manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Brunswick
  6. •
  7. Laboratory Equipment
  8. •
  9. Brunswick BIOFLO 3000 User manual

Brunswick BIOFLO 3000 User manual

Other manuals for BIOFLO 3000

1

Other Brunswick Laboratory Equipment manuals

Brunswick Innova 4000 User manual

Brunswick

Brunswick Innova 4000 User manual

Brunswick BIOFLO 3000 User manual

Brunswick

Brunswick BIOFLO 3000 User manual

Popular Laboratory Equipment manuals by other brands

Radleys Carousel 6 Plus instructions

Radleys

Radleys Carousel 6 Plus instructions

Hemotherm 400MR Operation & technical manual

Hemotherm

Hemotherm 400MR Operation & technical manual

mrclab DVO Series Operation manual

mrclab

mrclab DVO Series Operation manual

CoolLED pE-800 quick start guide

CoolLED

CoolLED pE-800 quick start guide

Luminex Guava easyCyte Maintenance Guide

Luminex

Luminex Guava easyCyte Maintenance Guide

IKA A10 basic operating instructions

IKA

IKA A10 basic operating instructions

Belden HIRSCHMANN RPI-P1-4PoE installation manual

Belden

Belden HIRSCHMANN RPI-P1-4PoE installation manual

Koehler K1223 Series Operation and instruction manual

Koehler

Koehler K1223 Series Operation and instruction manual

Globe Scientific GCM-12 quick start guide

Globe Scientific

Globe Scientific GCM-12 quick start guide

Getinge 86 SERIES Technical manual

Getinge

Getinge 86 SERIES Technical manual

CORNING Everon 6000 user manual

CORNING

CORNING Everon 6000 user manual

Biocomp GRADIENT MASTER 108 operating manual

Biocomp

Biocomp GRADIENT MASTER 108 operating manual

Cole Parmer Stuart SB3/1 instruction manual

Cole Parmer

Cole Parmer Stuart SB3/1 instruction manual

BEVS 2501/2A user manual

BEVS

BEVS 2501/2A user manual

Qiagen DML 3000 user manual

Qiagen

Qiagen DML 3000 user manual

Queensgate NANOSCAN OP400 Quick start instructions

Queensgate

Queensgate NANOSCAN OP400 Quick start instructions

Parr Instrument 4560 Operating instructions manual

Parr Instrument

Parr Instrument 4560 Operating instructions manual

Integra DOSE IT operating instructions

Integra

Integra DOSE IT operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Great Bay Community College Document Number: 1.16.1
320 Corporate Drive Revision Number: 5
Portsmouth, NH 03801 Effective Date: 20May09
Page 1 of 20
Title: New Brunswick BioFlo 3000 Bioreactor SOP
Approvals:
Preparer: ___Kari Britt_______________________________ Date _____15May09______
Reviewer: __Bob O’Brien_____________________________ Date _____15May09______
1. Purpose: Operation of the New Brunswick BioFlo 3000 Bioreactor.
2. Scope: Applies to growing yeast or bacteria in the New Brunswick BioFlo 3000 Bioreactor.
3. Responsibilities:
3.1. It is the responsibility of the course instructor /lab assistant to ensure that this SOP is
performed as directed and to update the procedure when necessary.
3.2. It is the responsibility of the students/technicians to follow the SOP as described and to
inform the instructor about any deviations or problems that may occur while performing
the procedure.
4. References:
4.1. New Brunswick BioFlo 3000 Bioreactor instrumentation manual
4.2. BioCommand Lite Software Manual
4.3. autoclave SOP
5. Definitions: N/A
6. Precautions: N/A
7. Materials:
7.1. pH 4.0 buffer standard
7.2. pH 7.0 buffer standard
7.3. laboratory tissues
7.4. silicone tubing
7.5. PharMed tubing
7.6. clamps
7.7. aluminum foil
7.8. air filters
7.9. glass wool
7.10. probe stand with clamps
7.11. glycerol
7.12. laboratory glassware detergent
7.13. deionized water
7.14. DO electrolyte
7.15. 3M potassium chloride (KCl) solution for pH probe storage
7.16. bleach
7.17. Optional: antifoaming agent (such as Antifoam A)
7.18. personal computer to run BioCommand Lite Software and Microsoft Excel
7.19. Bioreactor
7.19.1. pH probe
7.19.2. DO probe
7.19.3. RTD temperature probe
7.19.4. harvest tube
7.19.5. sparger tube
7.19.6. thermowell tube
11
Great Bay Community College Document Number: 1.16.1
320 Corporate Drive Revision Number: 5
Portsmouth, NH 03801 Effective Date: 20May09
Page 2 of 20
Title: New Brunswick BioFlo 3000 Bioreactor SOP
7.19.7. glass vessel
7.19.8. condenser
7.19.9. feed solution flask assembly
7.19.10. sampler assembly
7.19.11. heat exchanger base
7.19.12. baffle
