on some of the main features:
DA-2006-VL (Basic Features):
SIMPLE TO OPERATE
PULSED BIPHASIC, BIPHASIC, AND MONOPHASIC COMPATIBLE
FULLY AED COMPATIBLE
5000 V, 1000 J CAPACITY
SMART LOADS, NO SETTINGS TO CHANGE IN DA-2006 OR DA-2006P
BC BIOMEDICAL
DA-2006-VL
DEFIBRILLATOR ANALYZER VARIABLE LOAD

2
This section looks at the layout of a DA-2006-VL and gives descriptions of the elements
that are present.
LAYOUT
Large Defib Plates:
With banana jack
connectors for
paddle free use.
Load Selection Knob
25-200 Ohm loads with
25 Ohm increments
DA-2006 series interface connections
Banana connections for Sternum/Apex
jumpers. Mini din connector for
interface control.

3
Connecting to a DA-2006 series analyzer
This unit is connected to the DA-2006 series analyzer through three included cables,
BC20-00139. The two black jumper cables run from the Sternum/Apex jacks on the top of
the DA-2006-VL to the corresponding defib plates on the DA-2006 Series Analyzer as
shown below. The Communication cable runs from the Mini-din connector on the top of the
DA-2006-VL to the printer port on the rear of the DA-2006 Series Analyzer.
Once connected, the display on the DA-2006 Series Analyzer will show the current variable
load setting corresponding to the dial on the side of the DA-2006-VL
INSTRUCTIONS FOR USE
Load Setting
The DA-2006 Series Analyzer
automatically detects the DA-2006-VL
load setting (requires firmware
DT7395CH or later)

4
Revision # Revisions Made
Rev 01 Origination
MANUAL REVISIONS
LIMITED WARRANTY
WARRANTY: BC GROUP INTERNATIONAL, INC. WARRANTS ITS NEW PRODUCTS TO BE FREE
FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY
ARE INTENDED. THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF
SHIPMENT.
EXCLUSIONS: THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR
IMPLIED, INCLUDING, BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY
OR FITNESS FOR A PARTICULAR PURPOSE.
BC GROUP INTERNATIONAL, INC. IS NOT LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL
DAMAGES.
NO PERSON OTHER THAN AN OFFICER IS AUTHORIZED TO GIVE ANY OTHER WARRANTY OR
ASSUME ANY LIABILITY.
REMEDIES: THE PURCHASER'S SOLE AND EXCLUSIVE REMEDY SHALL BE: (1) THE REPAIR OR
REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS, WITHOUT CHARGE. (2) AT THE OPTION
OF BC GROUP INTERNATIONAL, INC., THE REFUND OF THE PURCHASE PRICE.

5
DA-2006-VL VARIABLE LOAD
MAXIMUM POWER
12W
10 Defibrillator pulses of 360 J every 5 minutes
RESISTANCE
25 – 200 Ω, in 25 Ω steps
TOLERANCE
±1 %
INDUCTANCE
< 10 μH
DA-2006 HIGH RANGE
DA-2006 LOW RANGE
PEAK VOLTAGE
≤ 5000 V
≤ 1000 V
PEAK CURRENT
≤ 100 A
≤ 20 A
ENERGY
≤ 1000 J
≤ 50 J
ACCURACY
50 Ω
≤ 100 J
± 2 J
≤ 20 J
± 0.4 J
> 100 J
± 2% of
reading
> 20 J
± 2% of
reading
ACCURACY
25, 75-200 Ω
≤ 100 J
± 3 J
≤ 20 J
± 0.6 J
> 100 J
± 3% of
reading
> 20 J
± 3% of
reading
PHYSICAL & ENVIRONMENTAL
SIZE
9.80 x 8.10 x 4.75 Inches
(248.9 x 205.7 x 120.7 mm)
WEIGHT
5 Lbs (2.27 kg)
CONSTRUCTION
ENCLOSURE
ABS Plastic
FACEPLATE
Polycarbonate,
Back printed
OPERATING RANGE
15° to 40° C (59° to 104° F)
STORAGE RANGE
-20° to 65° C (-4° to 149° F)
SPECIFICATIONS