manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. BIONIME
  6. •
  7. Blood Glucose Meter
  8. •
  9. BIONIME Rightest Elsa User manual

BIONIME Rightest Elsa User manual

Intended Use
The RIGHTEST ELSA Blood Glucose Monitoring System is designed for in vitro diagnostic (for testing
outside the body) use only and can be used by home user and healthcare professional.
The system can test glucose concentration in fresh capillary whole blood (drawn from fingertip, palm
and forearm).
The glucose result displayed is calibrated into the plasma glucose testing equivalent.
The system is not intended for screening or diagnoses of diabetes mellitus.
RIGHTEST ELSA Blood Glucose Test Strips are designed for use only with the RIGHTEST ELSA Blood
Glucose Meter.
Test Procedure
Preparing the Lancing Device
1) Hold the depth adjustable cap in one hand and hold the hub in the other hand. Bend the cap towards
the down side, until a gap appears between the cap and hub.
2) Pull the cap and hub off in opposite directions, remove the cap.
3) Insert a new disposable lancet into the lancet carrier. Make sure it is held securely in place.
4) Twist off and set aside the protective cover of the disposable lancet.
5) Reattach the depth adjustable cap.
6) Rotate the clear top part of the cap to adjust the depth. Check the number of lines visible in the
window. More lines corresponds to a greater depth. Try " " for soft or thin skin, " " for average
skin, or " " for thick or calloused skin.
7) Pull back the plunger until you hear a click. The device is now primed. Let go of the plunger. It will
return to its original position.
Performing a Test
1) Wash your hands with warm soapy water
and dry thoroughly.
2) Take one test strip from the vial. Re-cap
the vial cap immediately.
3) Insert the strip into the test strip port of the
meter.
Test Result
- Blood glucose test results are shown on the meter in mg/dL or mmol/L, depending on the preset of
your meter.
- If your blood glucose result is unusually high or low and you have doubts about the accuracy, repeat
the test with a new test strip. You can also run a Quality Control Test with the RIGHTEST Control
Solution GC570 to check your meter and test strip. If the test result still remains unusually high or low,
contact your healthcare professional immediately.
- If you are experiencing symptoms that are not consistent with your blood glucose test results and you
have followed all the instructions in this manual, contact your healthcare professional immediately.
- The RIGHTEST ELSA Meter displays results between 10 and 600 mg/dL or 0.6 and 33.3 mmol/L. If
your test result is below 10 mg/dL (0.6 mmol/L) , " Lo " will appear on the screen. Please repeat the test
with a new strip. If " Lo " still appears, contact your healthcare professional immediately.
- If your test result is above 600 mg/dL (33.3 mmol/L), " Hi " will appear on the screen. Please repeat the
test with a new strip. If " Hi " still appears, contact your healthcare professional.
Limitations
- The meter readings of the blood glucose may be significantly lower than " true glucose levels " in the
hyperglycemic-hyperosmolar state, with or without ketosis. Critically ill patients should not be tested by
the RIGHTEST System, or tested with extreme caution.
- Caution is advised in interpreting glucose values below 50 mg/dL (2.8 mmol/L) or above 250 mg/dL
(13.9 mmol/L). Consult a healthcare professional as soon as possible if these values are obtained.
- Healthcare professionals should occasionally check their glucose testing technique and the technique
of their patients. It is recommended to compare the meter result with a laboratory result acquired using
the same blood sample. A well characterized clinical laboratory method using hexokinase or glucose
oxidase should be used as the comparative method.
- Flouoride should not be used as a preservative when collecting blood glucose samples.
