BIOCLEAR Blaster BC-B5000 User manual

INSTRUCTION MANUAL
Air Polisher
BC-B5000 / BC-B5010 / BC-B5020 / BC-B5030 / BC-B5040
INSTRUCTION MANUAL
Air Polisher
BC-B5000 / BC-B5010 / BC-B5020 / BC-B5030 / BC-B5040

Thank you for purchasing the air polisher “The BLASTER”.
For optimum performance of your polisher, read this instruction manual carefully and thoroughly before use.
Mishandling of this instrument may cause mechanical trouble and accidents during treatment.
Keep this instruction manual for future reference.
Specifi cations in this manual may be changed for improvement and modifi cation without prior notice.
* The BLASTER is indicated as “polisher” in this instruction manual.
* MULTIfl ex® is a registered trademark of Kaltenbach & Voigt GmbH & Co. Kg, Germany.
Roto Quick® is a registered trademark of W&H Dentalwerk Bürmoos GmbH, Austria.
Sirona® is a registered trademark of Sirona Dental Systems GmbH, Germany.
Unifi x® is a registered trademark of Bien-Air Dental S.A., Switzerland.
Mach® and Phatelus® are registered trademarks of NAKANISHI INC., Japan.
Thank you for purchasing the air polisher “The BLASTER”.
For optimum performance of your polisher, read this instruction manual carefully and thoroughly before use.
Mishandling of this instrument may cause mechanical trouble and accidents during treatment.
Keep this instruction manual for future reference.
Specifi cations in this manual may be changed for improvement and modifi cation without prior notice.
* The BLASTER is indicated as “polisher” in this instruction manual.
* MULTIfl ex® is a registered trademark of Kaltenbach & Voigt GmbH & Co. Kg, Germany.
Roto Quick® is a registered trademark of W&H Dentalwerk Bürmoos GmbH, Austria.
Sirona® is a registered trademark of Sirona Dental Systems GmbH, Germany.
Unifi x® is a registered trademark of Bien-Air Dental S.A., Switzerland.
Mach® and Phatelus® are registered trademarks of NAKANISHI INC., Japan.

CONTENTS
1. PREFACE------------------------------------------------------------------------------------------------------------------------------------ 1
2. CAUTIONS FOR USE --------------------------------------------------------------------------------------------------------------------- 2
3. INTENDED USE----------------------------------------------------------------------------------------------------------------------------- 4
4. CHARACTERISTICS----------------------------------------------------------------------------------------------------------------------- 4
5. DELIVERY PACKAGE-------------------------------------------------------------------------------------------------------------------- 5
6. PRODUCT COMPONENTS------------------------------------------------------------------------------------------------------------- 6
7. HOW TO USE ------------------------------------------------------------------------------------------------------------------------------ 7
7.1 Before use---------------------------------------------------------------------------------------------------------------------------- 7
7.2 Air pressure setting (at fi rst use)----------------------------------------------------------------------------------------------- 8
7.3 Attaching head-------------------------------------------------------------------------------------------------------------------10
7.4 Attaching powder cap --------------------------------------------------------------------------------------------------------11
7.5 Connecting to dental unit ----------------------------------------------------------------------------------------------------13
7.6 Water removal in channels (idle run) -------------------------------------------------------------------------------------15
7.7 Powder fi lling-----------------------------------------------------------------------------------------------------------------------16
7.8 Confi rming correct operation and water volume--------------------------------------------------------------------19
7.9 Treatment ---------------------------------------------------------------------------------------------------------------------------20
7.10 Other remarks on use ---------------------------------------------------------------------------------------------------------24
8. MAINTENANCE AND CHECKS------------------------------------------------------------------------------------------------------26
8.1 Regular maintenance and checks ----------------------------------------------------------------------------------------26
8.2 Cleaning at troubles in air and powder spray -------------------------------------------------------------------------29
9. CLEANING AND STERILIZATION ----------------------------------------------------------------------------------------------------31
9.1 Cleaning and disinfection ----------------------------------------------------------------------------------------------------31
9.2 Sterilization --------------------------------------------------------------------------------------------------------------------------31
9.3 Storage-------------------------------------------------------------------------------------------------------------------------------32
10.REPLACEMENT OF CONSUMABLE PARTS, DISPOSAL ----------------------------------------------------------------------34
10.1 Replacement of head--------------------------------------------------------------------------------------------------------34
10.2 Replacement of head o-ring-----------------------------------------------------------------------------------------------34
10.3 Replacement of powder cap ---------------------------------------------------------------------------------------------34
10.4 Replacement of powder seal----------------------------------------------------------------------------------------------35
10.5 Disposal ----------------------------------------------------------------------------------------------------------------------------35
11.TROUBLE SHOOTING------------------------------------------------------------------------------------------------------------------36
12.SPECIFICATIONS -----------------------------------------------------------------------------------------------------------------------38
13.COMPONENT PARTS------------------------------------------------------------------------------------------------------------------39
CONTENTS
1. PREFACE------------------------------------------------------------------------------------------------------------------------------------ 1
2. CAUTIONS FOR USE --------------------------------------------------------------------------------------------------------------------- 2
3. INTENDED USE----------------------------------------------------------------------------------------------------------------------------- 4
4. CHARACTERISTICS----------------------------------------------------------------------------------------------------------------------- 4
5. DELIVERY PACKAGE-------------------------------------------------------------------------------------------------------------------- 5
6. PRODUCT COMPONENTS------------------------------------------------------------------------------------------------------------- 6
7. HOW TO USE ------------------------------------------------------------------------------------------------------------------------------ 7
7.1 Before use---------------------------------------------------------------------------------------------------------------------------- 7
7.2 Air pressure setting (at fi rst use)----------------------------------------------------------------------------------------------- 8
7.3 Attaching head-------------------------------------------------------------------------------------------------------------------10
7.4 Attaching powder cap --------------------------------------------------------------------------------------------------------11
7.5 Connecting to dental unit ----------------------------------------------------------------------------------------------------13
7.6 Water removal in channels (idle run) -------------------------------------------------------------------------------------15
7.7 Powder fi lling-----------------------------------------------------------------------------------------------------------------------16
7.8 Confi rming correct operation and water volume--------------------------------------------------------------------19
7.9 Treatment ---------------------------------------------------------------------------------------------------------------------------20
7.10 Other remarks on use ---------------------------------------------------------------------------------------------------------24
8. MAINTENANCE AND CHECKS------------------------------------------------------------------------------------------------------26
8.1 Regular maintenance and checks ----------------------------------------------------------------------------------------26
8.2 Cleaning at troubles in air and powder spray -------------------------------------------------------------------------29
9. CLEANING AND STERILIZATION ----------------------------------------------------------------------------------------------------31
9.1 Cleaning and disinfection ----------------------------------------------------------------------------------------------------31
9.2 Sterilization --------------------------------------------------------------------------------------------------------------------------31
9.3 Storage-------------------------------------------------------------------------------------------------------------------------------32
10.REPLACEMENT OF CONSUMABLE PARTS, DISPOSAL ----------------------------------------------------------------------34
10.1 Replacement of head--------------------------------------------------------------------------------------------------------34
10.2 Replacement of head o-ring-----------------------------------------------------------------------------------------------34
10.3 Replacement of powder cap ---------------------------------------------------------------------------------------------34
10.4 Replacement of powder seal----------------------------------------------------------------------------------------------35
10.5 Disposal ----------------------------------------------------------------------------------------------------------------------------35
11.TROUBLE SHOOTING------------------------------------------------------------------------------------------------------------------36
12.SPECIFICATIONS -----------------------------------------------------------------------------------------------------------------------38
13.COMPONENT PARTS------------------------------------------------------------------------------------------------------------------39

