functions are working properly
without interference before
proceeding to the next step.
(2) Secure the power unit onto the foot
board using hanging hooks.
(3) Firmly connect the mattress' air hose
couplings to the quick release
couplers on the power unit's air
outlet.
(4) Plug the power unit into a wall outlet
and turn on the main power rocker
switch on the right side panel to
inflate the mattress system.
(5) Turn the comfort knob on the power
unit's control panel to maximum
setting to shorten the initial inflation
time if needed.
(6) Wait for approx. 15 minutes or less
for the mattress to inflate fully.
7.0 Program Settings
(1) Center the patient on the mattress.
Adjust the mattress' internal pressure
according to the patient weight by
using the comfort knob on the
control panel of the power unit. If
the patient feels the mattress is too
soft or firm, increases or decreases
the mattress' internal pressure by fine
turning the comfort knob one
increment at a time and wait for the
system to stabilize before making
another change until a comfortable
state is achieved.
(2) Caregivers should always perform a
hand check by placing their hands
underneath patient's pelvis area to
check if there is sufficient air support
to ensure the patient is not bottoming
out.
(3) IMPORTANT: TUCKING THE
SHEET IN TIGHTLY REDUCES
THE EFFECTIVENESS OF THE
SYSTEM.
8.0 Patient Transfer
and Transport
Transfer
It is recommended to have the mattress
system at maximum weight setting
during transfer. Make sure the bed is
secured before proceeding.
Transport
If a transport occasion comes up, two
options are available: (1) Detach the
mattress' air hose couplings from the
power unit's quick release couplers and
connect the two air hose couplings
together to retain air in the mattress and
the mattress will stay inflated for approx.
2 hours. (2) Unplug the power unit's
power cord from the wall outlet and the
mattress should stay inflated for approx.
40 - 50 minutes. To resume normal
operation, please follow 7.0 Program
Setting.
9.0 Cleaning
Instruction
Air mattress and power unit must be
cleaned thoroughly between patients to
English
OPERATION MANUALCOZINY 100/ 200/ 300 6