GE DigitalFlow XGM868i User guide

GE
Measurement & Control Flow
910-197U Rev. F1
February 2016
DigitalFlow™ XGM868i
General-Purpose Gas Flow Transmitter (1 & 2 Channel)
Startup Guide


www.gemeasurement.com
©2016 General Electric Company. All rights reserved.
Technical content subject to change without notice.
DigitalFlow™ XGM868i
General-Purpose Gas Flow Transmitter (1 & 2 Channel)
Startup Guide
910-197U Rev. F1
February 2016

ii
[no content intended for this page]

DigitalFlow™ XGM868i Startup Guide iii
Preface
Information Paragraphs
Note: These paragraphs provide information that provides a deeper understanding of the situation, but is not
essential to the proper completion of the instructions.
IMPORTANT: These paragraphs provide information that emphasizes instructions that are essential to proper
setup of the equipment. Failure to follow these instructions carefully may cause unreliable
performance.
Safety Issues
Auxiliary Equipment
Local Safety Standards
The user must make sure that he operates all auxiliary equipment in accordance with local codes, standards,
regulations, or laws applicable to safety.
Working Area
CAUTION! This symbol indicates a risk of potential minor personal injury and/or severe damage
to the equipment, unless these instructions are followed carefully.
WARNING! This symbol indicates a risk of potential serious personal injury, unless these
instructions are followed carefully.
WARNING! It is the responsibility of the user to make sure all local, county, state and national
codes, regulations, rules and laws related to safety and safe operating conditions are met for each
installation. The safety of any system incorporating the equipment is the responsibility of the
assembler of the system.
WARNING! Auxiliary equipment may have both manual and automatic modes of operation. As
equipment can move suddenly and without warning, do not enter the work cell of this equipment
during automatic operation, and do not enter the work envelope of this equipment during manual
operation. If you do, serious injury can result.
WARNING! Make sure that power to the auxiliary equipment is turned OFF and locked out before
you perform maintenance procedures on the equipment.

Preface
iv DigitalFlow™ XGM868i Startup Guide
Qualification of Personnel
Make sure that all personnel have manufacturer-approved training applicable to the auxiliary equipment.
Personal Safety Equipment
Make sure that operators and maintenance personnel have all safety equipment applicable to the auxiliary
equipment. Examples include safety glasses, protective headgear, safety shoes, etc.
Unauthorized Operation
Make sure that unauthorized personnel cannot gain access to the operation of the equipment.
Environmental Compliance
Waste Electrical and Electronic Equipment (WEEE) Directive
GE Measurement & Control is an active participant in Europe’s Waste Electrical and Electronic Equipment
(WEEE) take-back initiative, directive 2012/19/EU.
The equipment that you bought has required the extraction and use of natural resources for its production. It
may contain hazardous substances that could impact health and the environment.
In order to avoid the dissemination of those substances in our environment and to diminish the pressure on
the natural resources, we encourage you to use the appropriate take-back systems. Those systems will reuse or
recycle most of the materials of your end life equipment in a sound way.
The crossed-out wheeled bin symbol invites you to use those systems.
If you need more information on the collection, reuse and recycling systems, please contact your local or
regional waste administration.
Visit www.gemeasurement.com/environmental-health-safety-ehs for take-back instructions and more
information about this initiative.