7.19.13. thumb screws
8. Procedure:
Note: Refer to Section 9 (Attachments) for diagrams of the bioreactor parts as needed throughout
this procedure.
8.1. Clean All Bioreactor Parts
8.1.1. Gather all bioreactor vessel parts, including the DO probe, pH probe and
calibrating buffers.
8.1.2. Clean all bioreactor vessel parts (except pH and DO probes) with a soft cloth,
warm water, and glassware detergent. Rinse thoroughly with warm tap water, and
then with several changes of deionized water.
Note: A soft brush may be used on difficult to clean parts of the bioreactor vessel.
8.1.3. Gently clean the pH and DO probes using a soft cloth and deionized water. Be
careful not to harm the sensory tips of these probes.
8.1.4. Rinse the inside of the sparger tube to remove clogs and prevent damage to the
bioreactor.
8.2. Assemble the Vessel
Note: All fittings are hand-tightened except the DO probe adapter.
8.2.1. Attach the DO, pH, and temperature cables to their appropriate outlets on the
front side of the cabinet of the BioFlo 3000.
8.2.2. If necessary, insert the plastic plugs into the holes in the glass vessel.
8.2.2.1. Make sure that the top steel ring is lifted above the holes before inserting the
plugs or the headplate will not be able to be attached.
8.2.2.2. Insert the plastic bolt through the hole from the inside of the glass vessel.
Place the rubber plug over the end of the bolt on the outside of the vessel
(with the tapered side facing into the vessel), secure the plug with a plastic
nut, and tighten the nut. Repeat this process for each of the four plugs as
needed.
8.2.3. If the plugs are already attached to the vessel, make sure that they are tightly
secured or the vessel will leak.
8.2.4. Place the glass vessel onto the heat exchanger base, making sure that the vessel is
centered within the bottom steel ring to prevent leaking.
8.2.5. Secure the glass vessel to the heat exchanger base by connecting the bottom steel
ring to the heat exchanger base using the small thumb screws. Tighten the screws
in a crosswise manner.
8.2.6. Check for leaks by pouring approximately 3L of water into the vessel and
allowing it to sit for at least 5 minutes while observing the vessel for a leak.
8.2.6.1. If a leak is observed, detach the vessel and the heat exchanger base, and
repeat steps 8.2.3. through 8.2.6., as needed, until no leaks are observed.
12
Great Bay Community College Document Number: 1.16.1
320 Corporate Drive Revision Number: 5
Portsmouth, NH 03801 Effective Date: 20May09
Page 3 of 20
Title: New Brunswick BioFlo 3000 Bioreactor SOP
8.2.7. If no leaks are observed, remove the water from the vessel by lifting the vessel
(still attached to the heat exchanger base) and pouring the water out into a
sink or container.
8.2.8. Place the baffle into the glass vessel.
8.3. Assemble the Headplate - Underside (Figure 3)
8.3.1. If necessary, attach the impellers to the agitator shaft. The lower impeller should
be flush with the base of the agitator shaft, and the upper impeller should be
approximately 6cm above the lower impeller.
8.3.2. Inspect the integrity of the O-rings on the thermowell, harvest tube and sparger
tube. Replace if worn or cracked.
8.3.3. Insert the thermowell into the thermowell port from the underside of the
headplate.
8.3.4. Adjust the height of the thermowell so that the base is approximately 2 cm above
the lower impeller.
8.3.5. Secure the thermowell in place with a nut by lowering the nut over the top of the
thermowell and hand-tightening it to the threads on the topside of the headplate.
8.3.6. Insert the harvest tube into the harvest tube port from the bottom of the headplate.
8.3.7. Adjust the height of the harvest tube so that the base is directly below the lower
impeller. Rotate the lower impeller by hand to make sure that there is enough
clearance for rotation without collision with the harvest tube.
8.3.8. Secure the harvest tube in place with a nut by lowering the nut over the top of the
harvest tube and hand-tightening it to the threads on the topside of the headplate.
8.3.9. Insert the sparger tube into the sparger tube port from the bottom of the headplate.
8.3.10. Adjust the height of the sparger tube so that the base is directly below the lower
