If Control Solution results are out of the range, your GM100 BGMS may not
be working properly. Repeat the Quality Control Test. If your Control Solution
results outside the range still exist, do not use the GM100 BGMS to test your
blood glucose. And contact Bionime authorized representative or Customer
Service.
Performing a Quality Control Test 12.
1. Take one test strip from vial and Re-cap the
vial cap immediately.
2. Insert the test strip with view window, facing
up, into test strip port.
3. While the blood drop symbol flashing, press and hold the main button for
over 5 seconds until the " " symbol appears.
Use with Control Solution
4. You will see blinking " " symbol and " "
symbol on the screen prompting you to
apply Control Solution.
5. Shake the bottle of Control Solution well
before opening the cap. Then open the cap
and put it on the table.
6. Drip a drop of Control Solution on the top of
the cap.
7. Gently touch sample entry of the strip with
the Control Solution on the top of the cap.
CLOSE
8. When you hear a beep, leave the meter on
the table while waiting for the test result.
The screen will display the countdown.
9. Tightly replace the cap on the Control
Solution bottle.
10. The Control Solution result appears. Compare your Quality Control Test
result to the Control Solution Range printed on the test strip vial label.
Battery Error
1. The " " symbol is blinking when the battery
power is low. Please change battery as soon
as you can. You can still do the test.
2. The " " and " " symbols are blinking
when the battery is too low. Meter can not do the
strip test. Please change the battery immediately.
3. After changing the battery, perform a Quality Control Test.
Temperature Error
In order to get accurate test result, testing between
(50 ~ 104 ˚F).
perform 10 ~40 ˚C
1. When the ambient temperature is 0 ~ 9 ˚C (32 ~ 49 ˚F) or
41 ~ 50 ˚C (105 ~ 122 ˚F) the " " warning symbol will
be blinking, you still can do the test but the test result is
only for reference because the test result under these
ranges of temperature might not be correct. Repeat the test
at an area with temperature between operating range. (10 ~
40˚C or 50 ~ 104˚F)
2. When you move from the area with temperature outside the
operating range of test strip to another area with
temperature inside operating range of test strip, please wait
for 30 minutes before you do the test.
3. When the temperature is below 0 ˚C (32 ˚F) or over 50 ˚C
(122 ˚F), Meter can not do the test and the " " symbol
will blink in this condition. Please move the meter to
environment with temperature between 10 ~ 40 ˚C (50 ~
104 ˚F) and repeat the test after 30 minutes.
Sampling Error
Before " " appears. Please don't apply the blood to the entry
of the strip because the meter is still doing internal check. If
you do so, the meter will show " " and " ". Please
remove the test strip and insert a new one to do the test again.
Meter Malfunction
If you find the meter can't be switched on, please follow the steps below for
quick inspection.
1. Open the battery cover, take out the battery.
2. Wait for 5 minutes and reload the new battery.
The meter should be work normally after finishing above steps. If not, please
contact Bionime authorized representative or Customer Service.
PRECAUTION
- Check the expiration date printed on the package every time you use
a test strip. Do not use expired test strips.
- Use each test strip immediately after removing from the vial.
- Do not reuse test strips.
- Do the test at least 30 minutes after moving into a different location
with significant change in temperature.
- Place the blood drop only on the sample entry of the
test strip.
- Please don't drip or inject the blood sample directly by
syringe to the entry port of test strip. Doing this might
contaminate the meter or cause damage and is not recommended.
The GM100 meter displays results between 0.6 and 33.3
mmol/L. If your test result is below 0.6 mmol/L , " " will
appear on the screen. Please repeat your test again by a new
test strip. If you still get " " result, you should immediately
contact your healthcare professional.
Blood glucose test results are shown on the meter as ,
depending on which unit of measurement you have chosen. ( This meter
for some countries is fixed at mmol/L or mg/dL and is not able to be
changed freely.) Consult your healthcare professional before making any
changes to your diabetes medication program.
If your blood glucose result is unusually high or low, or if you question your
testing results, repeat the test with a new test strip. You can also run a
Quality Control Test with to check your meter and test strip. If the test result
still remains unusually high or low, contact your healthcare professional
immediately.