DigitalFlow™ XGM868i Startup Guide v
Contents
Chapter 1. Installation
1.1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.2 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.3 Site Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
1.3.1 Electronics Enclosure Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
1.3.2 Flowcell Location. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
1.3.3 Transducer Location. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
1.3.4 Cable Lengths. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
1.3.5 Temperature and Pressure Transmitters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
1.3.6 Transducer Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
1.4 Installing a Flowcell. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
1.5 Installing Temperature and Pressure Transmitters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
1.6 Mounting the XGM868i Electronics Enclosure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
1.7 Making the Electrical Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
1.7.1 Wiring the Line Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
1.7.2 Wiring the Transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
1.7.3 Wiring Std 0/4-20 mA Analog Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.7.4 Wiring the Serial Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.7.5 Wiring the Option Cards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Chapter 2. Initial Setup
2.1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
2.2 Programming Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
2.3 The XGM868i Enclosure Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
2.4 Entering Data in the Global Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
2.4.1 Entering Global System Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
2.5 Activating a Channel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
2.6 Entering System Data for the Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
2.7 Entering Transducer and Pipe Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
2.7.1 Special Transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
2.7.2 Pipe Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Chapter 3. Operation
3.1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
3.2 Powering Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
3.3 The LCD Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
3.4 The Optional PanaView Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
3.5 Taking Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
3.5.1 Programming the LCD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
3.5.2 Using the LCD Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
3.5.3 PanaView Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
3.5.4 Pausing Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53

Contents
vi DigitalFlow™ XGM868i Startup Guide
Chapter 4. Specifications
4.1 General Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
4.1.1 Hardware Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
4.1.2 Environmental . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
4.1.3 Velocity Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
4.1.4 Velocity Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
4.1.5 Rangeability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
4.1.6 Repeatability. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
4.2 Electrical Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
4.2.1 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
4.2.2 Power Consumption. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
4.2.3 Operating Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
4.2.4 European Compliance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
4.2.5 Input/Output Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
4.2.6 Preamplifier. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
4.3 Flow Transducer Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
4.3.1 Physical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
4.3.2 Area Classifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
4.4 Flowcell Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59
4.4.1 Spoolpiece. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59
4.4.2 Cold Tap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59
4.4.3 Pipe Size & Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59
Appendix A. CE Mark Compliance
A.1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
A.2 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
Appendix B. Data Records
B.1 Available Option Cards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63
B.2 Option Cards Installed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64
B.3 Setup Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65
Appendix C. Measuring P and L Dimensions
C.1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
C.2 Measuring P and L. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69

DigitalFlow™ XGM868i Startup Guide 1
Chapter 1. Installation
Chapter 1. Installation
1.1 Introduction
To ensure safe and reliable operation of the Model XGM868i Ultrasonic Flow Transmitter, the system must be installed
in accordance with the guidelines established by GE engineers. Those guidelines, which are explained in detail in this
chapter, include the following specific topics:
•Unpacking the Model XGM868i system
•Selecting suitable sites for the electronics enclosure and the flowcell/transducers
•Installing the flowcell/transducers
Note: See the enclosed Transducer Installation Guide for detailed instructions on transducer installation.
•Installing optional temperature and pressure transmitters
•Installing the electronics enclosure
•Wiring the electronics enclosure
WARNING! The Model XGM868i flow transmitter can measure the flow rate of many gases, some of which
are potentially hazardous. In such cases, the importance of proper safety practices cannot be
overemphasized.
Be sure to follow all applicable local safety codes and regulations for installing electrical
equipment and working with hazardous gases or flow conditions. Consult company safety
personnel or local safety authorities to verify the safety of any procedure or practice.
!ATTENTION EUROPEAN CUSTOMERS!
To meet CE Mark requirements, all cables must be installed as described in Appendix A,
CE Mark Compliance.
1.2 Unpacking
Carefully remove the electronics enclosure, the transducers, and the cables from the shipping containers. Before
discarding any of the packing materials, account for all components and documentation listed on the packing slip. The
discarding of an important item along with the packing materials is all too common. If anything is missing or damaged,
contact the factory immediately for assistance.

