BOXIANGLAI ELECTRONIC SCI-TECH BXL-V10 User manual

郑州博祥来电子科技有限公司
ZHENGZHOU BOXIANGLAI ELECTRONIC SCI-TECH CO. LTD.
ADD: No.36 Dongfeng Road, Jinshui District, Zhengzhou City, China. Zip:450003
Tel: +86-159 3870 1512 Email: kevin@boxianglai.com Website: www.bxlvet.com
User Manual
Full Digital Veterinary B-Ultrasound Device
(Model: BXL-V10)

郑州博祥来电子科技有限公司
ZHENGZHOU BOXIANGLAI ELECTRONIC SCI-TECH CO. LTD.
—2—
Important Statement
1. After purchasing this product, the customer is fully responsible for the maintenance and management
of this product.
2. Even during the quality assurance period, the quality assurance does not include the following
(1) Damage or loss caused by wrong or rough use.
(2) Damage or loss caused by force majeure such as fire, earthquake, flood or lightning.
(3) Failure to meet the operating conditions specified by this system, such as damage or loss caused
by insufficient power supply, incorrect installation or environmental conditions that do not meet the
requirements.
(4) There is no damage or loss caused by using the system in the region where it was originally
purchased.
(5) Damage or loss of the system not purchased from our company or its authorized distributors or
agents.
3. Only qualified medical personnel with professional qualification certificates can use this system.
4. It is forbidden to modify the software or hardware of this product without authorization.
5. Under any circumstance, the company will not be responsible for any problems, damages or losses
caused by the reinstallation, modification or maintenance of the system by non-designated personnel
of the company.
6. This system aims to provide doctors with data needed for clinical diagnosis.The doctor is responsible
for the diagnosis process. The company is not responsible for the diagnosis process.
7. Be sure to back up important data to an external storage medium, such as a notebook.
8. The company is not responsible for any loss of data stored in the system due to operator errors or
abnormal conditions.
9. This instruction manual contains warnings about foreseeable potential risks. At all times, we must
maintain a high level of vigilance against unexplained risks. The company is not responsible for any
damage or loss caused by negligence or ignorance of the preventive measures specified in this
manual.
10. Please keep this manual with the machine to ensure that management and operating personnel can
refer to this manual at any time. Once the administrator of this system changes, this manual must be
handed over.
11. Please dispose of the product in accordance with local laws and regulations. The internal maintenance
and maintenance of the product must be completed by a professionally trained engineer or contact our
company.

郑州博祥来电子科技有限公司
ZHENGZHOU BOXIANGLAI ELECTRONIC SCI-TECH CO. LTD.
—3—
Safety Tips
1. Definition Of Warning Symbols
In the instruction manual, warning symbols are used to indicate safety and other important matters, and
their specific meanings are as follows. Before reading this instruction manual, please carefully understand
and remember the meaning of these words.
2. Safety Symbol
Symbols
Definition
Danger
Indicates that there will be an imminent dangerous situation that, if not avoided, will
cause death or serious injury to personnel.
Warning
It means that there will be a potentially dangerous situation, if it is not avoided, it may
cause personal injury.
Attention
Indicates that there will be potential dangers or unexpected use. If not avoided, it may
cause minor injuries to personnel or property damage or affect use.
Symbols
Meaning
Symbols
Meaning
Type B Application Part
Upward
DC (Direct Current)
Keep Dry
Power Switch
Fragile
Power indicator
Stacking Layers Limit
Chg
Battery Charging Indicator
Temp. Limit
(storage & transportation)
Adapter Connection Indicator
Humidity Limit
(storage & transportation)
Restricted Use Sign
Of Hazardous Substances
(electrical & electronic products)
Atmospheric Pressure Limit
(storage & transportation)
Recyclable