impeller. Rotate the lower impeller by hand to make sure that there is enough
clearance for rotation without collision with the sparger tube.
8.3.11. Secure the sparger tube in place with a nut by lowering the nut over the top of the
sparger tube and hand-tightening it to the threads on the topside of the headplate.
8.4. Pour the cell culture media that will be used for the run into the vessel (refer to the
process SOP for media compostion).
8.5. Attach the Headplate to the Vessel
8.5.1. Place the headplate onto the vessel.
8.5.2. Secure the headplate to the top steel ring using the large thumb screws. Tighten
the screws in a crosswise manner.
8.6. Assemble the Headplate – Top Side (Figure 4)
8.6.1. Insert the tube of the sampler assembly into the sampler assembly port. Lower
the tube until the base is approximately half way between the upper and lower
impellers. Hand tighten the nut at the top of the tube.
8.6.2. Inspect the integrity of the O-ring at the bottom of the condenser. Replace if worn
or cracked.
8.6.3. Insert the condenser into the condenser port and secure in place by hand-
tightening the nut to the threads on the topside of the headplate. Make sure that
the barbed connectors are pointing to the left. Refer to Figure 1 for proper
orientation of the condenser.
13
Great Bay Community College Document Number: 1.16.1
320 Corporate Drive Revision Number: 5
Portsmouth, NH 03801 Effective Date: 20May09
Page 4 of 20
Title: New Brunswick BioFlo 3000 Bioreactor SOP
8.6.4. Lower the foam probe into the foam probe port as far as it will go and then hand-
tighten the nut to the threads on the topside of the headplate.
8.7. Connect Bioreactor to Cabinet
8.7.1. Place the vessel and heat exchanger base on the platform in front
of the cabinet so that the heat exchanger base fits on the three steel plugs. The
heat exchanger connectors should face the front left corner of the platform. See
Figure 1 for proper orientation of the heat exchanger.
8.7.2. Make sure that the two-way valve of the sampling assembly is closed. The valve
is closed when the position of the two-way valve handle is perpendicular to the
sample tube, as shown in Figure 2.
8.7.3. Connect the “Jacket Water Out” water line to the top heat exchanger connector
using the quick connects adaptors.
8.7.4. Connect the “Jacket Water In” water line to the bottom heat exchanger connector
using the quick connects adaptors.
8.7.5. Connect the “Condenser Water Out” water line to the top condenser connector
using the quick connects adaptors.
8.7.6. Connect the “Condenser Water In” water line to the bottom condenser connector
using the quick connects adaptors.
8.7.7. Turn on the water. The water valve is located under the lab bench where the
BioFlo 3000 is located. The pressure gauge should read between 15 and 20psi
when the water is turned on.
8.8. Calibrate the pH Probe
8.8.1. Turn on the BioFlo 3000.
8.8.2. Select fermentation (2) and press the ENTER key.
8.8.3. Check the pH probe for any damage. Replace the probe if damaged.
8.8.4. Remove the shorting cap from the top of the pH probe and connect the pH cable
to the pH probe.
8.8.5. Press the SCREEN key to select the <Calibration> screen and then press the
ENTER key.
8.8.6. Remove the protective cap from the bottom of the pH probe. Carefully pour the
3M potassium chloride storage solution into a test tube with a screw cap. Label
the tube as: Bioflo 3000 pH Probe Storage Solution, 3M potassium chloride. Store
in a test tube rack at room temperature until needed.
8.8.7. Carefully immerse the pH and RTD temperature probes into a pH 7 standard
buffer solution. Allow a few minutes for the electrode to equilibrate.
8.8.8. Use the arrow keys to select the box in the “pH” row of the “Function” column.
This box should be displaying “Read” initially.
8.8.9. Change “Read” to “Zero” by pressing the ALTER key until “Zero” appears. Then
press the ENTER key.
8.8.10. Use the arrow keys to select the box located in the “pH” row of the “Zero”
column. Enter “7.0” using the number key pad and then press the ENTER key.
Do this step even if “7.0” was already displayed when selecting the box.
8.8.11. Rinse the pH and RTD temperature probes thoroughly with deionized water. Pat
them dry with a laboratory tissue.
14
Great Bay Community College Document Number: 1.16.1