Chapter 1. Installation
2 DigitalFlow™ XGM868i Startup Guide
1.3 Site Considerations
Because the relative location of the flowcell and the electronics enclosure is important, use the guidelines in this
section to plan the XGM868i installation. Figure 1 shows a typical XGM868i enclosure and spoolpiece ready for
insertion into a process line.
Figure 1: A Typical Model XGM868i System
1.3.1 Electronics Enclosure Location
The standard Model XGM868i electronics enclosure is a powder-coated, cast-aluminum, Type-7/4X explosion-proof
enclosure; an optional stainless-steel enclosure is available. Typically, the enclosure is mounted as close as possible to
the transducers. When choosing a site, make sure the location permits easy access to the electronics enclosure for
programming, maintenance and service.
Note: For compliance with the European Union’s Low Voltage Directive (2006/95/EC), this unit requires an external
power disconnect device such as a switch or circuit breaker. The disconnect device must be marked as such,
clearly visible, directly accessible, and located within 1.8 m (6 ft) of the Model XGM868i.

DigitalFlow™ XGM868i Startup Guide 3
Chapter 1. Installation
1.3.2 Flowcell Location
The pipeline flowcell consists of the flow rate transducers and any pressure and/or temperature transducers employed
as part of the flow measurement system. Ideally, choose a section of pipe with unlimited access to the flowcell; for
example, a long stretch of pipe that is above ground. However, if the flowcell is to be mounted on an underground pipe,
dig a pit around the pipe to facilitate installation of the transducers.
1.3.3 Transducer Location
For a given fluid and pipe, the Model XGM868i’s accuracy depends primarily on the location and alignment of the
transducers. In addition to accessibility, when planning for transducer location, adhere to the following guidelines:
1. Locate the transducers so that there are at least 20 pipe diameters of straight, undisturbed flow upstream and 10
pipe diameters of straight, undisturbed flow downstream from the measurement point. To ensure undisturbed flow,
avoid sources of turbulence in the fluid such as valves, flanges, expansions and elbows; and dips or low spots in
which condensed liquid may collect.
2. Because condensate or sediment at the bottom of the pipe may cause attenuation of the ultrasonic signal, locate the
transducers on the side of a horizontal pipe, when possible. If limited pipe access necessitates top-mounted
transducers and the sound beam path includes a reflection, shift the transducers to at least 10° off top dead center.
This will minimize the influence of any sediment on the reflected ultrasonic signals.
1.3.4 Cable Lengths
Locate the electronics enclosure as close as possible to the flowcell/transducers, preferably directly on the flowcell.
However, the factory can supply transducer cables up to 1,000 ft (300 m) in length for remote location of the
electronics enclosure. If longer cables are required, consult the factory for assistance.
1.3.5 Temperature and Pressure Transmitters
When installing temperature and/or pressure transmitters in the flowcell, locate them downstream of the transducers.
These transmitters should be positioned no closer to the Model XGM868i transducers than 2 pipe diameters and no
further away from the transducers than 20 pipe diameters.
1.3.6 Transducer Cables
When installing the transducer cables, always observe established standard practices for the installation of electrical
cables. Specifically, do not route transducer cables alongside high-amperage AC power lines or any other cables that
could cause electrical interference. Also, protect the transducer cables and connections from the weather and corrosive
atmospheres.
Note: When using non-GE cables to connect the flow rate transducers to the Model XGM868i electronics enclosure,
the cables must have electrical characteristics identical to the GE cables. Type RG62 a/u coaxial cable should
be used, and each cable must be the same length (within ±4 in.).