If you are experiencing symptoms that are not consistent with your blood
glucose test results and you have made sure to follow all instructions of this
manual, contact your healthcare professional immediately.
mg/dL or mmol/L
*
If your test result is above the high end of the system's detection
range 33.3 mmol/L, " " will appear on the screen. Please
repeat your test again by a new test strip. If you still get
" " result, you should immediately contact your healthcare
professional.
Recalling Test Result & Average10.
Insufficient blood sample Enough blood sample
View window appearance
9.
Make sure your blood sample covers the whole area of the View Window to
get an accurate test result. Insufficient blood sample will cause inaccurate
test result. Repeat the test with a new test strip.
PRECAUTION
Each time you open a new bottle of
Control Solution, write the discard date
on the label. Control Solution is good
for 3 months after opening the bottle,
or until the expiration date printed on
the label, whichever comes first.
Exp. Date 2017-12
When should perform a Quality Control Test ?
Before doing a blood glucose test with your meter for the first time.
When you open and start using a new test strip vial.
When your meter is dropped or splashed with liquids.
Whenever you think your test result does not consistent with the way you
feel.
Whenever you want to check if your system is working properly or not.
Whenever you want to practice testing and check correct procedure.
Please use Control Solution tested with GM100 BGMS under Control
Solution Mode. If the test result is within the Control Solution Range printed
on the strip vial label, the GM100 BGMS passes Quality Control Test. That
means your GM100 BGMS is working correctly.
Entering Control Solution Mode
Insert the strip into the strip port on meter. While the blood drop symbol
flashing, press and hold the main button for over 5 seconds. Then you will
see " " symbol blinking on the screen indicating that you've successfully
entered the Control Solution Mode.
Quality Control Test11.
PRECAUTION
- Your Control Solution results will not be calculated for average
reading but still can be recalled. The Control Solution Test result will
be shown with " " symbol on the screen.
Example of Control Solution Range
printed on your test strip vial label.
Control Solution Range
4.4-5.9 mmol/L
Maintenance
Keep your meter and test strip free of dust, water or any other liquid. Store
the meter in the carrying case when not in use. If meter is dropped or
damaged, perform a Quality Control Test before doing a blood glucose test.
Cleaning Meter
Clean the outside of the meter with a damp cloth and mild soap/detergent.
Keep the test strip port from getting wet.
- Don't drip the Control Solution to sample entry of the strip directly.
The reagent on strip might be sucked into the bottle of Control
Solution and might cause the degeneration of Control Solution.
Doing this might contaminate the meter via the
test strip port as well.
- Don't touch the Control Solution. If you have touched it,
please clean up with water.
- Before " " and " " appears, please don't touch the Control
Solution to the sample entry on strip because the meter is still in an
internal check. If you do so, the meter will show " " and " ".
Measurement Technology
Sample
Minimum Sample Volume
Measuring Range
Operating Temperature
Operating Relative Humidity
Hematocrit
Power Supply
Battery Life
Meter Dimension
Meter Weight
Monitor
Display Area
Meter Storage Conditions
Test Strip Storage Conditions
Specification
Oxidase Electrochemical Sensor
Capillary
1.4 microliters (
0.6 - 33.3 mmol/L
whole blood
µL )
Error Message and Trouble Shooting13.
Er1 - The inserted strip has been used or
damaged. Please use a new test strip
from vial.
test
Er2 - Meter has malfunctioned. Do the Quality
Control Test or reinstall the battery to see
if the meter works properly.
Er3 - Signal transmission is disrupted, repeat the test.
Above, if error screen still appears contact Bionime
authorized representative or Customer Service.
4~30 ˚C(39~86 ˚F), < 90 % relative humidity
Limitations
- GM100 BGMS is not intended for serum or plasma test.
- Inaccurate test results may be obtained at high altitude more than about
10,000 feet ( 3,048 meters ) above sea level.
- Severe dehydration and excessive water loss may cause inaccurately low results.
- GM100 BGMS has not been validated for use on neonates.
- The glucose test may be interfered under abnormal concentration of
Ascorbic Acid ≧ 5 mg/dL ( 0.28 mmol/L )
Dopamine Hcl ≧ 2 mg/dL ( 0.11 mmol/L )
L-Dopa ≧ 3 mg/dL ( 0.15 mmol/L )
Tolazamide ≧ 15 mg/dL ( 0.48 mmol/L )
Cholesterol ≧ 500 mg/dL ( 12.93 mmol/L )
Glutathione reduced ≧ 60 mg/dL ( 1.95 mmol/L )
Hemoglobin ≧ 6,000 mg/dL ( 0.94 mmol/L )
Uric acid ≧ 9 mg/dL ( 0.54 mmol/L )
Customer Service
We sincerely like to provide complete, considerate services to our customers.
