Camille Bauer SINEAX AM2000 User manual

PM 1000162 000 14 Device handbook SINEAX AMx000 2/118
Legal information
Warning notices
In this document warning notices are used, which you have to observe to ensure personal safety and to prevent
damage to property. Depending on the degree of danger the following symbols are used:
If the warning notice is not followed death or severe personal injury
will result.
If the warning notice is not followed damage to property or severe
personal injury may result.
If the warning notice is not followed the device may be damaged or
may not fulfill the expected functionality.
Qualified personnel
The product described in this document may be handled by personnel only, which is qualified for the respective
task. Qualified personnel have the training and experience to identify risks and potential hazards when working
with the product. Qualified personnel are also able to understand and follow the given safety and warning
notices.
Intended use
The product described in this document may be used only for the application specified. The maximum electrical
supply data and ambient conditions specified in the technical data section must be adhered. For the perfect and
safe operation of the device proper transport and storage as well as professional assembly, installation,
handling and maintenance are required.
Disclaimer of liability
The content of this document has been reviewed to ensure correctness. Nevertheless it may contain errors or
inconsistencies and we cannot guarantee completeness and correctness. This is especially true for different
language versions of this document. This document is regularly reviewed and updated. Necessary corrections
will be included in subsequent version and are available via our webpage https://www.camillebauer.com.
Feedback
If you detect errors in this document or if there is necessary information missing, please inform us via e-mail to:
customer-support@camillebauer.com

PM 1000162 000 14 Device handbook SINEAX AMx000 3/118
Contents
1. Introduction ....................................................................................................................................................5
1.1 Purpose of this document..........................................................................................................................5
1.2 Scope of supply .........................................................................................................................................5
1.3 Further documents.....................................................................................................................................5
2. Safety notes....................................................................................................................................................6
3. Device overview .............................................................................................................................................6
3.1 Brief description .........................................................................................................................................6
3.2 Device overview.........................................................................................................................................7
3.3 Available measurement data .....................................................................................................................8
4. Mechanical mounting ....................................................................................................................................9
4.1 Panel cutout...............................................................................................................................................9
4.2 Panel mounting of the device ....................................................................................................................9
4.3 Hat rail mounting AM1000 .......................................................................................................................10
4.4 Demounting of the device........................................................................................................................10
5. Electrical connections .................................................................................................................................11
5.1 General safety notes................................................................................................................................11
5.2 Terminal assignments of the I/O extensions of AM2000/3000................................................................12
5.3 Possible cross sections and tightening torques ......................................................................................13
5.4 Inputs .......................................................................................................................................................14
5.5 Power supply ...........................................................................................................................................26
5.6 Relays ......................................................................................................................................................26
5.7 Digital inputs ............................................................................................................................................26
5.8 Digital outputs ..........................................................................................................................................28
5.9 Analog outputs.........................................................................................................................................29
5.10 Modbus interface RS485 .........................................................................................................................29
5.11 Fault current detection .............................................................................................................................31
5.12 Temperature inputs..................................................................................................................................33
5.13 Uninterruptible power supply (UPS), AM3000 only.................................................................................33
5.14 GPS time synchronization........................................................................................................................34
6. Commissioning ............................................................................................................................................36
6.1 Parametrization of the device functionality ..............................................................................................37
6.2 Operating LED of AM1000 without display..............................................................................................37
6.3 Installation check .....................................................................................................................................37
6.4 Ethernet installation .................................................................................................................................39
6.4.1 Settings ............................................................................................................................................39
6.4.2 Connection of the standard interface ...............................................................................................42
6.4.3 Connection of the IEC61850 interface .............................................................................................42
6.4.4 Connection of the PROFINET interface...........................................................................................42
6.4.5 MAC addresses................................................................................................................................43
6.4.6 Resetting the communication settings of an AM1000 without display .............................................44
6.5 Communication tests ...............................................................................................................................44
6.6 IEC 61850 interface .................................................................................................................................44
6.7 PROFINET IO interface...........................................................................................................................45
6.7.1 General stations description file (GSD)............................................................................................45
6.7.2 Parameterization of the device.........................................................................................................46
6.7.3 Validity of measurements.................................................................................................................47
6.7.4 PROFINET state ..............................................................................................................................48
6.8 Simulation of analog / digital outputs.......................................................................................................49
6.9 Security system .......................................................................................................................................49
6.9.1 RBAC management .........................................................................................................................50
6.9.2 User log in / out via website .............................................................................................................53
6.9.3 User log in / out via local display......................................................................................................54
6.9.4 Simple password protection .............................................................................................................55
6.9.5 Whitelisting clients............................................................................................................................56
6.9.6 Secure communication using https ..................................................................................................56