Chapter 1. Installation
4 DigitalFlow™ XGM868i Startup Guide
1.4 Installing a Flowcell
A flowcell is the section of pipe where the transducers are mounted. It can be created either by mounting the
transducers on the existing pipeline or by mounting them on a spoolpiece. A spoolpiece is a separately manufactured
pipe section, matched to the existing pipe, which contains ports for mounting the transducers. This approach allows the
transducers to be aligned and calibrated before inserting the spoolpiece into the pipeline.
Figure 1 on page 2 shows a typical Model XGM868i spoolpiece, with a mounting bracket to hold the electronics
enclosure. For detailed instructions on installing the transducers and/or spoolpiece, refer to the supplied drawings and
the enclosed GE Gas Transducer Installation Guide (916-049).
1.5 Installing Temperature and Pressure Transmitters
Optional temperature and pressure transmitters may be installed near the ultrasonic transducer ports as part of the
flowcell. Be sure to observe the siting requirements given earlier in this chapter. These transmitters should send a
0/4-20 mA signal to the Model XGM868i. In turn, the Model XGM868i must be fitted with a suitable option card to
process the signals and to provide the required 24 VDC power to the transmitters. Any desired transmitters or sensors
may be used, but they must have an accuracy equal to 0.5% of the reading or better.
Note: Resistive Thermal Devices (RTDs) are a good choice for measuring the temperature.
Typically, a 1/2” or 3/4” NPT female threaded port is used to mount the transmitters on the flowcell. If the pipeline is
insulated, the coupling may need to be extended to provide convenient access. Of course, other types of mounting
ports, including flanged ports, may be used for the transmitters.
Figure 2 shows a typical mounting arrangement for the pressure and temperature transmitters. The temperature sensor
should protrude 1/4 to 1/2 way into the pipe.
Figure 2: Typical Temperature/Pressure Transmitter Mounting
Temperature Transmitter
RTD
Thermowell
Coupling
Pressure Transmitter
Isolation Valve
Flowcell Wall

DigitalFlow™ XGM868i Startup Guide 5
Chapter 1. Installation
1.6 Mounting the XGM868i Electronics Enclosure
The standard Model XGM868i electronics package is housed in a Type-4X weather-resistant enclosure suitable for
indoor or outdoor use. Refer to Figure 8 on page 23 for the mounting dimensions and the weight of this enclosure.
The Model XGM868i electronics enclosure is fitted with a mounting boss that has a single 3/4” NPTF threaded hole at
its center and four 1/4-20 tapped holes at its corners. By using this mounting boss, the electronics enclosure may be
mounted by any of the typical techniques shown in Figure 10 on page 25.
WARNING! Proper grounding of the XGM868i chassis is required to prevent the possibility of electric shock.
See Figure 10 on page 25 for the location of the chassis grounding screw.
IMPORTANT: Since the IREX keys do not work properly in direct sunlight, be sure to position the XGM868i with an
optional sunshield or out of direct sunlight.
1.7 Making the Electrical Connections
This section contains instructions for making all the necessary electrical connections to the Model XGM868i flow
transmitter. Refer to Figure 10 on page 25 for a complete wiring diagram.
Except for the power connector, all electrical connectors are stored on their terminal blocks during shipment and may
be removed from the enclosure for more convenient wiring. Simply, feed the cables through the conduit holes on the
side of the enclosure, attach the wires to the appropriate connectors and plug the connectors back onto their terminal
blocks.
!ATTENTION EUROPEAN CUSTOMERS!
To meet CE Mark requirements, all cables must be installed as described in Appendix A,
CE Mark Compliance.
Refer to Figure 3 on page 6 and prepare the Model XGM868i for wiring by completing the following steps:

Chapter 1. Installation
6 DigitalFlow™ XGM868i Startup Guide
1.7 Making the Electrical Connections (cont.)
WARNING! Always disconnect the line power from the Model XGM868i before removing either the front
cover or the rear cover. This is especially important in a hazardous environment.
1. Disconnect any existing power line from its source.
2. Loosen the set screw on the rear cover.
3. Place a rod or long screwdriver across the cover in the slots provided, and rotate the cover counterclockwise until it
comes free from the enclosure.
4. Install any required cable clamps in the appropriate conduit holes around the side of the enclosure.
5. Note the labels inside the rear cover to assist in wiring the power and option card connections. Also, Figure 11 on
page 26 shows the rear cover option card labels for every available option card configuration.
Proceed to the appropriate section of this chapter to make the desired wiring connections.
Figure 3: Rear Cover with Connection Labels
Power ConnectionsOption Card Connections
Wiring Label Slots
Set Screw
Inside View Outside View