- Hands and fingers contaminated with sugar from foods or beverages may cause falsely high results.
- Storage of strips near bleach as well as bleach containing products will affect the results of RIGHTEST
Test Strips.
- RIGHTEST Blood Glucose Test Strips are designed for use with capillary whole blood samples.
Do not use serum or plasma samples.
- Incorrect test results may be obtained at high altitudes of more than approximately 3,048 meters
(10,000 feet) above sea level.
- Severe dehydration and excessive water loss may cause inaccurately low results.
- RIGHTEST Blood Glucose Monitoring System has not been validated for use on neonates. DO NOT
use it test for neonates.
Precautions
- Check the expiration date printed on the strip vial. Do not use expired test strips.
- Close the vial cap immediately after taking a test strip out of the vial.
- Do not perform quality control tests with expired control solution.
- Do not bend or twist the test strip. Damaged test strips may give incorrect results.
- Do not reuse test strips and lancets.
- Discard the used disposable lancet and strip into an appropriate sharps biohazard container.
- If the RIGHTEST Meter and Test Strips are exposed to temperature environments outside the operating
range of 10 - 40°C (50 - 104°F), wait at least 30 minutes before testing.
- To purchase new control solutions, please contact your authorized Bionime representative.
1
4) The drop symbol will flash continuously. This signifies that a blood sample should be applied within 2
minutes.
5) Place the lancing device against the pad of your fingertip and press the release button.
9) Remove the cap from the lancing device. Do not touched the
used lancet. Press the lancet tip into the protective cover.
Hold down the release button and pull back the plunger to
eject the lancet. Discard the lancet into a suitable sharps
biohazard container.
Warning
- Keep the test strips and vial cap away from children. They may cause a choking hazard. If a test strip
or vial cap is swallowed, contact a healthcare professional immediately.
0.75 1.0 3μL μL 1.5 μL 2.0 μL .0 μLUse a minimum blood sample volume of 0.75 μL for
blood glucose testing. Samples larger than 3.0 μL
may contaminate the meter.
Sample Size Example
For more information on how to use your meter and lancing device and how to understand your
test results, see the User's Manual.
6) Touch and hold the blood droplet to the test strip sample entry port until you hear a " beep " and the
Viewing Window is filled with blood. If the Viewing Window is not completely filled with blood or the
test does not start, please discard the test strip and repeat the test with a new test strip.
7) You will see a countdown on the screen. After 5 seconds, the test result will appear.
8) Remove the test strip from the meter. Please follow local regulations and discard the used strip
properly.
2
4
5
7
8
(1)
Expected values
GLUCOSE LEVEL
70 to 99 mg/dL (3.9 to 5.5 mmol/L)
100 to 125 mg/dL (5.6 to 6.9 mmol/L)
126 mg/dL (7.0 mmol/L) and above on more than one testing occasion
INDICATION
Normal fasting glucose
Pre-diabetes (Impaired fasting glucose)
Diabetes
Fasting Blood Glucose
3
6a 6b 6c
ELSA BLOOD GLUCOSE TEST STRIP INSERT
7654
32
1
Alternative site testing - palm or forearm blood sampling
- To perform a test using samples obtained from alternative sites, install the
clear cap on the lancing device. (For more information on how to install,
see the Instructions for the Lancing Device.)
- Massage the target area on the palm or forearm for a few seconds to
increase blood flow.
- Immediately after massaging, press and hold the lancing device against the
target area.
- Press the release button.
- After pressing the release button continue holding the lancing device
against the skin. Gradually push the device against your skin with
increasing pressure until the blood droplet is of sufficient size.
