Bionizer OxiBOSS User manual

DOWN UP
AUTO /
OFF
CANCEL
PROGRAM /
SET
TM
C h l o r i n e - F r e e F r e s h w a t e r P o o l s
TM
INSTRUCTION BOOK
Liquid Oxidizer & Sanitiser Control System
Please read before you call us for advice. Version 2.1a, 19-Sep-2019
bionizer.com.au bionizer.com
Bionizer highly
recommends that a
pH BOSS automatic
pH Controller is used
together with the
OxiBOSS Controller.
When using OxiBOSS
for liquid chlorine,
an anti-crystallising
agent should be
used with very new
drum of Chlorine.
C h l o r i n e - F r e e F r e s h w a t e r P o o l s

CONTENTS
2
TABLE OF CONTENTS
Page
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
OxiBOSS DISPLAY INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
OxiBOSS ILLUSTRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
SAFETY INSTRUCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
CHEMICAL SAFETY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
DRUM WEIGHT HOSE INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
PROGRAMMING YOUR pH BOSS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
RESETTING THE SQUEEZE TUBE REMINDER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
BIONIZER CONTACT INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
CONTROL BOX INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
PUMP SQUEEZE TUBE LUBRICATION AND REPLACEMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
TEE PIECE INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
TEE PIECE HOSE INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
AUTO / OFF BUTTON . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
CHANGING THE pH SET POINT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
ORP PROBE ORIENTATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
SANITISER PUMP TUBE INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
DOSE SETTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
MANUAL DOSE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
ALARM FUNCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
ORP PROBE CALIBRATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
FITTING NEW SUPPLY AND DELIVERY TUBES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
ORP PROBE INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
OxiBOSS INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

INTRODUCTION
IMPORTANT : SEE CHEMICAL SAFETY & WARRANTY SECTION.
PLEASE TAKE THE TIME TO READ THESE INSTRUCTIONS CAREFULLY AND NOTE HOW
THEY APPLY TO YOUR PARTICULAR POOL AND EQUIPMENT.
When Chlorine is the primary sanitiser in a
swimming pool application, an ORP level of around
650 mV is adequate to maintain a good level of
sanitising power all year round.
Please note that when running Chlorine at under
1.0 ppM, we highly recommend that an Eco-
Oxidizer is also installed. The Eco-Oxidizer will
prevent accumulation of combined chlorine (such
as chloramines), which are cause red eyes, itchy
skin and that “chlorine smell”. The Eco-Oxidizer will
oxidize combined chlorine back to free chlorine
continuously, whenever the pool filtration pump is
running.
ORP is not specifically a measurement of the ppM
of Chlorine, so there is no conversion from one to
the other. In many ways, ORP is a much easier way
to maintain a pool than attempting to maintain a
specific ppM of Chlorine. For example, a heated
pool with high bather load would need around
4ppM of Chlorine to maintain a chemical activity of
650 mV ORP. A pool that is not being used during
winter with low water temperature may need
1ppM of Chlorine to maintain the same level of
650mV of ORP.
WHAT IS “ORP” ?
“ORP” stands for “Oxidation Reduction Potential”.
It is also often referred to as “Redox”. In simple
terms, ORP is a measurement of chemical activity
in a body of water. The unit of measurement is
millivolts (”mV”).
When Chlorine is used as a backup sanitiser, for
example when a Bionizer is installed, an ORP level
of around 550 mV will help to maintain a crystal
clear pool, but with a lower Chlorine level.
When using the OxiBOSS, the only requirement for
testing chlorine is a regular test with a chlorine test
kit. The frequency of this should be once per week
in a commercial pool with heavy bather load, down
to once per month for a domestic swimming pool
with a small bather load.
So an ORP system can be used with the SAME set
point all year round, and will simply add the correct
amount of Chlorine to maintain a suitable level of
sanitising effectiveness. EASY ! The graph below
illustrates this point.
3
INTRODUCTION
SUMMER AUTUMN WINTER SPRING
Required Chlorine, ppM
Required ORP, mV

DOWN UP
AUTO /
OFF
CANCEL
PROGRAM /
SET
TM
C h l o r i n e - F r e e F r e s h w a t e r P o o l s
4OxiBOSS ILLUSTRATION
OxiBOSS ILLUSTRATION
DISPLAY
AUTO / OFF
PROGRAM / SET
STATUS INDICATOR
UP & DOWN SETTING ADJUSTMENT
PERISTALTIC DOSING PUMP
BNC ORP PROBE SOCKET
ORP PROBE
DC POWER SOCKET
Press to select:
Press this button to cancel without saving changes when adjusting
settings in the programming menus.
OFF..................Unit will not add any sanitiser automatically.
Auto ................Normal automatic sanitiser control
Press to scroll through the six user setting menus.
Press to save changes when adjusting settings.
16 x 2 Line alphanumeric LCD with LED backlight.
Press to makes changes to user settings.
Precision, high quality dosing
pump. 160mL/min flow rate
while running.
Connect ORP Probe by
pushing on, then
clockwise 1/4 turn to
lock into place.
Use only genuine
power adaptor
supplied by Bionizer to
avoid damage to your
OxiBOSS. Tee Piece, ORP Probe
and one way valve
One Way
Valve
FLOW DIRECTION
Tee Piece
Indicates the current status of the OxiBOSS, for example if it is
adding chemical, or waiting for the added chemical to mix in the
pool.
See page 5 for full information.

See page 13 for ORP Probe calibration instructions.
The OxiBOSS will show the ORP Probe calibration
reminder, 12 months after the last successful
calibration.
When the peristaltic dosing pump has operated for
a total of 50 hours, the OxiBOSS will show the
squeeze tube reminder.
See page 15 for instructions on how to lubricate or
change the Squeeze Tube.
Cal Age is the number of days since the last
successful ORP probe calibration.
Tube Age displays how many hours the peristaltic
dosing pump has pumped sanitiser with the current
squeeze tube.
OxiBOSS DISPLAY INFORMATION
OxiBOSS DISPLAY INFORMATION
The 32 character display in your OxiBOSS provides all the information you need to help ensure that the unit is
maintaining the sanitizer level in your pool accurately, and that you know when it is time to carry out some
basic maintenance. Additionally, the OxiBOSS will warn you when it detects that something is not working as it
should. This page is a summary of the information that your OxiBOSS may display at various times.
Main OxiBOSS Display
5
The current ORP of the Pool.
The Set Point at which the OxiBOSS will
maintain the ORP level. See page 12.
The arrow is also animated whenever
the dosing pump is adding chemical.
The Ç Arrow here indicates that the
OxiBOSS is set to increase ORP with
sanitiser when the ORP reading is too
low.
Status Indicator
Wait ....OxiBOSS is waiting for added sanitiser to mix
into the pool before adding more, if required.
Alrm ....ORP has been below the Set Point for 4
continuous hours of operation, without ever
reaching the Set Point. See page 12.
Add .....OxiBOSS is in Auto mode, reading is below the
Set Point, and dosing pump is adding sanitiser.
OK .......OxiBOSS is in Auto mode, and reading is within
the Set Point. No action required at this time.
“Wait” is also displayed at start-up while the
system is waiting for the ORP to stabilise.
Cool.....OxiBOSS has switched dosing pump off for 1
minute to cool the squeeze tube.
Off.......OxiBOSS dosing pump is switched off. No
sanitiser will be added, regardless of reading.
See page 11.
Alarm and Reminder Messages
When the ORP reading has been below the Set
Point for 4 hours, the display alternates between
the alarm screen, and “Alrm” in the status, and the
beeper sounds. See page 12 for how to reset the
alarm and rectify the cause.