DigitalFlow™ XGM868i Startup Guide 7
Chapter 1. Installation
1.7.1 Wiring the Line Power
The Model XGM868i may be ordered for operation with power inputs of 100-120 VAC, 220-240 VAC, or 12-28 VDC.
The label on the side of the electronics enclosure lists the meter’s required line voltage and power rating. The fuse size
is listed in Chapter 4, Specifications. Be sure to connect the meter only to the specified line voltage.
Note: For compliance with the European Union’s Low Voltage Directive (2006/95/EC), this unit requires an external
power disconnect device such as a switch or circuit breaker. The disconnect device must be marked as such,
clearly visible, directly accessible, and located within 1.8 m (6 ft) of the Model XGM868i.
Note: Only use Class 2 Rated Power supply for line power connection to DC instrument.
Refer to Figure 10 on page 25 to locate terminal block TB5 and connect the line power as follows:
WARNING! Improper connection of the line power leads or connecting the meter to the incorrect line
voltage may damage the unit. It may also result in hazardous voltages at the flowcell and
associated piping as well as within the electronics enclosure.
1. Prepare the line power leads by trimming the line and neutral AC power leads (or the positive and negative DC
power leads) to a length 0.5 in. (1 cm) shorter than the ground lead. This ensures that the ground lead is the last to
detach if the power cable is forcibly disconnected from the meter.
2. Install a suitable cable clamp in the conduit hole indicated in Figure 10 on page 25. If possible, avoid using the
other conduit holes for this purpose, to minimize any interference in the circuitry from the AC power line.
!ATTENTION EUROPEAN CUSTOMERS!
To meet CE Mark requirements, all cables must be installed as described in Appendix A,
CE Mark Compliance.
3. Strip 1/4 in. of insulation from the end of each of the three power line leads.
4. Route the cable through the conduit hole and connect the line power leads to terminal block TB5, using the pin
number assignments shown in Figure 10 on page 25.
5. Leaving a bit of slack, secure the power line with the cable clamp.
WARNING! Make sure both covers, with their o-ring seals, are installed and the set screws tightened before
applying power in a hazardous environment.
CAUTION! The transducers must be properly wired before applying power to the meter.
Proceed to the next section to continue the initial wiring of the Model XGM868i flow transmitter.

Chapter 1. Installation
8 DigitalFlow™ XGM868i Startup Guide
1.7.2 Wiring the Transducers
Before wiring the XGM868i transducers, complete the following steps:
•disconnect the main power from the electronics enclosure
•remove the rear cover and install all required cable clamps
Based on the location of the electronics enclosure, proceed to the appropriate sub-section for detailed instructions.
1.7.2a Flowcell-Mounted Enclosure
For an electronics enclosure mounted directly on the flowcell, wire the transducers as follows:
WARNING! Before connecting the transducers, take them to a safe area and discharge any static buildup by
shorting the center conductor of the transducer cables to the metal shield on the cable
connector.
Note: For transducer frequencies below 2 MHz, the cable lengths given for a pair of transducers must be within 4 in.
(10 cm) of each other. If the transducer frequency exceeds 2 MHz, the cable lengths must be within 0.5 in. (1.25
cm) of each other.
1. Locate the CH1 transducer cables and connect them to the two CH1 transducers in the flowcell. Feed the free ends
of the cables through the selected conduit hole in the electronics enclosure.
2. Refer to the wiring diagram in Figure 10 on page 25 and connect the transducer cables to terminal block J3. Then,
secure the cable clamp.
Note: The RED cable leads are the SIG(+) leads and the BLACK cable leads are the RTN(-) leads.
!ATTENTION EUROPEAN CUSTOMERS!
To meet CE Mark requirements, all cables must be installed as described in Appendix A,
CE Mark Compliance.
3. For a 2-Channel XGM868i, repeat steps 1-2 to connect the CH2 transducers to terminal block J4. It is not required
that both channels of a 2-Channel unit be connected.
4. If wiring of the unit has been completed, reinstall the rear cover on the enclosure and tighten the set screw.
Note: A channel must be activated before it can begin taking measurements. See Chapter 2, Initial Setup, for
instructions.

