BREWFERM COBRA User manual

COBRA
pH pen
MANUAL


COBRA
pH pen
EN

4
I. INTRODUCTION
Congratulations on your purchase of the
Cobra. Please read the manual completely
before first use. Keep the manual for fu-
ture reference. It is recommended to soak
the electrode for at least 30 minutes be-
fore use to make the electrode wet.
Features:
• IP65 Waterproof housing
• Dual display with ATC
• Data hold to freeze readings
• Compact size, easily fits in your pocket
• Low battery indicator
• Auto power off
• °C/°F unit switchable
• Power by 2 CR2032 batteries
• Multi point calibration
• One-touch-only for calibration
II. MATERIAL SUPPLIED
This package contains:
• The meter x 1
• CR2032 button battery x 2
• Manual
III. POWER SUPPLY
The meter is powered by 2 CR2032 bat-
teries. Check the battery:
1. before first use;
2. when the battery symbol appears on
the LCD screen;
3. when the meter can not be switched
on.

5
To install the batteries:
1. Turn off the meter.
2. Losen the battery cover in clockwise di-
rection with a coin (DON’T discard the
black washer!).
3. Replace the old batteries with 2 new
button cells CR2032.
4. Make sure the batteries are in place
and the polarity is correct.
5. Put back the battery cover and turn it
in counter-clockwise direction to lock.
NOTE:
1. Please re-calibrate the meters after
changing batteries.
2. Remove the battery from the Cobra if
you do not plan to use it for a month or
more.

6
IV. LCD DISPLAY
• The 1st display shows the measured pH
reading.
• The 2nd display shows the temp.
• Cal = Calibration mode
• ATC = Auto Temp. Compensation
• H= Data hold
V. KEYPAD
VI. HOW TO OPERATE
1. Remove the probe storage bottle from
meter to expose the electrode. It is nor-
mal to find white crystals on the cap or
electrode assembly.
WARNING: always make sure the bottle
is filled with a storage buffer to keep
the electrode wet and in a good stor-
age condition.
CAL
H
ATC
F
C
O
O

7
2. Dip the electrode into the test solution.
Press “ ⌽” to switch on the meter and
stir it to get a stable reading.
3. A small dot “ .” flashes while the meter
is in measurement mode. The LCD not
only shows the measured pH value but
also displays the temperature.
4. Press “ HOLD ” to freeze current read-
ings. The “ H” icon will appear on the
LCD and the small dot will not flash.
Press “ HOLD ” again to release the hold
mode.
ATC
C
O
Rotate the
storage bottle in
counter-clockwise
direction and pull out the
bottle completely to expose the electrode.
Electrode

8
5. Turn off the meter by pressing “ ⌽”
button.
6. Once you have finished the measure-
ment, clean the electrode, put back the
bottle and store the meter below 50°C.
7. Please see page 9 for the calibration
procedures.
VII. AUTO POWER OFF (SLEEP FUNCTION)
This meter will turn off automatically af-
ter 20 minutes of inactivity. If you want to
operate for a longer time, you can disable
the sleep mode.
To disable the auto power off:
Before turning the power on,
press “ ⌽” + “ HOLD ” keys si-
multaneously until an “n” ap-
peares on the screen. Then
release the keys to return to
normal mode.
Note: This setting (sleep function off) will
be disabled when you switch the power
off.
H
ATC
F
O

9
VIII. AUTOMATIC TEMPERATURE COM-
PENSATION (ATC) & TEMP. UNIT SET-
TING
The Cobra is capable of measuring with
Automatic Temperature Compensation.
“ATC” will show in the middle top of the
LCD.
Temp. Unit Setting of the Cobra:
To select the temp. unit (°C or °F), turn
off the meter first. When the meter is off,
press “ ⌽” and “ CAL ” at the same time
until the C or F appears on the LCD. Press
“ HOLD ” to select the preferred unit and
then press “ CAL ” to save. “ SA ” will ap-
pear on the LCD for one second and then
the LCD screen will go back to normal.
IX. CALIBRATION MODE (CAL)
Calibration is necessary and should be
done regularly. It is recommended to cali-
brate every day if the meter is used often.
The unique calibration design of the me-
ter features an automatic buffer recogni-
tion to avoid errors.
o
C

