Biocare 3060 User manual

Owner’s Manual
Model : 3060
Blood Pressure Monitor
Wrist Strap Blood Pressure Monitor
Dear 3060 System Owner :
Thank you for purchasing the 3060 Blood Pressure Monitoring System. This manual provides
important information to help you operate this system correctly. Before using this product,
please read the following contents thoroughly and carefully.
With the compact size and easy operation of this 3060 Blood Pressure Monitoring System, you
can easily monitor your blood pressure by yourself at any time or place. In addition, this
system can help you and your healthcare professionals to monitor and adjust your treatment
plans, and keep your blood pressure under control.
If you have other questions regarding this product, please contact the place of purchase.
IMPORTANT SAFETY PRECAUTIONS READTHIS BEFORE USE
1. Use this device ONLY for the intended use described in this manual.
2. Do NOT use accessories which are not specied by the manufacturer.
3. Do NOT use the device if it is not working properly or damaged.
4. Do NOT use the equipment where aerosol sprays are being used, or where oxygen is being
administered.
5. Do NOTuse under any circumstances on newborns.
6. This device does NOT serve as a cure for any symptoms or diseases.The data measured are
for reference only. Always consult your doctor to have the results interpreted.
7. Keep the equipment and its exible cord away from hot surfaces.
8. Do NOT apply the cu to areas other than the place directed.
9. Do NOT remove plastic wrapping membrane to ensure the waterproof function.
10.proper maintenance and periodically calibration are essential to the longevity of your
device. If you are concerned about your accuracy of measurement, please contact local
customer service for help.
KEEP THESE INSTRUCTIONS
BEFORE YOU BEGIN
⊙Intended Use
The 3060 system is a system designed to measure the systolic and diastolic blood pressure and
pulse rate by using a non-invasive technique in which an inatable cu is wrapped on the
wrist. It is intended for individuals age 18 and above for use at home and in clinical settings.
The device is not to be used for the diagnosis or screening of hypertension or for testing on
newborns.
⊙ Test Principle
Blood pressure is measured non-invasively at the wrist based on oscillometric method.
This device is NOT able to take measurements in the presence of common
arrhythmia, such as atrial or ventricular premature beats or atrial brillation. It may
produce reading error.
MONITOR OVERVIEW
Do not use the accessory or equipments which are not certified by Manufacture. It
might increased emissions or decreased immunity of this system.
DISPLAY SCREEN
SETTING THE MONITOR
Before using your monitor for the rst time, check and update these settings. Make sure you
complete the steps below and save the desired settings.
Entering the Setting Mode
Start with the monitor o. Press and rmly hold M/SET for 3 seconds until the“2012” ashes on
the upper right corner.
If the meter is idle for 1.5 minutes, it turns to no time setting:“--/--/--:--”.
Setting the Date and the Time
With the year ashing, press M/SET button until the correct year appears.
Press CAM button to set.
With the month ashing, press M/SET button until the correct month appears.
Press CAM button to set.
With the day ashing, press M/SET button until the correct day appears.
Press CAM button to set.
With the hour ashing, press M/SET button to select the desired time system 24hr or 12hr,
AM/PM only displays while the 12hr system is selected. Press CAM button to set.
With the minute ashing, press M/SET button until the correct minute appears.
Press CAM button to set and exit from date/time setting.
NOTE
1. If the meter is idle for 1.5 minutes during the setting mode, it will switch o automatically
Setting the Alarm
With device in standby status, press and rmly hold M/SET
button and CAM button at the same time for 2 seconds until
an alarm icon ashes on the upper right corner.
Release the both buttons and then “1” will ash on the screen,
press M/SET button to switch between alarm“1” and“2”.
Press CAM button with the hour ashing, press M/SET button
until the desired hour appears. Press CAM button to set.
With the minute ashing, press M/SET button until the desired
minute appears.
Press CAM button to set. The meter will switch o automatically.