DigitalFlow™ XGM868i Startup Guide 9
Chapter 1. Installation
1.7.2b Remote-Mounted Enclosure
For a remote mounted enclosure, refer to the wiring diagram in Figure 10 on page 25 and the remote transducer wiring
in Figure 12 on page 27, and complete the following steps:
WARNING! Before connecting the transducers, take them to a safe area and discharge any static buildup by
shorting the center conductor of the transducer cables to the metal shield on the cable
connector.
1. Using the pair of coaxial cables with BNC to BNC connectors supplied by the factory (or equivalent cables),
connect both CH1 transducers to the preamplifier.
CAUTION! As part of maintaining the FM/CSA environmental rating (TYPE 4) on the remote preamplifier,
thread sealant is required on all conduit entries.
2. If an optional lightning protector is being installed, connect it to the preamplifier as shown.
3. Using the supplied coaxial cables with BNC to flying lead connectors (or equivalent cables), route the cables
through one of the conduit holes in the electronics enclosure and connect the preamplifier to terminal block J3.
Secure the cable clamp.
Note: The RED cable leads are the SIG(+) leads and the BLACK cable leads are the RTN(-) leads.
!ATTENTION EUROPEAN CUSTOMERS!
To meet CE Mark requirements, all cables must be installed as described in Appendix A,
CE Mark Compliance.
4. For a 2-Channel XGM868i, repeat steps 3-5 to connect the CH2 transducers to terminal block J4 in the electronics
enclosure. It is not required that both channels of a 2-Channel unit be connected.
5. If wiring of the unit has been completed, reinstall the rear cover on the enclosure and tighten the set screw.
Note: A channel must be activated before it can begin taking measurements. See Chapter 2, Initial Setup, for
instructions.
Proceed to the next section to continue the initial wiring of the unit.

Chapter 1. Installation
10 DigitalFlow™ XGM868i Startup Guide
1.7.3 Wiring Std 0/4-20 mA Analog Outputs
The standard configuration of the Model XGM868i flow transmitter includes two isolated 0/4-20 mA analog outputs
(designated as outputs 1 and 2). Connections to these outputs may be made with standard twisted-pair wiring, but the
current loop impedance for these circuits must not exceed 600 ohms.
To wire the analog outputs, complete the following steps:
1. Disconnect the main power and remove the rear cover.
2. Install the required cable clamp in the chosen conduit hole on the side of the electronics enclosure.
3. Refer to Figure 10 on page 25 for the location of terminal block J1 and wire the analog outputs as shown. Secure
the cable clamp.
Note: Analog outputs 1 and 2 in the wiring diagram correspond to analog outputs A and B in Slot 0 in the XGM868i
software.
!ATTENTION EUROPEAN CUSTOMERS!
To meet CE Mark requirements, all cables must be installed as described in Appendix A,
CE Mark Compliance.
4. If wiring of the unit has been completed, reinstall the rear cover on the enclosure and tighten the set screw.
WARNING! Make sure both covers, with their o-ring seals, are installed and the set screws tightened before
applying power in a hazardous environment.
Note: Prior to use, the analog outputs must be set up and calibrated. See Chapter 1, Calibration, of the Service
Manual for detailed instructions.
Proceed to the next section to continue the initial wiring of the unit.

