Chino IR-M22 SERIES User manual

INST No INE-403-0P0
IR-M2
□□□
Series
INFRARED MOISTURE METER
DESIGNED AS A FIBER OPTIC
(Model :IR-M21
□□
,IR-M22
□□
)
Please be sure to deliver these instructions
with the unit to the end user.
Always keep these instruction with the unit.

INST No INE-403-0P0
−C1−
Request to designers, instrument controllers, and sale agents
Delivery this instruction manual to the operator of this instrument without fail.
Request to the operator of this instrument
This instruction manual is necessary for maintenance, too. Keep this manual with due
care until the instrument is discarded.
■
FOR SAFE USE
For the purpose of using this unit safely, observe the following cautions for safety during
the operation, maintenance, and repair of this unit without fail.
Take safety measures separately against wrong operation, troubles, and other circumstances
of this unit.
●The following warning mark is struck to this unit to enable users to use
this unit safely.
●This instruction manual uses the following symbol marks for safe use.
■
PREFACE
Requests and notices
Thank you for your purchase of IR-M21□□□□, IR-M22□□□□series Infrared
moisture meter designed as a fiber optic. Please read this instruction manual without fail
for using this instrument correctly and safely and also preventing troubles in advance.
1.For the contents of this manual, alteration is reserved without notice in the future.
2.This manual has been prepared by making assurance doubly sure about its contents.
However, if any question arises or if any error, an omission, or other deficiencies
were found, please inform your nearest CHINO’s sales agent of them.
3.You are requested to understand that CHINO is not responsible for any operation
results.
Note
This warning mark is struck to the portions being subject to an electric shock
acc
ident. Turn off the power supply of the feed source without fail before starting
wiring, maintenance, and repair.
Observe this caution item strictly for avoiding a danger, otherwise
negligence of this caution maypossibly cause troubles of this unit and
accidents
Observe this caution item strictly, otherwise negligence of this caution
may cause the damage of the unit or measuring errors.
This mark is described in each title covering or
instructions.
!
Warning
Caution
!Warning
DANGER
Reference
Information that you can use as a reference.

INST No INE-403-0P0
−C2−
●
Power supply
Make sure that the power voltage of this unit meets the feed power voltage without fail
before feeding power to this unit.
●
Protective grounding
Perform the protective grounding without fail before turning on the power supply of this
unit for the purpose of preventing an electric shock accident.
●
Necessity of protective grounding
Don’t cut the internal or external protective grounding wire of this unit nor disconnect
the wiring at the protective grounding terminal.
●
A defect of protective functions
Don’t operate this unit, if the protective grounding, fuse, or other protective functions
seem to be defective. Check the protective functions for defects before operating this
unit.
●
Use in a gaseous atmosphere
Never operate this unit at a place where a combustible gas, an explosive gas or their
vapors exist., otherwise a very dangerous accident will occur.
●
Never touch the interior by hand.
This unit is internally provided with high-voltage circuits. Never touch the interior by
hand when the voltage is applied to this unit.
●
External connections
Perform the protective grounding securely without fail before connecting this unit to a
measuring object and/or an external control circuit.
■
CAUTIONS FOR SAFETY