SAFETY INSTRUCTIONS
SAFETY INSTRUCTIONS - READ
PLEASE FOLLOW THESE IMPORTANT SAFETY INSTRUCTIONS WHEN HANDLING ANY
CHEMICALS.
CHEMICAL SAFETY
4. The life of the unit will be extended (as with most
other pool equipment) if it is placed indoors or in
an enclosed structure. This is recommended as
sunlight can damage the Liquid Crystal Display on
the face of the control unit. The LCD is covered
within the warranty on a pro-rata basis.
IMPORTANT :
3. Keep all cables and tubes visible and do not bury
them. Also, position them so that they do not get
damaged by lawn mowers, hedge trimmers and
other equipment.
4. WAR NING - To prevent electrical shock,
sanitiser spillage or incorrect dosing, replace
damaged cords or tubes immediately.
1. Ensure you read and follow all instructions.
CAUTION :
Your warranty could become invalid should you fail
to observe the following :
3. Turn off the power to the OxiBOSS before
replacing the peristaltic dosing pump squeeze
tube.
It is important to follow basic safety precautions
when installing and using this electrical equipment,
including the following :
5. Always power the OxiBOSS from a genuine
Bionizer power adaptor. An incorrect adaptor
could damage the OxiBOSS and will void your
warranty.
5. Read these instructions and keep the instruction
manual in a safe place.
2. Damage due to chemical spillage is not covered
by warranty.
2. WARNING - Do not allow children to use this
product unless they are closely supervised at all
times. This is to prevent accidental injury.
1. The OxiBOSS must be installed and operated as
specified.
Liquid Chlorine does not need to be diluted for any size of pool.
When decanting Liquid Chlorine (or any other chemical) from one container to another :
1. ALWAYS wear safety glasses and gloves.
2. ALWAYS decant chemicals in a well ventilated area, preferably outdoors if practical.
5. ALWAYS have a source of fresh water nearby, such as a hose, so you can wash down accidental spillages.
6. ALWAYS ensure that the sanitiser drum is BELOW the level of the OxiBOSS Tee, to ensure that the
sanitiser cannot syphon from the drum to the pool when the pool pump is switched off.
7. In case of sanitiser spillage onto clothing, eyes, skin, concrete etc, wash down with large amounts of fresh
water immediately. SEEK MEDICAL ADVICE IF NECESSARY.
6

NOTE THAT THE 10cm SPACINGS SHOWN ABOVE ARE FROM THE EDGES OF THE PODS, NOT
THE CENTRE OF THE PODS.
OxiBOSS INSTALLATION
READ THE DO’S AND DON’TS AND TAKE THE TIME TO INSTALL CORRECTLY.
OxiBOSS INSTALLATION
3. Be sure to install the Tee piece in the correct
direction of the water flow, as per the arrow on
the Tee.
The Tee piece should be installed so that the
O
ORP Probe is within 30 of vertical. (See next
page)
1. Installing your OxiBO SS takes around 30
minutes.
4. The Probe/Injection Tee must be fitted on the
pipe that sends water back to the pool after it has
been filtered (the “Return Line”).
The OxiBOSS can be installed onto a new pool or an
existing pool with minimal pool preparation. Ensure
that the pool is correctly balanced, in particular the
pH.
TEE PIECE INSTALLATION
3. If installing it yourself you will need :
1. For 50mm (2") pipe, cut out a 6cm (2.5") section
of pipe sufficient for the Probe/Injection Tee
piece. For 40mm (1.5") pipe, where the reducing
bushes will be used, cut out a 7cm (2.75") section
of pipe.
NOTE: Turn off all valves prior to fitting the T-piece
(avoid leaks), and turn them back on again before
switching pool pump back on.
!PVC pressure glue and primer.
!A fine tooth saw.
2. Alternatively engage a competent person to do
the job.
!If a pH BOSS, Eco-Oxidizer and/or Bionizer is
installed, fit the OxiBOSS Tee LAST, AFTER the
pH Tee, Eco-Oxidizer Pod and Bionizer Pod .
!If a Salt Chlorinator is installed, fit the OxiBOSS
tee AFTER the Salt Chlorinator cell.
2. If the Tee piece diameter is larger than the pipe
then use the 50mm to 40mm reducing bushes
provided by glueing them in place in the Tee
piece before fitting. The pipe should now fit
snugly into the Tee piece.
Filter
For sand filters,
glass or silica sand
media provides
optimum results.
Filter
Pressure
Pool
Pump
pH Tee ORP Tee
10cm
min 10cm
min
Oxidizing
Pod Ionizing
Pod
If a heater is
installed, then fit
the heater FIRST.
FLOW DIRECTION FLOW DIRECTION
7
10cm
min

ORP Probe Orientation.
When installing the OxiBOSS Tee, it is critical that
the probe be within 30 from vertical.
DO NOT FIT THE TEE TO A VERTICAL PIPE, AS
THIS WILL MAKE THE ORP PROBE HORIZONTAL,
WHICH WILL RENDER THE ORP PROBE
INOPERABLE.
Correctly Installed Tee
Incorrectly Installed Tee
Do not mount on vertical section of pipe, which makes the probe horizontal :
OxiBOSS INSTALLATION
FLOW DIRECTION
8

