Betriebsanleitung Temperatur-Kalibrierset
Operating instruction temperature calibrating set
Mode d’emploi jeu de calibrage de la température
KERN DBS-A01
Version 1.1
02/2013
DBS_A01-BA-def-1311
BIS S.r.l.
Via Trieste, 31
20080 Bubbiano MI – Italia
Tel.: +39 02 90834207
Fax: +39 02 90870542
e-mail: [email protected]
P.IVA e C.F.: 03774900967
www.BilanceOnLine.it

2 DBS-A01-BA-def-1311
D
KERN DBS-A01
Version 1.102/2013
Betriebsanleitung
Temperatur-Kalibrierset
1 Technische Daten
Einheiten Celsius [°C]
Fahrenheit [°F]
Messbereich -50°C – +300°C
-58°F - +574°F
Auflösung Anzeige 0,1 °
Auflösung Alarm 1°
Messgenauigkeit Bereich
-50°C - +150°C
-58°F – +302°F
± 1°C, sonst ± 4°C
± 2°F, sonst ± 8°F
Batterie 1.5 V AAA
Umgebungstemperatur Umgebungstemperatur
-10°C - +50°C / -14°F - +122°FC
2 Geräteübersicht
Messfühler
Anzeigegerät

DBS-A01-BA-def-1311 3
2.1 Anzeigenübersicht
Anzeige Beschreibung
MAX Kennzeichnet den höchsten Messwert einer Messung
MIN Kennzeichnet den niedrigsten Messwert einer Messung
HOLD Kennzeichnet einen festgehaltenen Anzeigenwert
HI OFF Alarm für oberen Temperatur-Grenzwert ausgeschaltet
LO OFF Alarm für unteren Temperatur-Grenzwert ausgeschaltet
HI XXX Oberer Temperatur-Grenzwert
Lo XXX Unterer Temperatur-Grenzwert
2.2 Tastatur- und Funktionsübersicht
Beschreibung
ON
OFF Gerät ein- und ausschalten
Während der Anzeige eines MAX/MIN oder HOLD-Wertes kann das
Gerät nicht ausgeschaltet werden.
UP Wert verringern
HOLD Nach Tastendruck wird der aktuelle Anzeigewert festgehalten, der Indikator
„HOLD“ blinkt. Er wird solange angezeigt, bis er durch erneuten Tastendruck
gelöscht wird.

4 DBS-A01-BA-def-1311
DOWN Wert erhöhen
MAX/MIN Anzeige Max. / Min.-Temperatur
Nach Tastendruck wird der höchste Messwert angezeigt, der Indikator „MAX“
blinkt.
Nach erneutem Tastendruck wird der niedrigste Messwert angezeigt, der Indi-
kator „MIN“ blinkt.
Nach erneutem Tastendruck kehrt das Gerät zurück in den Messmodus.
Zum Löschen der MAX/MIN-Werte Gerät mit ON/OFF aus- und wieder ein-
schalten.
SET Alarm für Temperatur-Grenzwerte einstellen
[SET] drücken „HI XXX“ blinkt. Mit [UP] und [DOWN] oberen Grenzwert ein-
stellen. Mit jedem Tastendruck erhöht/verringert sich der Wert. Durch Dauer-
druck auf die Taste wird der Wert automatisch erhöht/verringert. Mit [ON/OFF]
kann der Grenzwert ausgeschaltet werden, „HI OFF“ wird angezeigt. Oberen
Grenzwert mit [SET] bestätigen, „Lo XXX“ blinkt. Unteren Grenzwert auf die-
selbe Art wie beim oberen Grenzwert beschrieben einstellen.
Ist die Temperatur außerhalb der Grenzwerte wird der Alarm ausgelöst. Mit
jeder beliebigen Taste wird der Alarm ausgeschaltet. Der Indikator [HI] bzw.
[LO] blinkt jedoch weiter solange sich die Temperatur außerhalb der Grenz-
werte befindet.
°C / °F Umschaltung
[SET] gedrückt halten die Anzeige wechselt von °C ⇔°F.
3 Batterie wechseln
Wenn die Batteriespannung nicht mehr ausreicht, verdunkelt sich die Anzeige.
+Gerät ausschalten.
+Deckel des Batteriefachs durch Lösen der 4 Schrauben öffnen.
+Batterie tauschen.
Auf die richtige Polung achten.
+Den Deckel wieder verschließen.