At the time of your alarm, the meter will beep and automatically
turn on. You can press every button to silence the alarm. When
the alarm “1” or ”2”ashing, press CAM button to switch o the
alarm, the display “ON” changed to be“OFF”
USER NUMBER SELECTION
This system stores blood pressure measurements for up to two users.
Each user has 60 memories stored separately.
1. With device in standby status, user 1 is the default user. Press START
button and M/SET button for 1 second till User“1” icon hides and User“2” icon shows.
2. When User“2” icon displays, Press START button and M/SET button for 1 second again, User
“1” icon displays and hides User “2”icon.
➊
➋
➍
➎
➏ ➐
➑ ➒
➓
➌
Owner’s Manual
TESTING YOUR BLOOD PRESSURE
BEFORE MEASUREMENT
⊙Avoid caeine, tea, alcohol and tobacco for at least 30 minutes before measurement.
⊙Wait 30 minutes after exercising or bathing before measurement.
⊙Sit or lie down for at least 10 minutes before measurement.
⊙Do not measure when feeling anxious or tense.
⊙Take a 5-10 minute break between measurements. This break can be longer if necessary,
depending on your physical condition
⊙Keep the records for your doctor as reference.
⊙Blood pressure naturally varies between each arm. Always measure your blood pressure on
the same wrist.
CORRECT APPLICATION OF THE PRESSURE CUFF
Slide the cu onto your wrist. Wrap the cu comfortably
around your wrist.
The top margin of the meter should be about 0.6 cm to
1.3 cm (1/4 to 1/2 inch) below the ball of thumb. When the
cu is positioned correctly, fold the loose end back and
secure it in place by pressing the pile material rmly
against the hook material.
Warning!!
Always apply the wrist cu before turning on the device. Failure
to do so will damage the device.
TESTING YOUR BLOOD PRESSURE
Step1 Sit down for at least 10 minutes before measuring
Step2 Apply the wrist cu
Step3 Press the POWER button
and place your arm in front of your chest. Relax and make sure
the device is at the same height as your heart. Holding the
elbow of the arm wearing the wrist cu with your other hand
helps you hold still during measurement. Remain still and do
not talk or move during the measurement.
TAKING MEASUREMENTS
TAKING A SINGLE MEASUREMENT
Always apply the pressure cu before turning on the monitor.
1.Touch the Power button to turn the monitor on, move the device
to be the same height as the heart. When the position is correct, a
red bar and a yellow bar ash on the display and device starts
automatically. Otherwise, two red bars ashing indicate the position
is too high while two green bars ashing to prompt too low.
2. “0” displays on LCD which means device is under auto zero resetting.
3. Pump starts to inate and pressure in cu increases
to working pressure.
4. The heart symbol“ ” will show silently when pulse is
detected during the ination. When signal of pulse becomes
strong,“ ” icon starts to ash.
Owner’s Manual
5. After the measurement, the monitor displays
the systolic pressure, diastolic pressure and pulse
rate with a long beeper sounds. The reading is
classied by color bar which divided according
to WHO classication of blood pressure.
If IHB is detected, LCD displays the measurement
with“ ” icon ashing 8 times to remind users of
irregularheartbeat.
Note
⊙If you press Power button during measurement, the monitor will be turned o.
⊙If the pulse rate symbol is shown as“ ” instead of “ ”, this indicates that the monitor
has detected an irregular heart beat.
PROSMEASUREMENT MODE
Always apply the pressure cu before turning on the
monitor.
1.Press CAM button to begin. move the deviceto be
same height as the heart, when the position is correct,
a red bar and a yellow bar ash on the display and device
starts automatically. Otherwise, two red bars ashing
indicates the position is too high while two green bars
ashing means too low.“0” displays on LCD which means
device is under auto zero resetting. LCD shows in
Pros mode;“1” means 1st measurement.
2. Pump starts to inate and pressure in cu increases to
working pressure.
3. The heart symbol“ ” will show silently when pulse is
detected during the ination. When signal of pulse
becomes strong, “ ” icon starts to ash and a beep
is heard for every heartbeat.
4. After the 1st measuring, cu deates completely and
device comes into 60 seconds count down.