PM 1000162 000 14 Device handbook SINEAX AMx000 4/118
6.9.7 Audit log (SYSLOG) ........................................................................................................................ 57
7. Operating the device................................................................................................................................... 59
7.1 Operating elements ................................................................................................................................ 59
7.2 Selecting the information to display........................................................................................................ 59
7.3 Measurement displays and used symbols.............................................................................................. 60
7.4 Resetting measurement data ................................................................................................................. 62
7.5 Configuration .......................................................................................................................................... 62
7.5.1 Configuration at the device ............................................................................................................... 62
7.5.2 Configuration via web browser.......................................................................................................... 64
7.6 Alarming.................................................................................................................................................. 66
7.6.1 Limit values on base quantities ....................................................................................................... 66
7.6.2 Monitoring fault-currents ................................................................................................................. 67
7.6.3 Temperature monitoring .................................................................................................................. 68
7.6.4 Monitoring functions ........................................................................................................................ 69
7.6.5 Summary alarm ............................................................................................................................... 70
7.7 Data recording ........................................................................................................................................ 71
7.7.1 Periodic data ................................................................................................................................... 71
7.7.2 User-defined events ........................................................................................................................ 74
7.7.3 Disturbance recorder (AM1000 / AM3000 only).............................................................................. 76
7.7.4 Micro SD card (AM2000/3000 only) ................................................................................................ 79
7.8 Measurement information in file format .................................................................................................. 80
7.8.1 Creating periodic file data ............................................................................................................... 80
7.8.2 Accessing file information via webpage .......................................................................................... 81
7.8.3 Periodical sending to a SFTP Server .............................................................................................. 82
7.9 Timeouts ................................................................................................................................................. 82
8. Service, maintenance and disposal .......................................................................................................... 83
8.1 Calibration and new adjustment ............................................................................................................. 83
8.2 Cleaning.................................................................................................................................................. 83
8.3 Battery..................................................................................................................................................... 83
8.4 Disposal .................................................................................................................................................. 83
9. Technical data ............................................................................................................................................. 84
10. Dimensional drawings ................................................................................................................................ 91
Annex .................................................................................................................................................................. 93
ADescription of measured quantities .......................................................................................................... 93
A1 Basic measurements .............................................................................................................................. 93
A2 Harmonic analysis .................................................................................................................................. 97
A3 System imbalance .................................................................................................................................. 98
A4 Mean values and trend ........................................................................................................................... 99
A5 Meters ................................................................................................................................................... 100
BDisplay matrices........................................................................................................................................ 101
B0 Used abbreviations for the measurements........................................................................................... 101
B1 Display matrices for single phase system ............................................................................................ 106
B2 Display matrices for split-phase (two-phase) systems ......................................................................... 107
B3 Display matrices for 3-wire system, balanced load .............................................................................. 108
B4 Display matrices for 3-wire system, balanced load, phase shift........................................................... 109
B5 Display matrices for 3-wire systems, unbalanced load ........................................................................ 110
B6 Display matrices for 3-wire systems, unbalanced load, Aron............................................................... 111
B7 Display matrices for 4-wire system, balanced load .............................................................................. 112
B8 Display matrices for 4-wire systems, unbalanced load ........................................................................ 113
B9 Display matrices for 4-wire system, unbalanced load, Open-Y............................................................ 114
B10 Common display matrices .................................................................................................................... 115
CLogic functions.......................................................................................................................................... 116
DFCC statement ........................................................................................................................................... 117
INDEX ................................................................................................................................................................ 118