320 Corporate Drive Revision Number: 5
Portsmouth, NH 03801 Effective Date: 20May09
Page 5 of 20
Title: New Brunswick BioFlo 3000 Bioreactor SOP
8.8.12. Immerse the pH and RTD temperature probes into a pH 4 standard buffer
solution. Allow a few minutes for the electrode to equilibrate.
8.8.13. Use the arrow keys to choose the box located in the “pH” row of the “Span”
column. Enter “4.0” using the number keypad and then press the ENTER key. Do
this step even if “4.0” was already displayed when selecting the box.
8.8.14. Rinse the pH and RTD temperature probes with deionized water and pat dry with
a laboratory tissue.
8.8.15. Disconnect the pH cable from the pH probe, and replace the shorting cap.
8.8.16. Inspect the integrity of the internal O-ring of pH probe port. Replace if worn or
cracked.
8.8.17. Apply a small amount of deionized water to the pH probe and then insert it into
the pH probe port.
8.8.18. Ensure that the pH probe electrode is not touching the baffle.
8.9. Install Dissolved Oxygen Probe
8.9.1. Remove the protective cap from the bottom of the DO probe and inspect the
screen. Replace the screen if damaged.
8.9.2. Remove the bottom housing that encases the DO probe tip.
8.9.3. Inspect the integrity of the O-ring and replace if worn or cracked.
8.9.4. Replenish the DO electrolyte in the bottom housing.
8.9.5. Replace the bottom housing.
8.9.6. Carefully insert the DO probe into the DO port of the headplate. Tighten the
adaptor with a wrench.
8.9.7. Ensure that the DO probe is not touching the baffle.
8.10. Attach Tubing
8.10.1. Attach silicone tubing to the harvest port, bend the tube in half and secure with a
clamp. Wrap glass wool and aluminum foil around the exposed end.
8.10.2. Place a piece of PharmMed tubing (approximately 1.5 meters in length) onto each
addition port, bend the tubes in half and secure with clamps. Wrap glass wool and
aluminum foil around the exposed ends.
8.10.3. Place silicone tubing on the top outlet of the condenser and connect an air filter.
Wrap glass wool and aluminum foil around the exposed end of the filter.
8.10.4. Place silicone tubing on the top of the sparger tube and connect an air filter.
Bend the tube in half and secure with a clamp. Wrap glass wool and aluminum
foil around the exposed end of the tubing.
8.10.5. Remove the rubber bulb from the sampler assembly. Insert glass wool into the
bulb and reattach it to the sampler assembly.
8.11. Autoclave the vessel
8.11.1. Turn off the BioFlo 3000.
8.11.2. Turn off the water valve (under the lab bench).
8.11.3. Disconnect the water lines to the heat exchanger.
8.11.4. Disconnect the water lines to the condenser.
8.11.5. Double check all tubes and ports to ensure that the vessel is completely sealed.
8.11.6. Autoclave the entire vessel and heat exchanger assembly for at least 30 minutes
per autoclave SOP.
15
Great Bay Community College Document Number: 1.16.1
320 Corporate Drive Revision Number: 5
Portsmouth, NH 03801 Effective Date: 20May09
Page 6 of 20
Title: New Brunswick BioFlo 3000 Bioreactor SOP
8.12. Prepare for Operation
8.12.1. Remove the clamp from the tubing attached to the top outlet of the condenser.
8.12.2. Place the vessel and heat exchanger base on the platform in front
of the cabinet so that it fits on the three steel plugs. The heat exchanger
connectors should face the front left corner of the platform. See Figure 1 for
proper orientation of the heat exchanger base.
8.12.3. It is advantageous to begin polarization of the DO probe at this point in the
procedure. Follow directions in step 8.14.1 to do this, and then return to step
8.12.4. to connect the water lines.
8.12.4. Connect the “Jacket Water Out” water line to the top heat exchanger connector
using the quick connects adaptors.
8.12.5. Connect the “Jacket Water In” water line to the bottom heat exchanger connector
using the quick connects adaptors.
8.12.6. Connect the “Condenser Water Out” water line to the top condenser connector
using the quick connects adaptors.
8.12.7. Connect the “Condenser Water In” water line to the bottom condenser connector
using the quick connects adaptors.
8.12.8. Turn on the water. The water valve is located under the lab bench. The pressure
gauge should read between 15 and 20psi.