INST No INE-403-0P0
−C3−
CONTENTS
1.
General
・・・・・・・・・・・・・・・・・・・・・・・・・・・・
1
1.1
Summary
・・・・・・・・・・・・・・・・・・・・・・・・・・・
1
1.2
Configuration
・・・・・・・・・・・・・・・・・・・・・・・・
1
2.
Models and accessories
・・・・・・・・・・・・・・・・・
2
2.1
Models・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2
2.1.1 Detector unit
・・・・・・・・・・・・・・・・・・・・・・・・
2
2.1.2Reflectivetypefiver
・・・・・・・・・・・・・・・・・・・
2
2.1.3 Output checker plate for reflective
type fiver optics
・・・・・・・・・・・・・・・・・・・・・・
2
2.1.4 Reflective type fiver optics with lens attachment 2
2.1.5 Output checker plate for reflective
type fiver optics with lens attachment
・・・・・・・・
2
2.1.6 Transparent type fiver optics
・・・・・・・・・・・・・
2
2.1.7 Measuring cell for liquids
・・・・・・・・・・・・・・・
2
2.1.8 Connection cable
・・・・・・・・・・・・・・・・・・・・・
2
2.2
Accessories
・・・・・・・・・・・・・・・・・・・・・・・・・・
2
3.
Setting
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
3
3.1
Setting of dispswitches ・・・・・・・・・・・・・・・・・・ 3
3.2
Installation of reflective type fiver optics ・・・・・・ 5
3.2.1 Installation of tip part・・・・・・・・・・・・・・・・・・・ 5
3.2.2 Air purge(Only IR-WCRN)・・・・・・・・・・・・・・ 6
3.2.3 Connections with detector unit ・・・・・・・・・・・・ 6
3.2.4 Optical path ・・・・・・・・・・・・・・・・・・・・・・・・・ 7
3.2.5 Measuring place
(for objects of pulverized or granular substance) 7
3.2.6 For the objects in sheet state such as paper, etc
.
7
3.3
Installation of transparent type fiver optics ・・・・ 8
3.3.1 Installation of cell for liquids ・・・・・・・・・・・・・・ 8
3.3.2 Connections with detector unit ・・・・・・・・・・・・ 9
3.3.3 Disassemble cleaning ・・・・・・・・・・・・・・・・・・・ 9
3.4
Installation ofdetector unit ・・・・・・・・・・・・・・・10
3.4.1 Outside and mounting dimensions of
detector unit ・・・・・・・・・・・・・・・・・・・・・・・・・10
4. Connections
・・・・・・・・・・・・・・・・・・・・・・・・・
11
4.1
Connections
・・・・・・・・・・・・・・・・・・・・・・・・・・
11
4.2
Layout of connection cable
・・・・・・・・・・・・・・・
11
4.3
Multi-detector connections
・・・・・・・・・・・・・・・
12
4.3.1 Connections through relay box
・・・・・・・・・・・・・
12
4.3.2 Connections without relay box
・・・・・・・・・・・・・
13
5
.
Names and functions of component parts
・・・
14
5.1
Names and functions of detector unit
・・・・・・・・
14
5.2
Names and functions of setting display part
・・・・
15
6. Operation
・・・・・・・・・・・・・・・・・・・・・・・・・・・・
16
6.1
Confirmation before operation
・・・・・・・・・・・・
16
6.2
Setting of detector unit number
・・・・・・・・・・・・
16
6.3
Setting ofcalibration curve data
・・・・・・・・・・・・
16
6.3.1 Setting data
・・・・・・・・・・・・・・・・・・・・・・・・・・
17
6.3.2 Setting of calibration curve data
・・・・・・・・・・・
18
6.4
Preparation of calibration curbe
by knee point line approximation
・・・・・・・・・・・
18
6.4.1 Setting of knee point data
・・・・・・・・・・・・・・・・
18
6.4.2 Knee point line setting data
・・・・・・・・・・・・・・・
19
6.4.3 Setting ofcalibrationcurbedata
for knee point line approximation
・・・・・・・・・・
20
6.5 Setting of operating conditions
・・・・・・・・・・・・・
21
6.5.1 Setting data
・・・・・・・・・・・・・・・・・・・・・・・・・・
21
6.5.2 Setting
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
23
6.6
Calibration
・・・・・・・・・・・・・・・・・・・・・・・・・・
23
6.7
Setting of calibration curve number・・・・・・・・・24
6.8
Correction of calibration curve
・・・・・・・・・・・・
24
7.
Preparationof calibration curve
・・・・・・・・・・
25
7.1
Sample preparation ・・・・・・・・・・・・・・・・・・・・25
7.1.1 Powder or guranular
・・・・・・・・・・・・・・・・・・・
25
7.1.2Paper(sheetsubstance)
・・・・・・・・・・・・・・・・・・
26
!!!
!
!
!
!
!
!