DBS-A01-BA-def-1311 5
4 Temperaturwert des Feuchtebestimmers KERN DBS justieren
Wir empfehlen den Temperaturwert des Gerätes gelegentlich mit dem optionalen
Temperatur-Kalibrierset DBS-A01 zu prüfen. Die Temperatur wird an zwei Punkten
(100°C u. 180°C) gemessen und kann an diesen korrigiert werden.
Zuvor sollte sich das Gerät nach der letzten Heizphase auf Raumtemperatur abgekühlt
haben.
Vorbereitung:
Messfühler lt. Abb. fixieren.
Am Feuchtebestimmer Probenschale
und Schalenträger entfernen.
Temperatur-Kalibrierset lt. Abb.
installieren.
Heizhaube schließen.
Ggf. wird die Fehlermeldung „ol“ ange-
zeigt, kann aber in diesem Fall ignoriert
werden.
Temperatur-Kalibrierset mit ON/OFF-
Taste einschalten.

6 DBS-A01-BA-def-1311
Menü aufrufen:
Menü mit der Menu-Taste aufrufen, der erste
Menüpunkt „PRoGRM“ wird angezeigt.
Mit den Navigationstasten Menüpunkt
„CAL“ anwählen.
Mit ENTER-Taste bestätigen, die aktuelle Ein-
stellung wird angezeigt. Falls nötig mit den Na-
vigationstasten „TEMP“ anwählen.
Mit ENTER-Taste bestätigen, der erste zu prü-
fende Temperaturpunkt wird angezeigt.
START-Taste drücken, die erste Aufheizphase
wird gestartet.
•
Während der Aufheizphase die Heizhaube nicht länger als 1 Min. ge-
öffnet lassen, sonst erscheint die Fehlermeldung „ERR.100“.In diesem
Fall ESC-Taste drücken und Menüpunkt „TEMP“ erneut starten.
•
Vorsicht einzelne Teile wie z.B. Lüftungsgitter und Sichtfenster können
sich während des Betriebs stark erwärmen.
•
Die Korrektur bei blinkender Temperaturanzeige muss binnen 15 Min.
erfolgen, sonst wird die Temperaturjustierung abgebrochen (AboRT
wird angezeigt).
Nach 15 Min ist die Temperaturkalibrierung des
1. Punktes beendet. Bei blinkender Tempera-
turanzeige den angezeigten Temperaturwert am
Thermometer mit dem des Feuchtebestimmers
vergleichen. Stimmen die beiden Werten nicht
überein, mit den Navigationstasten korri-
gieren und mit ENTER -Taste bestätigen.
Numerische Eingabe siehe Betriebsanleitung
KERN DBS, Kap. 7.1.1.

DBS-A01-BA-def-1311 7
Der zweite zu prüfende Temperaturpunkt wird
angezeigt.
Mit ENTER-Taste bestätigen, die zweite Auf-
heizphase wird gestartet.
Nach 15 Min ist die Temperaturkalibrierung des
2. Punktes beendet. Bei blinkender Tempera-
turanzeige den angezeigten Temperaturwert am
Thermometer mit dem des Feuchtebestimmers
vergleichen. Stimmen die beiden Werten nicht
überein, mit den Navigationstasten korri-
gieren und mit ENTER -Taste bestätigen.
Numerische Eingabe siehe Betriebsanleitung
KERN DBS, Kap. 7.1.1.
Bei Anzeige „END“ ist die Justierung abge-
schlossen. Das Gerät kehrt automatisch zurück
ins Menü.
Bei Anschluss eines optionalen Druckers und
aktivierter GLP-Funktion wird die Ausgabe des
Justierprotokolls gestartet
(siehe Betriebsanleitung KERN DBS Kap. 6.3).
Mit ESC-Taste zurück in den Feuchtebestim-
mungsmodus.
•
Bei Anschluss eines optionalen Druckers und aktivierter GLP-Funktion
(siehe Betriebsanleitung KERN DBS, Kap. 6.3) erfolgt die Ausgabe
eines Justierprotokolls.