PM 1000162 000 14 Device handbook SINEAX AMx000 5/118
1. Introduction
1.1 Purpose of this document
This document describes the universal measurement devices for heavy-current quantities of the SINEAX
AM series. It is intended to be used by:
•Installation personnel and commissioning engineers
•Service and maintenance personnel
•Planners
Scope
This handbook is valid for all hardware versions of the AMx000. Some of the functions described in this
document are available only, if the necessary optional components are included in the device.
Required knowledge
A general knowledge in the field of electrical engineering is required. For assembly and installation of the
device knowledge of applicable national safety regulations and installation standard is required.
1.2 Scope of supply
•Measurement device
•Safety instructions (multiple languages)
•Mounting set: 2 mounting clamps
•Only AM3000: Battery pack (optional, if device with UPS)
1.3 Further documents
The following documents are provided electronically via our homepage https://www.camillebauer.com/ :
•Safety instructions SINEAX AM2000 / SINEAX AM3000
•Safety instructions SINEAX AM1000
•Data sheet SINEAX AM1000/AM2000/AM3000
•Modbus basics: General description of the communication protocol
•Modbus interface SINEAX AMx000: Register description of Modbus communication
•IEC61850 interface SINEAX AMx000/DM5000, LINAX PQx000, CENTRAX CUx000
•Camille Bauer certificate for encrypted HTTPS communication

PM 1000162 000 14 Device handbook SINEAX AMx000 6/118
2. Safety notes
Device may only be disposed in a professional manner!
The installation and commissioning should only be carried out by trained personnel.
Check the following points before commissioning:
– that the maximum values for all the connections are not exceeded, see "Technical data"
section,
– that the connection wires are not damaged, and that they are not live during wiring,
– that the power flow direction and the phase rotation are correct.
The instrument must be taken out of service if safe operation is no longer possible (e.g. visible
damage). In this case, all the connections must be switched off. The instrument must be
returned to the factory or to an authorized service dealer.
It is forbidden to open the housing and to make modifications to the instrument. The instrument
is not equipped with an integrated circuit breaker. During installation check that a labeled switch
is installed and that it can easily be reached by the operators.
Unauthorized repair or alteration of the unit invalidates the warranty.
3. Device overview
3.1 Brief description
The SINEAX AM-series devices are compact instruments to measure and monitor in heavy current grids.
They excel in display quality and intuitive operation. The devices provide a wide range of functionalities
which may even be extended by optional components. The nameplate on the device gives further details
about the present version. They are connected to the process environment by communication interfaces,
via digital I/Os, analog outputs or relays.
The devices have been designed for universal use in industrial plants, building automation or in energy
distribution. Nominal voltages of up to 690 V and measurement category CATIII can be directly connected
in low voltage systems. The universal measuring system permits the direct use of the devices in any type
of grid, from single-phase mains through to 4-wire unbalanced load systems.
A comprehensive security concept protects the devices from unauthorized access, eavesdropping of
communication or data manipulations. Implemented security mechanisms are Role-Based Access Control
(RBAC), encrypted data transmission via HTTPS, logging of all activities in an Audit log with Syslog
protocol support, a client whitelist for limiting computers with access authorization and digitally signed
firmware files for secure updates.
The AM series devices may be completely adapted to the requirements on site via TFT display. Versions
with an Ethernet interface permit webpage configuration. No special software is required for device
configuration or measurement display.