CONTENTS
1. PREFACE------------------------------------------------------------------------------------------------------------------------------------ 1
2. CAUTIONS FOR USE --------------------------------------------------------------------------------------------------------------------- 2
3. INTENDED USE----------------------------------------------------------------------------------------------------------------------------- 4
4. CHARACTERISTICS----------------------------------------------------------------------------------------------------------------------- 4
5. DELIVERY PACKAGE-------------------------------------------------------------------------------------------------------------------- 5
6. PRODUCT COMPONENTS------------------------------------------------------------------------------------------------------------- 6
7. HOW TO USE ------------------------------------------------------------------------------------------------------------------------------ 7
7.1 Before use---------------------------------------------------------------------------------------------------------------------------- 7
7.2 Air pressure setting (at fi rst use)----------------------------------------------------------------------------------------------- 8
7.3 Attaching head-------------------------------------------------------------------------------------------------------------------10
7.4 Attaching powder cap --------------------------------------------------------------------------------------------------------11
7.5 Connecting to dental unit ----------------------------------------------------------------------------------------------------13
7.6 Water removal in channels (idle run) -------------------------------------------------------------------------------------15
7.7 Powder fi lling-----------------------------------------------------------------------------------------------------------------------16
7.8 Confi rming correct operation and water volume--------------------------------------------------------------------19
7.9 Treatment ---------------------------------------------------------------------------------------------------------------------------20
7.10 Other remarks on use ---------------------------------------------------------------------------------------------------------24
8. MAINTENANCE AND CHECKS------------------------------------------------------------------------------------------------------26
8.1 Regular maintenance and checks ----------------------------------------------------------------------------------------26
8.2 Cleaning at troubles in air and powder spray -------------------------------------------------------------------------29
9. CLEANING AND STERILIZATION ----------------------------------------------------------------------------------------------------31
9.1 Cleaning and disinfection ----------------------------------------------------------------------------------------------------31
9.2 Sterilization --------------------------------------------------------------------------------------------------------------------------31
9.3 Storage-------------------------------------------------------------------------------------------------------------------------------32
10.REPLACEMENT OF CONSUMABLE PARTS, DISPOSAL ----------------------------------------------------------------------34
10.1 Replacement of head--------------------------------------------------------------------------------------------------------34
10.2 Replacement of head o-ring-----------------------------------------------------------------------------------------------34
10.3 Replacement of powder cap ---------------------------------------------------------------------------------------------34
10.4 Replacement of powder seal----------------------------------------------------------------------------------------------35
10.5 Disposal ----------------------------------------------------------------------------------------------------------------------------35
11.TROUBLE SHOOTING------------------------------------------------------------------------------------------------------------------36
12.SPECIFICATIONS -----------------------------------------------------------------------------------------------------------------------38
13.COMPONENT PARTS------------------------------------------------------------------------------------------------------------------39
CONTENTS
1. PREFACE------------------------------------------------------------------------------------------------------------------------------------ 1
2. CAUTIONS FOR USE --------------------------------------------------------------------------------------------------------------------- 2
3. INTENDED USE----------------------------------------------------------------------------------------------------------------------------- 4
4. CHARACTERISTICS----------------------------------------------------------------------------------------------------------------------- 4
5. DELIVERY PACKAGE-------------------------------------------------------------------------------------------------------------------- 5
6. PRODUCT COMPONENTS------------------------------------------------------------------------------------------------------------- 6
7. HOW TO USE ------------------------------------------------------------------------------------------------------------------------------ 7
7.1 Before use---------------------------------------------------------------------------------------------------------------------------- 7
7.2 Air pressure setting (at fi rst use)----------------------------------------------------------------------------------------------- 8
7.3 Attaching head-------------------------------------------------------------------------------------------------------------------10
7.4 Attaching powder cap --------------------------------------------------------------------------------------------------------11
7.5 Connecting to dental unit ----------------------------------------------------------------------------------------------------13
7.6 Water removal in channels (idle run) -------------------------------------------------------------------------------------15
7.7 Powder fi lling-----------------------------------------------------------------------------------------------------------------------16
7.8 Confi rming correct operation and water volume--------------------------------------------------------------------19
7.9 Treatment ---------------------------------------------------------------------------------------------------------------------------20
7.10 Other remarks on use ---------------------------------------------------------------------------------------------------------24
8. MAINTENANCE AND CHECKS------------------------------------------------------------------------------------------------------26
8.1 Regular maintenance and checks ----------------------------------------------------------------------------------------26
8.2 Cleaning at troubles in air and powder spray -------------------------------------------------------------------------29
9. CLEANING AND STERILIZATION ----------------------------------------------------------------------------------------------------31
9.1 Cleaning and disinfection ----------------------------------------------------------------------------------------------------31
9.2 Sterilization --------------------------------------------------------------------------------------------------------------------------31
9.3 Storage-------------------------------------------------------------------------------------------------------------------------------32
10.REPLACEMENT OF CONSUMABLE PARTS, DISPOSAL ----------------------------------------------------------------------34
10.1 Replacement of head--------------------------------------------------------------------------------------------------------34
10.2 Replacement of head o-ring-----------------------------------------------------------------------------------------------34
10.3 Replacement of powder cap ---------------------------------------------------------------------------------------------34
10.4 Replacement of powder seal----------------------------------------------------------------------------------------------35
10.5 Disposal ----------------------------------------------------------------------------------------------------------------------------35
11.TROUBLE SHOOTING------------------------------------------------------------------------------------------------------------------36
12.SPECIFICATIONS -----------------------------------------------------------------------------------------------------------------------38
13.COMPONENT PARTS------------------------------------------------------------------------------------------------------------------39
1
1. PREFACE
The defi nitions in this instruction manual of the words preceded by the symbol are as follows. To avoid dangers, observe the
instructions preceded by these words.
: This indicates not specifi c general attentions, cautions and dangers.
CONTRAINDICATION AND RESTRICTION:
This indicates exemptions from responsibilities such as improper use and excess of design limit.
DANGER: This indicates potential direct dangers that may cause serious injuries, death, property damages such as total loss
of the instrument or fi re disaster.
CAUTION: This indicates potential indirect dangers that may cause serious injuries, death, property damages such as total
loss of the instrument or fi re disaster.
ATTENTION: This indicates potential dangers that may cause light to middle injuries, property damages such as partial loss of
the instrument or loss of computer data.
The drawings and illustrations in this instruction manual is BC-B5000 unless otherwise specifi ed.
1
1. PREFACE
The defi nitions in this instruction manual of the words preceded by the symbol are as follows. To avoid dangers, observe the
instructions preceded by these words.
: This indicates not specifi c general attentions, cautions and dangers.
CONTRAINDICATION AND RESTRICTION:
This indicates exemptions from responsibilities such as improper use and excess of design limit.
DANGER: This indicates potential direct dangers that may cause serious injuries, death, property damages such as total loss
of the instrument or fi re disaster.
CAUTION: This indicates potential indirect dangers that may cause serious injuries, death, property damages such as total
loss of the instrument or fi re disaster.
ATTENTION: This indicates potential dangers that may cause light to middle injuries, property damages such as partial loss of
the instrument or loss of computer data.
The drawings and illustrations in this instruction manual is BC-B5000 unless otherwise specifi ed.

