
BNP 6012 and 7212 SUCTION BLAST CABINETS Page 13
© 2016 CLEMCO INDUSTRIES CORP. www.clemcoindustries.com Manual No. 27723, Rev A, 07/16
5.4 Reclaimer Static Pressure
5.4.1 Correct static pressure varies with size of
reclaimer and the size, weight and type of media.
5.4.2 Adjust static pressure by opening (handle inline
with air flow) or closing (handle perpendicular to air flow)
the dust collector ventilation damper. Refer to the dust
collector owner’s manual, the damper is located on the
inlet on CDC-1 dust collectors, and on the exhauster
outlet on RPC and RPH dust collectors. If the damper is
not opened far enough, the reclaimer will not remove
fines, resulting in dusty media, poor visibility, and possible
media blockage in the conveying hose. If the damper is
opened too far, it may cause carryover (usable media
carried into the dust collector) and result in excessive
media consumption. Open only as far as necessary to
obtain a balance of dust removal without media carryover.
5.4.3 A manometer is useful when adjusting or
monitoring static pressure. The optional manometer kit is
listed under Optional Accessories in Section 9.1. Refer
to Section 5.8 for manometer operation. The following
are static pressure starting points for given media. Static
pressure may need to be lower with finer media and
higher with coarser media. Run the media through
several blast cycles allowing the reclaimer to function
with these settings. Inspect the media in the reclaimer
and fines in the dust collector as noted in Paragraph
5.4.2. Continue adjusting static pressure until optimum
media cleaning without carryover is attained.
Glass Bead No. 4 to 7 .................................. 3" to 3-1/2"
Glass Bead No. 8 to 13 .....................................2-1/2 - 3"
Alox. 60 & coarser ...................................................4 - 5"
Alox. 80 & finer ..................................................2-1/2 - 3"
Steel Grit ..................................................................6 - 7"
5.4.4 If the damper has been adjusted and carryover
or excessive dust in the media continues, adjusting the
vortex cylinder, may help by fine-tuning media
separation. Refer to Section 5.5 to adjust the cylinder.
5.5 Optional Externally-Adjustable Vortex Cylinder
NOTE: The externally adjustable vortex is an option
when the cabinet is provided with a CDC-1 Dust
collector. The vortex is standard with reclaimers when
the cabinet is provided from the factory with an RPC or
RPH dust collector.
The vortex cylinder fine-tunes media separation. Before
adjusting the vortex cylinder, adjust the damper on the
dust collector to increase or decrease static pressure
per Section 5.4. Once the damper is adjusted, adjust the
cylinder.
5.5.1 The vortex cylinder is located atop the reclaimer
where the flex hose connects. Adjustments are made by
loosening the handle's tensioning knob and moving the
handle to achieve the correct setting. When the correct
setting is established, tighten the locking knob to prevent
movement. Start with the lever slightly to the right (about
one o’clock as shown in Figure 19) of the vertical position.
5.5.2 To Remove More Fines: (Too much dust in
media) Raise the cylinder by moving the lever left toward
"COARSE", in 1/4" increments at the indicator plate.
Allow the media to go through blast several cycles
before determining if further adjustment is needed.
5.5.3 To Remove Less Fines: (Too much usable
media is carried to the dust collector) Lower the vortex
cylinder by moving the lever right toward "FINE", in 1/4"
increments at the indicator plate. Run the media through
a cycle and check the dust container to determine if
further adjustment is required. NOTE: If the cylinder is
lowered too far, the reclaimer will again begin to allow
usable media to be carried over, and cause abnormally
high static pressure.
Figure 19
5.5.4 When using media finer than 180-mesh, the
reclaimer inlet baffle may need to be removed. Refer to
Section 1.8.6.
5.6 Cabinet Air-Inlet Damper
5.6.1 Once the inlet is initially set per Section 2.6, it
seldom requires readjustment. The initial setting
produces approximately .5to .75of static pressure in
the cabinet enclosure. Do not confuse cabinet static
pressure with reclaimer static pressure, which is
controlled by the dust collector ventilation-damper
as noted in Section 5.4. Reclaimer pressure must be
set before cabinet pressure. In rare circumstances,
cabinet pressure may need to be slightly higher or lower.
5.6.2 A manometer (as noted in Section 5.8 and listed
in Section 9.1) is the most accurate method of
monitoring and adjusting cabinet pressure. Following the
instructions packed with the manometer, start the
exhauster and insert the needle into a glove, and adjust
pressure using the cabinet’s air-inlet damper. Open the
Remove more fines
from media by
moving the handle
farther to the left
Remove less
fines by moving
the handle farther
to the right
Tensioning Knob