8.12.9. Add approximately 1mL of glycerol to the thermowell and insert the RTD
temperature probe into the thermowell.
8.12.10. Remove the shorting cap from the top of the pH probe and connect the pH cable
to the pH probe.
8.12.11. Connect the air line from the sparger tube to the sparger gas port located on the
front of the cabinet, near the top.
8.12.12. Turn on the air supply.
8.12.13. The air pressure gauge should not exceed 10psi (5psi is the ideal pressure level).
8.12.14. Remove the agitator cover from the top of the headplate and place the agitator
motor securely in the agitator shaft.
8.12.15. Turn on the BioFlo 3000.
8.12.16. Select Fermentation mode (2) and press the ENTER key.
8.12.17. If the Master screen does not automatically appear use the SCREEN key to select
the <Master> screen and then press the ENTER key.
8.12.18. Set the temperature control mode to “Prime” by using the arrow keys to select
the box in the “Control” row of the “Temp 1” column. Press the ALTER key
until “Prime” is displayed and then press the ENTER key.
8.12.19. Wait 5 minutes before setting the working temperature. Refer to the process
SOP for the desired working temperature.
8.12.19.1. Set the working temperature by using the arrow keys to select the box in
the “Set” row of the “Temp 1” column. Use the number keypad to enter
the desired temperature and then press the ENTER key. Do this step even
if the desired temperature was already displayed when selecting the box.
16
Great Bay Community College Document Number: 1.16.1
320 Corporate Drive Revision Number: 5
Portsmouth, NH 03801 Effective Date: 20May09
Page 7 of 20
Title: New Brunswick BioFlo 3000 Bioreactor SOP
8.12.20. Set the control mode to P-I-D by selecting the box in the “Control” row of the
“Temp 1” column. Press ALTER until “P-I-D” appears and then press the
ENTER key.
8.12.21. Set the agitation loop to P-I-D control by selecting the box in the “Control” row
of the “Agit 1” column. Press the ALTER key until “P-I-D” appears and then
press the ENTER key.
8.12.22. Set the agitation setpoint to 100rpm by selecting the box in the “Set” row of the
“Agit 1” column. Use the number keypad to enter 100 and then press the ENTER
key.
8.12.23. Set up the feed solution flasks on the appropriate feed pumps, as indicated by the
process SOP. Acidic or basic solutions should be set up on Feed 1. Refer to the
process SOP for indication of an acidic or basic solution.
8.12.23.1. Unclamp the tube attached to an addition port and remove the glass wool
and aluminum foil.
8.12.23.2. Thread the free end of the tube through the bottom of the feed pump,
counter clockwise around the pump, and out the top.
8.12.23.3. Place the free end of the tube on the top of the glass pipet of the feed
solution flask.
Note: The pump moves in a clockwise direction. Therefore the solution
will flow from right to left in the tube.
8.12.24. Repeat step 8.12.23 for additional feed solution flasks, if needed.
8.12.25. Set the pH setpoint to the desired value. Refer to the process SOP for the desired
pH.
8.12.25.1. Set the pH setpoint by selecting the box in the “Set” row of the “pH 1”
column. Use the number key pad to enter the desired pH and then press the
ENTER key. Do this step even if the desired pH was already displayed
when selecting the box.
8.12.26. Set the pH loop to P-I-D control by selecting the box in the “Control” row of the
“pH 1” column. Press the ALTER key until “P-I-D” appears and then press the
ENTER key.
8.12.27. Designate Feed 1 as acid or base. Refer to the process SOP to determine if the
Feed 1 solution is acidic or basic.
8.12.27.1. Designate Feed 1 by using the arrow keys to select the box in the “Loop
Name” row of the “pH 1” column. This box should be displaying the
column heading “pH 1”. Press the ALTER key until the “Feed 1” column
heading appears then press the ENTER key.
8.12.27.2. Use the arrow keys to select the box in the “Control” row of the “Feed 1”
column. Use the ALTER key to select either “Base” or “Acid” and then
press the ENTER key.
8.12.27.3. Set the setpoint to 100 by choosing the box in the “Set” row of the “Feed
1” column. Use the number keypad to enter “100” and then press the
ENTER key.
8.12.28. Return to the pH loop display by selecting the “Feed 1” column heading. Press