2
2. CAUTIONS FOR USE
CONTRAINDICATION AND RESTRICTION
1. Non-eligible patients
1) Never use this product on patients
- - on a low salt diet (due to hypertension, hypernatremia, edema, pregnancy toxemia, etc.),
- - having a serious ulcer in digestive organs,
- - having a renal dysfunction,
- - having cardiac dysfunction,
- - having asthma or chronic pulmonary disease,
- - having electrolyte abnormality due to hypochloric alkalosis.
2) Avoid using this product on patients below, other than those stated in above 1):
- - having an intraoral injury or disease,
- - having a hyperemia, infl ammation or bleeding in the intraoral cavity,
- - easily get oral mucosa infl ammation or bleeding,
- - having allergies,
- - wearing contact lenses.
3) Depending on the patient’s physical condition, allergy symptoms or stomatitis may occur. In such a case, stop using this
product.
2. Usage
1) This product must be used only by qualifi ed dental professionals and only for dental treatment.
2) Never dismantle or modify this product.
3) In case of trouble, do not attempt to repair. Return to dealer for repair.
4) In case of deterioration or malfunction, do not use this product.
5) Do not use this product when in an unmaintained condition.
6) To operate this instrument correctly and safely, keep the instruction manual.
7) Use sodium hydrogen carbonate powder for dental prophylaxis only. Other powder (for example, alumina powder) will cause
a decrease in cleaning effi ciency, powder clogging, or damage to the polisher.
2
2. CAUTIONS FOR USE
CONTRAINDICATION AND RESTRICTION
1. Non-eligible patients
1) Never use this product on patients
- - on a low salt diet (due to hypertension, hypernatremia, edema, pregnancy toxemia, etc.),
- - having a serious ulcer in digestive organs,
- - having a renal dysfunction,
- - having cardiac dysfunction,
- - having asthma or chronic pulmonary disease,
- - having electrolyte abnormality due to hypochloric alkalosis.
2) Avoid using this product on patients below, other than those stated in above 1):
- - having an intraoral injury or disease,
- - having a hyperemia, infl ammation or bleeding in the intraoral cavity,
- - easily get oral mucosa infl ammation or bleeding,
- - having allergies,
- - wearing contact lenses.
3) Depending on the patient’s physical condition, allergy symptoms or stomatitis may occur. In such a case, stop using this
product.
2. Usage
1) This product must be used only by qualifi ed dental professionals and only for dental treatment.
2) Never dismantle or modify this product.
3) In case of trouble, do not attempt to repair. Return to dealer for repair.
4) In case of deterioration or malfunction, do not use this product.
5) Do not use this product when in an unmaintained condition.
6) To operate this instrument correctly and safely, keep the instruction manual.
7) Use sodium hydrogen carbonate powder for dental prophylaxis only. Other powder (for example, alumina powder) will cause
a decrease in cleaning effi ciency, powder clogging, or damage to the polisher.

2
2. CAUTIONS FOR USE
CONTRAINDICATION AND RESTRICTION
1. Non-eligible patients
1) Never use this product on patients
- - on a low salt diet (due to hypertension, hypernatremia, edema, pregnancy toxemia, etc.),
- - having a serious ulcer in digestive organs,
- - having a renal dysfunction,
- - having cardiac dysfunction,
- - having asthma or chronic pulmonary disease,
- - having electrolyte abnormality due to hypochloric alkalosis.
2) Avoid using this product on patients below, other than those stated in above 1):
- - having an intraoral injury or disease,
- - having a hyperemia, infl ammation or bleeding in the intraoral cavity,
- - easily get oral mucosa infl ammation or bleeding,
- - having allergies,
- - wearing contact lenses.
3) Depending on the patient’s physical condition, allergy symptoms or stomatitis may occur. In such a case, stop using this
product.
2. Usage
1) This product must be used only by qualifi ed dental professionals and only for dental treatment.
2) Never dismantle or modify this product.
3) In case of trouble, do not attempt to repair. Return to dealer for repair.
4) In case of deterioration or malfunction, do not use this product.
5) Do not use this product when in an unmaintained condition.
6) To operate this instrument correctly and safely, keep the instruction manual.
7) Use sodium hydrogen carbonate powder for dental prophylaxis only. Other powder (for example, alumina powder) will cause
a decrease in cleaning effi ciency, powder clogging, or damage to the polisher.
2
2. CAUTIONS FOR USE
CONTRAINDICATION AND RESTRICTION
1. Non-eligible patients
1) Never use this product on patients
- - on a low salt diet (due to hypertension, hypernatremia, edema, pregnancy toxemia, etc.),
- - having a serious ulcer in digestive organs,
- - having a renal dysfunction,
- - having cardiac dysfunction,
- - having asthma or chronic pulmonary disease,
- - having electrolyte abnormality due to hypochloric alkalosis.
2) Avoid using this product on patients below, other than those stated in above 1):
- - having an intraoral injury or disease,
- - having a hyperemia, infl ammation or bleeding in the intraoral cavity,
- - easily get oral mucosa infl ammation or bleeding,
- - having allergies,
- - wearing contact lenses.
3) Depending on the patient’s physical condition, allergy symptoms or stomatitis may occur. In such a case, stop using this
product.
2. Usage
1) This product must be used only by qualifi ed dental professionals and only for dental treatment.
2) Never dismantle or modify this product.
3) In case of trouble, do not attempt to repair. Return to dealer for repair.
4) In case of deterioration or malfunction, do not use this product.
5) Do not use this product when in an unmaintained condition.
6) To operate this instrument correctly and safely, keep the instruction manual.
7) Use sodium hydrogen carbonate powder for dental prophylaxis only. Other powder (for example, alumina powder) will cause
a decrease in cleaning effi ciency, powder clogging, or damage to the polisher.
3
CONTRAINDICATION AND RESTRICTION
8) Do not reuse the powder.
9) The powder must be fi lled into the powder cap. Filling into the chamber unit may cause a premature wear of the screw
threads and insuffi cient sealing and powder scattering.
10) Never spray into the subgingival, root canal, or extraction socket. It can cause accidental symptoms such as emphysemas.
11) Never spray directly into the intraoral soft tissues (gingiva, mucosa). It can cause accidental symptoms such as surface
damages.
12) This product must not be repaired by the user in case of malfunction. Contact your dealer for repair.
3
CONTRAINDICATION AND RESTRICTION
8) Do not reuse the powder.
9) The powder must be fi lled into the powder cap. Filling into the chamber unit may cause a premature wear of the screw
threads and insuffi cient sealing and powder scattering.
10) Never spray into the subgingival, root canal, or extraction socket. It can cause accidental symptoms such as emphysemas.
11) Never spray directly into the intraoral soft tissues (gingiva, mucosa). It can cause accidental symptoms such as surface
damages.
12) This product must not be repaired by the user in case of malfunction. Contact your dealer for repair.