DigitalFlow™ XGM868i Startup Guide 11
Chapter 1. Installation
1.7.4 Wiring the Serial Port
The Model XGM868i flowmeter is equipped with an RS232 or an RS485 serial interface. An RS485 option is also
available with MODBUS capability. When the MODBUS option is present, the XGM868i may also have the standard
RS232 serial interface.
The serial port is used to transmit stored data and displayed readings to a personal computer by connecting the meter’s
serial interface to the serial port of the PC. In addition, the Model XGM868i can receive and execute remote
commands, using the Instrument Data Manager or PanaView software via this link.
For more information on serial communications refer to your EIA-RS Serial Communications manual (916-054).
Proceed to the appropriate sub-section for wiring instructions.
!ATTENTION EUROPEAN CUSTOMERS!
To meet CE Mark requirements, all cables must be installed as described in Appendix A,
CE Mark Compliance.
1.7.4a Wiring the RS232 Interface
Use the serial port to connect the Model XGM868i flow transmitter to a printer, an ANSI terminal or a personal
computer. The RS232 interface is wired as Data Terminal Equipment (DTE). Table 1 lists the standard cables available
from the factory for this purpose.
Each of the cables listed in the table above is available in several standard lengths. However, a user-supplied cable may
be used, if preferred. In either case, wire the XGM868i end of the serial cable in accordance with the pin designations
listed in Table 2 on page 12.
Table 1: GE Serial Cables
Part Number PC Connector XGM868i Connector
704-659 DB-25 Male Flying Leads (5)
704-660 DB-9 Male Flying Leads (5)
704-661 DB-25 Female Flying Leads (5)
704-662 DB-9 Female Flying Leads (5)

Chapter 1. Installation
12 DigitalFlow™ XGM868i Startup Guide
1.7.4a Wiring the RS232 Interface (cont.)
Refer to Figure 10 on page 25 to complete the following steps:
WARNING! Dangerous voltages exist within the electronics enclosure. Do not attempt to wire the unit until
the main power has been disconnected
1. Disconnect the main power and remove the rear cover.
WARNING! The XGM868i must be moved to a safe environment before removing either cover.
2. Install the required cable clamp in the chosen conduit hole on the side of the electronics enclosure.
3. Use the information in Table 2 to construct a suitable cable for connecting the Model XGM868i to the external
device. If desired, an appropriate cable may be purchased from the factory.
4. Feed the cable’s flying leads through the conduit hole and wire it to terminal block J1. Connect the other end of the
cable to the external serial device and secure the cable clamp.
After the wiring has been completed, consult the User’s Manual for the external device to configure it for use with the
XGM868i.
Table 2: RS232 Connection to DCE or DTE Device
J1
Pin #
Signal
Description
DCE
DB25
Pin #
DCE
DB9
Pin #
DTE
DB25
Pin #
DTE
DB9
Pin #
5DTR (Data Terminal
Ready) 20 4 20 4
6 CTS (Clear to Send) 4 7 5 8
7 COM (Ground) 7 5 7 5
8 RX (Receive) 2 3 3 2
9 TX (Transmit) 3 2 2 3
Table of contents
Other GE Measuring Instrument manuals

GE
GE XL Vu+ VideoProbe User manual

GE
GE DigitalFlow GF868 Owner's manual

GE
GE DigitalFlow DF868 Owner's manual

GE
GE PanaFlow HT User manual

GE
GE Hydran M2-X User manual

GE
GE AquaTrans AT600 User manual

GE
GE DB-16 User manual

GE
GE TransPort PT878GC User manual

GE
GE RHM 30 User manual

GE
GE Druck ADTS 405 User manual
Popular Measuring Instrument manuals by other brands

Endress+Hauser
Endress+Hauser Deltapilot S FMB70 operating instructions

Apogee Instruments
Apogee Instruments QUANTUM MQ-301X owner's manual

BBB
BBB MARK-B 1 user manual

Haltian
Haltian Thingsee COUNT User and installation guide

PCB Piezotronics
PCB Piezotronics Imi Sensors 622B11 Installation and operating manual

Extech Instruments
Extech Instruments 461750 user manual