TiePie GMTO ATS5004D User manual

ATS5004D
User manual

ATTENTION!
Measuring directly on the line voltage can be very dangerous.
Copyright c
2016 TiePie engineering.
All rights reserved.
Revision 2.14, January 2016
Despite the care taken for the compilation of this
user manual, TiePie engineering can not be held
responsible for any damage resulting from errors
that may appear in this manual.

Contents
1 Safety 1
2 Declaration of conformity 3
3 Introduction 5
3.1 Differential input .................... 5
3.1.1 Differential attenuators ............ 7
3.1.2 Differential test lead .............. 9
3.2 Sampling ........................ 9
3.3 Sample frequency .................... 10
3.3.1 Aliasing ..................... 11
3.4 Digitizing ........................ 13
3.5 Signal coupling ..................... 13
4 Driver installation 15
4.1 Introduction ....................... 15
4.2 Where to find the driver setup ............ 15
4.3 Executing the installation utility ........... 15
5 Hardware installation 21
5.1 Power the instrument ................. 21
5.1.1 External power ................. 21
5.2 Connect the instrument to the computer ....... 22
5.2.1 Found New Hardware Wizard ......... 22
5.3 Plug into a different USB port ............ 25
6 Front panel 27
6.1 Channel input connectors ............... 27
6.2 Power indicator ..................... 27
7 Rear panel 29
7.1 Power .......................... 29
7.1.1 USB power cable ................ 30
7.1.2 Power adapter ................. 31
7.2 USB ........................... 31
7.3 Extension Connector .................. 31
8 Specifications 33
Contents I

8.1 Acquisition system ................... 33
8.2 BNC inputs CH1 to CH4 ............... 33
8.3 Trigger system ..................... 34
8.4 Physical ......................... 34
8.5 Power .......................... 34
8.6 Interface ......................... 34
8.7 I/O connectors ..................... 34
8.8 System requirements .................. 35
8.9 Environmental conditions ............... 35
8.10 Certifications and Compliances ............ 35
8.11 Measure leads ...................... 35
8.12 Differential attenuators ................. 36
8.13 Package contents .................... 36
II

Safety 1
When working with electricity, no instrument can guaran-
tee complete safety. It is the responsibility of the person
who works with the instrument to operate it in a safe way.
Maximum security is achieved by selecting the proper in-
struments and following safe working procedures. Safe
working tips are given below:
•Always work according (local) regulations.
•Work on installations with voltages higher than 25 VAC or
60 VDC should only be performed by qualified personnel.
•Avoid working alone.
•Observe all indications on the ATS5004D before connecting
any wiring
•Check the probes/test leads for damages. Do not use them
if they are damaged
•Take care when measuring at voltages higher than 25 VAC or
60 VDC.
•Do not operate the equipment in an explosive atmosphere or
in the presence of flammable gases or fumes.
•Do not use the equipment if it does not operate properly.
Have the equipment inspected by qualified service personal.
If necessary, return the equipment to GMTO for service and
repair to ensure that safety features are maintained.
Safety 1

2Chapter 1

Declaration of conformity 2
TiePie engineering
Koperslagersstraat 37
8601 WL Sneek
The Netherlands
EC Declaration of conformity
We declare, on our own responsibility, that the product
ATS5004D
for which this declaration is valid, is in compliance with
EN 55011:2009/A1:2010 IEC 61000-6-1/EN 61000-6-1:2007
EN 55022:2006/A1:2007 IEC 61000-6-3/EN 61000-6-3:2007
according the conditions of the EMC standard 2004/108/EC
and also with
Canada: ICES-001:2004 Australia/New Zealand: AS/NZS
Sneek, 1-11-2010
ir. A.P.W.M. Poelsma
Declaration of conformity 3