PM 1000162 000 14 Device handbook SINEAX AMx000 7/118
3.2 Device overview

PM 1000162 000 14 Device handbook SINEAX AMx000 8/118
3.3 Available measurement data
The SINEAX AM3000 provides measurements in the following subcategories:
a) Instantaneous values: Present TRMS values and associated min/max values
b) Energy: Power mean-values with trend and history as well as energy meters. With the data logger
option “periodic data” mean-value progressions (load profiles) and periodic meter readings are
available as well.
c) Harmonics: Total harmonic distortion THD/TDD, individual harmonics and their maximum values
d) Phasor diagram: Overview of all current and voltage phasors and phase sequence check
e) Waveform of current and voltage inputs
f) Events: State list of monitored alarms. With the data logger option also chronological lists of events
and alarms as well as operator events are available.

PM 1000162 000 14 Device handbook SINEAX AMx000 9/118
4. Mechanical mounting
►The AM3000 is designed for panel mounting
Please ensure that the operating temperature limits are not exceeded when determining the
place of mounting (place of measurement).
By installing, the device becomes part of an electrical power installation that must be
designed, operated and maintained in accordance with country-specific regulations so that
the installation is safe and provides prevention against fire and explosion as far as possible.
It is the task of this installation to ensure that dangerous connections of the device cannot
be touched during operation and that the spread of flames, heat and smoke from the interior
is prevented. This may be done by providing an enclosure (e.g. case, cabinet) or using a
room accessible to qualified personal only and compliant with local fire safety standards.
4.1 Panel cutout
AM2000 / AM3000 AM1000
Dimensional drawings: See chapter 10
4.2 Panel mounting of the device
The device is suitable for panel widths up to 8mm (AM2000/AM3000) resp. 10mm (AM1000), below
shown for an AM2000/3000.
a) Slide the device into the cutout from
the outside. Orientation as shown.
b) From the side slide in the mounting
clamps into the intended openings and
pull them back about 2 mm
c) Tighten the fixation screws until the
device is tightly fixed with the panel

PM 1000162 000 14 Device handbook SINEAX AMx000 10/118
4.3 Hat rail mounting AM1000
The AM1000 can be clipped onto a top-hat rail according to EN 60715, orientation as shown below.
Dimensional drawing
AM1000:
See section 10
The AM1000 with display for hat-rail
mounting can also be mounted that the front of
the device protrudes through a cut-out in the
enclosure. This way the operating buttons and
the display become accessible. With centric
mounting using the below maximum cut-out a
gap between enclosure and device results,
which does not exceed 2.5mm on each side.
4.4 Demounting of the device
The demounting of the device may be performed only if all connected wires are out of service. Remove all
plug-in terminals and all connections of the current and voltage inputs. Pay attention to the fact, that
current transformers must be shorten before removing the current connections to the device. Then
demount the device in the opposite order of mounting.

PM 1000162 000 14 Device handbook SINEAX AMx000 11/118
5. Electrical connections
Ensure under all circumstances that the leads are free of potential
when connecting them!
5.1 General safety notes
Please observe that the data on the type plate must be adhered to!
The national provisions have to be observed in the installation and material selection of electric lines,
e.g. in Germany VDE 0100 “Erection of power installations with nominal voltages up to 1000 V“!
Nameplate of a AM3000 with
•Ethernet interface
•Modbus/RTU interface
•4 relay outputs
•4 analog outputs
•Data logger
Nameplate of a AM1000 with
•Ethernet interface
•Modbus/RTU interface
•2 relay outputs
•Data logger
Hint: For the device version with
display for hat-rail mounting the
nameplate is divided into three
plates