DOWN U P
AUTO /
OFF
CANCEL
PROGRAM /
SET
TM
C h l o r i n e - F r e e F r e s h w a t e r P o o l s
OxiBOSS INSTALLATION
SET THE TIMER FOR THIS OUTLET TO THE
SAME SETTING AS THE POOL PUMP OUTLET.
2. Make sure that the unit is undercover and that the
Display is not in direct sunlight as it is not covered
by warranty if it is damaged by sunlight.
2) Plug the optional AC/DC power adaptor into
one of the two mains power outlets provided on
the Bionizer unit.
4. Connect the low voltage plug to the OxiBOSS DC
Power Socket (see Page 4). Use only a genuine
Bionizer power adaptor or warranty may be void.
1) Connect the short DC to DC adaptor lead from
one of the Bionizer DC outlet sockets to the
OxiBOSS DC Power Socket.
3) When connecting to a pool control system,
connect the AC/DC adaptor to a suitable power
outlet on the control system, if available.
CONTROL BOX INSTALLATION
The control box MUST be mounted vertically, with
the cables at the bottom, in order to be weather
resistant. Any water ingress due to the unit being
incorrectly mounted is not covered by warranty.
1. Choose a position that will allow you to see the
OxiBOSS Display easily. Fix the control box in
position using screws at each corner of the box.
NOTE: Make sure that the OxiBOSS is positioned
near enough to a power outlet and to the T-piece
and sanitiser drum to allow the power connecting
and sensor cables to reach without straining.
3. The OxiBOSS should only be powered on when the
pool pump is running. This can be done in 3 ways :
For USA installations, when there is no spare
outlet socket, cut the 110V plug off the AC/DC
adaptor cable, and hard wire to the control system
terminals.
DO NOT SWITCH THE POWER ON UNTIL
INSTALLATION HAS BEEN COMPLETED.
SANITISER PUMP TUBE INSTALLATION.
Drum weight
inside sanitiser drum.
Drum Label
(e.g. Chlorine) Ensure that your Sanitiser
Drum has the correct label
with applicable safety
information
Delivery tube
firmly fixed
into one way
valve in Tee.
See page 10.
Supply tube fed
through 7.0mm
(1/4") hole, drilled
into sanitiser drum lid.
Supply tube
on left side
of sanitiser
pump
Delivery tube
on right side
of sanitiser pump
FLOW DIRECTION
9

OxiBOSS INSTALLATION
2. Drill a 7.0mm or 1/4 inch hole into the top of the
drum lid.
3. Push the end of the drum weight hose through
the drum lid.
1. Remove the lid from the chemical drum that you
will use.
DRUM WEIGHT HOSE INSTALLATION
4. Fit the hose onto the drum weight tube fitting in
the same way as described in the Tee Piece Hose
Installation section above and illustration below.
1. Fit Teflon tape onto the thread of the ORP Probe. Carefully remove the
protective cap from the tip of the probe.
ORP PROBE INSTALLATION
2. Screw the ORP Probe into the Tee piece thread. Tighten firmly by hand. A
tool may be used for another ½ to 1 turn, if the fitting is leaking when the
pump is running. DO NOT OVER-TIGHTEN.
3. After the ORP Probe is installed in the Tee piece, connect plug on the ORP
Probe cable into the probe socket on the OxiBOSS. Push the plug into place
and 1/4 turn clockwise to lock it into place. See the OxiProbe socket
diagram, Page 4.
AFTER ALL HOSE FITTINGS ARE COMPLETED, AND THE PROBE IS FITTED, TEST
THE SYSTEM WITH FRESH WATER, USING THE MANUAL DOSE FUNCTION
(PAGE 12). CHECK CAREFULLY FOR ANY LEAKS AND CORRECT AS NEEDED.
1. Remove the nut from the one way valve fitting on
the Tee piece.
2. Slip the Nut over the hose.
TEE PIECE HOSE INSTALLATION
3. Push the hose firmly onto the tube fitting.
4. Tighten the nut over the hose. Tighten firmly
with fingers. DO NOT USE A TOOL.
10

AUTO / OFF BUTTON
O F F : Unit will not
dose any sanitiser
again until this setting
is changed.
AUTO / OFF is also the “CANCEL” button to exit
without saving changes in programming screens.
AUTO: Unit will dose
sanitiser when ORP is
outside the set point.
Press to alternatively select Auto or Off : This range of this
setting is 2 to 25
minutes.
!Use the p and q buttons to enter setting 2 to 25.
!Press PROGRAM / SET save the new setting.
As a guide, use the following setting as a STARTING
point when the unit is first installed :
2....Up to 30,000 L / 8,000 Gallons
5....30,000 to 75,000 L / 8,000 to 20,000 Gallons
15..150,000 to 250,000 L / 40,000 to 65,000 Gallons
DOSE SETTING
Alternatively, press AUTO / OFF to cancel without
saving.
The dose setting is used to set the sanitiser pump
addition and waiting times, and are related to how
quickly the pool turns over the water. This is related
to both pool size, and pump/pipe setup.
In general, a smaller pool would use a lower Dose
Setting than a large pool. However other factors can
affect this, such as pool turnover time.
20..250,000 to 500,000 L / 65,000 to 130,000 Gallons
25..>500,000 L / 130,000 Gallons
10..75,000 to 150,000 L / 20,000 to 40,000 Gallons
If the OxiBOSS does not keep up with sanitiser
demand, and the Alarm sounds, increase the dose
setting by one step and check in 2 days.
If the alarm is tripped due to high ORP, reduce the
dose setting by one step and check in 2 days.
PROGRAMMING YOUR OxiBOSS
!If no key has been pressed for 30 seconds, the unit will return to the normal readout display. Any changes
will not be saved.
PROGRAMMING YOUR OxiBOSS
!Press the PROGRAM / SET button to move through each of the five user settings.
!Use the p and q keys to change settings.
!When changing a setting, press PROGRAM / SET after the desired setting is programmed to save the
changes and return to the normal readout display.
!To quit any screen without saving changes, press AUTO / OFF to cancel.
11