INST No INE-403-
−C4−
7.1.3 Liquid
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
26
7.2
Sample measuremernt ・・・・・・・・・・・・・・・・・・ 27
7.2.1 Powder or granular
・・・・・・・・・・・・・・・・・・・・
27
7.2.2Paper(sheetsubstance)
・・・・・・・・・・・・・・・・・・
28
7.2.3 Liquid
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
28
7.3
Preparation of calibration curves・・・・・・・・・・・ 29
7.3.1 Computation of regression expression
・・・・・・・・
29
7.3.2Confirmation ofregression expression
・・・・・・・
30
8.
Correction function by sample temprature
from external input (option)
・・・・・・・・・・・・
31
8.1
Summary ・・・・・・・・・・・・・・・・・・・・・・・・・・・ 31
8.2
Model・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 31
8.3
Wiring of detector unit ・・・・・・・・・・・・・・・・・・ 31
8.4
Setting of dipswitch ・・・・・・・・・・・・・・・・・・・・ 31
8.5
Settingof correctionenabled/disabled
by sample temperature ・・・・・・・・・・・・・・・・・ 32
8.5.1 Setting at the detector unit
・・・・・・・・・・・・・・・・
32
8.5.1 Setting at the display unit
・・・・・・・・・・・・・・・・
32
8.6
Settingof sampletemperature scaling
(Only 4 to 20mA DC) ・・・・・・・・・・・・・・・・・・・ 33
8.6.1 Setting at the detector unit
・・・・・・・・・・・・・・・・
33
8.6.2 Setting at the display unit
・・・・・・・・・・・・・・・・
33
8.7
Setting of sample temperature ・・・・・・・・・・・・ 33
8.8
Settingofcalibration curvedata ・・・・・・・・・・・ 34
8.9
Measurement ・・・・・・・・・・・・・・・・・・・・・・・・ 35
8.10
Display of sample temperature Ts ・・・・・・・・・ 35
8.10.1 Display at the detector unit
・・・・・・・・・・・・・・
35
8.10.2 Display at the setting display unit
・・・・・・・・・・
35
9.
Inspection and maintenance
・・・・・・・・・・・・・
35
9.1
Periodical inspection ・・・・・・・・・・・・・・・・・・・ 35
9.2
Self-diagnosis function ・・・・・・・・・・・・・・・・・・ 36
9.3 Measures against troubles not included in
self-diagnosis ・・・・・・・・・・・・・・・・・・・・・・・・・ 37
9.4
Maintenance parts・・・・・・・・・・・・・・・・・・・・・ 38
9.4.1 List of maintenance parts
・・・・・・・・・・・・・・・・
38
9.4.2 Replacement of light source lamp
・・・・・・・・・・・
38
9.4.3 Replacement of motor
・・・・・・・・・・・・・・・・・・・
39
9.4.4 Replacement of lithium batterty
・・・・・・・・・・・・
40
9.4.5 Replacement of power supply unit
・・・・・・・・・・
41
9.4.5 Ram clearing
・・・・・・・・・・・・・・・・・・・・・・・・・
41
10.
Specifications
・・・・・・・・・・・・・・・・・・・・・・・
42
10.1
Detector unit ・・・・・・・・・・・・・・・・・・・・・・・・42
10.2
Fiver optics ・・・・・・・・・・・・・・・・・・・・・・・・・43
10.2.1 Reflective type fiver optics IR-WCRN
・・・・・・・
43
10.2.2 Output checker plate for reflective type
fiver optics IR-WCNB
・・・・・・・・・・・・・・・・・
43
10.2.3 Raflective type fiver optics with lens IR-WCRE 43
10.2.4 Output checker plate for reflective type
fiver optics with lens IR-WCREB
・・・・・・・・・・
43
10.2.5Measuringcell for liquid IR-WCC1
・・・・・・・・
44
10.2.6 Transparent type fiver optics IR-WCT
・・・・・・
44
10.3
Accessories ・・・・・・・・・・・・・・・・・・・・・・・・・44
10.3.1 Power supply unit (conforming to CE) IR-WEP 44
10.3.2 Connection cable IR-WERT
・・・・・・・・・・・・・
44
CONTENTS
!
!
!
!