Palm
Forearm
Rotate
Push the
release button
Pressing for
a few seconds
References
1) Diabetes Information - American Association for Clinical Chemistry (AACC)〔Electronic Version〕
Retrieved Jan. 26, 2021 from www.labtestsonline.org/understanding/analytes/glucose/test.html
2) In vitro Diagnostics in Diabetes: Meeting the Challenge. Clinical Chemistry 45:9, 1596-1601 (1999).
- Do not perform blood glucose tests at temperatures below 10°C (50°F) or above 40°C (104°F), or below
10% or above 90% relative humidity. The suggested temperature range for the control solution test is
15 - 40°C (59 -104°F).
Troubleshooting and Customer Service
For more information on error messages and trouble shooting, please refer to the Error Messages and
Trouble Shooting section of the User's Manual.
If you have any questions or issues with your RIGHTEST products, please contact your local Bionime
distributor or send an email to [email protected] .
Additional Information for Healthcare Professionals
Quality Control Section
Please refer to the Quality Control section of the User's Manual.
Storage and Handling
- Store the strips in the original capped vial at temperatures between 4°C to 30°C (39°F to 86°F) and 10
to 90% relative humidity. Do not freeze.
- Close the vial cap immediately and tightly after taking a test strip out. Do not leave the vial open. If
strips are exposed to the air for too long, it will absorb moisture and cause inaccurate test results.
- Every time when you open a new vial of test strips, please write the opening date on the label. Use test
strips within 3 months after opened or until the expiration date printed on the label (whichever comes
first).
Measurement Range
The measurement range of the RIGHTEST ELSA is 10 to 600 mg/dL or 0.6 to 33.3 mmol/L.
NOTE
- Suggest not to use this meter close to source of strong electromagnetic radiation, to avoid
interference with proper operation.
- Suggest to keep meter free of dust, water or any liquid.
Lay User Evaluation
A total of 104 users were enrolled. Each user tested their fingertip blood samples with 3 lots of ELSA strip
and ELSA meter. Collected blood samples were then centrifuged immediately after collection to obtain
plasma. Plasma analysis was performed using a lab instrument (YSI 2300 analyzer). 100% of the ELSA
BGM values were within ± 15% of YSI values at glucose concentrations ≧ 100mg/dL(5.55 mmol/L) and
within ± 15 mg/dL (0.83 mmol/L) at glucose concentrations < 100 mg/dL (5.55 mmol/L).
Hematocrit (Hct)
Hematocrit (Hct) should be between 30 - 57%. If you do not know your hematocrit, ask your healthcare
professional.
Interfering Substances
26 substances were tested in low and high glucose concentrations using the blood glucose measuring
device. Only 3 substances may interfere with glucose measurement:
Ascorbic acid ≧ 6 mg/dL (0.34 mmol/L)
Glutathione reduced ≧ 70 mg/dL (2.28 mmol/L)
Uric Acid ≧ 16 mg/dL (0.95 mmol/L)
Other 23 substances within specified concentration may not interfere with glucose measurement :
Acetaminophen ≤ 20 mg/dL (1.32 mmol/L); Dopamine ≤ 2.5 mg/dL (0.13 mmol/L); EDTA ≤ 0.1 mg/dL
(0.003 mmol/L); Gentisic Acid ≤ 7.5 mg/dL (0.49 mmol/L); Heparin ≤ 18.75 U/mL; Ibuprofen ≤ 50
mg/dL (2.42 mmol/L); L-Dopa ≤ 3 mg/dL (0.15 mmol/L); Methyldopa ≤ 1.5 mg/dL (0.06 mmol/L);
Pralidoxime Iodide≤ 4 mg/dL (0.15 mmol/L); Salicylic Acid ≤ 60 mg/dL (4.34 mmol/L); Tetracycline ≤
1.5 mg/dL (0.03 mmol/L); Tolazamide ≤ 15 mg/dL (0.48 mmol/L); Tolbutamide ≤ 64 mg/dL (2.37
mmol/L); Bilirubin ≤ 50 mg/dL (0.86 mmol/L); Cholesterol ≤ 700 mg/dL (18.10 mmol/L); Creatinine ≤ 10
mg/dL (0.67 mmol/L); Hemoglobin ≤ 6000 mg/dL (0.94 mmol/L); Triglycerides ≤ 3000 mg/dL (99.22
mmol/L); Maltose ≤ 200 mg/dL (5.55 mmol/L); Xylose ≤ 40 mg/dL (2.66 mmol/L); Galactose ≤ 200
mg/dL (11.10 mmol/L); Lactose ≤ 50 mg/dL (1.46 mmol/L); Icodextrin ≤ 500 mg/dL (30.84 mmol/L).
Reagents
Each Blood Glucose Test Strip contains the following reagents:
Glucose Oxidase (GOD) 18.8%
Potassium Ferricyanide 37.7%
Non-reactive Ingredients 43.5%
Accuracy
The accuracy of the RIGHTEST ELSA Blood Glucose Monitoring System was tested by comparing
fingertip whole blood (plasma equivalent) glucose values measured by the RIGHTEST ELSA with plasma
glucose values obtained from a YSI 2300 reference instrument.
The YSI 2300 was calibrated with NIST (SRM) 917c reference. The results are shown in the tables below:
(ii) Control solution:
P-01
300
41.2 (2.3)
2.4 (0.13)
5.8%
P-02
300
86.9 (4.8)
2.7 (0.15)
3.1%
P-03
300
145 (8.1)
3.4 (0.19)
2.4%
(i) Venous whole blood sample:
Glucose levels
(1) Total test numbers (n)
(2) Mean mg/dL (mmol/L)
(3) SD mg/dL (mmol/L)
(4) CV (%)
Glucose levels
(1) Total test numbers (n)
(2) Mean mg/dL (mmol/L)
(3) SD mg/dL (mmol/L)
(4) CV (%)
(2)
Detection Principle
The glucose oxidase and potassium ferricyanide in the strip react with glucose in the sample to produce
an electrical current which is proportional to the amount of glucose in the sample. The meter measures
the current and converts it to the corresponding glucose concentration.
Performance Characteristics
Precision
The precision was evaluated by including (i) venous whole blood samples - the blood samples were
collected over a span of time so as not to exceed one day per meter and reagent lot combination and
(ii) 3 control solutions of different glucose concentrations (across a period of 10 days with 10 meters and
3 batches of strips).
P-04
300
198.8 (11)
5.8 (0.32)
2.9%
P-05
300
370.1 (20.6)
6.5 (0.36)
1.8%
CS-L
300
47.3 (2.6)
1.7 (0.09)
3.5%
CS-N
300
85.6 (4.8)
2.8 (0.25)
3.3%
CS-H
300
247.5 (13.8)
6.4 (0.36)
2.6%
Difference range: RIGHTEST
ELSA and YSI
The percent (and number) of samples was the difference between
RIGHTEST ELSA and the YSI value within the following intervals.
Bias range: RIGHTEST ELSA
and YSI
Within ± 5%
Within ± 10%
Within ± 15%
The percent (and number) of samples was the difference between
RIGHTEST ELSA and the YSI value within the following intervals.
*Acceptance criteria in ISO15197 : 2013 are that 95% of all differences in glucose values should be
within ± 15 mg/dL (0.83 mmol/L) at glucose concentrations < 100 mg/dL (5.55 mmol/L), and within
± 15% at glucose concentrations ≧ 100 mg/dL (5.55 mmol/L).
Note: For glucose concentrations < 100 mg/dL (5.55 mmol/L), ranges are expressed in mg/dL (mmol/L),
and for glucose concentrations ≧ 100 mg/dL (5.55 mmol/L), ranges are expressed as a percentage.
Fingertip
66.9% (265/396)
92.4% (366/396)
100.0% (396/396)
Palm
62.1% (246/396)
90.7% (359/396)
99.2% (393/396)
Forearm
56.8% (225/396)
85.4% (338/396)
97.7% (387/396)
Fingertip
82.1% (197/240)
100.0% (240/240)
100.0% (240/240)
Palm
79.6% (191/240)
97.1% (233/240)
100.0% (240/240)
Forearm
65.0% (156/240)
97.1% (233/240)
100.0% (240/240)
Within ± 5 mg/dL (0.28 mmol/L)
Within ± 10 mg/dL (0.56 mmol/L)
Within ± 15 mg/dL (0.83 mmol/L)
Table 3: Represents samples for glucose results ≧ 100 mg/dL (5.55 mmol/L).
Table 2: Represents samples for glucose results < 100 mg/dL (5.55 mmol/L).
101-3GS551-070
EN
Rev. Date:2021-06
Within ± 15 mg/dL
(0.83 mmol/L) or within ± 15%
Test range in mg/dL (mmol/L) 29 - 468 (1.61 - 26.0)
636/636 (100.0%)
29 - 464 (1.61 - 25.78)
636/636 (100.0%)
31 - 471 (1.72 - 26.17)
636/636 (100.0%)
Table 1: Accuracy basic information
Fingertip Palm Forearm
CE Mark with number of Notified Body
humidity limitation
Importer
LOT
Expiry date
Storage temperature limitation
Lot number
IVD For in vitro diagnostic use
For single use only
Manufacturer
EU Representative
Consult instructions for use
Biological risks
BIONIME CORPORATION
No. 100, Sec. 2, Daqing St., South Dist.,
Taichung City 40242, Taiwan
Tel: +886 4 2369 2388
Fax: +886 4 2261 7586
Email: [email protected]
http://www.bionime.com
Emergo Europe
Prinsessegracht 20
2514 AP The Hague
The Netherlands
Email: EmergoVigilance@ul.com