10
The Cobra Calibration:
1. Switch on the meter.
2. Place the electrode into a 7pH buffer
solution (afterwards place the elec-
trode into a 4pH or a 10pH buffer solu-
tion for better accuracy).
3. Press “ CAL ” to enter pH calibration
mode. The icon “ CAL ” will appear on
the LCD screen for one second and then
pH value (4, 7 or 10) will be displayed
on the LCD screen.
4. If the probe or buffer display an error:
• the buffer is incorrectly inserted;
• the probe is damaged
• the probe can not detect the buffer in
below voltage range.
The meter will exit calibration mode
automatically after 10
seconds. “ End ” will ap-
pear on the LCD screen
in one second and then
will go back to normal
status.
Acceptable voltage range of each point:
pH 4.00 : 97mV ... 250mV
pH 7.00 : - 60mV ... 60mV
pH 10.00 : - 250mV ... -97mV
5. If the probe recognizes the buffer:
If the probe successfully recognizes the
CALATC
F
O

11
buffer, the buffer pH value (4, 7 or 10)
will appear on the display in 2 seconds.
If the calibration buffer is not 4,7 or 10,
but another value, such as 7.01, just
press “ HOLD ” to change the value.
6. The adjustable cal. point range for
4.0pH is from 3.50 to 4.50. For 7.0pH
is from 6.50 to 7.50. For 10.0pH is from
9.50 to 10.50.
7. To save the calibration value:
When the electrode reads a stable
value and the user is not pressing any
key, the meter will automatically save
the value and then exit the calibration
mode.
8. Rinse the probe with de-ionised water
or a rinse solution (e.g. tap water) after
each measurement to prolong the me-
ter’s life.
9. Repeat the steps above until the point
calibration is finished.
CALATC
F
O

12
X. MAINTENANCE
• Always keep the pH glass bulb wet by
using the storage bottle to protect and
store the electrode.
• Always rinse the pH electrode in de-ion-
ised water or rinse solution (e.g. tap
water) before the next use.
• Never touch or rub the glass bulb in or-
der to prolong the pH electrode’s life.
• Make sure the electrode is clean. Be-
tween measurements, rinse the elec-
trode with de-ionised water. If the
eletrode has been exposed to a solvent
immiscible with water, clean it with a
solvent miscible with water (e.g. etha-
nol or acetone) and rinse carefully with
water.
• Store the electrode carefully.
XI. TROUBLESHOOTING
• Power on but no display
1. Make sure you press the power key
more than 100 mS.
2. Check for whether the batteries are
in place, they make contact and that
the polarity is correct.
3. Replace the batteries and try again.
4. Remove the batteries for one minute
and then put them back in again.
• Slow response
Clean the probe by immersing the elec-
trode in tap water for 10-15 minutes,
then rinse thoroughly with distilled wa-
ter or use a general purpose electrode
cleaner.

13
• LCD display “ - - - ”
Out of pH range, too
acidic/or too alkaline.
• “ H” or “ L”
Out of temp. range, too
cold or too hot.
• pH value fluctuate quickly
This is normal when the electrode is not
immersed in water, but exposed to air.
XII. SPECIFICATIONS
• Operating Temp.: 0°~50°C (32~122°F)
• Operating Humidity: 0~80% RH
• pH calibration buffer suggested:
ATC
C
O
ATC
C
O

14
• Standard USA buffers
• Standard NIST buffers
• Standard DIN buffers
• Battery Life: >100 hrs continuous use
XIII. WARRANTY
The meter is warranted to be free from
defects in material and workmanship for
a period of one year from the date of pur-
chase. This warranty covers normal oper-
ation but does not cover battery, misuse,
abuse, alteration, tampering, neglect, im-
proper maintenance, or damage resulting
from leaking batteries. Proof of purchase
is required for warranty repairs. Warranty
is void if the meter used has been taken
apart.
XIV. RETURN AUTHORIZATION
Authorization must be obtained from the
supplier before returning items for any
reason. When requiring an RA (Return Au-
thorization), please include data regard-
ing the defective reason. The meters must
be returned properly packaged to prevent
any damage in shipment and be insured
against possible damage or loss.