4
3. INTENDED USE
This product is intended to remove dental plaque, stain, and discoloration from tooth surface etc. by dental professionals during
dental treatment.
4. CHARACTERISTICS
1) The compact handpiece form allows easy movement between dental units and takes up little space for storage.
2) The quick-disconnect system permits easy replacement with turbines and immediate use. No need for cumbersome installation
and allows easy integration into your treatment system.
3) The contra-angled head rotates 360 degrees, allowing excellent access to all areas.
4) The detachable head allows easy maintenance in case of powder clogging.
5) Driven by compressed air and water from the dental unit, polisher neither effects other medical devices and is not effected by
electromagnetic waves.
4
3. INTENDED USE
This product is intended to remove dental plaque, stain, and discoloration from tooth surface etc. by dental professionals during
dental treatment.
4. CHARACTERISTICS
1) The compact handpiece form allows easy movement between dental units and takes up little space for storage.
2) The quick-disconnect system permits easy replacement with turbines and immediate use. No need for cumbersome installation
and allows easy integration into your treatment system.
3) The contra-angled head rotates 360 degrees, allowing excellent access to all areas.
4) The detachable head allows easy maintenance in case of powder clogging.
5) Driven by compressed air and water from the dental unit, polisher neither effects other medical devices and is not effected by
electromagnetic waves.

4
3. INTENDED USE
This product is intended to remove dental plaque, stain, and discoloration from tooth surface etc. by dental professionals during
dental treatment.
4. CHARACTERISTICS
1) The compact handpiece form allows easy movement between dental units and takes up little space for storage.
2) The quick-disconnect system permits easy replacement with turbines and immediate use. No need for cumbersome installation
and allows easy integration into your treatment system.
3) The contra-angled head rotates 360 degrees, allowing excellent access to all areas.
4) The detachable head allows easy maintenance in case of powder clogging.
5) Driven by compressed air and water from the dental unit, polisher neither effects other medical devices and is not effected by
electromagnetic waves.
4
3. INTENDED USE
This product is intended to remove dental plaque, stain, and discoloration from tooth surface etc. by dental professionals during
dental treatment.
4. CHARACTERISTICS
1) The compact handpiece form allows easy movement between dental units and takes up little space for storage.
2) The quick-disconnect system permits easy replacement with turbines and immediate use. No need for cumbersome installation
and allows easy integration into your treatment system.
3) The contra-angled head rotates 360 degrees, allowing excellent access to all areas.
4) The detachable head allows easy maintenance in case of powder clogging.
5) Driven by compressed air and water from the dental unit, polisher neither effects other medical devices and is not effected by
electromagnetic waves.
5
5. DELIVERY PACKAGE
The components and accessories below are included in the delivery kit. When you fi rst open the package, confi rm all components
are included in the package. If something is missing, contact your dealer immediately.
No powder is included in the delivery package.
Powder seal (spare) Head o-ring (spare)
Cleaning wire
BC-B5000
BC-B5020BC-B5010
BC-B5040BC-B5030
The BLASTER
Contents
The BLASTER
BC-B5000/B5010/
B5020/B5030/B5040
1
Cleaning wire 1
Head o-ring (spare) 1
Powder seal (spare) 1
Instruction manual 1
5
5. DELIVERY PACKAGE
The components and accessories below are included in the delivery kit. When you fi rst open the package, confi rm all components
are included in the package. If something is missing, contact your dealer immediately.
No powder is included in the delivery package.
Powder seal (spare) Head o-ring (spare)
Cleaning wire
BC-B5000
BC-B5020BC-B5010
BC-B5040BC-B5030
The BLASTER
Contents
The BLASTER
BC-B5000/B5010/
B5020/B5030/B5040
1
Cleaning wire 1
Head o-ring (spare) 1
Powder seal (spare) 1
Instruction manual 1

6
6. PRODUCT COMPONENTS
Head coupling
Nozzle
Head o-ring
Chamber unit coupling
Head
Powder cap
Chamber unit
Grip
Joint
Powder seal
Powder pipe Powder chamber
Powder stirring cap
Mounting position mark
Triangle symbol
6
6. PRODUCT COMPONENTS
Head coupling
Nozzle
Head o-ring
Chamber unit coupling
Head
Powder cap
Chamber unit
Grip
Joint
Powder seal
Powder pipe Powder chamber
Powder stirring cap
Mounting position mark
Triangle symbol

6
6. PRODUCT COMPONENTS
Head coupling
Nozzle
Head o-ring
Chamber unit coupling
Head
Powder cap
Chamber unit
Grip
Joint
Powder seal
Powder pipe Powder chamber
Powder stirring cap
Mounting position mark
Triangle symbol
6
6. PRODUCT COMPONENTS
Head coupling
Nozzle
Head o-ring
Chamber unit coupling
Head
Powder cap
Chamber unit
Grip
Joint
Powder seal
Powder pipe Powder chamber
Powder stirring cap
Mounting position mark
Triangle symbol
7
7. HOW TO USE
7.1 Before use
CAUTION
Do not impact this product (dropping, for example). If the polisher gets an impact, stop using and ask your dealer for inspection
and repair of the main body and the powder cap together. If the main body and/or powder cap has got a crack (including a
small crack invisible to the naked eye) through an impact, the main body may break and cause accidents.
ATTENTION
1) Use this product for the prescribed purpose according to the specifi ed usage only.
2) When using other products which are compatible (such as peripheral devices and instruments) to this product, observe
instructions for use of such devices or instruments.
1) Opening the package
When unpacking the package fi rst time, confi rm all contents in “5. DELIVERY PACKAGE” are included in the package.
2) Confi rmation of defect
Confi rm polisher has no defect or damage such as cracks.
3) Sterilization (before fi rst use and when a sterilization is required)
Clean with alcohol and sterilize in an autoclave according to instructions in “9. CLEANING AND
STERILIZATION”.
ATTENTION
1) This product is not sterilized in the delivery condition. Clean with alcohol and sterilize in an autoclave before fi rst use.
2) Do not lubricate this product. Lubricant will be mixed into the compressed air and cause powder clogging.
7
7. HOW TO USE
7.1 Before use
CAUTION
Do not impact this product (dropping, for example). If the polisher gets an impact, stop using and ask your dealer for inspection
and repair of the main body and the powder cap together. If the main body and/or powder cap has got a crack (including a
small crack invisible to the naked eye) through an impact, the main body may break and cause accidents.
ATTENTION
1) Use this product for the prescribed purpose according to the specifi ed usage only.
2) When using other products which are compatible (such as peripheral devices and instruments) to this product, observe
instructions for use of such devices or instruments.
1) Opening the package
When unpacking the package fi rst time, confi rm all contents in “5. DELIVERY PACKAGE” are included in the package.
2) Confi rmation of defect
Confi rm polisher has no defect or damage such as cracks.
3) Sterilization (before fi rst use and when a sterilization is required)
Clean with alcohol and sterilize in an autoclave according to instructions in “9. CLEANING AND
STERILIZATION”.
ATTENTION
1) This product is not sterilized in the delivery condition. Clean with alcohol and sterilize in an autoclave before fi rst use.
2) Do not lubricate this product. Lubricant will be mixed into the compressed air and cause powder clogging.