PROGRAMMING YOUR OxiBOSS
12
!Press PROGRAM / SET to start pumping.
!Use the p and q buttons to enter 1 to 25 Litres.
Please note that the OxiBOSS factory preset default
Set Point is 650mV. This can be adjusted up or down,
based on pool balance tests.
!Press PROGRAM / SET to save.
!Take care when using this function, to not
overdose the pool with sanitiser.
T h e ra n g e o f this
setting is from 0 to 25
Litres.
Use this function to add a measured dose of sanitiser.
CHANGING THE ORP SET POINT
!Use the Use the p and q buttons to enter a new
Set Point value.
MANUAL DOSE
Alternatively, press AUTO / OFF to quit without
saving changes.
Problem ...ORP Probe is broken or faulty.
Fix.............Check all tube and hose fittings for
leakage. Re-fit or replace as necessary.
!Press PROGRAM / SET to save.
When the alarm has been triggered, the OxiBOSS
will sound the beeper and also display will
alternative between an alarm screen and “ ” in Alrm
the status message :
!Use the Use the p and q buttons to select On or
Off.
Use PROGRAM / SET to select Alarm.
You must press AUTO / OFF to reset the alarm and
return to normal operation.
THE SANITISER PUMP WILL BE DISABLED AND
THE OxiBOSS WILL NOT PUMP MORE SANITISER
UNTIL THE ALARM HAS BEEN RESET.
Problem ...Dose Setting is too low or too high.
Fix.............See page 11 for information on how to
change the Dose Setting.
ALARM FUNCTION
When the Alarm is set to ON, the OxiBOSS will sound
the beeper and disable the sanitiser pump when the
ORP is below the set point for more than 4 hours.
The reason for the alarm being triggered should be
found and rectified too. It could be due to any of the
following :
Problem ...ORP Reading is incorrect
!Press AUTO / OFF to quit without saving changes.
Fix.............Refill the sanitiser drum.
Fix.............Re-calibrate the ORP Probe. See page 13.
Fix.............If re-calibration is not possible, then
replace the ORP Probe.
When the Alarm is set to OFF, the alarm will never be
triggered, regardless of the ORP reading. Only
advanced users should set the alarm to OFF, and only
temporarily while the cause of the alarm beeping is
being rectified.
A replacement ORP Probe can be
ordered by telephone or online at
www.bionizer.com.au
Problem ...Sanitiser drum is empty.
Problem ...Pump is not pumping sanitiser correctly.
Check squeeze tube inside pump for
leakage and replace if necessary. See
page 15.
A replacement Squeeze Tube can be
ordered by telephone or online at
www.bionizer.com.au

