Connecting the probes
To connect a probe to the instrument with option E, first place the TP-BNC-T50G
grounded terminator at the Ch1 input of the scope. When using a Handyscope
HS6 DIFF-1000 or WiFiScope WS6 DIFF-1000, make sure to connect the extra
ground lead to the grounding terminal next to the Ch1 input. When using a Handy-
scope HS6-1000 or WiFiScope WS6-1000, the extra ground lead is not required
and can be left unconnected.
Next connect the BNC connector of the TP-BNC-SMA-200 flexible antenna cable to
the grounded terminator. Insert the female SMA connector of the TP-BNC-SMA-
200 through the ring of the TP-3P-15 tripod, from the side with the recessed hex
shape.
Connect the male SMA connector of one end of the TP-SMA-20 semi flexible an-
tenana cable to the flexible antenna cable. Use the supplied 8 mm wrench to
tighten the nut to avoid the semi flexible cable to rotate. Do not over tighten,
maximum torque is 0.3 - 0.6 Nm (3 - 5 in lbf). Finally, connect the required probe
to the other end of the semi flexible antenna cable. Using the semi flexible cable
and the tripod, the probe can be positioned at the exact location near the test
subject.