8
4) Checking your air compressor
To supply polisher with clean and dry air, drain air fi lter of the compressor in your dental offi ce. Do not supply compressed air
containing moisture. For drainage and regular maintenance of your compressor, please refer to your air compressor’s instruction
manual. Confi rm the compressor is maintained according to its instructions for use.
7.2 Air pressure setting (at fi rst use)
1) Preparation of polisher
(a) If polisher is connected to a turbine hose, disconnect it from the turbine hose.
(b) Attach the head and powder cap to the chamber unit according to “7.3 Attaching head” and “7.4 Attaching powder cap”. Do
not fi ll the powder yet.
2) Connection
(a) Connect your coupling (separately sold) to your turbine hose. Confi rm secure connection.
(b) Connect a pressure gauge to your coupling on your turbine hose until you hear a locking sound. Confi rm the gauge is fi rmly
attached. (If you don’t have a pressure gauge, consult with your dental dealer.)
(c) Connect polisher to the coupling of the pressure gauge until you hear a locking sound. Confi rm the gauge is fi rmly seated.
8
4) Checking your air compressor
To supply polisher with clean and dry air, drain air fi lter of the compressor in your dental offi ce. Do not supply compressed air
containing moisture. For drainage and regular maintenance of your compressor, please refer to your air compressor’s instruction
manual. Confi rm the compressor is maintained according to its instructions for use.
7.2 Air pressure setting (at fi rst use)
1) Preparation of polisher
(a) If polisher is connected to a turbine hose, disconnect it from the turbine hose.
(b) Attach the head and powder cap to the chamber unit according to “7.3 Attaching head” and “7.4 Attaching powder cap”. Do
not fi ll the powder yet.
2) Connection
(a) Connect your coupling (separately sold) to your turbine hose. Confi rm secure connection.
(b) Connect a pressure gauge to your coupling on your turbine hose until you hear a locking sound. Confi rm the gauge is fi rmly
attached. (If you don’t have a pressure gauge, consult with your dental dealer.)
(c) Connect polisher to the coupling of the pressure gauge until you hear a locking sound. Confi rm the gauge is fi rmly seated.

8
4) Checking your air compressor
To supply polisher with clean and dry air, drain air fi lter of the compressor in your dental offi ce. Do not supply compressed air
containing moisture. For drainage and regular maintenance of your compressor, please refer to your air compressor’s instruction
manual. Confi rm the compressor is maintained according to its instructions for use.
7.2 Air pressure setting (at fi rst use)
1) Preparation of polisher
(a) If polisher is connected to a turbine hose, disconnect it from the turbine hose.
(b) Attach the head and powder cap to the chamber unit according to “7.3 Attaching head” and “7.4 Attaching powder cap”. Do
not fi ll the powder yet.
2) Connection
(a) Connect your coupling (separately sold) to your turbine hose. Confi rm secure connection.
(b) Connect a pressure gauge to your coupling on your turbine hose until you hear a locking sound. Confi rm the gauge is fi rmly
attached. (If you don’t have a pressure gauge, consult with your dental dealer.)
(c) Connect polisher to the coupling of the pressure gauge until you hear a locking sound. Confi rm the gauge is fi rmly seated.
8
4) Checking your air compressor
To supply polisher with clean and dry air, drain air fi lter of the compressor in your dental offi ce. Do not supply compressed air
containing moisture. For drainage and regular maintenance of your compressor, please refer to your air compressor’s instruction
manual. Confi rm the compressor is maintained according to its instructions for use.
7.2 Air pressure setting (at fi rst use)
1) Preparation of polisher
(a) If polisher is connected to a turbine hose, disconnect it from the turbine hose.
(b) Attach the head and powder cap to the chamber unit according to “7.3 Attaching head” and “7.4 Attaching powder cap”. Do
not fi ll the powder yet.
2) Connection
(a) Connect your coupling (separately sold) to your turbine hose. Confi rm secure connection.
(b) Connect a pressure gauge to your coupling on your turbine hose until you hear a locking sound. Confi rm the gauge is fi rmly
attached. (If you don’t have a pressure gauge, consult with your dental dealer.)
(c) Connect polisher to the coupling of the pressure gauge until you hear a locking sound. Confi rm the gauge is fi rmly seated.
9
3) Air pressure setting
(a) Turn off the water supply of your dental unit. Direct polisher
to your cuspidor and depress your foot control to activate the
dental unit.
(b) Adjust the air pressure of the unit until the gauge shows 0.25
to 0.4 MPa (about 2.5 to 4.1kgf/cm2).
(c) Detach the pressure gauge and polisher after pressure
adjustment.
CAUTION
1) When connecting your coupling to this product, confi rm secure connection after hearing a locking sound. Make sure the
head and powder cap are fi rmly screwed in. A failure to attach correctly may cause injury due to instrument becoming
disconnected or powder being scattered.
2) Never pull the handpiece out of your coupling while polisher is in operation. This may cause injury due to disconnection of
the polisher. (For BC-B5000 or BC-B5020)
3) Never release your coupling from the handpiece while polisher is in operation. This may cause injury due to disconnection
of the polisher. (For BC-B5010, BC-B5030 or BC-B5040)
4) The head and powder cap must not be detached during operation. This may cause injuries due to instrument becoming
disconnected, or powder scattering.
5) The air pressure at the coupling must be set within 0.25 to 0.4 MPa (about 2.5 to 4.1kgf/cm2). Excess air pressure will
damage tooth surface and cause patient pain. It may also cause trouble during use such as injuries due to instrument damage,
or powder scattering.
Air pressure
0.25 to 0.4 MPa
9
3) Air pressure setting
(a) Turn off the water supply of your dental unit. Direct polisher
to your cuspidor and depress your foot control to activate the
dental unit.
(b) Adjust the air pressure of the unit until the gauge shows 0.25
to 0.4 MPa (about 2.5 to 4.1kgf/cm2).
(c) Detach the pressure gauge and polisher after pressure
adjustment.
CAUTION
1) When connecting your coupling to this product, confi rm secure connection after hearing a locking sound. Make sure the
head and powder cap are fi rmly screwed in. A failure to attach correctly may cause injury due to instrument becoming
disconnected or powder being scattered.
2) Never pull the handpiece out of your coupling while polisher is in operation. This may cause injury due to disconnection of
the polisher. (For BC-B5000 or BC-B5020)
3) Never release your coupling from the handpiece while polisher is in operation. This may cause injury due to disconnection
of the polisher. (For BC-B5010, BC-B5030 or BC-B5040)
4) The head and powder cap must not be detached during operation. This may cause injuries due to instrument becoming
disconnected, or powder scattering.
5) The air pressure at the coupling must be set within 0.25 to 0.4 MPa (about 2.5 to 4.1kgf/cm2). Excess air pressure will
damage tooth surface and cause patient pain. It may also cause trouble during use such as injuries due to instrument damage,
or powder scattering.
Air pressure
0.25 to 0.4 MPa