Environmental considerations
This section provides information about the environmental impact
of the ATS5004D.
ATS5004D end-of-life handling
Production of the ATS5004D required the extraction and use of
natural resources. The equipment may contain substances that
could be harmful to the environment or human health if improperly
handled at the ATS5004D’s end of life.
In order to avoid release of such substances into the environment
and to reduce the use of natural resources, recycle the ATS5004D
in an appropriate system that will ensure that most of the materials
are reused or recycled appropriately.
The symbol shown below indicates that the ATS5004D complies
with the European Union’s requirements according to Directive
2002/96/EC on waste electrical and electronic equipment (WEEE).
Restriction of Hazardous Substances
The ATS5004D has been classified as Monitoring and Control equip-
ment, and is outside the scope of the 2002/95/EC RoHS Directive.
4Chapter 2

Introduction 3
Before using the ATS5004D first read chapter 1about safety.
Many technicians investigate electrical signals. Though the
measurement may not be electrical, the physical variable is of-
ten converted to an electrical signal, with a special transducer.
Common transducers are accelerometers, pressure probes, current
clamps and temperature probes. The advantages of converting the
physical parameters to electrical signals are large, since many in-
struments for examining electrical signals are available.
The ATS5004D is a portable four channel measuring instru-
ment with differential inputs. The native resolution is 12 bits, but
user selectable resolutions of 14 and 16 bits are available too, with
adjusted maximum sampling frequency:
resolution Maximum sampling frequency
12 bit 50 MS/s
14 bit 3.125 MS/s
16 bit 195 kS/s
Table 3.1: Maximum sampling frequencies
With the accompanying software the ATS5004D can be used as
an oscilloscope, a spectrum analyzer, a true RMS voltmeter or a
transient recorder. All instruments measure by sampling the input
signals, digitizing the values, process them, save them and display
them.
3.1 Differential input
Most oscilloscopes are equipped with standard, single ended inputs,
which are referenced to ground. This means that one side of the
input is always connected to ground and the other side to the point
of interest in the circuit under test.
Introduction 5

Figure 3.1: Single ended input
Therefore the voltage that is measured with an oscilloscope with
standard, single ended inputs is always measured between that
specific point and ground.
When the voltage is not referenced to ground, connecting a
standard single ended oscilloscope input to the two points would
create a short circuit between one of the points and ground, possi-
bly damaging the circuit and the oscilloscope.
A safe way would be to measure the voltage at one of the two
points, in reference to ground and at the other point, in reference to
ground and then calculate the voltage difference between the two
points. On most oscilloscopes this can be done by connecting one
of the channels to one point and another channel to the other point
and then use the math function CH1 - CH2 in the oscilloscope to
display the actual voltage difference.
There are some disadvantages to this method:
•a short circuit to ground can be created when an input is
wrongly connected
•to measure one signal, two channels are occupied
•by using two channels, the measurement error is increased,
the errors made on each channel will be combined, resulting
in a larger total measurement error
•The Common Mode Rejection Ratio (CMRR) of this method
is relatively low. If both points have a relative high voltage,
but the voltage difference between the two points is small,
the voltage difference can only be measured in a high input
range, resulting in a low resolution
A much better way is to use an oscilloscope with a differential
input.
6Chapter 3

Figure 3.2: Differential input
A differential input is not referenced to ground, but both sides
of the input are ”floating”. It is therefore possible to connect one
side of the input to one point in the circuit and the other side of
the input to the other point in the circuit and measure the voltage
difference directly.
Advantages of a differential input:
•No risk of creating a short circuit to ground
•Only one channel is required to measure the signal
•More accurate measurements, since only one channel intro-
duces a measurement error
•The CMRR of a differential input is high. If both points have
a relative high voltage, but the voltage difference between the
two points is small, the voltage difference can be measured
in a low input range, resulting in a high resolution
3.1.1 Differential attenuators
To increase the input range of the ATS5004D, it comes with a
differential 1:10 attenuator for each channel. This differential at-
tenuator is specially designed to be used with the ATS5004D.
Figure 3.3: Differential attenuator
For a differential input, both sides of the input need to be at-
tenuated.
Introduction 7

