Caution
•When operating the handpiece always consider the safety of the patient.
•The handpiece is designed only for dental clinical use.
•Do not attempt to disassemble the handpiece or tamper with the mechanism.
•Check the vibration, noise and overheating outside the patient's oral cavity before use. If any abnormalities are found, stop using
immediately and contact Brasseler USA.
•Should the handpiece function abnormally, cease operation immediately and return the handpiece to Brasseler USA for repair.
•Depressing the push-button while handpiece bur is in rotation will result in OVERHEATING of the handpiece head. Special caution
must be exercised during use to keep cheek tissue AWAY from the push-button of the handpiece. Contact with cheek tissue may
cause the push-button to depress and injury to the patient may occur.
•Do not allow any impact on the sheath. Do not drop the handpiece.
•Do not use bent, damaged, or sub-standard burs. The shank could be bent or broken even within the recommended speed.
•Always keep the bur shank clean. Entry of debris into the chuck could cause bur run-out.
•Do not exceed the speed recommended by the bur manufacturers.
•Do not operate the handpiece without the bur or test-bur
•Remove the handpiece or the bur after the air supply is stopped completely.