10
7.3 Attaching head
1) Polisher must be disconnected from a turbine
hose of dental unit when attaching and
detaching the head. If it is connected,
disconnect from the turbine hose fi rst.
2) Confi rm the o-ring of the head is not
damaged.
3) If water, oil, or powder adheres to head or
inside the chamber unit’s coupling, wipe off
with a cotton swab or dry cloth.
4) Align the head to the thread of the chamber
unit’s coupling, turn the head in the direction
of the arrow in the illustration until it stops.
Confi rm secure attachment.
Remove water,
oil, and powder.
Confirm the o-ring of the head
is not damaged.
10
7.3 Attaching head
1) Polisher must be disconnected from a turbine
hose of dental unit when attaching and
detaching the head. If it is connected,
disconnect from the turbine hose fi rst.
2) Confi rm the o-ring of the head is not
damaged.
3) If water, oil, or powder adheres to head or
inside the chamber unit’s coupling, wipe off
with a cotton swab or dry cloth.
4) Align the head to the thread of the chamber
unit’s coupling, turn the head in the direction
of the arrow in the illustration until it stops.
Confi rm secure attachment.
Remove water,
oil, and powder.
Confirm the o-ring of the head
is not damaged.

10
7.3 Attaching head
1) Polisher must be disconnected from a turbine
hose of dental unit when attaching and
detaching the head. If it is connected,
disconnect from the turbine hose fi rst.
2) Confi rm the o-ring of the head is not
damaged.
3) If water, oil, or powder adheres to head or
inside the chamber unit’s coupling, wipe off
with a cotton swab or dry cloth.
4) Align the head to the thread of the chamber
unit’s coupling, turn the head in the direction
of the arrow in the illustration until it stops.
Confi rm secure attachment.
Remove water,
oil, and powder.
Confirm the o-ring of the head
is not damaged.
10
7.3 Attaching head
1) Polisher must be disconnected from a turbine
hose of dental unit when attaching and
detaching the head. If it is connected,
disconnect from the turbine hose fi rst.
2) Confi rm the o-ring of the head is not
damaged.
3) If water, oil, or powder adheres to head or
inside the chamber unit’s coupling, wipe off
with a cotton swab or dry cloth.
4) Align the head to the thread of the chamber
unit’s coupling, turn the head in the direction
of the arrow in the illustration until it stops.
Confi rm secure attachment.
Remove water,
oil, and powder.
Confirm the o-ring of the head
is not damaged.
11
7.4 Attaching powder cap
* Water due to sterilization must be expelled
by an idle run beforehand. Do not fi ll the
powder yet.
1) The powder cap must be attached or detached when
polisher is not connected to a turbine hose. If
connected, detach from the turbine hose.
2) If water, oil, or powder adheres to inside chamber unit,
powder cap, or powder seal, wipe off with a dry cloth.
3) Confi rm the powder seal (blue) is fully engaged in the
groove of chamber unit shown in the illustration.
4) Turn the powder cap in the direction of the arrow in
the illustration until it stops. Confi rm the powder cap is
fi rmly attached and the triangle symbol on the powder
cap lies within the range of the mounting position mark.
In case the triangle symbol lies outside of the mounting
position mark, replace your powder seal with new one
according to “10.4 Replacement of powder seal”.
Remove water, oil,
and powder.
Enlarged sectional view
Groove
Powder seal
Mounting position mark
Triangle symbol
11
7.4 Attaching powder cap
* Water due to sterilization must be expelled
by an idle run beforehand. Do not fi ll the
powder yet.
1) The powder cap must be attached or detached when
polisher is not connected to a turbine hose. If
connected, detach from the turbine hose.
2) If water, oil, or powder adheres to inside chamber unit,
powder cap, or powder seal, wipe off with a dry cloth.
3) Confi rm the powder seal (blue) is fully engaged in the
groove of chamber unit shown in the illustration.
4) Turn the powder cap in the direction of the arrow in
the illustration until it stops. Confi rm the powder cap is
fi rmly attached and the triangle symbol on the powder
cap lies within the range of the mounting position mark.
In case the triangle symbol lies outside of the mounting
position mark, replace your powder seal with new one
according to “10.4 Replacement of powder seal”.
Remove water, oil,
and powder.
Enlarged sectional view
Groove
Powder seal
Mounting position mark
Triangle symbol

12
CAUTION
1) Powder cap and powder seal must be attached fi rmly. An unstable attachment will cause the powder cap to undo during use
and may cause injuries or powder scattering.
2) Make sure the powder stirring cap is attached. Improper attachment will lower the cleaning effi ciency.
3) Remove powder remaining in/on the screw threads of the main body, powder cap, and sealing surface of the powder seal.
If the powder adheres fi rmly and is hard to remove, clean the remaining powder according to the instructions in “8.1 3)
Cleaning and disinfection”. As the remaining powder may cause premature wear of the threads and insuffi cient sealing
which may cause powder scattering.
4) Before each use, check if the screw threads of the main body and powder cap have any damage such as wear, crack,
deformation, etc. If the main body or powder cap has a damage, do not use and ask your dealer for repair together.
5) If the triangle symbol on the powder cap lies outside of the mounting position mark while the polisher is in use, this may
cause trouble like crack, damage or premature wear of the threads of the main body and powder cap.
ATTENTION
Remove water, oil, or powder remaining in/on the head, coupling of chamber unit, powder chamber, and powder cap.
Otherwise, the powder will get damp and cause clogging.
12
CAUTION
1) Powder cap and powder seal must be attached fi rmly. An unstable attachment will cause the powder cap to undo during use
and may cause injuries or powder scattering.
2) Make sure the powder stirring cap is attached. Improper attachment will lower the cleaning effi ciency.
3) Remove powder remaining in/on the screw threads of the main body, powder cap, and sealing surface of the powder seal.
If the powder adheres fi rmly and is hard to remove, clean the remaining powder according to the instructions in “8.1 3)
Cleaning and disinfection”. As the remaining powder may cause premature wear of the threads and insuffi cient sealing
which may cause powder scattering.
4) Before each use, check if the screw threads of the main body and powder cap have any damage such as wear, crack,
deformation, etc. If the main body or powder cap has a damage, do not use and ask your dealer for repair together.
5) If the triangle symbol on the powder cap lies outside of the mounting position mark while the polisher is in use, this may
cause trouble like crack, damage or premature wear of the threads of the main body and powder cap.
ATTENTION
Remove water, oil, or powder remaining in/on the head, coupling of chamber unit, powder chamber, and powder cap.
Otherwise, the powder will get damp and cause clogging.