Figure 3.4: Differential input
Standard oscilloscope probes and attenuators only attenuate
one side of the signal path. These are not suitable to be used with
a differential input. Using these on a differential input will have
a negative effect on the CMRR and will introduce measurement
errors.
Figure 3.5: Differential input
The Differential Attenuator and the inputs of the Handy-
scope HS4 Diff are differential, which means that the outside
of the BNC’s are not grounded, but carry life signals.
When using the attenuator, the following points have to be
taken into consideration:
•do not connect other cables to the attenuator than the ones
that are supplied with the instrument
•do not touch the metal parts of the BNC’s when the atten-
uator is connected to the circuit under test, they can carry
8Chapter 3

a dangerous voltage. It will also influence the measurements
and create measurement errors.
•do not connect the outside of the two BNC’s of the attenuator
to each other as this will short circuit a part of the internal
circuit and will create measurement errors
•do not connect the outside of the BNC’s of two or more at-
tenuators that are connected to different channels of the Han-
dyscope HS4 Diff to each other
•do not apply excessive mechanical force to the attenuator in
any direction (e.g. pulling the cable, using the attenuator as
handle to carry the ATS5004D, etc.)
3.1.2 Differential test lead
The ATS5004D comes with a special differential test lead. This
test lead is specially designed to ensure a good CMRR.
The special heat resistant differential test lead provided with
the ATS5004D is designed to be immune for noise from the sur-
rounding environment.
3.2 Sampling
When sampling the input signal, samples are taken at fixed inter-
vals. At these intervals, the size of the input signal is converted to a
number. The accuracy of this number depends on the resolution of
the instrument. The higher the resolution, the smaller the voltage
steps in which the input range of the instrument is divided. The
acquired numbers can be used for various purposes, e.g. to create
a graph.
Introduction 9

Figure 3.6: Sampling
The sine wave in figure 3.6 is sampled at the dot positions. By
connecting the adjacent samples, the original signal can be recon-
structed from the samples. You can see the result in figure 3.7.
Figure 3.7: ”connecting” the samples
3.3 Sample frequency
The rate at which the samples are taken is called the sampling
frequency, the number of samples per second. A higher sampling
frequency corresponds to a shorter interval between the samples.
As is visible in figure 3.8, with a higher sampling frequency, the
original signal can be reconstructed much better from the measured
samples.
10 Chapter 3

Figure 3.8: The effect of the sampling frequency
The sampling frequency must be higher than 2 times the high-
est frequency in the input signal. This is called the Nyquist fre-
quency. Theoretically it is possible to reconstruct the input signal
with more than 2 samples per period. In practice, 10 to 20 sam-
ples per period are recommended to be able to examine the signal
thoroughly.
3.3.1 Aliasing
When sampling an analog signal with a certain sampling frequency,
signals appear in the output with frequencies equal to the sum and
difference of the signal frequency and multiples of the sampling
frequency. For example, when the sampling frequency is 1000 Hz
and the signal frequency is 1250 Hz, the following signal frequencies
will be present in the output data:
Multiple of sampling frequency 1250 Hz signal -1250 Hz signal
...
-1000 -1000 + 1250 = 250 -1000 - 1250 = -2250
0 0 + 1250 = 1250 0 - 1250 = -1250
1000 1000 + 1250 = 2250 1000 - 1250 = -250
2000 2000 + 1250 = 3250 2000 - 1250 = 750
...
Table 3.2: Aliasing
As stated before, when sampling a signal, only frequencies lower
than half the sampling frequency can be reconstructed. In this
case the sampling frequency is 1000 Hz, so we can we only observe
Introduction 11

signals with a frequency ranging from 0 to 500 Hz. This means
that from the resulting frequencies in the table, we can only see
the 250 Hz signal in the sampled data. This signal is called an
alias of the original signal.
If the sampling frequency is lower than twice the frequency of
the input signal, aliasing will occur. The following illustration
shows what happens.
Figure 3.9: Aliasing
In figure 3.9, the green input signal (top) is a triangular signal
with a frequency of 1.25 kHz. The signal is sampled with a fre-
quency of 1 kHz. The corresponding sampling interval is 1/1000Hz
= 1ms. The positions at which the signal is sampled are depicted
with the blue dots. The red dotted signal (bottom) is the result
of the reconstruction. The period time of this triangular signal
appears to be 4 ms, which corresponds to an apparent frequency
(alias) of 250 Hz (1.25 kHz - 1 kHz).
To avoid aliasing, always start measuring at the highest sam-
pling frequency and lower the sampling frequency if required.
12 Chapter 3