郑州博祥来电子科技有限公司
ZHENGZHOU BOXIANGLAI ELECTRONIC SCI-TECH CO. LTD.
— 4 —
Contents
Chapter 1 Overview ................................................................................................................................... 6
1.1 Introduction ..................................................................................................................................... 6
1.2 Scope of application ....................................................................................................................... 6
1.3 Appearance .................................................................................................................................... 6
1.4 Technical specification .................................................................................................................... 7
1.5 Working principle ............................................................................................................................ 8
1.6 The structure of the device .............................................................................................................. 8
1.7 Electromagnetic compatibility statement ........................................................................................ 9
Chapter 2 Installation ...................................................................................................................... 9
2.1 Use-environment requirement ........................................................................................................ 9
2.2 Battery ............................................................................................................................................ 9
2.3 Probe install and remove .............................................................................................................. 10
2.4 Power supply ................................................................................................................................ 11
2.4.1 Time and date settings ........................................................................................................... 11
Chapter 3 Operation Keyboard ............................................................................................................... 12
3.1 Screen Display .............................................................................................................................. 12
3.2 Keyboard function ......................................................................................................................... 12
3.3 Instructions ................................................................................................................................... 13
Chapter 4 Function Operation ................................................................................................................ 15
4.1 Power-on ...................................................................................................................................... 15
4.2 Diagnosis ...................................................................................................................................... 16
4.3 Imaging parameter adjustment ..................................................................................................... 16
4.3.1 Frequency adjustment ............................................................................................................ 16
4.3.2 Gain adjustment ..................................................................................................................... 17
4.3.3 Brightness and contrast adjustment ....................................................................................... 17
4.3.4 Image focus adjustment ......................................................................................................... 17
4.3.5 Frame related adjustment ...................................................................................................... 17
4.3.6 Gamma correction adjustment ............................................................................................... 17
4.3.7 Probe settings ........................................................................................................................ 17
4.3.8 Default settings ...................................................................................................................... 18
4.4 Comment ...................................................................................................................................... 18
4.4.1 Image Comment ..................................................................................................................... 18
4.4.2 Date and time adjustment ...................................................................................................... 18
4.4.3 Posture markings ................................................................................................................... 19
4.4.4 Language switch .................................................................................................................... 19
4.5 Image Processing ......................................................................................................................... 19
4.5.1 Cine Loop function ................................................................................................................. 20

郑州博祥来电子科技有限公司
ZHENGZHOU BOXIANGLAI ELECTRONIC SCI-TECH CO. LTD.
— 5 —
4.5.2 Store image ............................................................................................................................ 20
4.5.3 Browse images ....................................................................................................................... 21
4.5.4 Image smoothing adjustment ................................................................................................. 21
4.5.5 Image up/down flip ................................................................................................................. 21
4.5.6 Image left/right flip .................................................................................................................. 22
4.5.7 Pseudo-color function ........................................................................................................... 22
4.5.8 Cleanup of images ................................................................................................................. 22
4.6 Measurement Operation ............................................................................................................... 23
4.6.1 General measurement............................................................................................................ 23
4.6.2 Volume measurement............................................................................................................. 25
4.6.3 Histogram statistics function................................................................................................... 26
4.6.4 Heart rate measurement......................................................................................................... 26
4.7 Obstetric calculation ..................................................................................................................... 27
4.8 Image upload for computer ........................................................................................................... 33
4.9 Power-off ...................................................................................................................................... 33
Chapter 5 Storage and Transportation ................................................................................................... 34
5.1 Storage and transportation environment requirement .................................................................. 34
5.2 Transportation ............................................................................................................................... 34
5.3 Storage ......................................................................................................................................... 34
Chapter 6 Inspection and Maintenance ................................................................................................. 34
6.1 Life span ....................................................................................................................................... 34
6.2 Host maintenance ......................................................................................................................... 35
6.3 Probe maintenance ....................................................................................................................... 35
6.4 Battery........................................................................................................................................... 37
Chapter 7 Basic Fault Inspection and Elimination ................................................................................ 38
7.1 Inspection ..................................................................................................................................... 38
7.2 Fault Elimination............................................................................................................................ 38