5. Pump starts to inate and pressure in cu increases to
working pressure.
6. The heart symbol“ ” will show silently when pulse is
detected during the ination. When signal of pulse
becomes strong,“ ”icon starts to ash and a beep is
heard for every heartbeat.
Note
⊙ If the reading meet the condition of |SYS BP1-SYS BP 2| <14 mmHg and |Dia BP1 - Dia BP2|
<9 mmHg as well as pulse rate> 60 for 1st and 2nd measurement, the whole Pros measure-
ment nished after the two measurements and the result shows on LCD.
⊙ If the reading does not meet the mentioned condition, it will continue to the third turn
measurement.
7 . After the 2nd measuring, cu deates completely and
device comes into 60 seconds count down.
8. Pump starts to inate and pressure in cu increases to
working pressure.
Owner’s Manual
1. WHO classification indicator
2. User Number
3. Night-time Average
4. Day-time Average
5. Number of memory
6. Memory recall
7. CAM Mode
8. Date
9. Time
10. Daily Alarm
11. Irregular heartbeat indicator
12. Pulse rate
13. Low battery indicator
14. Diastolic pressure reading
15. Systolic pressure reading
➊ AC/DC Adaptor Socket
➋M/SET Buttton
➌ CAM Button
➍ Wrist Cuff
➎ Display Screen
➏ Power Button
➐ Battery compartment

TROUBLESHOOTING
1. If no display appears after pressing the POWER button :
2. If the heart rate is higher/lower than user’s average :
3. If the result is higher/lower than user’s average measurement :
4. If the cu inates again while measuring :
DETAILED INFORMATION
REFERENCE VALUES
Human blood pressure naturally increases after reaching middle age.This symptom is a result
of continuous ageing of the blood vessels. Further causes include diabetes, lack of exercise
and cholesterol (LDL) adhering to the blood vessels. Rising blood pressure accelerates
hardening of the arteries, and the body becomes more susceptible to apoplexy and coronary
infarction.
Denitions and Classication of blood pressure levels according to 2007 ESH-ESC Practice
Guidelines for the Management of Arterial Hypertension :
Isolated systolic hypertension should be graded (1, 2, 3) according to systolic blood pressure values in the
ranges indicated, provided that diastolic values are < 90mmHg.
Source: The European Society of Hypertension and European Society of CardiologyTask Force Members. 2007
ESH-ESC Practice Guidelines for the Management of Arterial Hypertension. J Hypertens 2007; 25: 1751-1762.
SYMBOL INFORMATION
Owner’s Manual
SPECIFICATIONS
SYSTEM PERFORMANCE
BLOOD PRESSURE MEASUREMENT PERFORMANCE
This device has been tested to meet the electrical and safety requirements of: IEC/EN 60601-1, IEC/EN
60601-1-2
Reference to Standards:
• EN 1060-1 /-3, NIBP-requirements
• IEC60601-1 General requirement for safety
• IEC60601-1-2 Requirements for EMC
• EN1060-4, NIBP clinical investigation
•AAMI / ANSI / IEC 80601-2-30, ANSI/AAMI/ISO 81060-2, NIBP requirements
Owner’s Manual
9. The heart symbol“ ” will show silently when pulse
is detected during the ination.
When signal of pulse becomes strong, “ ”icon starts to
ash and a beep is heard for every heartbeat.
Note
If the reading meet the condition of |SYS BP1-SYS BP 2| >15mmHg or |Dia BP1 - Dia BP2| >10
mmHg as well as pulse rate> 60 for 1st and 2nd measurement, the whole Pros function
nished with 3 times measurement and the result shows on LCD.
10. Pros averaged values and the corresponding
color bars----WHO classication of blood pressure
---- will be displayed on LCD.
⊙ If arrhythmia is detected, LCD displays the measurement
with “ ” icon ashing 8 times to remind users of irregular
heartbeat (show IHB icon whenever pulse less
than 42 beat / min-instead of former 48)
Note
If the meter is idle for 1.5 minutes, it will switch o automatically.