3.4 Digitizing
When digitizing the samples, the voltage at each sample time is
converted to a number. This is done by comparing the voltage
with a number of levels. The resulting number is the number cor-
responding to the level that is closest to the voltage. The number
of levels is determined by the resolution, according to the following
relation: LevelCount = 2Resolution.
The higher the resolution, the more levels are available and
the more accurate the input signal can be reconstructed. In figure
3.10, the same signal is digitized, using two different amounts of
levels: 16 (4-bit) and 64 (6-bit).
Figure 3.10: The effect of the resolution
The ATS5004D measures at e.g. 12 bit resolution (212=4096
levels). The smallest detectable voltage step depends on the input
range. This voltage can be calculated as:
V oltageStep =F ullInputRange/LevelCount
For example, the 200 mV range ranges from -200 mV to +200
mV, therefore the full range is 400 mV. This results in a smallest
detectable voltage step of 0.400V/4096 = 97.65 µV.
3.5 Signal coupling
The ATS5004D has two different settings for the signal coupling:
AC and DC. In the setting DC, the signal is directly coupled to the
Introduction 13

input circuit. All signal components available in the input signal
will arrive at the input circuit and will be measured.
In the setting AC, a capacitor will be placed between the input
connector and the input circuit. This capacitor will block all DC
components of the input signal and let all AC components pass
through. This can be used to remove a large DC component of the
input signal, to be able to measure a small AC component at high
resolution.
When measuring DC signals, make sure to set the signal
coupling of the input to DC.
14 Chapter 3

Driver installation 4
Before connecting the ATS5004D to the computer, the
drivers need to be installed.
4.1 Introduction
To operate a ATS5004D, a driver is required to interface between
the measurement software and the instrument. This driver takes
care of the low level communication between the computer and
the instrument, through USB. When the driver is not installed, or
an old, no longer compatible version of the driver is installed, the
software will not be able to operate the ATS5004D properly or even
detect it at all.
The installation of the USB driver is done in a few steps. Firstly,
the driver has to be pre-installed by the driver setup program. This
makes sure that all required files are located where Windows can
find them. When the instrument is plugged in, Windows will detect
new hardware and install the required drivers.
4.2 Where to find the driver setup
The driver setup program and measurement software can be found
in the download section on TiePie engineering’s website and on the
CD-ROM that came with the instrument. It is recommended to
install the latest version of the software and USB driver from the
website. This will guarantee the latest features are included.
4.3 Executing the installation utility
To start the driver installation, execute the downloaded driver
setup program, or the one on the CD-ROM that came with the
instrument. The driver install utility can be used for a first time
Driver installation 15

installation of a driver on a system and also to update an existing
driver.
The screen shots in this description may differ from the ones
displayed on your computer, depending on the Windows version.
Figure 4.1: Driver install: step 1
When drivers were already installed, the install utility will re-
move them before installing the new driver. To remove the old
driver successfully, it is essential that the ATS5004D is discon-
nected from the computer prior to starting the driver install utility.
When the ATS5004D is used with an external power supply, this
must be disconnected too.
16 Chapter 4
Table of contents
Other TiePie Test Equipment manuals

TiePie
TiePie Handyscope HS3 User manual

TiePie
TiePie WiFiScope WS6 User manual

TiePie
TiePie Handyscope HS6 DIFF Series User manual

TiePie
TiePie WiFiScope WS6 User manual

TiePie
TiePie TP-EMI-HS6 User manual

TiePie
TiePie Handyprobe HP3 User manual

TiePie
TiePie Handyscope HS5 series User manual

TiePie
TiePie Handyscope TP450 User manual

TiePie
TiePie Handyscope TP450 User manual

TiePie
TiePie Handyscope HS4 User manual