Please review all the instructions to make sure you are performing the steps
correctly. If you have any questions or problems of GM100 products,
Warranty
Bionime Corporation warrants that your GM100 Meter will be free from
defects in materials and workmanship for five years from the date of
purchase.
This warranty does not apply to the performance of a GM100 Meter that has
been altered, misused, tampered with or abused in any way.
This warranty applies only to the original purchaser of the meter.
Please complete and return the enclosed warranty card to Bionime
authorized representative.
Different models have different specifications. Some of the models are not
included with the warranty card.
Parts of Critical Component
Blood Glucose Meter, Test Strip, Control Solution and Lancing Device
Manufacturer: Bionime Corp.
Product complied with In Vitro Diagnostic Medical Device Directive 98/79/EC.
(CE 0197)
EU Rep: BIONIME GmbH; Tramstrasse 16, 9442 Berneck , Switzerland
For in vitro diagnostic use
Temperature limitation Manufacturer
LOT
IVD Lot number
Use by
Do not reuse
Method of sterilization using irradiation
10 ~ 40 ˚C (50 ~ 104 ˚F)
10 - 90 %
35 - 48 %
One CR2032 battery
About 1,000 tests
95.0 mm x 43.8 mm x 13.0 mm
43.0 g with batteries
LCD display
38.0 mm x 29.0 mm
-10 ~ 60 ˚C (14 ~ 140 ˚F)
Disposable Sterile Lancets
Manufacturer: SteriLance Medical (SuZhou) Inc.
No.68 LiTangHe RD, XiangCheng, SuZhou, JiangSu 215133, P.R. China
EC- Rep:
EMERGO EUROPE, Molenstraat 15, 2513 BH The Hague, The Netherlands
Product complied with Medical Device Directive 93/42/EEC ( CE0197 ).
150 blood glucose test results with date and time
8 seconds
Power Saving Turn off automatically after 2 minutes no use.
Or press the " " button for 2 seconds.
OPEN
The GM100 meter is able to store 150 test results with time and date
automatically. If your meter has stored 150 results, which is the maximum
memory of the meter, the newest test result will replace the oldest one.
To recall your test memory, start with the meter without test strip inserted.
3. Quick Searching: you could also quick press the main button
twice, and then it will automatically display the average and
memory by sequence. Anytime you want to stop it, you could just press
the main button again. It will stop and display the reading right at the
moment you press the main button. You could just apply the way above;
especially when you find a particular reading you want to check with.
4. Reading right after test: if you just finish the test and press the main button
to review. The display sequence will be the latest reading, 1-day, 7-day, 14-
day, 30-day, 90-day average, and the memory 2 reading until the oldest
one (memory 150 reading).
1. Press the main button to switch to screen to memory function,
you will see ”AVG” symbol on the up left corner of the screen.
The display sequence will be 1-day, 7-day,14-day, 30-day, 90-
day average, and then memory 1 reading, memory 2
reading,…until memory 150 reading which will show ”MEM”
on the up right corner of the screen.
2. Under average screen, the number on the down right corner
of the screen means how many days of readings have been
calculated as the average while the number on the down left
corner of the screen means how many readings have been
counted in. Under memory screen, it will show the sequence
number on the down left corner and then display date and time.
PRECAUTION
- You have to set the time and date to activate the average function.
- The " " , " " results, the Control Solution results and the test
result made out of normal temperature range (<10 ˚C, >40 ˚C) are
not calculated in the average.
Test Str ips
The your Control Solution results are out of the range :
Y
dure correctly.
Malfunction of the meter.
possible reasons
- our Control Solution has expired or after 3 months since opened.
- Your test strip has expired.
- You leave the cap off the vial of test strips or Control Solution for a long time.
- You didn't perform the test proce
-
Bionime GmbH
Tramstrasse 16
9442 Berneck
Switzerland
EU representative
Control Solution Range
CE-mark (with No. of notified body)
Version: August 2017
Test Time
Memory Capacity
Consult the instruction for use Biological risks
WEEE