Ozone SAM
I-2022
0 to 0.75
PPM (mg/Liter)
Battery Replacement
To ensure that the instrument is waterproof:
• seal ring (A) must be in position
• battery compartment cover (B) must be fixed with the four screws
To Set Zero
1. Press the ON/OFF key.
2. The display will show “O3”.
3. Insert the ZERO ampoule (supplied in Vacu-vials®kit), flat end first, into the
sample cell compartment (with mild downward pressure), making sure that it is
fully seated.
4. Place the light shield over the ZERO ampoule.
5. Press the Zero/Test key. The “O3” symbol will flash for approximately 8 sec-
onds, then the display will show “SEt”.
To Generate Reagent Blank Value
1. Insert an unsnapped ozone Vacu-vial ampoule, flat end first, into the sample
cell compartment (with mild downward pressure), making sure that it is fully
seated.
2. Place the light shield over the ampoule.
3. Press the Zero/Test key. The “O3” symbol will flash for approximately 8 sec-
onds, then the display will show “tESt“.
Note: The instrument has stored a reagent blank value for this ampoule.
To Obtain Test Result
1. Follow the Test Procedure in the Ozone Vacu-vials test kit (Cat. # K-7433) using
the ampoule from which the reagent blank value was generated above.
2. Insert the resulting Ozone Vacu-vial ampoule, flat end first, into the sample cell
compartment (with mild downward pressure), making sure that it is fully seated.
3. Place the light shield over the test ampoule.
4. Press the Zero/Test key. The “O3” symbol will flash for approximately 3 sec-
onds, then the sample test result will appear in the display as ppm (mg/Liter).
5. To perform the next test, remove the test ampoule from the sample cell com-
partment. Press the Zero/Test key. The display will show “SEt”. Return to To
Generate Reagent Blank Value.
Operating Tips
• Upon startup, the photometer automatically proceeds to the zeroing process.
Every time the photometer powers on, it must be re-zeroed.
• To re-zero the photometer, it must be turned off and back on again.
• A series of readings can be taken without re-zeroing, as long as the photome-
ter stays on during the series.
• Protect photometer from extreme humidity, corrosive fumes and dusty areas.
Store in a cool, dry place.
• Remove the batteries when photometer is not in use.
• Press the !key to turn the display back light on or off.
• When moving the photometer from one temperature extreme to another, wait
at least 10 minutes before use to allow photometer to come to temperature
equilibrium.
• Contamination of the optics in the sample chamber will result in incorrect meas-
urements. The windows in the sample chamber should be checked at regular
intervals and cleaned as necessary. Use a soft moist cloth or cotton swab for
cleaning purposes.
• If the sample cell adapter has been removed, it must be replaced with proper
orientation, aligning the triangle on the adapter with the triangle on the pho-
tometer.
Displays and Troubleshooting
E01: Light absorption too great (dirty optics)
E20/E21: Too much light reaching detector
E22 or Battery Icon: Battery should be replaced
E27/E28/E29: Instrument zeroed incorrectly, misaligned adapter, vial not
properly seated, dirty optics or failing light source.
Hi/E03: Measuring range exceeded or excessive turbidity
Lo: Test result has a negative value (less than 0 ppm)
Specifications
Auto Shutoff: After 15 minutes of non-use
Optics: 610 nm LED/interference filter and photosensor in transparent sample
chamber
Operating Temp.: 5 to 40°C (41 to 104°F)
Battery: 4 AAA batteries (approx. 5,000 tests or 17 hours)
Waterproof: Floating, IP68 (1 hour at 0.1 meter)
Wavelength Accuracy: ± 1 nm
Photometric Accuracy*: 3% full scale (T = 20 - 25° C / 68 - 77° F)
Photometric Resolution: 0.01 A
Ambient Conditions: Temperature 5 - 40° C / 41 - 104° F
Rel. humidity 30 - 90 % (non-condensing)
CE: Certificate of Declaration of CE-Conformity available upon request.
* Measured with Standard Solutions
Instrument
Back
Batteries
Notch
Screws
Seal Ring (A)
Battery
Compartment
Cover (B)
www.chemetrics.com
4295 Catlett Road, Midland, VA 22728 U.S.A.
Phone: (800) 356-3072; Fax: (540) 788-4856
E-Mail: or[email protected]
Sept. 14, Rev. 1
Menu Selection
1. Hold the Mode key down and press the On/Off key. Allow the 3 decimal points
to appear in the display before releasing the Mode key.
2. Use the !key to toggle through 2 menu options:
Store = recall stored data
Date & Time = setting the date and time.
3. Press the Mode key to select the menu option that is indicated by the arrow.
Note: If the instrument is already on, press and hold the !key for more than
4 seconds to access the recall menu.
Recall of Stored Data
The colorimeter shows the last 16 data sets in the following format:
Sample Number: nXX (XX: 16, 15, 14, ....1)
Year: XXXX (e.g. 2009)
Date: mm.dd (e.g. 07.31)
Time: hh:mm (e.g. 12:05)
Result: XX or XXX
The Zero/Test key repeats the current data set.
The Mode key scrolls through all stored data.
Quit the menu by pressing the !key.
Setting Date and Time
The setting starts with the year (XXXX), then month (XX), then day (XX), then
hour (XX), then minutes (XX).
Increase value by pressing the Mode key.
Decrease value by pressing the Zero/Test key.
Proceed to the next setting by pressing the !key.
After setting the minutes and pressing the !key, the display will show "IS SET"
and return to measurement mode.

Popular Test Equipment manuals by other brands

Hanna Instruments Foodcare HI981034 instruction manual

Hanna Instruments

Hanna Instruments Foodcare HI981034 instruction manual

SDI Solo Series quick start guide

SDI

SDI Solo Series quick start guide

Hanna Instruments Foodcare HI981031 user manual

Hanna Instruments

Hanna Instruments Foodcare HI981031 user manual

Ecom Instruments Magnet-Ex 12 operating instructions

Ecom Instruments

Ecom Instruments Magnet-Ex 12 operating instructions

Fluke 709 Technical data

Fluke

Fluke 709 Technical data

RJG CoPilot AP 32C Hardware Installation Quick Start Guide

RJG

RJG CoPilot AP 32C Hardware Installation Quick Start Guide

Agilent Technologies E6630A Getting started guide

Agilent Technologies

Agilent Technologies E6630A Getting started guide

HBM PW20i Mounting instructions

HBM

HBM PW20i Mounting instructions

Sealey BT103 instructions

Sealey

Sealey BT103 instructions

TiePie TP-EMI-HS6 quick start guide

TiePie

TiePie TP-EMI-HS6 quick start guide

Klein 3000 Series quick start guide

Klein

Klein 3000 Series quick start guide

Aeroflex 8800 Series Operation manual

Aeroflex

Aeroflex 8800 Series Operation manual

Tektronix TBS1000B Series user manual

Tektronix

Tektronix TBS1000B Series user manual

GRAS 42AP instruction manual

GRAS

GRAS 42AP instruction manual

Slingmax CHECK FAST Plus Instruction & safety manual

Slingmax

Slingmax CHECK FAST Plus Instruction & safety manual

Rohde & Schwarz RTP044B Getting started

Rohde & Schwarz

Rohde & Schwarz RTP044B Getting started

TiePie GMTO ATS5004D user manual

TiePie

TiePie GMTO ATS5004D user manual

EXFO RTU-2 user guide

EXFO

EXFO RTU-2 user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.