AVERAGING RESULT MODE
Press M/SET button, the average of all storage shows on LCD.
Press M/SET button again, the average of storage measured in day shows on LCD.
Press M/SET button one more, the average of storage measured at night shows on LCD.
MEMORY RECALL AND DELETION
The device stores 90 most recent blood
pressure test results along with respective dates and times in the monitor memory for each
user.
To recall the memory, touch the Power button toturn the monitor on.
Reviewing Test Results
1. Press M/SET button, the single memory reading appears.
Per the setting, the larger memory number indicates the
newer record.
Note
will show on LCD if the measurement is taken
under CAM mode.
2. Press M/SET button again to display the next memory.
Review all test results stored in the monitor.
3. Delete the memory
With device in standby status, hold down M/SET button for 7
seconds until “CL”ashes on LCD, press M/SET button to
conrm with two short beeps and all memories removed.
4. Exit the memory
Press Power button and the monitor will turn o.
NOTE
The deleting operation will delete all the readings in memory, not only single reading.
MAINTENANCE
BATTERY
Your monitor comes with one 4.2V Li-on rechargeable battery.
LOW BATTERY SIGNAL
The monitor will display the message below to alert you when the monitor power is getting
low.
The symbol appears along with display messages:The monitor is functional and the
result remains accurate, but it is time to recharge the battery.
RECHARGE THE BATTERY
1. Battery charge display
It is time to charge the battery. Plug USB in to enter battery charge mode.
NOTE
⊙ Recharging of the battery does not aect the test results stored in memory.
⊙ Recharging of the battery does not aect any function other than blood pressure easuring.
⊙ Measuring during battery recharging in prohibited.
CARING FOR YOUR MONITOR
To avoid the monitor attracting dirt, dust or other contaminants, wash and dry your hands
thoroughly before use.
Owner’s Manual Owner’s Manual
CLEANING
⊙To clean the monitor exterior, wipe it with a cloth moistened with tap water or a mild
cleaning agent, then dry the device with a soft dry cloth. Do NOT ush with water.
⊙Do NOT use organic solvents to clean the monitor.
⊙Do NOT wash the pressure cu.
⊙Do NOT iron the pressure cu.
MONITOR STORAGE
⊙ Storage condition: -20oC to 55oC (-4oF to 131oF), below 95% relative humidity.
⊙Always store or transport the monitor in its original storage case.
⊙Avoid dropping or heavy impact.
⊙Avoid direct sunlight and high humidity.
No modications of 3060 Blood Pressure Monitoring System allowed by persons other
than the manufacturer.
DETAILED INFORMATION
REFERENCE VALUES
Clinical studies show that adult diabetes is often accompanied by elevated blood pressure.
People with diabetes can reduce their heart risk by managing their blood pressure along with
diabetes treatment*4.
Monitoring your routine blood pressure trend helps you to know your body condition. Human
blood pressure naturally increases after reaching middle age.This symptom is a result of
continuous ageing of the blood vessels. Further causes include obesity, lack of exercise and
cholesterol (LDL) adhering to the blood vessels. Rising blood pressure accelerates hardening
of the arteries, and the body becomes more susceptible to apoplexy and coronary infarction.
The recommended blood pressure range is as below :
*2: National Heart, Lung, and Blood Institute, Diseases and Conditions
*3: The Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation,
and Treatment of High Blood Pressure. NIH Publication. 2003. No. 03-5233
*4: American Diabetes Association: The Diabetes-Heart Disease Link Surveying Attitudes,
Knowledge and Risk (2002)
SYSTEM TROUBLESHOOTING
If you follow the recommended action but the problem persists, or error messages other than
the ones below appear, please call your local customer service. Do not attempt to repair by
yourself and never try to disassemble the monitor under any circumstances.