郑州博祥来电子科技有限公司
ZHENGZHOU BOXIANGLAI ELECTRONIC SCI-TECH CO. LTD.
— 6 —
Chapter 1 Overview
1.1 Introduction
(1) The instrument uses microcomputer control and digital scan converter (DSC), adopts digital beam
synthesis (DBF), real-time dynamic aperture imaging (RDA), and real-time dynamic aperture
imaging (RDA).Time dynamic sound velocity tracking (DRA), real-time point-by-point dynamic
receiving and focusing (DRF), numerical control dynamic frequency scanning (DFS), frame
correlation and other technologies, the whole machine image is clear, stable, and high resolution..
(2) It can realize the image real-time, freezing, storage, recall, up and down flip, left and right flip and
large-capacity cine loop; with multi-level scanning depth, dynamic Range, frame correlation
coefficient adjustment and focus number, focus position adjustment and other functions. 16 kinds
of animal body marks..
(3) Date, clock display, notes, distance, perimeter, area, volume measurement.
(4) The high-speed USB2.0 interface can upload real-time ultrasound images to the computer.
(5) 11.1V lithium ion battery power supply mode, special power saving mode, make battery power
supply more durable.
(6) Injection molding shell, ergonomic hand-held structure, comfortable hand feeling, ease of
operation fatigue, easy to carry out the diagnosis of large and small animals
(7) The equipment consists of two parts: the host and the probe. Standard configuration
CXA50R/3.5MHz convex array, optional 6.5MHz animal rectal probe, CXA20R/5.0MHz Micro
convex probe, LNA40/7.5MHz high frequency linear array probe.
1.2 Scope of application
Suitable for pregnancy diagnosis of animals such as swine, horse, bovine, sheep, cats, dogs, etc.
1.3 Scope of application
Figure 1-1. Front View
Function button
Power on/off
(Freeze button)

郑州博祥来电子科技有限公司
ZHENGZHOU BOXIANGLAI ELECTRONIC SCI-TECH CO. LTD.
— 7 —
Figure 1-2. Backplane Interface View
Figure 1-3. Bottom view
1.4 Scope of application
Model Name
BXL-V10
Probe
LNA40/6.5MHZ
Rectal Linear
CXA50R/3.5MHZ
Convex
CXA20R/5.0MHZ
Micro-Convex
Depth Range (mm)
≥80
≥140
≥90
Resolution (mm)
Sideways
≤1(Depth≤60)
≤3(Depth≤80)
≤5(80<Depth≤130)
≤3(Depth≤60)
Axial
≤1(Depth≤80)
≤1(Depth≤80)
≤1(Depth≤60)
Blind Area(mm)
≤3
≤6
≤5
Geometric position
accuracy(%)
Horizontal
≤5
≤7.5
≤7.5
Vertical
≤5
≤5
≤5
Display
5.6” HD LCD Screen (640*480 Pixel)
Couplant slot
Battery compartment

郑州博祥来电子科技有限公司
ZHENGZHOU BOXIANGLAI ELECTRONIC SCI-TECH CO. LTD.
— 8 —
1.5 Working principle
The BXL-V10 veterinary B-type ultrasonic diagnostic instrument is based on different animal tissues
and organs with different densities and different ultrasonic propagation speeds, that is, different
The characteristic of the acoustic impedance (the product of the medium density and the speed of
sound), the piezoelectric chip (transducer) is excited by a certain regular electric pulse, so that it
produces a certain frequency of ultrasonic waves, and the ultrasonic waves (sound energy) are
injected into In animals, reflected echoes are generated through the interfaces of different organs in
the body; the reflected echoes of different sizes are received by the original ultrasonic transducer,
which in turn converts the received echo (sound energy) into electrical pulses. This kind of electric
pulse is amplified, detected and digitally scanned and transformed to form a standard video signal,
which displays the cross-sectional image of the organ on the monitor screen.
1.6 The structure of the device
Standard Kit
(1) Host device (Include build-in Li-On battery)
(2) 3.5MHz Convex Prob (CXA50R)
(3) Power adapter (Include power cable)
(4) User manual
(5) Couplant (According to shipping condition)
Optional Accessories
(1) 6.5MHz Rectal Linear Probe (LNA64)
(2) 5.0MHz Micro-convex Probe (CXA20R)
(3) 7.5MHz Linear Probe (LNA40)
Warning: Accessories of the specified models listed above should be selected for use. If the
accessories of non-specified models are arbitrarily selected, the company will not bear the risk
for the unexpected degradation of safety & electromagnetic compatibility performance caused.
Imaging Modes
B / 2B / 4B / BM / M
Image Gray Scale
256
Host Storage
64 frames
Cine Loop
≥400 frames
Scan Depth Adjustment
70mm-240mm
Image Flip Function
up/down left/right
Body Remark
16 kinds
Image Processing
Pseudo-color / Grayscale correction adjustment
Image smoothing adjustment / Histogram statistics
Frequency Adjustment
3 gears
Frame Related Adjustment
Support
Measurement
Distance / Circumference / Area / Volume
Gestational age / Expected birth date
Characters and comments
Date / Time / Full-screen character editing
Output Port
USB2.0
Continuous Working Time
≥3.0 Hours
Battery Capacity
3000mAh