the ALTER key until “pH 1” appears and then press the ENTER key.
17
Great Bay Community College Document Number: 1.16.1
320 Corporate Drive Revision Number: 5
Portsmouth, NH 03801 Effective Date: 20May09
Page 8 of 20
Title: New Brunswick BioFlo 3000 Bioreactor SOP
8.12.29. Wait for the pH value to reach its setpoint before proceeding.
8.13. Activate additional feed loops at the appropriate time as indicated by the process
SOP.
8.13.1. Activate a feed loop by selecting the box in the “Loop Name” row of the “pH 1”
column. This box should be displaying the column heading “pH 1”. Press the
ALTER key until the “Feed 2” column heading appears (or the appropriate feed
column heading if not “Feed 2”).
8.13.2. Enter the appropriate setpoint for the feed solution, as indicated by the process
SOP, by choosing the box in the “Set” row of the “Feed 2” column. Use the
number keypad to enter the appropriate setpoint value and then press the ENTER
key.
8.13.3. In the “Control” row of the “Feed 2” column press the ALTER key until “on” is
displayed, then press ENTER.
8.14. Calibrate the Dissolved Oxygen Probe
Note: The DO probe should not be calibrated until the desired working temperature has been
reached.
8.14.1. Polarize the DO probe for 30 minutes prior to performing calibration.
8.14.1.1. If necessary, connect the DO cable to the DO electrical outlet on the front of
the cabinet.
8.14.1.2. Remove the protective cap from the top of the DO probe and connect the
DO Cable to the DO probe.
8.14.1.3. If necessary, turn on the BioFlo 3000.
8.14.1.4. Allow the probe to sit for at least 30 minutes.
8.14.1.5. Verify that the DO level is above 0. If the DO display reads 0 or below there
may be a problem with the DO probe. Contact the lab manager if this
occurs.
8.14.2. Use the SCREEN key to choose the <Calibration> screen and then press the
ENTER key.
8.14.3. Use the arrow keys to choose the box in the “DO” row of the “Function” column.
This box should be displaying “Read”.
8.14.4. Change “Read” to “Zero” by pressing the ALTER key until “Zero” appears and
then pressing the ENTER key.
8.14.5. Use the arrow keys to choose the box in the “DO” row of the “Zero” column.
Enter “0.0” using the number keypad and then press the ENTER key.
8.14.6. Use the SCREEN key to choose the <Master> screen and then press the ENTER
key.
8.14.7. Use the arrow keys to choose the box in the “Set” row of the “Agit 1” column.
8.14.8. Enter “1000” (for 1000rpm) using the number keypad and then press the ENTER
key. This will turn on the agitation motor.
8.14.9. Press the SCREEN key and then use the arrow keys to select the <Gases> screen
and then press the ENTER key.
8.14.10. Set the gases mode to “Manual” by selecting the box in the “Mode” row (it is the
only box available in the “Mode” row). Press the ALTER key until “Manual” is
displayed, then press the ENTER key.
18
Great Bay Community College Document Number: 1.16.1
320 Corporate Drive Revision Number: 5
Portsmouth, NH 03801 Effective Date: 20May09
Page 9 of 20
Title: New Brunswick BioFlo 3000 Bioreactor SOP
8.14.11. Allow ten to thirty minutes for the vessel to equilibrate.
8.14.12. Press the SCREEN key and then use the arrow keys to select the <Calibration>
screen and then press the ENTER key.
8.14.13. Use the arrow keys to choose the box in the “DO” row of the “Function” column.
This box should be displaying “Read”.
8.14.14. Change “Read” to span by pressing the ALTER key until “Span” is displayed
and then press the ENTER key.
8.14.15. Choose the box in the “DO” row of the “Span” column. Enter “100.0” using the
number keypad and then press the ENTER key.
8.15. There are three possible control modes for dissolved oxygen. Refer to the process
SOP for the desired mode.
8.15.1. Controlling by Agitation Only
8.15.1.1. Use the SCREEN key to choose the <Master> screen and then press the
ENTER key.
8.15.1.2. Set the DO loop to P-I-D by selecting the box in the “Control” row of the
“DO2 1” column. Press the ALTER key until “P-I-D” appears and then
press the ENTER key.
8.15.1.3. Temporarily set the agitation loop to P-I-D by selecting the box in the
“Control” row of the “Agit 1” column. Press the ALTER key until “P-I-D”
appears and then press the ENTER key.