! Unplug the ORP Probe from the OxiBOSS first.
(This will avoid the cable getting twisted when the
ORP Probe is unscrewed. Un s crew t he pro be
completely from the Tee.
! When the reading is stable and the “ü” is
displayed, press PROGRAM / SET to calibrate.
5. Calibration Procedure
! Rinse the probe well in clean water. De-ionized or
distilled water is ideal, but tap water will suffice.
4. Place OxiProbe into 230mV ORP Standard
! Press the PROGRAM / SET button until the ORP
Calibration screen is displayed, as per the example
below. The “Current” reading may be 220 or
another value, depending on the condition of the
ORP Probe and current calibration setting.
! Reconnect the ORP Probe connector to the
OxiBOSS.
! Adjust the “Cal ORP” value to exactly 230. The
flashing “X”, indicates that the reading is not yet
stable. DO NOT calibrate until the “X” changes to a
“ü”.
3. Remove the ORP Probe from the Tee
! Switch the OxiBOSS power on. If it is connected to
a Bionizer, set the Bionizer to Manual:On mode to
switch on the power outlet.
! Press the PROGRAM / SET button twice to bypass
the start-up screen delay.
! If you are using a Bionizer to control the power to
your pool pump and OxiBOSS, unplug the pool
pump from the Bionizer Control Unit.
ORP PROBE CALIBRATION
he OxiBOSS will display a reminder to calibrate the
ORP Probe approximately every 6 months:
! (Connect the OxiBOSS to a power outlet socket,
but do not switch on yet.)
! (If you are powering the OxiBOSS and Pool Pump
from an external power source, please ensure that
the pool pump is switched off and unplugged
during the entire calibration procedure.)
The ORP Probe should be calibrated to ensure that
the readout is correct. Calibration is simply the
process of placing the ORP Probe into a solution of
known ORP, and adjusting the readout to that value.
2. Prepare 230mV ORP Calibration Standard
! Leave the OxiBOSS connected to the Bionizer.
1. Prepare for OxiBOSS for Calibration
MAINTENANCE
A LITTLE PREVENTATIVE MAINTENANCE WILL HELP YOU GET THE MOST FROM YOUR OxiBOSS
MAINTENANCE 13
30
20
10
100
50
!There are TWO bottles of
ORP Calibration solution
(ZoBell Part A & Part B).
! Empty the entire contents
of small Part B bottle into
the larger Part A bottle.
! Replace the lid onto the
Part A bottle a shake
gently to mix the contents
well.
!Dispose of solution in
normal drain such as a
kitchen sink. Solution
expires within 4 weeks of
being made up.
! Unplug the ORP Probe from the OxiBOSS first.
(This will avoid the cable getting twisted when the
ORP Probe is unscrewed. Un s crew t he pro be
completely from the Tee.
! When the reading is stable and the “ü” is
displayed, press PROGRAM / SET to calibrate.
5. Calibration Procedure
! Rinse the probe well in clean water. De-ionized or
distilled water is ideal, but tap water will suffice.
4. Place OxiProbe into 230mV ORP Standard
! Press the PROGRAM / SET button until the ORP
Calibration screen is displayed, as per the example
below. The “Current” reading may be 220 or
another value, depending on the condition of the
ORP Probe and current calibration setting.
! Reconnect the ORP Probe connector to the
OxiBOSS.
! Adjust the “Cal ORP” value to exactly 230. The
flashing “X”, indicates that the reading is not yet
stable. DO NOT calibrate until the “X” changes to a
“ü”.
3. Remove the ORP Probe from the Tee
! Switch the OxiBOSS power on. If it is connected to
a Bionizer, set the Bionizer to Manual:On mode to
switch on the power outlet.
! Press the PROGRAM / SET button twice to bypass
the start-up screen delay.
! If you are using a Bionizer to control the power to
your pool pump and OxiBOSS, unplug the pool
pump from the Bionizer Control Unit.
ORP PROBE CALIBRATION
he OxiBOSS will display a reminder to calibrate the
ORP Probe approximately every 6 months:
! (Connect the OxiBOSS to a power outlet socket,
but do not switch on yet.)
! (If you are powering the OxiBOSS and Pool Pump
from an external power source, please ensure that
the pool pump is switched off and unplugged
during the entire calibration procedure.)
The ORP Probe should be calibrated to ensure that
the readout is correct. Calibration is simply the
process of placing the ORP Probe into a solution of
known ORP, and adjusting the readout to that value.
2. Prepare 230mV ORP Calibration Standard
! Leave the OxiBOSS connected to the Bionizer.
1. Prepare for OxiBOSS for Calibration
100
50
! Leave the ORP Probe in this
ORP Standard for at least 10
seconds before proceeding
with the calibration steps
below.
!Place the ORP Probe into ORP
Standard, as far as the probe
will go into the neck.
! Unplug the ORP Probe from the OxiBOSS first.
(This will avoid the cable getting twisted when the
ORP Probe is unscrewed. Un s crew t he pro be
completely from the Tee.
! When the reading is stable and the “ü” is
displayed, press PROGRAM / SET to calibrate.
5. Calibration Procedure
! Rinse the probe well in clean water. De-ionized or
distilled water is ideal, but tap water will suffice.
4. Place OxiProbe into 230mV ORP Standard
! Press the PROGRAM / SET button until the ORP
Calibration screen is displayed, as per the example
below. The “Current” reading may be 220 or
another value, depending on the condition of the
ORP Probe and current calibration setting.
! Reconnect the ORP Probe connector to the
OxiBOSS.
! Adjust the “Cal ORP” value to exactly 230. The
flashing “X”, indicates that the reading is not yet
stable. DO NOT calibrate until the “X” changes to a
“ü”.
3. Remove the ORP Probe from the Tee
! Switch the OxiBOSS power on. If it is connected to
a Bionizer, set the Bionizer to Manual:On mode to
switch on the power outlet.
! Press the PROGRAM / SET button twice to bypass
the start-up screen delay.
! If you are using a Bionizer to control the power to
your pool pump and OxiBOSS, unplug the pool
pump from the Bionizer Control Unit.
ORP PROBE CALIBRATION
he OxiBOSS will display a reminder to calibrate the
ORP Probe approximately every 6 months:
! (Connect the OxiBOSS to a power outlet socket,
but do not switch on yet.)
! (If you are powering the OxiBOSS and Pool Pump
from an external power source, please ensure that
the pool pump is switched off and unplugged
during the entire calibration procedure.)
The ORP Probe should be calibrated to ensure that
the readout is correct. Calibration is simply the
process of placing the ORP Probe into a solution of
known ORP, and adjusting the readout to that value.
2. Prepare 230mV ORP Calibration Standard
! Leave the OxiBOSS connected to the Bionizer.
1. Prepare for OxiBOSS for Calibration
! Unplug the ORP Probe from the OxiBOSS first.
(This will avoid the cable getting twisted when the
ORP Probe is unscrewed. Un s crew t he pro be
completely from the Tee.
! When the reading is stable and the “ü” is
displayed, press PROGRAM / SET to calibrate.
5. Calibration Procedure
! Rinse the probe well in clean water. De-ionized or
distilled water is ideal, but tap water will suffice.
4. Place OxiProbe into 230mV ORP Standard
! Press the PROGRAM / SET button until the ORP
Calibration screen is displayed, as per the example
below. The “Current” reading may be 220 or
another value, depending on the condition of the
ORP Probe and current calibration setting.
! Reconnect the ORP Probe connector to the
OxiBOSS.
! Adjust the “Cal ORP” value to exactly 230. The
flashing “X”, indicates that the reading is not yet
stable. DO NOT calibrate until the “X” changes to a
“ü”.
3. Remove the ORP Probe from the Tee
! Switch the OxiBOSS power on. If it is connected to
a Bionizer, set the Bionizer to Manual:On mode to
switch on the power outlet.
! Press the PROGRAM / SET button twice to bypass
the start-up screen delay.
! If you are using a Bionizer to control the power to
your pool pump and OxiBOSS, unplug the pool
pump from the Bionizer Control Unit.
ORP PROBE CALIBRATION
he OxiBOSS will display a reminder to calibrate the
ORP Probe approximately every 6 months:
! (Connect the OxiBOSS to a power outlet socket,
but do not switch on yet.)
! (If you are powering the OxiBOSS and Pool Pump
from an external power source, please ensure that
the pool pump is switched off and unplugged
during the entire calibration procedure.)
The ORP Probe should be calibrated to ensure that
the readout is correct. Calibration is simply the
process of placing the ORP Probe into a solution of
known ORP, and adjusting the readout to that value.
2. Prepare 230mV ORP Calibration Standard
! Leave the OxiBOSS connected to the Bionizer.
1. Prepare for OxiBOSS for Calibration
! Unplug the ORP Probe from the OxiBOSS first.
(This will avoid the cable getting twisted when the
ORP Probe is unscrewed. Un s crew t he pro be
completely from the Tee.
! When the reading is stable and the “ü” is
displayed, press PROGRAM / SET to calibrate.
5. Calibration Procedure
! Rinse the probe well in clean water. De-ionized or
distilled water is ideal, but tap water will suffice.
4. Place OxiProbe into 230mV ORP Standard
! Press the PROGRAM / SET button until the ORP
Calibration screen is displayed, as per the example
below. The “Current” reading may be 220 or
another value, depending on the condition of the
ORP Probe and current calibration setting.
! Reconnect the ORP Probe connector to the
OxiBOSS.
! Adjust the “Cal ORP” value to exactly 230. The
flashing “X”, indicates that the reading is not yet
stable. DO NOT calibrate until the “X” changes to a
“ü”.
3. Remove the ORP Probe from the Tee
! Switch the OxiBOSS power on. If it is connected to
a Bionizer, set the Bionizer to Manual:On mode to
switch on the power outlet.
! Press the PROGRAM / SET button twice to bypass
the start-up screen delay.
! If you are using a Bionizer to control the power to
your pool pump and OxiBOSS, unplug the pool
pump from the Bionizer Control Unit.
ORP PROBE CALIBRATION
he OxiBOSS will display a reminder to calibrate the
ORP Probe approximately every 6 months:
! (Connect the OxiBOSS to a power outlet socket,
but do not switch on yet.)
! (If you are powering the OxiBOSS and Pool Pump
from an external power source, please ensure that
the pool pump is switched off and unplugged
during the entire calibration procedure.)