12
CAUTION
1) Powder cap and powder seal must be attached fi rmly. An unstable attachment will cause the powder cap to undo during use
and may cause injuries or powder scattering.
2) Make sure the powder stirring cap is attached. Improper attachment will lower the cleaning effi ciency.
3) Remove powder remaining in/on the screw threads of the main body, powder cap, and sealing surface of the powder seal.
If the powder adheres fi rmly and is hard to remove, clean the remaining powder according to the instructions in “8.1 3)
Cleaning and disinfection”. As the remaining powder may cause premature wear of the threads and insuffi cient sealing
which may cause powder scattering.
4) Before each use, check if the screw threads of the main body and powder cap have any damage such as wear, crack,
deformation, etc. If the main body or powder cap has a damage, do not use and ask your dealer for repair together.
5) If the triangle symbol on the powder cap lies outside of the mounting position mark while the polisher is in use, this may
cause trouble like crack, damage or premature wear of the threads of the main body and powder cap.
ATTENTION
Remove water, oil, or powder remaining in/on the head, coupling of chamber unit, powder chamber, and powder cap.
Otherwise, the powder will get damp and cause clogging.
12
CAUTION
1) Powder cap and powder seal must be attached fi rmly. An unstable attachment will cause the powder cap to undo during use
and may cause injuries or powder scattering.
2) Make sure the powder stirring cap is attached. Improper attachment will lower the cleaning effi ciency.
3) Remove powder remaining in/on the screw threads of the main body, powder cap, and sealing surface of the powder seal.
If the powder adheres fi rmly and is hard to remove, clean the remaining powder according to the instructions in “8.1 3)
Cleaning and disinfection”. As the remaining powder may cause premature wear of the threads and insuffi cient sealing
which may cause powder scattering.
4) Before each use, check if the screw threads of the main body and powder cap have any damage such as wear, crack,
deformation, etc. If the main body or powder cap has a damage, do not use and ask your dealer for repair together.
5) If the triangle symbol on the powder cap lies outside of the mounting position mark while the polisher is in use, this may
cause trouble like crack, damage or premature wear of the threads of the main body and powder cap.
ATTENTION
Remove water, oil, or powder remaining in/on the head, coupling of chamber unit, powder chamber, and powder cap.
Otherwise, the powder will get damp and cause clogging.
13
7.5 Connecting to dental unit
1) Cleaning of coupling
(a) Confi rm no water or oil remains in the coupling on your turbine hose. Remove with a dry cloth if remains.
* When a handpiece (air turbine, air scaler, etc.) is detached, water and oil remain at the joint.
Remove the water and oil without fail.
(b) Confi rm no water or oil remains in the joint of polisher. Remove with a cotton swab if remains.
ATTENTION
Remove water or oil remaining inside of the coupling on your turbine hose and joint of the chamber unit. Otherwise, the
powder will become damp and cause clogging.
Remove water
and oil.
Remove water and oil.
13
7.5 Connecting to dental unit
1) Cleaning of coupling
(a) Confi rm no water or oil remains in the coupling on your turbine hose. Remove with a dry cloth if remains.
* When a handpiece (air turbine, air scaler, etc.) is detached, water and oil remain at the joint.
Remove the water and oil without fail.
(b) Confi rm no water or oil remains in the joint of polisher. Remove with a cotton swab if remains.
ATTENTION
Remove water or oil remaining inside of the coupling on your turbine hose and joint of the chamber unit. Otherwise, the
powder will become damp and cause clogging.
Remove water
and oil.
Remove water and oil.

14
2) Connecting a coupling
(a) Connect a coupling (separately sold)
compatible with your polisher to your turbine
hose. Confi rm the coupling is fi rmly seated.
3) Connecting to turbine hose
(a) Insert your coupling on your turbine hose
into the joint of polisher until you hear a
locking sound. Confi rm the handpiece is
fi rmly seated.
(b) If your coupling has a light source, turn off
the illumination switch of your dental unit.
CAUTION
When connecting your coupling to this product, confi rm secure connection after hearing a locking sound. Make sure the head
and powder cap are fi rmly screwed in. A failure to attach correctly may cause injury due to instrument becoming disconnected
or powder being scattered.
ATTENTION
To avoid shortening the life of the light source (light bulb) of your coupling, turn off the illumination of your dental unit when
using this product.
Coupling Turbine hose
14
2) Connecting a coupling
(a) Connect a coupling (separately sold)
compatible with your polisher to your turbine
hose. Confi rm the coupling is fi rmly seated.
3) Connecting to turbine hose
(a) Insert your coupling on your turbine hose
into the joint of polisher until you hear a
locking sound. Confi rm the handpiece is
fi rmly seated.
(b) If your coupling has a light source, turn off
the illumination switch of your dental unit.
CAUTION
When connecting your coupling to this product, confi rm secure connection after hearing a locking sound. Make sure the head
and powder cap are fi rmly screwed in. A failure to attach correctly may cause injury due to instrument becoming disconnected
or powder being scattered.
ATTENTION
To avoid shortening the life of the light source (light bulb) of your coupling, turn off the illumination of your dental unit when
using this product.
Coupling Turbine hose