INST No. INE-403-0P0
−1−
1.1 Summary
The infrared moisture meter IR-M2 series is designed as a fiber optic moisture meter for
on-line moisture meter utilizing infrared absorption of moisture, and it consists of a detector
unit, a fiber optics,a setting display unit and accessories.
The detector unit is provided with 99 built-in calibration curves to output linear analog
signalsaccording to the moisture.
Since the detector unit provides keys, digital displays and communications function, it can be
used as a single unit or combined with a personal computer.
Two kinds of fiber optics, reflective type and transparent type, are available. The reflective
type fiber optics is for moisture measurement in powder state or in sheet state objects and the
transparent type fiber optics isfor moisture measurement in liquids by combining with a
measuring cell for liquids.
Up to 9 detector unit can be connected with the setting display unit which can set various
parameters of the detector unit and displays the moisture value digitally.
Accessories including cablesand output checker plate are available for various applications.
This manual describes about the detector unit, the fiber optics (reflective type and transparent
type), the measuring cell for liquids and accessories.
For the setting display unit, refer to a separate instructionmanual for it.
1.2 Configuration
1. General
C H DATA
Transparent type fiberoptics
IR-WCT
Reflectivetype fiberopticswithlensattachment IR-WCRE
Reflectivetype fiberoptics IR-WCRN
Measuring
cell forliquids
IR-WCC1
Detector:ModelforIR-M2
Analog output
IR-WERT
(connection
cable)
24VDC
IR-WEP
(powersupplyunit)
IR-GMEG1
(setting displayunit)
Optical
fiber
RS-485
Communication and
Analogoutput

INST No. INE-403-0P0
−2−
2.1 Models
2.1.1 Detector unit IR-M2
Measurement object1: Universal type
2:High moisture type
Fiber optics combined0:Reflective type
5:Transparent type
External input/output0: Moisture output 4 to 20mA (Standard)
5: Correction input 4 to 20mA (Option) *1
8:Correction input Pt100 (Option) *1
Communications interface
S :RS-485 (Standard) *2
R :RS-232C (Option)
A :RS-422A (Option) *1
Special specification
Blank: None
x: Other special specification
2.1.2 Reflective type fiber IR-WCRN
2.1.3 Output checker plate for IR-WCRNB
reflective type fiber optics
2.1.4 Reflective type fiber optics with lens attachment IR-WCRE
2.1.5 Output checker plate for reflective
type fiber optics with lens attachment IR-WCREB
2.1.6 Transparent type fiber optics IR-WCT
2.1.7 Measuring cell for liquids IR-WCC1
2.1.8 Connection cable IR-WERT
2.2 Accessories
Component name Quantity
Remarks
Power supply unit (IR-WEP) 1no. S82K-05024 (Omron)
Fuse 2 nos. SMP20 (Daito)
Instruction manual 1 copy This document
Communications instruction manual 1 copy Attached to IR-M2R/A only
Inspection certificate 1 copy
2. Models and accessories
*1:
Combination of “5”
or “8” of
optional
external I/O and
communications
interface “A” is not
possible.
*2:Specify “S” for
combination with the
setting display unit.

INST No. INE-403-0P0
−3−
3.1 Setting of dipswitches
Set the internal dipswitches before installation of the detector unit.
1) Open adoor, remove 4 screws for fixing a setting display part and then remove the setting
display part.
2) The setting display part is connected to the interior through cables. Lift the setting display
part with due care so as not to damage the cable and disconnect the connector for the
setting display part at the position shown in the figure.
3) Set the dipswitches according to the working conditions.
3. Setting
Turn off the power supply before removing the setting display part.Warning
Loosen thesefourscrews
CH DATA
△
△▽
SELENT
CH/
□
DIPswitch 1
Connectorfordisplay
!