The ORP Probe should be calibrated to ensure that
the readout is correct. Calibration is simply the
process of placing the ORP Probe into a solution of
known ORP, and adjusting the readout to that value.
2. Prepare 230mV ORP Calibration Standard
! Leave the OxiBOSS connected to the Bionizer.
1. Prepare for OxiBOSS for Calibration
! Unplug the ORP Probe from the OxiBOSS first.
(This will avoid the cable getting twisted when the
ORP Probe is unscrewed. Un s crew t he pro be
completely from the Tee.
! When the reading is stable and the “ü” is
displayed, press PROGRAM / SET to calibrate.
5. Calibration Procedure
! Rinse the probe well in clean water. De-ionized or
distilled water is ideal, but tap water will suffice.
4. Place OxiProbe into 230mV ORP Standard
! Press the PROGRAM / SET button until the ORP
Calibration screen is displayed, as per the example
below. The “Current” reading may be 220 or
another value, depending on the condition of the
ORP Probe and current calibration setting.
! Reconnect the ORP Probe connector to the
OxiBOSS.
! Adjust the “Cal ORP” value to exactly 230. The
flashing “X”, indicates that the reading is not yet
stable. DO NOT calibrate until the “X” changes to a
“ü”.
3. Remove the ORP Probe from the Tee
! Switch the OxiBOSS power on. If it is connected to
a Bionizer, set the Bionizer to Manual:On mode to
switch on the power outlet.
! Press the PROGRAM / SET button twice to bypass
the start-up screen delay.
! If you are using a Bionizer to control the power to
your pool pump and OxiBOSS, unplug the pool
pump from the Bionizer Control Unit.
ORP PROBE CALIBRATION
he OxiBOSS will display a reminder to calibrate the
ORP Probe approximately every 6 months:
! (Connect the OxiBOSS to a power outlet socket,
but do not switch on yet.)
! (If you are powering the OxiBOSS and Pool Pump
from an external power source, please ensure that
the pool pump is switched off and unplugged
during the entire calibration procedure.)
The ORP Probe should be calibrated to ensure that
the readout is correct. Calibration is simply the
process of placing the ORP Probe into a solution of
known ORP, and adjusting the readout to that value.
2. Prepare 230mV ORP Calibration Standard
! Leave the OxiBOSS connected to the Bionizer.
1. Prepare for OxiBOSS for Calibration
! Unplug the ORP Probe from the OxiBOSS first.
(This will avoid the cable getting twisted when the
ORP Probe is unscrewed. Un s crew t he pro be
completely from the Tee.
! When the reading is stable and the “ü” is
displayed, press PROGRAM / SET to calibrate.
5. Calibration Procedure
! Rinse the probe well in clean water. De-ionized or
distilled water is ideal, but tap water will suffice.
4. Place OxiProbe into 230mV ORP Standard
! Press the PROGRAM / SET button until the ORP
Calibration screen is displayed, as per the example
below. The “Current” reading may be 220 or
another value, depending on the condition of the
ORP Probe and current calibration setting.
! Reconnect the ORP Probe connector to the
OxiBOSS.
! Adjust the “Cal ORP” value to exactly 230. The
flashing “X”, indicates that the reading is not yet
stable. DO NOT calibrate until the “X” changes to a
“ü”.
3. Remove the ORP Probe from the Tee
! Switch the OxiBOSS power on. If it is connected to
a Bionizer, set the Bionizer to Manual:On mode to
switch on the power outlet.
! Press the PROGRAM / SET button twice to bypass
the start-up screen delay.
! If you are using a Bionizer to control the power to
your pool pump and OxiBOSS, unplug the pool
pump from the Bionizer Control Unit.
ORP PROBE CALIBRATION
he OxiBOSS will display a reminder to calibrate the
ORP Probe approximately every 6 months:
! (Connect the OxiBOSS to a power outlet socket,
but do not switch on yet.)
! (If you are powering the OxiBOSS and Pool Pump
from an external power source, please ensure that
the pool pump is switched off and unplugged
during the entire calibration procedure.)
The ORP Probe should be calibrated to ensure that
the readout is correct. Calibration is simply the
process of placing the ORP Probe into a solution of
known ORP, and adjusting the readout to that value.
2. Prepare 230mV ORP Calibration Standard
! Leave the OxiBOSS connected to the Bionizer.
1. Prepare for OxiBOSS for Calibration
! Unplug the ORP Probe from the OxiBOSS first.
(This will avoid the cable getting twisted when the
ORP Probe is unscrewed. Un s crew t he pro be
completely from the Tee.
! When the reading is stable and the “ü” is
displayed, press PROGRAM / SET to calibrate.
5. Calibration Procedure
! Rinse the probe well in clean water. De-ionized or
distilled water is ideal, but tap water will suffice.
4. Place OxiProbe into 230mV ORP Standard
! Press the PROGRAM / SET button until the ORP
Calibration screen is displayed, as per the example
below. The “Current” reading may be 220 or
another value, depending on the condition of the
ORP Probe and current calibration setting.
! Reconnect the ORP Probe connector to the
OxiBOSS.
! Adjust the “Cal ORP” value to exactly 230. The
flashing “X”, indicates that the reading is not yet
stable. DO NOT calibrate until the “X” changes to a
“ü”.
3. Remove the ORP Probe from the Tee
! Switch the OxiBOSS power on. If it is connected to
a Bionizer, set the Bionizer to Manual:On mode to
switch on the power outlet.
! Press the PROGRAM / SET button twice to bypass
the start-up screen delay.
! If you are using a Bionizer to control the power to
your pool pump and OxiBOSS, unplug the pool
pump from the Bionizer Control Unit.
ORP PROBE CALIBRATION
he OxiBOSS will display a reminder to calibrate the
ORP Probe approximately every 6 months:
! (Connect the OxiBOSS to a power outlet socket,
but do not switch on yet.)
! (If you are powering the OxiBOSS and Pool Pump
from an external power source, please ensure that
the pool pump is switched off and unplugged
during the entire calibration procedure.)
The ORP Probe should be calibrated to ensure that
the readout is correct. Calibration is simply the
process of placing the ORP Probe into a solution of
known ORP, and adjusting the readout to that value.
2. Prepare 230mV ORP Calibration Standard
! Leave the OxiBOSS connected to the Bionizer.
1. Prepare for OxiBOSS for Calibration
! Unplug the ORP Probe from the OxiBOSS first.
(This will avoid the cable getting twisted when the
ORP Probe is unscrewed. Un s crew t he pro be
completely from the Tee.
! When the reading is stable and the “ü” is
displayed, press PROGRAM / SET to calibrate.
5. Calibration Procedure
! Rinse the probe well in clean water. De-ionized or
distilled water is ideal, but tap water will suffice.
4. Place OxiProbe into 230mV ORP Standard
! Press the PROGRAM / SET button until the ORP
Calibration screen is displayed, as per the example
below. The “Current” reading may be 220 or
another value, depending on the condition of the
ORP Probe and current calibration setting.
! Reconnect the ORP Probe connector to the
OxiBOSS.
! Adjust the “Cal ORP” value to exactly 230. The
flashing “X”, indicates that the reading is not yet
stable. DO NOT calibrate until the “X” changes to a
“ü”.
3. Remove the ORP Probe from the Tee
! Switch the OxiBOSS power on. If it is connected to
a Bionizer, set the Bionizer to Manual:On mode to
switch on the power outlet.
! Press the PROGRAM / SET button twice to bypass
the start-up screen delay.
! If you are using a Bionizer to control the power to
your pool pump and OxiBOSS, unplug the pool
pump from the Bionizer Control Unit.
ORP PROBE CALIBRATION
he OxiBOSS will display a reminder to calibrate the
ORP Probe approximately every 6 months:
! (Connect the OxiBOSS to a power outlet socket,
but do not switch on yet.)
! (If you are powering the OxiBOSS and Pool Pump
from an external power source, please ensure that
the pool pump is switched off and unplugged
during the entire calibration procedure.)
The ORP Probe should be calibrated to ensure that
the readout is correct. Calibration is simply the
process of placing the ORP Probe into a solution of
known ORP, and adjusting the readout to that value.
2. Prepare 230mV ORP Calibration Standard
! Leave the OxiBOSS connected to the Bionizer.
1. Prepare for OxiBOSS for Calibration
! Unplug the ORP Probe from the OxiBOSS first.
(This will avoid the cable getting twisted when the
ORP Probe is unscrewed. Un s crew t he pro be
completely from the Tee.
! When the reading is stable and the “ü” is
displayed, press PROGRAM / SET to calibrate.
5. Calibration Procedure
! Rinse the probe well in clean water. De-ionized or
distilled water is ideal, but tap water will suffice.
4. Place OxiProbe into 230mV ORP Standard
! Press the PROGRAM / SET button until the ORP
Calibration screen is displayed, as per the example
below. The “Current” reading may be 220 or
another value, depending on the condition of the
ORP Probe and current calibration setting.
! Reconnect the ORP Probe connector to the
OxiBOSS.
! Adjust the “Cal ORP” value to exactly 230. The
flashing “X”, indicates that the reading is not yet
stable. DO NOT calibrate until the “X” changes to a
“ü”.
3. Remove the ORP Probe from the Tee
! Switch the OxiBOSS power on. If it is connected to
a Bionizer, set the Bionizer to Manual:On mode to
switch on the power outlet.
! Press the PROGRAM / SET button twice to bypass
the start-up screen delay.
! If you are using a Bionizer to control the power to
your pool pump and OxiBOSS, unplug the pool
pump from the Bionizer Control Unit.
ORP PROBE CALIBRATION
he OxiBOSS will display a reminder to calibrate the
ORP Probe approximately every 6 months:
! (Connect the OxiBOSS to a power outlet socket,
but do not switch on yet.)
! (If you are powering the OxiBOSS and Pool Pump
from an external power source, please ensure that
the pool pump is switched off and unplugged
during the entire calibration procedure.)
The ORP Probe should be calibrated to ensure that
the readout is correct. Calibration is simply the
process of placing the ORP Probe into a solution of
known ORP, and adjusting the readout to that value.
2. Prepare 230mV ORP Calibration Standard
! Leave the OxiBOSS connected to the Bionizer.
1. Prepare for OxiBOSS for Calibration
100
50