14
2) Connecting a coupling
(a) Connect a coupling (separately sold)
compatible with your polisher to your turbine
hose. Confi rm the coupling is fi rmly seated.
3) Connecting to turbine hose
(a) Insert your coupling on your turbine hose
into the joint of polisher until you hear a
locking sound. Confi rm the handpiece is
fi rmly seated.
(b) If your coupling has a light source, turn off
the illumination switch of your dental unit.
CAUTION
When connecting your coupling to this product, confi rm secure connection after hearing a locking sound. Make sure the head
and powder cap are fi rmly screwed in. A failure to attach correctly may cause injury due to instrument becoming disconnected
or powder being scattered.
ATTENTION
To avoid shortening the life of the light source (light bulb) of your coupling, turn off the illumination of your dental unit when
using this product.
Coupling Turbine hose
14
2) Connecting a coupling
(a) Connect a coupling (separately sold)
compatible with your polisher to your turbine
hose. Confi rm the coupling is fi rmly seated.
3) Connecting to turbine hose
(a) Insert your coupling on your turbine hose
into the joint of polisher until you hear a
locking sound. Confi rm the handpiece is
fi rmly seated.
(b) If your coupling has a light source, turn off
the illumination switch of your dental unit.
CAUTION
When connecting your coupling to this product, confi rm secure connection after hearing a locking sound. Make sure the head
and powder cap are fi rmly screwed in. A failure to attach correctly may cause injury due to instrument becoming disconnected
or powder being scattered.
ATTENTION
To avoid shortening the life of the light source (light bulb) of your coupling, turn off the illumination of your dental unit when
using this product.
Coupling Turbine hose
15
7.6 Water removal in channels (idle run)
1) Confi rm no powder remains within the powder chamber.
2) Direct the nozzle to your cuspidor. Depress your foot
control without water supply from the dental unit. Run
for 1 minute to remove water remaining in the channels
of polisher.
ATTENTION
Run polisher for 1 minute without powder before use. If water remains inside of the main body, powder will become damp and
cause clogging.
Stop water supply.
Idle run for 1 min.
15
7.6 Water removal in channels (idle run)
1) Confi rm no powder remains within the powder chamber.
2) Direct the nozzle to your cuspidor. Depress your foot
control without water supply from the dental unit. Run
for 1 minute to remove water remaining in the channels
of polisher.
ATTENTION
Run polisher for 1 minute without powder before use. If water remains inside of the main body, powder will become damp and
cause clogging.
Stop water supply.
Idle run for 1 min.

16
7.7 Powder fi lling
1) Remove polisher from the turbine hose before refi lling
with powder.
2) Detach the powder cap by turning in the direction of the
arrow in the illustration above.
3) Confi rm the powder stirring cap is attached in place.
4) If water, oil, or powder remains inside of powder chamber
or powder cap, remove it with a dry cloth.
5) Fill sodium hydrogen carbonate powder not exceeding 15g
into the powder cap.
* The powder must be placed into the powder
cap. Take care not to spill the powder over the
thread of the powder cap. If the powder is spilt
over the thread, wipe it off with a dry cloth.
Filling too much powder will make the powder
adhere to the powder seal and cause
premature wear of the powder seal or
incomplete sealing.
6) Hold the chamber unit downward and the powder cap
upward as shown in the illustration. Turn the powder cap
in the direction of the arrow until it stops.
Remove water, oil, and powder.
Fill the powder into
the powder cap.
Confirm the powder
stirring cap is
attached in place.
16
7.7 Powder fi lling
1) Remove polisher from the turbine hose before refi lling
with powder.
2) Detach the powder cap by turning in the direction of the
arrow in the illustration above.
3) Confi rm the powder stirring cap is attached in place.
4) If water, oil, or powder remains inside of powder chamber
or powder cap, remove it with a dry cloth.
5) Fill sodium hydrogen carbonate powder not exceeding 15g
into the powder cap.
* The powder must be placed into the powder
cap. Take care not to spill the powder over the
thread of the powder cap. If the powder is spilt
over the thread, wipe it off with a dry cloth.
Filling too much powder will make the powder
adhere to the powder seal and cause
premature wear of the powder seal or
incomplete sealing.
6) Hold the chamber unit downward and the powder cap
upward as shown in the illustration. Turn the powder cap
in the direction of the arrow until it stops.
Remove water, oil, and powder.
Fill the powder into
the powder cap.
Confirm the powder
stirring cap is
attached in place.

16
7.7 Powder fi lling
1) Remove polisher from the turbine hose before refi lling
with powder.
2) Detach the powder cap by turning in the direction of the
arrow in the illustration above.
3) Confi rm the powder stirring cap is attached in place.
4) If water, oil, or powder remains inside of powder chamber
or powder cap, remove it with a dry cloth.
5) Fill sodium hydrogen carbonate powder not exceeding 15g
into the powder cap.
* The powder must be placed into the powder
cap. Take care not to spill the powder over the
thread of the powder cap. If the powder is spilt
over the thread, wipe it off with a dry cloth.
Filling too much powder will make the powder
adhere to the powder seal and cause
premature wear of the powder seal or
incomplete sealing.
6) Hold the chamber unit downward and the powder cap
upward as shown in the illustration. Turn the powder cap
in the direction of the arrow until it stops.
Remove water, oil, and powder.
Fill the powder into
the powder cap.
Confirm the powder
stirring cap is
attached in place.
16
7.7 Powder fi lling
1) Remove polisher from the turbine hose before refi lling
with powder.
2) Detach the powder cap by turning in the direction of the
arrow in the illustration above.
3) Confi rm the powder stirring cap is attached in place.
4) If water, oil, or powder remains inside of powder chamber
or powder cap, remove it with a dry cloth.
5) Fill sodium hydrogen carbonate powder not exceeding 15g
into the powder cap.
* The powder must be placed into the powder
cap. Take care not to spill the powder over the
thread of the powder cap. If the powder is spilt
over the thread, wipe it off with a dry cloth.
Filling too much powder will make the powder
adhere to the powder seal and cause
premature wear of the powder seal or
incomplete sealing.
6) Hold the chamber unit downward and the powder cap
upward as shown in the illustration. Turn the powder cap
in the direction of the arrow until it stops.
Remove water, oil, and powder.
Fill the powder into
the powder cap.
Confirm the powder
stirring cap is
attached in place.
17
7) Confi rm the powder cap is fi rmly attached within the
range of the mounting position mark. If the triangle
symbol on the powder cap lies outside of the mounting
position mark, replace your powder seal with a new one
according to “10.4 Replacement of powder seal”.
CONTRAINDICATION AND RESTRICTION
1) Use sodium hydrogen carbonate powder for dental prophylaxis only. Other powder (for example, alumina powder) will cause
a decrease in cleaning effi ciency, powder clogging, or damage to the polisher.
2) Do not reuse the powder.
3) The powder must be fi lled into the powder cap. Filling into the chamber unit may cause a premature wear of the screw
threads and insuffi cient sealing and powder scattering.
Mounting position mark
Triangle symbol
17
7) Confi rm the powder cap is fi rmly attached within the
range of the mounting position mark. If the triangle
symbol on the powder cap lies outside of the mounting
position mark, replace your powder seal with a new one
according to “10.4 Replacement of powder seal”.
CONTRAINDICATION AND RESTRICTION
1) Use sodium hydrogen carbonate powder for dental prophylaxis only. Other powder (for example, alumina powder) will cause
a decrease in cleaning effi ciency, powder clogging, or damage to the polisher.
2) Do not reuse the powder.
3) The powder must be fi lled into the powder cap. Filling into the chamber unit may cause a premature wear of the screw
threads and insuffi cient sealing and powder scattering.
Mounting position mark
Triangle symbol