郑州博祥来电子科技有限公司
ZHENGZHOU BOXIANGLAI ELECTRONIC SCI-TECH CO. LTD.
— 9 —
1.7 Electromagnetic compatibility statement
Warning: When the instrument is working in a strong electromagnetic environment,
interference signals may appear in the image, which will affect the diagnosis. At this time, stop
using it to prevent misdiagnosis, and use it after removing the electromagnetic interference.
Warning: Unexpected electromagnetic compatibility problems may occur when the instrument
and other equipment overlap or work at a close distance; if it must approach , please carefully
verify whether any equipment is affected by unexpected electromagnetic coupling through
observation.
Warning: When replacing parts that do not meet specifications or connecting with other
equipment, unexpected electromagnetic compatibility problems may occur. Care should be
taken to verify whether there are unexpected electromagnetic coupling effects.
Chapter 2 Installation
2.1 Use-environment requirement
(1) Ambient temperature range: +5℃~+40℃
(2) Relative humidity range: 30%~75%
(3) Atmospheric pressure range: 70KPa~106KPa
(4) Charging adapter power supply: a.c. 100~240V, 50Hz/60Hz
(5) Keep the equipment away from strong electric field, strong magnetic field equipment and high
voltage equipment; avoid direct sunlight to the display screen. The room should be protected from
light to facilitate image observation, and keep it ventilated, moisture-proof and dust-proof..
Warning: After unpacking the equipment, check carefully according to the "packing list". After
confirming that there is no transportation damage, install it according to the requirements and
methods of "installation".
Warning: Unpacking inspection found that the equipment is damaged, to ensure safety, it is
forbidden to use it. The probe should not be dropped or collided. The manufacturer is not
responsible for the risks caused by this.
2.2 Battery
Battery installation: This host uses a built-in lithium battery for power supply. It is strictly forbidden to
disassemble and replace the battery during use (as shown in the figure below).

郑州博祥来电子科技有限公司
ZHENGZHOU BOXIANGLAI ELECTRONIC SCI-TECH CO. LTD.
— 10 —
Figure 2-1. Schematic diagram of build-in Li-On battery
2.3 Probe install and Remove
Installation of the probe:
The instrument has only one probe socket, which can be connected to the optional probe supported by
the instrument (as shown in the figure below).
Figure 2-3. Schematic diagram of the installation of the probe
Remove of the probe:
Remove is the reverse process of installation (as shown in the figure below)
Figure 2-4. Schematic diagram of the disassembly of the probe.
Build-in Li-on battery inside
Loosen the probe locking sleeve
and pull out the probe
Probe locking sleeve
Plug tail sleeve
Align the notch of the
probe plug with the
concave edge of the probe
socket and insert
Rotate and tighten the plug
locking sleeve socket and insert