INST No. INE-403-0P0
−4−
3.1 Setting of dipswitches SWDipswitch
No Function
OFF ON Defaults
1 Measurement/
Calibration mode Measurement
mode Calibration
mode OFF (Don’t change.)
2 Mathematical
operation format Logarithm
operation Ratio operation ON (Don’t change.)
3 No.2 calibration
enable/disable Disable Enable OFF
4 Connection
destination
Setting display
unit PC OFF :IR-M2 S
ON :IR-M2R
IR-M2A
5 Key/display
enable/disable Enable Disable OFF
6 Not used OFF
7
Correction input
enable/disable Disable Enable OFF
8 Correction input 4 to 20mADC Pt100 OFF
3. Setting
12345678
O
N
DIPswitch 1

INST No. INE-403-0P0
−5−
3.2 Installation of reflective type fiber optics
There are two types of reflective type fiber optics. One is “Without lens (IR-WCRN)” and
the other is “With lens (IR-WCRE)”.
3.2.1 Installation of tip part
Set an attached holder to the installation place first. Insert the tip of fiber optics into the hole
of the holder and fix it by tightening the tip fixing screw. Two types of holder, horizontal
mounting type and flange type are attached.
The tipofreflection
type fiberoptics with
lensattachment
The tip ofreflectiontype
fiberoptics withlensattachment
23 27
84
64
(Ø30)
(Ø
30)
64
80 12
30
42
12
70
28
2-
Ø9
Mounting hole
2-
Ø9
Mounting hole
Tipfixing screw
Tipfixing
screw
Place the tip of fiber optics with t
he following measurement distance from
an object. ・Reflective type fiber optics 15 to 50mm
・Reflective type fiber optics with lens 25 to 100mm
Warning
Fix the fiber optics with appropriate metal toolsat several places
for
avoiding affect with vibrations, etc. Make sure not to bend the fiber
optics
below its minimum bending radius (100R).
Warning
[ModelforIR-WCRN ]
50mm
15mm
Ø
50
Ø
20
Ø
6
[ModelforIR-WCRE]
100m
25mm
Ø
40
Ø
25
Ø
20
Tipfixing screw
17
14
40
50 16
7
The tipofreflectiontype fiberoptics
(Ø
16)
2-
Ø
6.6Mountinghole
(Ø16)
2-
Ø5.5
Mountinghole
Thetip ofreflection
typefiberoptics
44
8
27
16
48
38
25
Tipfixingscrew
1)Reflection-typefiber
2)Reflectiontypefiberoptics
withlensattachment
3. Setting
!

INST No. INE-403-0P0
−6−
Purgeair
inlet
Purge airflow:50 to 200Nl/min
Purge airpressure:Lowerthan200kPa
3.2.2 Air purge (Only IR-WCRN)
When air purging is required, supply dry air not
containing oil, dust,etc.from apurge air inlet.
Connect a nylon tube of φ6mm in outer diameter and
φ4mm in inner diameter to the purge air inlet.
(When the reflection type fiber optics is used in
ambient temperature more than 60°C, remove the
purge air inlet and pipeto the screw hole of RC1/8
directly.
Purge air flow: 5 to 20NI/min
Purge air pressure: Less than 200kPa (2kgf/cm2)
3.2.3 Connectionswith detector unit
Connect the side being separated into two parts of light projecting side and light receiving
side with the fiber optics connectorsof the detector unit.
The fiber optics has “Light projecting
side (D)”and “Light receiving side
(E)”and each one has a carved mark. Connect the fiber optics through the
rubber bush as shown in the following figure.
Warning
3. Setting
!
Connectorforfiber
[Lightprojectingside]
Connectorforfiber
[Lightreceivingside]
Dentdepartment
Protrusiondepartment
DE
Rubberbush
Bushfittingring
Connectorring
Opticalfiber
D:Connect tolightprojectingside E:Connecttolightreceivingside

