Quality Control Test
Example of Expiration Date:
Exp. Date 2015-06
Performing a Quality Control Test
OPEN
5
6
12
The your Control Solution results are out of the range :
Y
possible reasons
TM
- our Rightest Control Solution has expired or has been opened 3 months ago.
TM
- Your Rightest Blood Glucose Test Strip GS260 has expired.
TM TM
- You left the cap of the Rightest Blood Glucose Test Strip GS260 vial opened or the Rightest
Control Solution off for a long time.
- You didn't perform the test procedure correctly.
TM
- Malfunction of the Rightest Blood Glucose Meter GM260 or Test Strip GS260.
TM
If Control Solution results are out of range, your Rightest Blood Glucose Monitoring System GM260
may not be working properly. Repeat the Quality Control Test. If your Control Solution results outside
the range still exist, do not use the Blood Glucose Monitoring System GM260 to test your blood
glucose and contact your Bionime Customer service.
NOTE
TM
- If you want to purchase new Rightest Control Solution, please contact your authorized
Bionime representative.
PRECAUTION
Each time you open a new bottle of
Control Solution is good for 3 months after opening the
bottle, or until the expiration date printed on the label,
whichever comes first.
TM
Rightest Control
TM
Solution, write the discard date on the label. Rightest
1. Take one
ith view
window, facing up, into test strip port.
3. While the blood drop symbol flashing, press and hold the main
button for over 3 seconds until the " " symbol appears.
4. You will see blinking " " symbol and " " symbol on the screen,
prompting you to apply
Control
Solution on the top of the cap.
TM
Rightest Blood Glucose Test Strip GS260 from vial and
re-cap the vial cap immediately.
TM
2. Insert the Rightest Blood Glucose Test Strip GS260 w
TM
Rightest Control Solution.
TM
5. Shake the bottle of Rightest Control Solution well before opening
the cap. Then open the bottle and put the cap on the table.
TM
6. Drip a drop of Rightest Control Solution on the top of the cap.
TM
7. Gently touch sample entry of the strip with the Rightest
8. When you hear a beep (if volume is turned on) please wait for the test result. The
screen will display the countdown number from 5.
7
4