! Press the AUTO/OFF button on the OxiBOSS until
it reads “Pump is now in AUTO mode”.
If the calibration failure persists, the ORP probe will
most likely need to be replaced. This can be
purchased online at bionizer.com.au.
! Disconnect the ORP Probe from the OxiBOSS (to
prevent twisting the cable) and re-fit into the Tee.
Re-connect the ORP Probe to OxiBOSS when
done.
! Switch your pool equipment back on.
! Re-connect the pool pump to the Bionizer power
outlet, or the power output to which it was
connected before starting this procedure.
Calibration Failure
Clean the ORP probe as per the troubleshooting
section on page 17, then retry calibration.
! If using a Bionizer, ensure that it is set to “Auto”
mode.
If ORP calibration fails, the OxiBOSS will display this
message:
6. Re-fit ORP Probe and Power Connections
MAINTENANCE
14

15
PUMP SQUEEZE TUBE LUBRICATION AND
REPLACEMENT
! Remove and inspect the squeeze tube. If it is
round and not flattened, and shows no visible
signs of wear, then it can simply be lubricated as
per the instructions in this section.
The OxiBOSS monitors the running time of the dosing
pump and will display a reminder to replace the
squeeze tube when this time exceeds 50 hours.
If the squeeze tube appears a little flattened and
has wear marks, then it is best to replace it. A new
squeeze tube can be ordered at bionizer.com.au.
! Switch off the OxiBOSS and ensure that the pool
pump is also off.
At this time, the peristaltic pump squeeze tube either
needs to be lubricated or replaced as per the
instructions below.
Removing the Squeeze Tube
Damage caused by chemical leaking from a
broken squeeze tube is not covered by the
warranty, so it is your responsibility to check and
replace it as necessary.
Remove the Squeeze
tube. The rotor may
come out, but can be
pushed back in place.
Pull o f f the p u m p
cover.
H o l d t h e t o p a n d
bottom of the pump
cover.
Lift up the pump cover
from the bottom side.
MAINTENANCE
Apply Silicon
grease here.
Pull out the pump rotor a little, and place the
squeeze tube between the pump body and the two
rollers.
Push the squeeze tube and rotor into the pump body.
Ensure that the inlet and outlet are seated in the
guides on each side of the pump body.
Refitting the Pump Cover
Re-fit the pump cover until it clips into place. There
should be no gaps anywhere between the pump
body and cover. Ensure that the squeeze tube is not
pinched on either side, after the cover is replaced.
Refitting the Squeeze Tube
Lubricating the Squeeze Tube
DO NOT over-lubricate, as this will attract dirt and
grit, which will cause premature wear of the squeeze
tube and rollers.
Apply a THIN SMEAR of Silicon grease to the
squeeze tube and pump rotor shaft as shown above.
(Skip this step when replacing squeeze tube, as these
are supplied pre-lubricated.)
WARNING : Use only silicon grease. Petroleum gel
and other oil-based lubricants may cause the
squeeze tube to break down, resulting in corrosive
chemical spillage.

Replacement Squeeze Tubes are supplied with new
Supply and Delivery tubes pre-fitted for your
convenience. It will be necessary to connect these to
the drum weight and to the one-way valve in the Tee
piece.
Fitting New Supply and Delivery Tubes Resetting the Squeeze Tube Reminder
! Press the PROGRAM / SET button until the Tube
Change screen is displayed, as per the example
below. The Tube Hours reading is the total number
of hours that the dosing pump has actually been
running.
! Press p button to select “Yes”. Press PROGRAM /
SET to reset the Tube Hours back to zero.
See page 10 for full instructions on making these
tube connections.
Once the Squeeze Tube has been either lubricated or
replaced, the last step is to reset the Squeeze Tube
reminder to zero.
Replacement Squeeze Tubes are supplied with new
Supply and Delivery tubes pre-fitted for your
convenience. It will be necessary to connect these to
the drum weight and to the one-way valve in the Tee
piece.
Fitting New Supply and Delivery Tubes Resetting the Squeeze Tube Reminder
! Press the PROGRAM / SET button until the Tube
Change screen is displayed, as per the example
below. The Tube Hours reading is the total number
of hours that the dosing pump has actually been
running.
! Press p button to select “Yes”. Press PROGRAM /
SET to reset the Tube Hours back to zero.
See page 10 for full instructions on making these
tube connections.
Once the Squeeze Tube has been either lubricated or
replaced, the last step is to reset the Squeeze Tube
reminder to zero.
FLOW DIRECTION
16 MAINTENANCE