PM 1000162 000 14 Device handbook SINEAX AMx000 12/118
Symbol Meaning
Device may only be disposed of in a professional manner!
Double insulation, device of protection class 2
CE conformity mark. The device fulfills the requirements of the applicable EU
directives.
Products with this mark comply with both the Canadian (CSA) and the American
(UL) requirements.
Caution! General hazard point. Read the operating instructions.
General symbol: Input
General symbol: Output
CAT III Measurement category CAT III
5.2 Terminal assignments of the I/O extensions of AM2000/3000
Function Option 1 Option 2 Option 3 Option 4
2 relay outputs 1.1: 51,52,53
1.2: 55,56,57
2.1: 61,62,63
2.2: 65,66,67
4.1: 31,32,33
4.2: 35,36,37
2 analog outputs 1.1: 56(+), 57(-)
1.2: 55(+), 57(-)
2.1: 66(+), 67(-)
2.2: 65(+), 67(-)
3.1: 46(+), 47(-)
3.2: 45(+), 47(-)
4.1: 36(+), 37(-)
4.2: 35(+), 37(-)
4 analog outputs
1.1: 56(+), 57(-)
1.2: 55(+), 57(-)
1.3: 52(+), 53(-)
1.4: 51(+), 53(-)
2.1: 66(+), 67(-)
2.2: 65(+), 67(-)
2.3: 62(+), 63(-)
2.4: 61(+), 63(-)
3.1: 46(+), 47(-)
3.2: 45(+), 47(-)
3.3: 42(+), 43(-)
3.4: 41(+), 43(-)
4.1: 36(+), 37(-)
4.2: 35(+), 37(-)
4.3: 32(+), 33(-)
4.4: 31(+), 33(-)
4 digital inputs (active)
1.1: 51(-), 53(+)
1.2: 52(-), 53(+)
1.3: 55(-), 57(+)
1.4: 56(-), 57(+)
2.1: 61(-), 63(+)
2.2: 62(-), 63(+)
2.3: 65(-), 67(+)
2.4: 66(-), 67(+)
3.1: 41(-), 43(+)
3.2: 42(-), 43(+)
3.3: 45(-), 47(+)
3.4: 46(-), 47(+)
4.1: 31(-), 33(+)
4.2: 32(-), 33(+)
4.3: 35(-), 37(+)
4.4: 36(-), 37(+)
4 digital inputs (passive)
1.1: 51(+), 53(-)
1.2: 52(+), 53(-)
1.3: 55(+), 57(-)
1.4: 56(+), 57(-)
2.1: 61(+), 63(-)
2.2: 62(+), 63(-)
2.3: 65(+), 67(-)
2.4: 66(+), 67(-)
3.1: 41(+), 43(-)
3.2: 42(+), 43(-)
3.3: 45(+), 47(-)
3.4: 46(+), 47(-)
4.1: 31(+), 33(-)
4.2: 32(+), 33(-)
4.3: 35(+), 37(-)
4.4: 36(+), 37(-)
2 temperature inputs 1.1: 52,53
1.2: 56,57
2.1: 62,63
2.2: 66,67
3.1: 42,43
3.2: 46,47
4.1: 32,33
4.2: 36,37

PM 1000162 000 14 Device handbook SINEAX AMx000 13/118
5.3 Possible cross sections and tightening torques
Inputs L1(2), L2(5), L3(8), N(11), PE(16), I1(1-3), I2(4-6), I3(7-9), IN(10-12), power supply (13-14)
Single wire •1 x 0,5...6.0mm2or 2 x 0,5...2.5mm2
•1 x 20 AWG…9 AWG or 2 x 20 AWG…14 AWG
Multiwire with end splices •1 x 0,5...4.0mm2or 2 x 0,5...2.5mm2
•1 x 20 AWG…11 AWG or 2 x 20 AWG…14 AWG
Tightening torque •0.5…0.6Nm
•4.42…5.31 lbf in
I/O's, relays, RS485 connector (A, B, C/X)
Single wire •1 x 0.5 ... 2.5mm2or 2 x 0.5 ... 1.0mm2
•1 x 20 AWG…14 AWG or 2 x 20 AWG…17 AWG
Multiwire with end splices •1 x 0.5 ... 2.5mm2or 2 x 0.5 ... 1.5mm2
•1 x 20 AWG…14 AWG or 2 x 20 AWG…16 AWG
Tightening torque •0.5…0.6Nm
•4.42…5.31 lbf in

