SCOPE OF THIS GUIDE
This Quick Installation Guide covers the EL-FLOW® Select (including Metal Sealed) series Digital Mass Flow Meters and
Controllers for gases. These instruments are designed for laboratory and clean processing conditions and are suited to measure
and control flow ranges between 0,014…0,7 mln/min and 8…1670 ln/min, at pressures between vacuum and 400 bar.
EL-FLOW® Select instruments are equipped with a digital printed circuit board, featuring diagnostics, alarm and counter
functions, digital communication (RS232) and a PID controller. They offer great flexibility thanks to the "multibus" concept,
whereby the instruments can be equipped with an on-board fieldbus interface. The Multi Fluid / Multi Range option (to be
specified at the moment of ordering) allows the user to select flow ranges and gas types, maintaining high accuracy and
turndown ratio for measurement and control.
This Quick Installation Guide will help you start up your EL-FLOW® Select instrument in 10 steps, covering the following subjects:
1. Checking functional properties
2. Checking test pressure
3. Checking piping
4. Mounting/installing instrument
5. Leak check
6. Electrical connection
7. Analog/digital operation
8. Multifunctional switch operation
9. Purging
10. Adjusting zero point
Other applicable documents:
- Manual EL-FLOW® Select series (document no. 9.17.099)
- Custom bus & I/O configurations
·The instrument manual and hook-up diagrams can be downloaded from the EL-FLOW® Select product pages on the
Bronkhorst website: www.bronkhorst.com/products
·Other documents can be found on our general download page (www.bronkhorst.com/downloads)
Check functional properties
Before installing the instrument, check if the properties stated on the
instrument label match your requirements:
·Flow and/or pressure rate
·Media to be used in the instrument
·Upstream and downstream pressure(s)
·Operating temperature
·Valve type (N.C. - Normally Closed / N.O. - Normally Opened)
·Input and output signal
Check test pressure
Before installation, make sure that the test pressure is in accordance with
the safety factor of your application.
The test pressure is stated on the instrument with a red label. If this label is
missing, or if the test pressure is insufficient, the instrument must not be
mounted in the process line and should be returned to the factory.
Check operating conditions
EL-FLOW® instruments are suitable for use at temperatures between -10 and +70 °C. Please note that EL-FLOW®
instruments have an ingress protection rating of IP40, implying that the electronics housing and electrical
connection do not offer any protection against moist environments.