郑州博祥来电子科技有限公司
ZHENGZHOU BOXIANGLAI ELECTRONIC SCI-TECH CO. LTD.
— 11 —
Warning: Never pull or insert the probe connector plug while the power is on to avoid damage
to the probe and the host; once the probe is connected to the host, please do not unplug or
plug at will to avoid poor contact of the probe connector.
Warning: Do not touch the connecting pins of the probe connector.
Warning: The probe should be avoided from falling or collision. The manufacturer is not
responsible for the risks caused by this.
Warning: handle with care when using the instrument.
2.4 Power supply
The instrument is powered by a lithium battery, and the lithium battery of the instrument can be
charged through an external standard adapter.
2.4.1 Time and date settings
The battery indicator on the screen will display the current remaining power. When the battery is
almost exhausted, the device will make a “didi” sound, and the battery should be charged at this time.
(1) Check whether the adapter's input power cord plug matches the power supply socket.
(2) Check whether the power supply is within the specified power supply range, and check whether
the power cord is connected correctly and firmly.
(3) Check whether the adapter is working properly: Plug the power cord into the AC input socket, turn
on the power switch of the plug-in board, and the DC output indicator light is normal.
(4) Turn off the power switch of the plug-in board.
(5) Put the host in the shutdown state.
(6) Insert the DC output plug of the adapter into the DC 14V socket on the right side of the host, and
turn on the socket power switch.
(7) The battery indicator on the host lights up, the adapter charges the battery, and the battery
indicator turns from on to off, indicating that the battery is fully charged.
Warning: The external power supply is forbidden to use any power supply other than the
standard configuration adapter.
Warning: The host should be turned off when the battery is removed for charging.

郑州博祥来电子科技有限公司
ZHENGZHOU BOXIANGLAI ELECTRONIC SCI-TECH CO. LTD.
— 12 —
Chapter 3 Operation Keyboard
3.1 Screen Display
Figure 3-1. Screen display
3.2 Keyboard function
The keyboard of the whole machine is as shown in the figure
Figure 3-2. Schematic diagram of the operation panel
Grayscale bar
Scan direction
Time and Date
Battery
Probe Frequency
Number and Spacing
of Focus Point
Frame Correlation
Total Gain
Near-field / Far-field Gain
Brightness and Contrast
Gamma correction
Image smoothing
Frame rate
Prob Model

郑州博祥来电子科技有限公司
ZHENGZHOU BOXIANGLAI ELECTRONIC SCI-TECH CO. LTD.
— 13 —
3.3 Instructions
“ ” Image freeze key : Press this key to switch between freezing and real-time status
Tips: The " FREEZE " icon appears in the lower right corner of the screen, indicating that the
image is frozen.
M/0 ~BM/9 Such keys are multi-function keys
When used as a character key:
(1) In menu operation, used to select sub-menu
(2) It is used for digital input in the time input option of the main menu. (See below for detailed
operation)
(3) M/0、DEEP/1、MEASURE/2、B/7、BB/8、BM/9 The key is used to input letters and spaces
in the comment option in the main menu. (See below for detailed operation)
When used as a function key, the specific breakdown is as follows:
(1) M/0 Function menu key: Press this key, the function menu is displayed on the screen, press
the number key in front to enter the sub-menu operation, see the following chapter for
detailed operation, if you want to exit the function menu, press the CLEAR button
(2) DEEP/1 Image zoom key: In “B” model real-time scanning state, this key is used for image
magnification conversion. The magnification adjustment range is: 80~220, 8 levels
adjustable.
(3) MEASURE/2 Measurement mark key: Pressing this key in the frozen state will display a "+"
mark on the screen, which can be moved by the arrow keys. (For detailed operation, please
refer to the introduction of distance and perimeter area measurement in the later chapter)
~Arrow keys
During measurement, it is used to move the measurement cursor. Among them, the /
keys are used to change the current scan display depth in the real-time state of the "B" mode. Press
the key to increase the current display depth, press the key to reduce the current display
depth, and the current depth is displayed in the lower right corner of the screen. / keys are
used to select the parameters on the right side of the screen, and the currently selected parameter is
in the highlighted state. At this time, the selected parameters can be adjusted by pressing the /
keys.