PM 1000162 000 14 Device handbook SINEAX AMx000 14/118
5.4 Inputs
All voltage measurement inputs must originate at circuit breakers or fuses rated 5 Amps or
less. This does not apply to the neutral connector. You have to provide a method for
manually removing power from the device, such as a clearly labeled circuit breaker or a
fused disconnect switch in accordance with IEC 60947-2 or IEC 60947-3.
When using voltage transformers you have to ensure that their secondary connections
never will be short-circuited.
No fuse may be connected upstream of the current measurement inputs!
When using current transformers their secondary connectors must be short-circuited
during installation and before removing the device. Never open the secondary circuit under
load.
Further hints
•The connection of the inputs depends on the configured system (connection type).
•AM3000 only: In the connection diagrams on the next pages conventional voltage transformers are
used. If a voltage transformer with extra windings for measuring the homopolar voltage is applied,
connections should be as shown below.
In order for the homopolar voltage to be
measured, the item „Measure homopolar
voltage“ must be set to „Yes“ in the settings of
the measurement. This item is only available for
3-wire system types.
L1 L2 L3 NPE
16112 5 8
L1 L2 L3
x
u
X
U
X
U
X
U
x
u
x
u
5 A
(UL listed)
e
n
e
n
e
n

PM 1000162 000 14 Device handbook SINEAX AMx000 15/118
Single-phase AC mains
AM2000 / AM3000 AM1000
Direct connection
Maximum permissible rated voltage 300V to ground!
If current INor voltage UNE does not need to be measured, connection
of IN or PE can be omitted.
The connectors IN und PE are not available for the AM2000.
With current transformers
If current INdoes not need to be measured, the corresponding
transformer can be omitted. If voltage UNE does not need to be
measured, connection of PE can be omitted.
The connectors IN und PE are not available for the AM2000.
With current and voltage transformers
If current INor voltage UNE does not need to be measured, the
corresponding transformers can be omitted.
The connectors IN und PE are not available for the AM2000.
L1 L2 L3
N
I1
31
I2
4
I3
97
IN PE
610 12 16 11 2 5 8
L1
N
5 A
(UL listed)
PE
L1 L2 L3 N I1 I2 I3
l ll l kkk
31 4 97611258
L1
N
5 A
(UL listed)
L1 L2 L3
N
I1
31
I2
4
I3
97
IN PE
610 12 16 11 2 5 8
L1
N
5 A
(UL listed)
K
k
L
l
K
k
L
l
PE
L1 L2 L3 N I1 I2 I3
l ll l kkk
31 4 97611258
K
k
L
l
L1
N
5 A
(UL listed)
L1
N
K
k
L
l
L1 L2 L3
N
I1
31
I2
4
I3
97
IN PE
610 12 16 11 2 5 8
5 A
(UL listed)
PE
K
k
L
l
u
U
v
V
u
U
v
V
L1 L2 L3 N I1 I2 I3
l ll l kkk
31 4 97611258
K
k
L
l
L1
N
5 A
(UL listed)
U
u
V
v

PM 1000162 000 14 Device handbook SINEAX AMx000 16/118
Three wire system, balanced load, phase shift
current measurement: L1, voltage measurement: L1-L2
AM2000 / AM3000
AM1000
Direct connection
Maximum permissible rated voltage 300V to ground (520V ph
-ph)!
With current transformers
With current and voltage transformers
In case of current measurement via L2 or L3 connect the device according to the following table:
Terminals 1 3 2 5 8
Current meas. via L2 I2(k) I2(l) L2 L3 -
Current meas. via L3 I3(k) I3(l) L3 L1 -
L1 L2 L3
N
I1
31
I2
4
I3
97
IN PE
610 12 16 11 258
L1
L2
L3
5 A
(UL listed)
L1 L2 L3 N I1 I2 I3
l ll l kkk
31 4 97611258
L1
L2
L3
5 A
(UL listed)
L1 L2 L3
N
I1
31
I2
4
I3
97
IN PE
610 12 16 11 258
L1
L2
L3
K
k
L
l
5 A
(UL listed)
L1 L2 L3 N I1 I2 I3
l ll l kkk
31 4 97611258
K
k
L
l
L1
L2
L3
5 A
(UL listed)
L1 L2 L3
N
I1
31
I2
4
I3
97
IN PE
610 12 16 11 258
L1
L2
L3
K
k
L
l
5 A
(UL listed)
u
U
v
V
L1 L2 L3 N I1 I2 I3
l ll l kkk
31 4 97611258
K
k
L
l
L1
L2
L3
5 A
(UL listed)
U
u
V
v