INST No. INE-403-0P0
−7−
3.2.4 Optical path
Select a place where water drops, dust particles, smoke, water steam,etc. will not enter
between the tip of fiber optics and an object. If these substances cannot be prevented and
their influences are negligible, purgethem with air, or other remedial measure is necessary.
If direct current (DC) light like the natural light, which does not change its intensity
momentarily, comes to an object, it does not cause any disturbance in measurement, but
alternate current (AC) light like as electric bulb, flame,etc., causes disturbance. Further,
strong light such as direct sunlight also causes disturbance. In such cases, it is necessary to
shade the disturbing light to prevent it from direct shining on an object.
3.2.5 Measuring place (for objects of pulverizedor granular substance)
For installation of the tip of fiber optics on on-line, select a place where the moisture
difference is small between the surface and the inside of an object, like as at a hopper outlet
or just behind the conveyor belt transfer place. If such a place cannot be selected, face the
inside to the surface by using a baffle plate during measurement as shown in the figure.
3.2.6 For the objects in sheet state such as paper,etc.
Tilt the tip of fiber optics by 15°from the vertical face as shown in the figure.
3. Setting
Baffle plate
Measuring object
Beltconveyor
Paperor
othersheets
15°

INST No. INE-403-0P0
−8−
3. Installation of transparent type fiber optics
3.3.1 Installation of cell for liquids
Install a measuring cell to flow a sample liquid from down to up. Connect with the fiber optics
between the detector unit and the measuring cell. Make sure to connect the light projecting side
and the light receiving side of the fiber optics to the correct sides of the detector unit and the
measuring cell.
Fix the fiber optics with appropriate metal toolsat several places for
avoiding affect with vibrations, etc.
Make sure not to bend the fiber optics below its minimum bending
radius
(250R).
Warning
C H D ATA
Detector:ModelforIR-M2
Powersupplyunit:
ModelforIR-WEP
Setting displayunit:
ModelforIR-GMEG1
Measuring cell
Output
Inputof
sampleliquid
Lightreceiving side fiberoptics
Lightprojecting side fiberoptics
Fixthe fiberoptics with appropriatemetaltoolsatseveralplaces.
3. Setting !

INST No. INE-403-0P0
−9−
3.3.2 Connection with detector unit
Connect the fiber optics, as shown in the following figure,through the rubber bushattached
to the detector unit.
3.3.3 Disassemble cleaning
For accurate measurements, recommend periodical cleaning of the cover glass in a liquid
contacting part of the measuring cell.
1)Remove the fiber optics (both of light
projecting side and light receiving side),
from the measuring cell.
2)Next, remove 4 screws fixing aconnecting
part and remove the connecting part from
the liquid-contacting part.
3)Remove a fitting ring by using the tool A.
Then after screwing the tool B into the
cover glass, pull out the cover glass from
the liquid-contacting part.
4)Wipe off dirt on the glass surface with a
gauze,etc.
5)Confirm that an Oring is inserted in
liquid-contacting part and push in the
cover glass to the liquid-contacting part.
6)Match the positionsof E and D characters
of the connecting part and the liquid
contacting part,and fix the connecting
part.
7)After the cleaning of the cover glass, make
sure to re-calibrate your moisture meter.
3. Setting
OUT
IN
Connectingpart
4-Fixingscew
Liquidcontacting
part
ToolAToolB
Fittingring
Curler
Coverglass
Oring
Connectorforfiber
[Lightprojecting side ]
Connectorforfiber
[Lightreceivingside]
Dentdepartment
Protrusion department
DE
Rubberbush
Bushfitting ring
Connector ring
Opticalfiber
D:Connect tolightprojectingside E:Connect tolightreceivingside

INST No. INE-403-0P0
−10−
3.4 Installation of detector unit
The detector unit is a wall-mounting type with 4 mounting holes. Fix the detector unit to wall
surface,etc. with 4 pieces of M8 screw. The outside and mounting dimensions are shown in
the figure.
3.4.1 Outside and mounting dimensions of detector unit
115 2
270
180
3 03 0
98 10
12
2323
74.5
47
47
165
145
125
4-M8Mounting hole
Vibrations and impacts
Use the detector unit in a place where is free of any noticeable
vibrations
or impacts.
Warning
Induction
The detector unit is designed for anti-induction. However separate
the
detector unit from induction heating oscillatorsor power
lines as far as
possible.
Warning
Ambient temperature and humidity
The ideal ambient temperature of the detector unit is 10 to 30°C
and the
ambient humidity as low as possible is recommended. In high
humidity
environment,
you may have error in measurement due to condensation of
steam bychange in temperature.
Warning
3. Setting !