郑州博祥来电子科技有限公司
ZHENGZHOU BOXIANGLAI ELECTRONIC SCI-TECH CO. LTD.
— 14 —
B/7 B and 4B type display mode keys
In the real-time or frozen state, press this key to enter the B-mode display mode (single B mode is the
default when it is turned on). After pressing the key again, enter the 4B display mode by pressing the
key, and the screen will display four B-shaped images. One of them is a "real-time" image, and
the other three are "frozen" images; press the key repeatedly to realize the cyclic conversion
between the "freeze" and "real-time" states of the four images. Press BM/9 key repeatedly to switch
between B and 4B modes.
BB/8 Double B type display mode keys
In real-time or frozen state, press this key to enter double B mode. The screen displays two B-type
images. One of them is a "frozen" image and the other is a "real-time" image; press this key repeatedly
to switch between the "freeze" and "real-time" states of the left and right images. Press the key,
and both images are "freeze" images.
BM/9 B-type and M-type display mode keys
In real-time or frozen state, press this key to enter B/M mode. The screen displays both B-type and
M-type images (abbreviated as "B/M" or "B+M"). The left side of the screen is the B-type real-time
image, and the right side is the M-type real-time image.
There is a vertical line composed of equidistant points on the B-type image, which is called the
sampling line. Press the key to move the sampling line to the left, and press the key to
move the sampling line to the right. Press the key again to enter the M-type display mode, and an
M-type image will be displayed on the screen. Press this key repeatedly to switch between B+M and M
mode.
Figure 3-3. Five display mode images

郑州博祥来电子科技有限公司
ZHENGZHOU BOXIANGLAI ELECTRONIC SCI-TECH CO. LTD.
— 15 —
CATALOG Obstetrics table menu key
Press this key in the freezing state of "B" or "BB" mode to display the obstetrics menu, press the
number key on the front to execute the corresponding operation, see the following chapter for detailed
operation. To exit the obstetrics table menu, please press the CLEAR key.
CONFIRM Measurement auxiliary operation key
Cooperate with M/0、MEASURE/2 and the arrow keys to complete the measurement of distance,
circumference, area, volume and histogram. For detailed operations, see the following chapter.
MENU Comment menu key
Press this key to display the comment menu, press the number key in front to enter the sub-menu
operation, see the following chapter for detailed operation, if you want to exit the comment menu,
press CLEAR key.
CLEAR Clear screen key
Press this key to clear the measurement mark, comment information, measurement result or exit the
parameter adjustment state. After the menu is displayed, press this key to exit the menu.
RESET Reset key
When the instrument "crashes" (pressing any key does not work) due to accidents or operating errors,
press this key to restore the instrument to normal.
Chapter 4 Function Operation
4.1 Power-On
Turn on the power switch of the host, the power indicator on the panel is on, and the startup interface
appears. At this time, press any key (except keys and RESET ) to enter the scanning state.
Adjust the gain to make the image meet the visual requirements of the operator.
Figure 4-1. Startup interface

郑州博祥来电子科技有限公司
ZHENGZHOU BOXIANGLAI ELECTRONIC SCI-TECH CO. LTD.
— 16 —
Tips: The display effect of the LCD monitor is related to the viewing angle, and the operator
should adjust the viewing angle appropriately.
Warning: Flickering or dysfunction occurs when the machine is turned on. This is caused by
the unstable battery of the host or the equipment being subjected to external electromagnetic
interference. When this phenomenon is encountered, restart the equipment.
Warning: The heat dissipation holes on the back of the instrument cannot be covered,
otherwise the instrument may be damaged due to overheating.
4.2 Diagnosis
Apply an appropriate amount of medical ultrasound couplant to the diagnostic site, and closely contact
the acoustic window of the probe to the diagnostic site. The screen will display the cross-sectional
ultrasound image of the tissue, move the probe appropriately to find and determine the best position
for detection. Adjust the gain to obtain the best cross-sectional image of the diagnostic site.
Warning:
(1) When touching the probe to the diagnostic site, it is not advisable to apply excessive force
to avoid damage to the probe.
(2) A probe suitable for the tissue characteristics of the diagnostic site should be used for
diagnosis.
4.3 Imaging parameter adjustment
Image parameter adjustment includes probe frequency, focus number, frame correlation, image
smoothing, gamma correction, gain, brightness, and contrast items. Press the or keys in
real time, one of them will be lit, and the specific parameter values can be set through the and
keys, and it will be displayed on the upper right of the screen.
4.3.1 Frequency adjustment
In the real-time state, press the or key to light up the frequency at the top right of the
screen, and then adjust the frequency through the and keys. The working frequency
setting range of each probe is as follows:
6.5MHz Rectal Linear Probe (LNA64) ---------- 5.5MHz / 6.5MHz / 7.5MHz
3.5MHz Convex Prob (CXA50R) ---------- 2.5MHz / 3.5MHz / 5.0MHz
5.0MHz Micro-convex Probe (CXA20R) ---------- 4.5MHz / 5.0MHz / 5.5MHz
7.5MHz Linear Probe (LNA40) ---------- 6.5MHz / 7.5MHz / 8.5MHz