PM 1000162 000 14 Device handbook SINEAX AMx000 17/118
Three wire system, balanced load, phase shift
current measurement: L1, voltage measurement: L2-L3
AM2000 / AM3000
AM1000
Direct connection
Maximum permissible rated voltage 300V to ground (520V ph
-ph)!
With current transformers
.
With current and voltage transformers
.
In case of current measurement via L2 or L3 connect the device according to the following table:
Terminals 1 3 2 5 8
Current meas. via L2 I2(k) I2(l) - L3 L1
Current meas. via L3 I3(k) I3(l) - L1 L2
L1 L2 L3
N
I1
31
I2
4
I3
97
IN PE
610 12 16 11 258
L1
L2
L3
5 A
(UL listed)
L1L2 L3 NI1 I2 I3
ll
l l kk
k
3
1 4 976112 5 8
L1
L2
L3
5 A
(UL listed)
L1 L2 L3
N
I1
31
I2
4
I3
97
IN PE
610 12 16 11 258
L1
L2
L3
K
k
L
l
5 A
(UL listed)
L1 L2 L3 N I1 I2 I3
l ll l kkk
31 4 976112 5 8
K
k
L
l
L1
L2
L3
5 A
(UL listed)
L1 L2 L3
N
I1
31
I2
4
I3
97
IN PE
610 12 16 11 258
L1
L2
L3
K
k
L
l
5 A
(UL listed)
u
U
v
V
L1 L2 L3 N I1 I2 I3
l ll l kkk
31 4 976112 5 8
K
k
L
l
L1
L2
L3
5 A
(UL listed)
U
u
V
v

PM 1000162 000 14 Device handbook SINEAX AMx000 18/118
Three wire system, balanced load, phase shift
current measurement: L1, voltage measurement: L3-L1
AM2000 / AM3000
AM1000
Direct connection
Maximum permissible rated voltage 300V to ground (520V ph
-ph)!
With current transformers
With current and voltage transformers
In case of current measurement via L2 or L3 connect the device according to the following table:
Terminals 1 3 2 5 8
Current meas. via L2 I2(k) I2(l) L2 - L1
Current meas. via L3 I3(k) I3(l) L3 - L2
L1 L2 L3
N
I1
31
I2
4
I3
97
IN PE
610 12 16 11 258
L1
L2
L3
5 A
(UL listed)
L1 L2 L3 N I1 I2 I3
l ll l kkk
31 4 97611258
L1
L2
L3
5 A
(UL listed)
L1 L2 L3
N
I1
31
I2
4
I3
97
IN PE
610 12 16 11 258
L1
L2
L3
K
k
L
l
5 A
(UL listed)
L1L2L3NI1I2I3
l l
ll k
kk
31 4 9
76
11258
K
k
L
l
L1
L2
L3
5 A
(UL listed)
L1 L2 L3
N
I1
31
I2
4
I3
97
IN PE
610 12 16 11 258
L1
L2
L3
K
k
L
l
5 A
(UL listed)
v
V
u
U
L1 L2 L3 N I1 I2 I3
l ll l kkk
31 4 9
7611258
K
k
L
l
L1
L2
L3
5 A
(UL listed)
V
u
U
v