8.15.1.4. To set the minimum rpm select the box in the “Set” row of the “Agit 1”
column. Use the number keypad to enter the desired minimum rpm, as
indicated by the process SOP and then press the ENTER key.
8.15.1.5. Set the agitation loop to DO control mode by selecting the box in the
“Control” row of the “Agit 1” column. Press the ALTER key until “DO”
appears and then press the ENTER key.
8.15.1.6. To set the maximum rpm select the box in the “Set” row of the “Agit 1”
column. Use the number keypad to enter the desired maximum rpm, as
indicated by the process SOP and then press the ENTER key.
Note: The controller will automatically adjust the rpm, therefore after the
maximum value is entered the displayed set value may change immediately
afterward.
8.15.1.7. To input the DO setpoint select the box in the “Set” row of the “DO2 1”
column. Use the number keypad to enter the desired setpoint, as indicated
by the process SOP and then press the ENTER key.
8.15.2. Controlling by % Oxygen Sparged
8.15.2.1. Use the SCREEN key to choose the <Master> screen and then press the
ENTER key.
8.15.2.2. Set the DO loop to P-I-D by selecting the box in the “Control” row of the
“DO2 1” column. Press the ALTER key until “P-I-D” appears and then
press the ENTER key.
8.15.2.3. Press the SCREEN key and then use the arrow keys to choose the <Gases>
screen and then press the ENTER key.
19
Great Bay Community College Document Number: 1.16.1
320 Corporate Drive Revision Number: 5
Portsmouth, NH 03801 Effective Date: 20May09
Page 10 of 20
Title: New Brunswick BioFlo 3000 Bioreactor SOP
8.15.2.4. Set the gases mode to DO control by selecting the box in the “Mode” row (it
is the only box available in the “Mode” row). Press the ALTER key until
“DO” is displayed and then press the ENTER key.
8.15.2.5. Use the SCREEN key to choose the <Master> screen and then press the
ENTER key.
8.15.2.6. To input the DO setpoint select the box in the “Set” row of the “DO2 1”
column. Use the number keypad to enter the desired setpoint, as indicated
by the process SOP and then press the ENTER key.
8.15.3. Controlling by Agitation and % Oxygen
Note: In this mode agitation will be increased first until maximum rpm is reached and
more oxygen is required, then control by % oxygen will begin.
8.15.3.1. Use the SCREEN key to choose the <Master> screen and then press the
ENTER key.
8.15.3.2. Set the DO loop to P-I-D by selecting the box in the “Control” row of the
“DO2 1” column. Press the ALTER key until “P-I-D” appears and then
press the ENTER key.
8.15.3.3. Temporarily set the agitation loop to P-I-D by selecting the box in the
“Control” row of the “Agit 1” column. Press the ALTER key until “P-I-D”
appears and then press the ENTER key.
8.15.3.4. To set the minimum rpm select the box in the “Set” row of the “Agit 1”
column. Use the number keypad to enter the desired minimum rpm, as
indicated by the process SOP and then press the ENTER key.
8.15.3.5. Set the agitation loop to DO control mode by selecting the box in the
“Control” row of the “Agit 1” column. Press the ALTER key until “DO”
appears and then press the ENTER key.
8.15.3.6. To set the maximum rpm select the box in the “Set” row of the “Agit 1”
column. Use the number keypad to enter the desired maximum rpm, as
indicated by the process SOP and then press the ENTER key.
Note: The controller will automatically adjust the rpm, therefore after the
maximum value is entered the displayed set value may change immediately
afterward
8.15.3.7. Press the SCREEN key and then use the arrow keys to choose the <Gases>
screen and then press the ENTER key.
8.15.3.8. Set the gases mode to agitation and % oxygen control by selecting the box
in the “Mode” row (it is the only box available in the “Mode” row). Press
the ALTER key until “Ag02DO” is displayed and then press the ENTER
key.
8.15.3.9. Use the SCREEN key to choose the <Master> screen and then press the
ENTER key.
8.15.3.10. Navigate to the air monitor loop by selecting the column heading “Agit 1”
in the “Loop Name” row. Press the ALTER key until “Air 1” appears.
8.15.3.11. Input the desired air pressure by selecting the box in the “Set” row of the
“Air 1” column. Use the number keypad to enter the desired air flow rate
(usually between 1.5 and 2 L/min.) and then press the ENTER key.
20