郑州博祥来电子科技有限公司
ZHENGZHOU BOXIANGLAI ELECTRONIC SCI-TECH CO. LTD.
— 17 —
4.3.2 Frequency adjustment
In the real-time state, press the or key to light up the total gain, near field, and far field on
the upper right side of the screen, and then adjust the gain value of each section through the or
key.
4.3.3 Brightness and contrast adjustment
Press the or key to light up " , " at the top right of the screen, and then use the
or key to adjust the brightness and contrast values.
4.3.4 Image focus adjustment
Press the or key to light up the "focus" at the top right of the screen. At this time, press the
key to move the focus position up, and press the key to move the focus position down. The
number of focal points is automatically set by the system according to different probe modes.
4.3.5 Frame related adjustment
In the "B", "BB", "BM" mode, enter the real-time scanning state and press the or key to light
up the "frame correlation" at the upper right of the screen, and then press the or key to
adjust the frame correlation coefficient. The device provides 8 frame correlation coefficients, which are
0.25, 0.35, 0.45, 0.55, 0.65, 0.75, 0.85, and 0.95 respectively.
4.3.6 Gamma correction adjustment
Press the or button to light up the "GM" at the upper right of the screen. The device has 4
gears for selection, GM0, GM1, GM2, and GM3. Press the or button to cycle selection.
4.3.7 Probe settings
The device has the function of automatic probe recognition, and it can automatically recognize and
display the probe type after entering the scanning working state. The device supports 4 types of
probes. The current probe type is displayed on the upper right of the screen. The objects of probe type
and probe model are as follows:
6.5MHz Rectal Linear Probe (LNA64)
3.5MHz Convex Prob (CXA50R)
5.0MHz Micro-convex Probe (CXA20R)
7.5MHz Linear Probe (LNA40 / EL1-5)
Tips: to replace the probe, please turn off the device first and restart it before automatic
identification can be realized.

郑州博祥来电子科技有限公司
ZHENGZHOU BOXIANGLAI ELECTRONIC SCI-TECH CO. LTD.
— 18 —
4.3.8 Default settings
(1) Press the M/0 key, the screen displays the function menu
(2) Press the MEASURE/2 key to restore the system to the default settings
4.4 Comment
In the freezing state, press the MENU key, and the Comment menu (as shown in the figure below) will
be displayed on the screen, through which functions such as comment, time, body mark, and
language can be realized.
4.4.1 Image Comment
Press the M/0 key and select " 0. COMMENT " to enter the image comment function, a white cursor
appears on the screen, and 10 commonly used Chinese comment characters appear at the bottom of
the screen. The CATALOG and CONFIRM keys are used to page up and down to display other
characters. Use the arrow keys to move the white cursor to annotate any position in the image area,
and press the corresponding number key in front to display the corresponding character at the cursor.
Press the CLEAR key to delete characters. After inputting, press the MENU key to confirm and exit. If
you want to abandon the input, press the MENU key directly.
4.4.2 Date and time adjustment
Press the DEEP/1 key and select " 1. TIME " to modify the system clock and date. The date and time
input interface is as follows:
For example, the time is: 2018-01-01 16:35:29, the input method is as follows:
After inputting, press the MENU key to confirm and exit, if you want to abandon the input, press the
MENU key directly.
VA1.B
0. COMMENT
1. TIME
2. BODYMARK
3. LANGUAGE
YY-MM-DD
■
HH-MM-SS
YY-MM-DD
180101
HH-MM-SS
163529