PM 1000162 000 14 Device handbook SINEAX AMx000 19/118
Three wire system, balanced load, current measurement via L1
AM2000 / AM3000 AM1000
Direct connection
Maximum permissible rated voltage 300V to ground (520V ph
-ph)!
With current transformers
.
With current and voltage transformers
.
In case of current measurement via L2 or L3 connect the device according to the following table:
Terminals 1 3 2 5 8
Current meas. via L2 I2(k) I2(l) L2 L3 L1
Current meas. via L3 I3(k) I3(l) L3 L1 L2
By rotating the voltage connections the measurements U12, U23 and U31 will be assigned interchanged!
L1 L2 L3
N
I1
31
I2
4
I3
97
IN PE
610 12 16 11 258
L1
L2
L3
5 A
(UL listed)
L1L2 L3 N I1 I2 I3
lll l k
kk
31 4 97611258
L1
L2
L3
5 A
(UL listed)
L1 L2 L3
N
I1
31
I2
4
I3
97
IN PE
610 12 16 11 258
L1
L2
L3
K
k
L
l
5 A
(UL listed)
L1L2L3N I1 I2I3
lll l k
kk
3
1 4 97
611258
K
k
L
l
L1
L2
L3
5 A
(UL listed)
L1 L2 L3
N
I1
31
I2
4
I3
97
IN PE
610 12 16 11 258
L1
L2
L3
K
k
L
l
5 A
(UL listed)
u
U
v
V
u
U
v
V
L1 L2 L3 N I1 I2 I3
l ll l kkk
31 4 97611258
K
k
L
l
L1
L2
L3
5 A
(UL listed)
U
u
V
v
U
u
V
v

PM 1000162 000 14 Device handbook SINEAX AMx000 20/118
Four wire system, balanced load, current measurement via L1
AM2000 / AM3000 AM1000
Direct connection
Maximum permissible rated voltage 300V to ground!
If voltage UNE does not need to be measured, connection of PE can
be omitted. PE is not available for AM2000.
With current transformer
If voltage UNE does not need to be measured, connection of PE can
be omitted. PE is not available for AM2000.
With current and voltage transformers
If voltage UNE does not need to be measured, the corresponding
transformer can be omitted. PE is not available for AM2000.
In case of current measurement via L2 or L3 connect the device according to the following table:
Terminals 1 3 2 11
Current meas. via L2 I2(k) I2(l) L2 N
Current meas. via L3 I3(k) I3(l) L3 N
L1 L2 L3
N
I1
31
I2
4
I3
9
7
IN PE
610 12 16 11 258
L1
L2
L3
N
5 A
(UL listed)
PE
L1 L2 L3 N I1 I2 I3
l ll l kkk
31 4 97611258
L1
L2
L3
N
5 A
(UL listed)
L1 L2 L3
N
I1
31
I2
4
I3
97
IN PE
610 12 16 11 258
L1
L2
L3
N
K
k
L
l
5 A
(UL listed)
PE
L1 L2 L3 N I1 I2 I3
l ll l kkk
31 4 97611258
K
k
L
l
L1
L2
L3
N
5 A
(UL listed)
L1 L2 L3
N
I1
31
I2
4
I3
97
IN PE
610 12 16 11 258
L1
L2
L3
N
K
k
L
l
5 A
(UL listed)
u
U
v
V
u
U
v
V
PE
L1 L2 L3 N I1 I2 I3
l ll l kkk
31 4 97611258
K
k
L
l
L1
L2
L3
N
5 A
(UL listed)
U
u
V
v
Other manuals for SINEAX AM2000
1
This manual suits for next models
2
Table of contents
Other Camille Bauer Controllers manuals
Popular Controllers manuals by other brands

Teddington
Teddington TSC5 instruction manual

Mitsubishi Electric
Mitsubishi Electric MR-J2S B Series instruction manual

Tuson RV Brakes
Tuson RV Brakes DirecLink Installation and user guide

Texas Instruments
Texas Instruments TUSB3200EVM2 user guide

Carel
Carel kVent manual

Sun Microsystems
Sun Microsystems Sun StorageTek 2500 Series Removal and Replacement