ERROR MESSAGES
Classication
Hypotension*2
Normal*3
Pre-hypertension*3
Stage 1 Hypertension*3
Stage 2 Hypertension*3
Less than 90
Less than 120
120 – 139
140 – 159
160 or more
Less than 60
Less than 80
80 – 89
90 – 99
100 or more
Systolic Pressure (mmHg) Diastolic (mmHg)
MESSAGE CAUSE WHAT TO DO
Blood pressure
measurement error
No pulse detected
Problems with the
monitor
Relax and repeat the measurement.
The position is moved
when the measurement
is taking.
Improper movement
detected during
measurement.
Ination or pressure
error.
Relax and repeat the measurement.
Remain still and do not talk or move
during the measurement.
Ret cu tightly and correctly. Relax and
repeat the measurement. If error still
remains, contact local customer service
for help.
Ret cu tightly and correctly. Relax and
repeat the measurement. If error still
remains, contact local customer service
for help.
Ret cu tightly and correctly. If error still
remains, contact local customer service
for help.
Review the instructions and repeat the
test. If the monitor still does not work,
please contact the local customer service
for help.
POSSIBLE CAUSE
Batteries exhausted.
Batteries incorrectly installed or absent.
POSSIBLE CAUSE
Movement during measurement.
Measurement taken just after exercise.
POSSIBLE CAUSE
May not be in correct position while
measuring.
Blood pressure naturally varies from
time to time.
POSSIBLE CAUSE
Cu is not fastened.
If user’s blood pressure is higher than the pressure the device has inated, the
device will automatically increase the pressure and start to inate again. Stay
relaxed and wait for the measurement.
WHAT TO DO
Replace the batteries.
Check that the batteries are correctly
installed.
WHAT TO DO
Repeat measurement.
Rest at least 30 minutes beforehand.
WHAT TO DO
Adjust to the correct position to
measure.
Keep in mind for next measurement.
WHAT TO DO
Fasten the cu again.
Category
Optimal
Normal*3
High normal
Grade 1 hypertension
Grade 2 hypertension
Grade 3 hypertension
Isolated systolic hypertension
Systolic (mmHg)
< 120
120–129
130–139
140 – 159
160–179
≥ 180
≥ 140
and
and/or
and/or
and/or
and/or
and/or
and
Diastolic (mmHg)
< 80
80–84
85–89
90 – 99
100–109
≥ 110
< 90
Symbol Referent Symbol Referent
Consult instructions for use Author ized representative i n
the European Community
Manufacturer Class II
Humidity limitation
Water-proof grade
CE mark
Temperature limitation
Keep dryKeep away from sunlight
Serial number
Caution, consult
accompanying documents
Type BF Equipment
Dispose of in accordance with
the requirements of your
country.
Power source
Size of monitor w/o cuff
Cuff type
Memory
Power saving
System operating condition
Meter storage condition
Pressure range
Heart rate range
Systolic Measurement Range
Diastolic Measurement Range
Pulse Rate Measurement Range
Maximum inflation pressure
Accuracy of Pressure
Accuracy of Pulse rate
Measurement unit
4.2V Li-ion
82.5 (L) x 68.1 (W) x 29.8 (H) mm,134 with cuff
soft wide range 12-21 cm (4.7”-8.3” inches)
with air tube 70 cm
Maximum 60 memory records for each user
Automatic power off if system idle for 1.5 minutes
10oC to 40oC (50oF to 104oF), 15-95% RH
0 ~ 3000 meter height (700hPa-986hPa)
-20oC to 55oC (-4oF to 131oF), 15-95% RH
20 - 280 mmHg
40 - 200 beat per minute
50 mmHg -250 mmHg
30 mmHg -180 mmHg
40 -200 beats / minute
280 mmHg
±3 mmHg or ±2% of the reading
±4% of the reading
mmHg
Shanghai International Trading Corp.
GmbH(Hamburg)
Eiffestrasse 80, 20537 Hamburg, Germany
TEL:+49‐40‐2513175
FAX:+49‐40‐255726
Biocare Asia Co., LTD
No.260,Mayun Road, New District Suzhou,
Jiangsu,215129,P.R.C
TEL:+86‐512‐68091772
FAX:+86‐512‐68097761
www.bca-medical.com