Other manuals for Rightest Elsa

2

Other BIONIME Blood Glucose Meter manuals

BIONIME Rightest GM100 User manual

BIONIME

BIONIME Rightest GM100 User manual

BIONIME Rightest GM720 User manual

BIONIME

BIONIME Rightest GM720 User manual

BIONIME Rightest Elsa User manual

BIONIME

BIONIME Rightest Elsa User manual

BIONIME Rightest GT100 User manual

BIONIME

BIONIME Rightest GT100 User manual

BIONIME Rightest GM100 User manual

BIONIME

BIONIME Rightest GM100 User manual

BIONIME GM550 User manual

BIONIME

BIONIME GM550 User manual

BIONIME RIGHTEST Max User manual

BIONIME

BIONIME RIGHTEST Max User manual

BIONIME RIGHTEST GT333 User manual

BIONIME

BIONIME RIGHTEST GT333 User manual

BIONIME GM550 User manual

BIONIME

BIONIME GM550 User manual

BIONIME RIGHTEST Max User manual

BIONIME

BIONIME RIGHTEST Max User manual

BIONIME Rightest ALPHA User manual

BIONIME

BIONIME Rightest ALPHA User manual

BIONIME GM550 User manual

BIONIME

BIONIME GM550 User manual

BIONIME Rightest GM300 User manual

BIONIME

BIONIME Rightest GM300 User manual

BIONIME Rightest GM300 User manual

BIONIME

BIONIME Rightest GM300 User manual

BIONIME Rightest ALPHA User manual

BIONIME

BIONIME Rightest ALPHA User manual

BIONIME RIGHTEST GM280B User manual

BIONIME

BIONIME RIGHTEST GM280B User manual

BIONIME rightest GM260 User manual

BIONIME

BIONIME rightest GM260 User manual

BIONIME Rightest GT100 User manual

BIONIME

BIONIME Rightest GT100 User manual

BIONIME RIGHTEST GM280B User manual

BIONIME

BIONIME RIGHTEST GM280B User manual

BIONIME GM232B User manual

BIONIME

BIONIME GM232B User manual

BIONIME RIGHTEST HR310 User manual

BIONIME

BIONIME RIGHTEST HR310 User manual

BIONIME Rightest GM210 User manual

BIONIME

BIONIME Rightest GM210 User manual

BIONIME GM232B User manual

BIONIME

BIONIME GM232B User manual

BIONIME Rightest GM700 User manual

BIONIME

BIONIME Rightest GM700 User manual

Popular Blood Glucose Meter manuals by other brands

Lifescan OneTouch Profile blood glucose monitor owner's booklet

Lifescan

Lifescan OneTouch Profile blood glucose monitor owner's booklet

GlucoRx TD-4277 owner's manual

GlucoRx

GlucoRx TD-4277 owner's manual

iSens CareSens N Pop manual

iSens

iSens CareSens N Pop manual

OneTouch Verio IQ owner's booklet

OneTouch

OneTouch Verio IQ owner's booklet

OneTouch ONETOUCH ULTRASMART Abbreviated Procedure Guide

OneTouch

OneTouch ONETOUCH ULTRASMART Abbreviated Procedure Guide

GlucoRx HCT owner's manual

GlucoRx

GlucoRx HCT owner's manual

OneTouch ULTRA2 Getting started guide

OneTouch

OneTouch ULTRA2 Getting started guide

Smart Lab mini nG user manual

Smart Lab

Smart Lab mini nG user manual

iSens e-Checker owner's booklet

iSens

iSens e-Checker owner's booklet

CareTouch CT210 Quick reference manual

CareTouch

CareTouch CT210 Quick reference manual

iSens CareSens N Voice Quick reference guide

iSens

iSens CareSens N Voice Quick reference guide

OneTouch ONETOUCH SELECT - CONTROL SOLUTION Starter Guide

OneTouch

OneTouch ONETOUCH SELECT - CONTROL SOLUTION Starter Guide

FOR-A G20 owner's manual

FOR-A

FOR-A G20 owner's manual

B. Braun Omnitest plus manual

B. Braun

B. Braun Omnitest plus manual

Abbott FreeStyle Precision Neo owner's guide

Abbott

Abbott FreeStyle Precision Neo owner's guide

Sinocare iCan quick start guide

Sinocare

Sinocare iCan quick start guide

Dexcom G4 Platinum quick start guide

Dexcom

Dexcom G4 Platinum quick start guide

Acon On-Call OGM-212 user manual

Acon

Acon On-Call OGM-212 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.