8 DBS-A01-BA-def-1311
GB
KERN DBS-A01
Version 1.1 02/2013
Operating Instructions
Temperature calibration set
1 Technical data
Units Celsius [°C]
Fahrenheit [°F]
Measuring range -50°C – +300°C
-58°F - +574°F
Resolution display 0.1 °
Resolution alarm 1°
Accuracy of measure-
ment Range
-50°C - +150°C
-58°F – +302°F
± 1°C, otherwise ± 4°C
± 2°F, otherwise ± 8°F
Battery 1.5 V AAA
Ambient temperature Ambient temperature
-10°C - +50°C / -14°F - +122°FC
2 Appliance overview
Measuring sensor
Display Unit

DBS-A01-BA-def-1311 9
2.1 Overview of display
Display Description
MAX Means the highest measured value of a measurement
MIN Means the lowest measured value of a measurement
HOLD Means a recorded display value
HI OFF Alarm for the upper temperature limit value switched off
LO OFF Alarm for the lower temperature limit value switched off
HI XXX Upper temperature limit value
Lo XXX Lower temperature limit value
2.2 Keyboard and function overview
Description
ON
OFF Switch appliance on/off
During display of a MAX/MIN or a HOLD value the unit cannot be
switched off.
UP Decrease value
HOLD Press the button to record the current displayed value, the indicator „HOLD“
flashes. It remains displayed until it is deleted by pressing button again.

10 DBS-A01-BA-def-1311
DOWN Increase value
MAX/MIN Display Max. / Min. temperature
Press button, the highest measured value will be displayed, the indicator
„MAX“ flashes.
Press button again, the lowest measured value will be displayed, the indicator
„MIN“ flashes.
Press button anew, the appliance returns to measuring mode.
To delete the MAX/MIN values, switch appliance off and on using the ON/OFF
feature.
SET Setting alarm for temperature limit values
Press [SET], „HI XXX“ flashes. St the upper limit value using [UP] and
[DOWN]. Every time the button is pressed, the value is increased/reduced. If
the button is permanently, the value is automatically increased/reduced. Use
[ON/OFF] to switch off the limit value, „HI OFF“ is displayed. Confirm upper
limit value using [SET], „Lo XXX“ flashes. Adjust the lower limit value in the
some way as described at the upper limit value.
If the temperature is beyond the limit values, the alarm will be triggered. The
alarm can be switched off with any button. The indicator [HI] or [LO], however,
remains as long as the temperature is beyond the limit values.
Commutation °C / °F
Keep set [SET] pressed, the display changes from °C ⇔°F.
3 Change battery
If the battery tension is not more sufficient, the display darkens.
+Turn off unit.
+Open the lid of the battery compartment loosening the 4 screws.
+Replace battery.
Ensure correct polarisation.
+Close again the lid.

DBS-A01-BA-def-1311 11
4 Adjust temperature value of the moisture analyzer KERN DBS
We recommend sometimes to check the temperature value of the appliance using the
optional temperature calibrating set DBS-A01. The temperature is measured at two
points (100°C and 180°C) and can be adjusted at these two points.
The equipment should cool down first to room temperature after the last heating pe-
riod.
Preparation:
Fix the temperature sensor on the tem-
perature calibration set in accordance
with the illustration.
Remove sample dish and dish holder
from the moisture analyser.
Install the temperature-calibration set
acc. to fig.
Close heated cover.
If applicable, the error message “ol” will
be displayed, but can be ignored in this
case.
On the temperature calibration set
switch-on the digital thermometer by the
ON/OFF button.

12 DBS-A01-BA-def-1311
Call up menu:
Press the Menu button to access the menu and
the first menu item “PRoGRM” will be dis-
played.
Use navigation buttons to select menu
item “CAL”.
Acknowledge using ENTER key, the current
setting is displayed. If necessary, use naviga-
tion buttons to select “TEMP”.
Press ENTER to confirm and the first tempera-
ture point to be tested will be displayed.
Press START and the first heating-up period
will start.
•Do not leave the heated cover open for more than 1 minute during
the heating-up period, otherwise the error message “ERR.100” will
appear. In this case press ESC and restart menu item “TEMP”.
•Attention: some parts e.g. ventilation grilles and inspection windows
may become very hot during operation.
•The adjustment must be made within 15 minutes when the tempera-
ture display flashes, otherwise the temperature adjustment will be
aborted (AboRT will be displayed).
Temperature calibration for the first point takes
15 min. Compare the displayed temperature
value on the thermometer with that of the mois-
ture analyser whilst the temperature display is
flashing. If the two values do not match, correct
them using the navigation buttons and
confirm with the ENTER button. Numeric input
see operating instructions KERN DBS, chap.
7.1.1.