COBRA
pH pen
NL

16
I. INLEIDING
Gefeliciteerd met je aankoop van de Cobra.
Lees de gebruiksaanwijzing volledig door
voordat je het product in gebruik neemt. Be-
waar de gebruiksaanwijzing voor later ge-
bruik. Het wordt aanbevolen om de elektro-
de voor gebruik vochtig te maken door deze
ten minste 30 minuten onder te dompelen.
Kenmerken:
• IP65 waterdichte behuizing
• Dubbel display met ATC
• HOLD-functie om uitlezingen vast te
houden
• Compacte afmetingen, past gemakkelijk
in je broekzak
• Indicatie batterijniveau laag
• Automatische uitschakelfunctie
• Wisselen tussen temperatuureenheid °C
of °F
• Werkt op 2 CR2032-batterijen
• Meerpuntskalibratie
• Slechts één aanraking om te kalibreren
II. INHOUD LEVERING
Dit pakket bevat:
• 1 x meter
• 2 x CR2032-knoopbatterij
• Gebruiksaanwijzing
III. VOEDING
De meter werkt op 2 CR2032-batterijen.
Controleer de batterij:
1. vóór het eerste gebruik;
2. alshetbatterij-pictogramwordtweergege-
ven op het LCD-scherm;
3. alshetnietluktomtemeterinteschakelen.

17
De batterijen plaatsen:
1. Zet de meter uit.
2. Draai het batterijdeksel met de klok
mee los met een muntje (gooi de ring
NIET weg!).
3. Vervang de oude batterijen door 2 nieu-
we CR2032-knoopbatterijen.
4. Controleer of de batterijen goed ge-
plaatst zijn en of de polariteit klopt.
5. Draai het batterijdeksel weer vast tegen
de klok in.
LET OP:
1. Kalibreer de meters opnieuw nadat de
batterijen zijn vervangen.
2. Haal de batterijen uit de Cobra als je
deze een maand of langer niet gaat ge-
bruiken.

18
III. LCD-SCHERM
• Het 1ste scherm geeft de gemeten pH-
waarde weer.
• Het 2de scherm geeft de temperatuur
weer.
• Cal = Kalibratiemodus
• ATC = Automatische temp. compensatie
• H= Gegevens vasthouden
V. TOETSEN
VI. BEDIENING
1. Haal de meter uit het bewaarflesje om
de elektrode bloot te leggen. De aan-
wezigheid van witte kristallen op de
dop of elektrode is normaal.
WAARSCHUWING: zorg dat het flesje
altijd gevuld is met buffervloeistof. Zo
blijft de elektrode vochtig en kan deze
langer bewaard worden.
CAL
H
ATC
F
C
O
O

19
2. Dompel de elektrode onder in de test-
vloeistof. Druk op “ ⌽” om de meter
aan te zetten en roer door de oplossing
om de uitlezing te stabiliseren.
3. Er knippert een stipje “ .” als de me-
ter zich in de meetmodus bevindt. Het
LCD-scherm geeft niet alleen de geme-
ten pH-waarde, maar ook de tempera-
tuur weer.
4. Druk op “ HOLD ” om de huidige uitlezing
vast te houden. Op het LCD-scherm
verschijnt er “ H” en het stipje knippert
niet. Druk opnieuw op “ HOLD ” om de
HOLD-functie te verlaten.
Draai het
bewaarflesje tegen
de klok in en trek
het flesje er volledig uit om
de elektrode bloot te leggen.
Elektrode
ATC
C
O

20
5. Zet de meter uit door op de knop “ ⌽”
te drukken.
6. Als je klaar bent met meten, maak de
elektrode dan schoon, plaats het flesje
terug en bewaar de meter onder de 50 °C.
7. Zie pagina 21 voor de kalibratieproce-
dures.
VII. AUTOMATISCHE UITSCHAKELFUNCTIE
(SLAAPMODUS)
Deze meter schakelt zichzelf na 20 minu-
ten van inactiviteit automatisch uit. Als je
de meter langer wilt gebruiken, kun je de
slaapmodus uitschakelen.
De automatische uitschakelfunctie uitscha-
kelen:
Houd voordat je het apparaat
aanzet de toetsen “ ⌽” + “
HOLD ” tegelijk ingedrukt tot
er een “ n ” verschijnt op het
scherm. Laat de toetsen los om
terug te gaan naar de normale
modus.
Let op: Deze instelling (slaapmodus UIT)
wordt uitgeschakeld zodra je het appa-
raat uitzet.
H
ATC
F
O
Table of contents
Languages:
Other BREWFERM Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

MONARCH INSTRUMENT
MONARCH INSTRUMENT Nova-Pro 100 User manual and installation guide

Bitzer
Bitzer OLC-D1-S technical information

Aqua-Scope
Aqua-Scope AQSWIE01 User and installation manual

Shimpo
Shimpo DT-365E instruction manual

YOKOGAWA
YOKOGAWA Powercert PR201 user manual

Precaster
Precaster ProRange-1200 quick start