郑州博祥来电子科技有限公司
ZHENGZHOU BOXIANGLAI ELECTRONIC SCI-TECH CO. LTD.
— 19 —
4.4.3 Posture markings
Press the MEASURE/2 key and select " 2. BODYMARK " to display the body mark, see the lower left
corner of the screen. Press the , , , and keys to realize 16 kinds of body mark
cyclic display.
Cat :
Dog :
Swine :
Sheep / Goat :
Bovine :
Horse :
4.4.4 Language switch
Press the key, select " 3. LANGUAGE " to select the system interface language, the language
selection interface is as follows:
Press DEEP/1 key to select Chinese and press MEASURE/2 key to select English.
Tips: When inputting " 0. COMMENT ", the device does not provide digital input. After calling up the
comment menu, if you want to exit, press the MENU key directly.
4.5 Image Processing
In the frozen state, press the M/0 key, the screen displays the image processing menu (as shown in
the figure below), through which the cine loop, storage, browse, up and down, left and right, pseudo
color processing, area volume, histogram could be realized Measurement and other functions.
The freeze status menu is as follows:
PLEASE ENTER LANGUAGE:
1. CHINESE 2. ENGLISH
VA1.B
0. CINELOOP 1. SAVE
2. SVLOAD 3. AREA-VOLM
4. HISTOGRAM 5. COLOR
6. ERASE

郑州博祥来电子科技有限公司
ZHENGZHOU BOXIANGLAI ELECTRONIC SCI-TECH CO. LTD.
— 20 —
4.5.1 Cine Loop function
The instrument provides continuous cine loop and single-frame cine loop functions for more than 400
images. After booting into the real-time mode, the device must first start the collection of the cine loop
image, and the collection time is 30 seconds.
(1) Put the device into freeze mode. Press the M/0 key to display the function menu, then press the
M/0 key to start cine loop, and the images are played in a loop.
(2) During cine loop, press the or key to enter the manual cine loop mode, press the
key to increase the image frame by frame, and press the key to decrease the image frame
by frame. Repeat step (1) to return to the automatic cine loop mode.
(3) When playing back in " Double B " mode, you can use the switch of " Double B " to play back in
different windows.
(4) To exit the cine loop mode, press the key.
4.5.2 Store image
The device provides 64 images without losing storage when power off.
(1) After freezing a satisfactory image, press the M/0 key to display the function menu and then press
the DEEP/1 key to save the current image, and the upper corner of the image area will
automatically display the serial number of the currently stored image. For example, "SAVING……
05", the storage is completed, and the serial number disappears. Press the key to return to
the real-time status.
(2) The device can store up to 64 images, and the stored images are automatically numbered
according to the order of storage. If there are images stored in 01-20, then the storage operation
will display the number 21; when the image storage area is full (that is, 64 frames of images are
stored), the following prompt will appear when re-saving images:
Prompt whether to overwrite the stored image numbered 01; press DEEP/1 key to overwrite the
current stored image; press MEASURE/2 key to abandon this image storage operation.
Select " 2. NO " to abandon the image storage, and then perform the storage operation, the system will
prompt you whether to overwrite the image number 02, and so on.
Tips: When the image storage area is full, if you first recall a stored image, and then perform
the image storage operation, the system will prompt whether to clear and store the numbered
image.
STORAGE IS FULL. ERASE NO.01?
1.YES 2.NO
Table of contents
Popular Industrial Equipment manuals by other brands

Terex
Terex RL4000D2 Operator, service & parts manual

ITT Industries
ITT Industries FLYGT 4610 Workshop manual

SECOP
SECOP 101N2130 instructions

HEUTE
HEUTE ProfilGate i45 aqua operating manual

woodmizer
woodmizer MP260 Original instructions

Gema
Gema OptiCenter OC08 Operating instructions and spare parts list

Rittal
Rittal VX 8617.510 Assembly instructions

Siemens
Siemens BD2-AK 02 M2 Series installation instructions

Philadelphia Scientific
Philadelphia Scientific HydroCart Max instructions

Riello
Riello FCU 09 N manual

Calder Brothers
Calder Brothers Mauldin Precision Spray Owner/Operator & Parts Manual

Emerson
Emerson Fisher ATST Desuperheater instruction manual