INST No. INE-403-0P0
−11−
4.1 Connections
The terminal board is accessible by opening the door of the detector unit. Lead in cables
through a small sized cable inlet at the lower part and connect them to terminals. Lead in the
exclusive cable IR-WERT through the large cable inlet.
4.2 Layout of connection cable
For connection of cables, be careful of the following points.
4. Connections
For preventing electric shock,
make sure to turn the power supply off
before wiring.
Warning
Caution
Use a cable conforming to the applicable outer diameter (
φ
4.5 to 6.5
mm)
of the small cable inlet; otherwise waterproofing of the
detector unit may
deteriorate. Tighten the nut at the cable inlet securely after connections.
・Separate the cables from induction heating oscillator and power lines.
・Keep the cables free of deposit of water, oil, etc.
・Don’tbend the cablesextremely or apply any excessive force to them.
・
For permanent layout, protect the cables with conduits, etc.
SA
(Black)
SB
(White)
SG
(Brown)
G*1
(Green)
P+
(Red)
P-
(Blue)
P-
(Blue)P+
(Red)G
(Green)
V+V-
LN
24VDC
Ground
Powersupply
(IR-WEP)
100-120V AC
200-240V AC
Terminal
board
Cableinlet(small):
Outerdiameterof
applicablecable
Ø
4.5to
Ø6
.5
Cableinlet(large):
[ForIR-WERT]
Outerdiameterof
applicablecable
Ø8
.5to
Ø10
.5
Connection cable
(IR-WERT)
Communication output
forIR-GMEG1,personal
computerand PLC.
Communication outputRS-485
(Optional:RS-232C,RS-422A)
*1:
Don'tconnect this cable usually,
butconnectit, if the unitisinterfared
withnoises.
SA
SB
SG
OUT
Measured value output :
4to 20mADC
Load resistance:Lowerthan 500Ω
(Optional:RS-232C,RS-422A)
P+
P-
24VDC
1.5A
SUPPLY
!

INST No. INE-403-0P0
−12−
4.3 Multi-detector connections
By using the setting display unit, you can connect the detector unit up to 9 sets.
Two wiring methods are available. One is wiring through a relay box (IR-WEE) and the other
is without the relaybox (IR-WEE).
4.3.1 Connections through relay box
4. Connection
Caution
For
both wiring methods, before connecting with the setting display unit or
before turning on the power supply of the setting display unit,
number
setting to the detector units is required. For the number setting, be careful
not to overlap any detector unit number of each detector unit.
Refer to [6.2 Setting of detector unit number] for the details.
Caution
1. Shorten the length of exclusive cable (IR-WERT * mark) as much as
possible.
2. For wiring to the relaybox, refer to “Instruction manual of relaybox”.
Detector
IR-WERT*
IR-WERT*
IR-WERT*
Relaying box(Powersupplyunitisbuiltin)
Relaying box(Powersupplyunitisbuiltin)
Relaying box(Powersupplyunitisbuiltin)
2-conductortwisted
cable(withshield)
2-conductortwisted
cable(withshield)
2-conductortwisted
cable(withshield)
100-240V AC
Detector
Detector
Setting diplayunit

INST No. INE-403-0P0
−13−
4.3.2 Connections without relay box
4. Connection
Detector
Detector
Detector
IR-WERT*
IR-WERT*
IR-WERT*
Setting diplayunit
100-240 V AC
2-conductortwisted
cable(withshield)
2-conductortwisted
cable(withshield)
2-conductortwisted
cable(withshield)
Taping
Taping
Taping
Connect the P+(Red),P-(Blue),
and G(externalshieldwire)of
IR-WERTto powersupply.
Andconnect the SA(Black),
SB(White)and SG(internalshield
brown)ofsignalwiretothe
2-conductortwistcable(withshield)
Fasten the signalwiresin
IR-WERT(SA,SB and SG)
and thesignalwirein 2-conductor
twistcable(SA,SB and SG)together
at the terminalboard of the detector.