DBS-A01-BA-def-1311 13
The second temperature point to be tested will
be displayed.
Confirm with the ENTER button, the second
heat-up phase is started.
Temperature calibration for the second point
takes 15 min. Compare the displayed tempera-
ture value on the thermometer with that of the
moisture analyser whilst the temperature dis-
play is flashing. If the two values do not match,
correct them using the navigation buttons
and confirm with the ENTER button. Numeric
input see operating instructions KERN DBS,
chap. 7.1.1.
The adjustment is completed when “END” is
displayed. The appliance returns to the menu
automatically.
The adjustment protocol output is started upon
connection to an optional printer and activated
GLP function. (see operating instructions KERN
DBS chap. 6.3).
Press ESC to return to Moisture analysis mode.
•The adjustment protocol output is started upon connection to an
optional printer and activated GLP function. (see operating instruc-
tions KERN DBS chap. 6.3).

14 DBS-A01-BA-def-1311
F
KERN DBS-A01
Version 1.102/2013
Mode d’emploi
Kit de calibrage de température
1 Caractéristiques techniques
Unités Celsius [°C]
Fahrenheit [°F]
Plage de mesure: -50°C – +300°C
-58°F - +574°F
Résolution affichage 0,1 °
Résolution alarme 1°
Exactitude de mesure Plage
-50°C - +150°C
-58°F – +302°F
± 1°C, sinon ± 4°C
± 2°F, sinon ± 8°F
Pile 1.5 V AAA
Température ambiante Température ambiante
-10°C - +50°C / -14°F - +122°FC
2 Vue d’ensemble de l’appareil
Capteur
Afficheur

DBS-A01-BA-def-1311 15
2.1 Vue d’ensemble des affichages
Afficheur Description
MAX Signifie la valeur mesurée plus haute d’une mesure
MIN Signifie la valeur mesurée plus basse d’une mesure
HOLD Signifie une valeur d’affichage enregistrée
HI OFF Alarme pour valeur limite supérieure de température éteinte
LO OFF Alarme pour valeur limite inférieure de température éteinte
HI XXX Valeur-limite supérieure pour la température
Lo XXX Valeur-limite inférieure pour la température
2.2 Vue d’ensemble du clavier et des fonctions
Description
ON
OFF Mise en marche et arrêt de l’appareil
Pendant l’affichage d’un MAX/MIN ou de la valeur HOLD, l’appareil
ne peut pas être mis hors fonctionnement.
UP Réduire la valeur
HOLD Enfoncer la touche pour enregistrer la valeur affichée actuelle, l’indicateur
„HOLD“ clignote. Elle continue d’être affichée jusqu’à ce que soit effacée en
appuyant autre fois sur la touche.

16 DBS-A01-BA-def-1311
DOWN Augmenter la valeur
MAX/MIN Affichage Température Max. / Min.
Enfoncer la touche, la valeur mesurée plus haute est affichée, l’indicateur
„MAX“ clignote.
Enfoncer la touche autre fois, la valeur mesurée plus basse est affichée,
l’indicateur „MIN“ clignote.
Enfoncer la touche une troisième fois, l’appareil retourne en mode de pesée.
Pour effacer les valeurs MAX/MIN, éteindre et rallumer l’appareil sur ON/OFF.
SET Régler l’alarme pour les valeurs limite de la température
Appuyer sur [SET] „HI XXX“ clignote. Régler la valeur limite supérieure sur
[UP] et [DOWN]. Lorsqu’on appuie sur une touche, la valeur est augmen-
tée/réduite. Enfoncer la touche de manière permanente, la valeur est automa-
tiquement augmentée/réduite. Avec [ON/OFF], la valeur limite peut être
éteinte, „HI OFF“ est affiché. Confirmer la valeur limite supérieure avec [SET],
„Lo XXX“ clignotera. Régler la valeur limite inférieure de la même manière
comme décrit avec la valeur limite supérieure.
Si la température est au dehors de valeurs limite, l’alarme est déclenchée.
Appuyer sur une touche quelconque pour éteindre l’alarme. Mais l’indicateur
[HI] ou [LO] continue à clignoter jusqu’à c que la température soit au dehors
des valeurs limite.
Commutation °C / °F
Tenir [SET] enfoncé, l’affichage change de °C ⇔°F.
3 Changement de la pile
Si la tension de la pile ne suffit plus, l’affichage devient sombre.
+Eteindre l’appareil.
+Ouvrir le couvercle du compartiment à piles en desserrant les quatre vis.
+Remplacer la batterie.
Veillez à la polarité correcte.
+Refermer le couvercle.