CHECK THIS SECTION BEFORE CALLING OR SENDING YOUR OxiBOSS FOR SERVICE.
TROUBLESHOOTING
Symptom Possible Causes and Remedies
Unstable or inaccurate
ORP readings
ORP reading cannot be
calibrated
ORP reads around 0 mV
without changing
Unit does not power up
Alarm sounds due to ORP
reading being too high
Sanitiser Pump turns, but
sanitiser is not flowing
Unit is not adding
sanitiser when dosing
pump is operating
ORP Probe is dirty.
!Screw probe back into pipe and re-connect probe BNC connector.
!Disconnect probe BNC connector from the OxiBOSS, then unscrew probe
from the pipe.
!Rinse the ORP Probe thoroughly in clean water.
!Wipe metal tip of the probe with a soft tissue soaked with methylated spirits.
!Gently sand the metal tip of the ORP Probe with 800 to 1200 grit wet-and-dry
sandpaper, using a little water as a lubricant on the sandpaper.
!Follow the ORP Probe cleaning procedure above.
ORP Probe is dirty.
Short circuit in ORP Probe BNC Plug or Socket
!Disconnect ORP Probe plug from socket. Dry out with a hair dryer or allow to
air dry. Reconnect and re-check.
!If problem persists, return ORP Probe to have connector replaced, or replace
the ORP Probe. Please contact your distributor or Bionizer.
!Replace power adaptor. Please contact your distributor or Bionizer.
Faulty AC/DC power adaptor
!Faulty OxiBOSS main unit. Return to Bionizer for repair.
!Reduce the Dose Setting by one level and check in 2 days.
Dose Setting is set too high
!Replace squeeze tube and/or rollers as needed (see page 15).
Leaking or worn squeeze tube and/or rollers.
!Tighten Sanitiser Pump fittings further or replace with spare fittings (page 9).
!Check that all suction and output hoses are fitted correctly, with no leakage.
!Check that the OxiBOSS sanitiser drum is not empty.
!Check that squeeze tube inside pump is not leaking.
WARNING : IF ANY SANITISER HAS LEAKED OUT OF HOSES, WASH DOWN
THE AREA WITH FRESH WATER IMMEDIATELY. TAKE THE LID OFF THE
PUMP, REMOVE SQUEEZE TUBE AND WASH OUT WITH FRESH WATER IF
SQUEEZE TUBE WAS LEAKING.
!Replace ORP Probe. Please contact your distributor or Bionizer.
ORP Probe is faulty.
!Re-calibrate ORP Probe as per details on page 13.
ORP Probe needs to be re-calibrated.
!Clean ORP Probe as per procedure above.
Pool water is very pure and has low conductivity.
!Add one bag of pool salt per 50,000 Litres (13,000 Gallons) of pool water.
IF YOU NEED TO RETURN YOUR OxiBOSS FOR SERVICE OR REPAIRS AT ANY TIME, PLEASE USE THE MANUAL DOSE
FUNCTION (PAGE 12) TO DOSE 1 LITRE (1 QUART) OF FRESH WATER THROUGH THE SYSTEM BEFORE SENDING BACK.
THIS WILL STOP SANITISER LEAKING FROM THE PUMP DURING TRANSPORT (WHICH WOULD DAMAGE YOUR UNIT
OR THE PACKAGING, VOIDING WARRANTY), AND IS ALSO A SAFETY HAZARD FOR OUR TECHNICIANS.
17
TROUBLESHOOTING

TROUBLESHOOTING
OxiBOSS ALERTS AND GENERAL MESSAGES.
Alert Messages
Reminder Messages
General Messages
Hardware Failure Messages
Possible Causes and Remedies
Action Required
Explanation of Message
Explanation of Message
!The OxiBOSS will not attempt any further sanitiser addition until the alarm is
cleared. Press AUTO / OFF to clear the alarm after rectifying the issue.
!Please refer to page 12 for a complete list of possible causes and remedies
when the Alarm message is flashing.
ORP reading has been below Set Point for more than 4 hours of operation.
!Calibrate the ORP Probe as per page 1.
!This reminder is displayed 6 months after the last ORP Calibration.
!Lubricate or change the Squeeze Tube as per page 15.
!This reminder is displayed when the peristaltic dosing pump has exceeded
50 hours of operation since the last time the Squeeze Tube was lubricated or
changed.
!This message is displayed each time the OxiBOSS is switched on.
!If you need to contact Bionizer for technical support, we may ask you the
software version number (“v2.1” in this example).
!On this screen, the OxiBOSS displays the accumulated pump Squeeze Tube
time, and the number of days since the last ORP probe calibration.
!This message is displayed each time the OxiBOSS is switched on, after the
version screen above.
!The OxiBOSS has not been able to obtain data from the internal clock chip.
!Switch the OxiBOSS off for 5 seconds, then back on again. If the problem
persists, return the OxiBOSS to Bionizer for repair.
!Switch the OxiBOSS off for 5 seconds, then back on again. If the problem
persists, return the OxiBOSS to Bionizer for repair.
!This message can be error number 1 to 9.
18 TROUBLESHOOTING

] Phosphates are food for algae. Bionizer PFP quickly and
safely removes Phosphates from your pool.
] Use less costly and harmful chemicals to maintain your
pool.
] Super concentrated Phosphate Remover. 550g Jar is
equivalent to 3.5 Litres of liquid Phosphate Remover sold
in pool shops.
PhosphateFreePool
] Helps to maintain crystal clear pool water.
] Ideal for less diligent pool owners who wish to spend less
time maintaining their pools.
] Reduces Copper Ion loss from various environmental
stressors.
] Use a double dose of Ionlife+ when going away for
extended holidays to help ensure that you return to a
pristine pool.
] May extend the life of Copper Ionizer Anodes.
+
Ionlife CopperIonStablizer
] Returns cloudy pool water to sparkling clarity.
] Compatible with all Ionized, Salt Chlorinated and
Chlorine Pools.
] Does not affect overall pool balance.
] Highly concentrated, can be used without pre-mixing.
PoolWaterClarifier
Blue
Bionizer Pool Supplements
bionizer.com.au
To order any of the Bionizer Pool Supplements, visit the Bionizer online store at
19
BIONIZER POOL SUPPLEMENTS

SAVE TIME & MONEY
SAVE THE ENVIRONMENT
SAVE YOUR LONG-TERM HEALTH
INSTRUCTION BOOK
Phone................1-480-696-2387
Scottsdale, AZ 85254
Bionizer Ltd
Email ...........relax@bionizer.com
16427 N Scottsdale Rd
Suite 410
USA.
Toll Free.............1-855-8-IONIZE
Web ........................bionizer.com
QLD 4558
Australia
Web ...................bionizer.com.au
1/173 Bradman Ave
Phone...................1800 10 62 74
Fax....................+61 7 5443 6700
Email ........info@bionizer.com.au
Bioniser Pty Ltd
Maroochydore
International .....+61 7 5479 6100
C h l o r i n e - F r e e F r e s h w a t e r P o o l s
Table of contents
Other Bionizer Lighting Equipment manuals