INST No. INE-403-0P0
−14−
5.1 Names and functions of detector unit
(1) Hole for installation of detector unit:Four M8 holes for installation of the detector unit
to wall surface, etc.
(2) Data display window:Displays calibration curve number and measuredvalue.
(3) Door:By opening the door, key switches and the terminal board are accessible.
(4) Cable inlet (small): Cable inlet for connection cable. Outside diameter of connection
cable applicable to this inlet:φ4.5 to 6.5mm.
(5) Cable inlet(large)Cable inlet forexclusive cable IR-WERT. Outside diameter:φ10mm.
(6) Serial No plate:Serial No plate of this detector.
(7) Connector on light projecting side: Connector for the light projecting side of the fiber
optics. Rubber bush is prepared for the connection.
(8) Connector on light receiving side:Connector for the light receiving side of the fiber
optics. Rubber bush is prepared for the connection.
(9) Setting display: Display part of this detector.
(10) Measurement part: Optical assembly, detector element and electric circuitsare built-in.
(11) Lamp housing: Lamp is located inside.
(12) Power supply part: Terminal board for wiring, power fuse, power supply unit and
output board are built in.
(13) Power fuse: A power fuse is mounted to the fuse holder.
Power fuse:Insertion type, 2A, time lag fusion type
(14) Terminal board:For wiring of connection cable. Terminus screw size: M4.
5. Names and functions of component parts
DATACH
SEL CH/・ ENT
DATACH
(1)Detector mounting
screwhole-2
(2)Data display
window(3)Door
(5)Connection cableinlet(small)
(6)Connection cableinlet(large)
(6) Manufacturing
nameplate
(8)Connectoron lightreceiving side
(7)Connectoron lightprojecting side
(11)Lamp(9)Setting display
part
(14)Terminalboard
(10) Measuring part
(12)Powersupplypart
(13)Power
supply
fuse
For preventing electric shock,
make sure to turn the power supply off
before wiring.
Warning
!

INST No. INE-403-0P0
−15−
5.2 Names and functions of setting display part
(1) CH display: Displays a calibration curve number in measurement mode and a parameter
item in setting mode.
(2) Data display Displays a measured value in the measurement mode and a parameter in
the setting mode.
(3) Setting keys
SEL : For entering into the setting mode and switching a setting screen.
CH/・: For setting a calibration curve number in the measurement mode and for
entering a decimal point in the setting mode. When the key is pressed again,
the decimal point disappears.
: For entering into a parameter setting in the setting mode and changing the digit
of the parameter.
:Forchanging the parameter in the setting mode.
:Forchanging the parameter in the setting mode.
ENT : For storing the parameter in the setting mode.
5. Names and functions of component parts
CH/・
SELENT
CHDATA
Settingdisplaypart
(1)CH Display(2)DataDisplay
(3)Setting
keys
This manual suits for next models
1
Table of contents
Other Chino Measuring Instrument manuals

Chino
Chino AH4000 User manual

Chino
Chino IM Series User manual

Chino
Chino KL4000 Technical specifications

Chino
Chino IRMA IRGMEG3 Series User manual

Chino
Chino LE5100 User manual

Chino
Chino AL3000 SERIES User manual

Chino
Chino IRGMEG2 Series User manual

Chino
Chino IM Series User manual

Chino
Chino IR-HA Series User manual

Chino
Chino AH4000 User manual
Popular Measuring Instrument manuals by other brands

COMAC CAL
COMAC CAL FLOW 32 Installation and technical conditions

LTB
LTB MNL X00 Series operating manual

BMS
BMS EASY-CHECK FN Operational manual

Svantek
Svantek SVAN 958 user manual

FLIR
FLIR EXTECH INSTRUMENTS PQ34-12 user guide

AMALGAMATED INSTRUMENT
AMALGAMATED INSTRUMENT RM4-FX Operation and instruction manual