DBS-A01-BA-def-1311 17
4 Ajuster la valeur de température de l’analyseur d’humidité KERN
DBS
Nous recommandons de vérifier occasionnellement la valeur de température de
l’appareil à l’aide du jeu de calibrage de la température en option DBS-A01. La tempé-
rature est mesurée à deux points (100°C et 180°C) et peut être corrigée au niveau de
ceux-ci.
Au préalable l’appareil devrait s’être refroidi après la dernière phase de chauffage à la
température ambiante.
Préparation:
Fixer le capteur de température confor-
mément à la figure au kit de calibrage
de la température.
Retirer la cuvette porte-échantillon et le
porte-coupelles de l’analyseur
d’humidité.
Installer le kit de calibrage de la tempé-
rature conformément à la figure.
Fermer la hotte chauffante.
Le cas échéant est affiché le message
d’erreur „ol“, mais il peut être ignoré
dans ce cas.
Mettre en marche le kit de calibrage du
thermomètre digital sur la touche
ON/OFF.

18 DBS-A01-BA-def-1311
Appel du menu :
Appel du menu sur la touche de menu, le pre-
mier point du menu „PRoGRM“ est affiché.
Sélectionner sur les touches de navigation
le point du menu „CAL“.
Valider sur la touche ENTER le point de menu
appelé, le réglage actuel est affiché. Le cas
échéant, sélectionner sur les touches de navi-
gation „TEMP“.
Valider sur la touche ENTER, le premier point
de température à contrôler est affiché.
Appeler la touche START, la première phase
de réchauffement est lancée.
•
En cours de phase de réchauffement ne pas laisser ouverte la hotte
chauffante plus d’1 min, sinon apparaît le message d’erreur
„ERR.100“.Dans ce cas appeler la touche ESC et redémarrer le point
du menu „TEMP“.
•
Attention certains éléments comme p. ex. grilles d’aération et regards
peuvent se réchauffer fortement en cours d’exploitation.
•
La correction doit s’opérer pendant l’affichage clignotant en l’espace de
15 min, sinon l’ajustage de la température est interrompu (AboRT est
affiché).
Après 15 min. le calibrage de la température
du premier point est terminé. En cours
d’affichage clignotant de la température com-
parer la valeur de température affichée sur le
thermomètre à celle de l’analyseur d’humidité.
Si ces deux valeurs ne coïncident pas corriger
sur les touches de navigation et valider
sur la touche ENTER. Saisie numérique voir
notice d’utilisation KERN DBS, chap. 7.1.1.

DBS-A01-BA-def-1311 19
Le deuxième point de température à contrôler
est affiché.
Valider sur la touche ENTER, la seconde phase
de réchauffement démarre.
Après 15 min le calibrage de la température du
deuxième point est terminé. En cours
d’affichage clignotant de la température com-
parer la valeur de température affichée sur le
thermomètre à celle de l’analyseur d’humidité.
Si ces deux valeurs ne coïncident pas corriger
sur les touches de navigation et valider
sur la touche ENTER. Saisie numérique voir au
chap. 7.1.1. Saisie numérique voir notice
d’utilisation KERN DBS, chap. 7.1.1.
Lorsque „END“ est affiché, l’ajustage est termi-
né. L’appareil retourne au menu automatique-
ment.
En cas de raccordement d’une imprimante en
option et après activation de la fonction GLP
(voir notice d’utilisation KERN DBS chap. 6.3)
démarre l’édition du protocole d’ajustage.
Retour au mode de détermination de l’humidité
sur la touche ESC.
•
En cas de raccordement d’une imprimante en option et fonction GLP
activée (voir notice d’utilisation KERN DBS au chap. 6.3) intervient
l’édition d’un protocole d’ajustage.
BIS S.r.l.
Via Trieste, 31
20080 Bubbiano MI – Italia
Tel.: +39 02 90834207
Fax: +39 02 90870542
e-mail: [email protected]
P.IVA e C.F.: 03774900967
www.BilanceOnLine.it
Table of contents
Languages:
Other BIS Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Kistler
Kistler 9281E Series instruction manual

BEKA
BEKA BA304C manual

EM-Technik
EM-Technik 4L Assembly and operation instructions

Fuji Electric
Fuji Electric ZKM Service manual

PCB Piezotronics
PCB Piezotronics ICP 034G05 Installation and operating manual

Hanna Instruments
Hanna Instruments HI96708C instruction manual