Clarity Platinum User manual

®
Clarity Planum Urine Analyzer
Operator’s Manual
We Support Beer Outcomes™

1. INTRODUCTION ................................................................................................... 1
Intended Use ...................................................................................................................... 1
GengStarted ................................................................................................................... 2
UnpackingtheClarityPlanumUrineAnalyzer.................................................................. 2
ClarityPlanumUrineAnalyzerHardwareOverview......................................................... 2
AssemblingtheClarityPlanumUrineAnalyzer ................................................................ 3
LoadingthePrinterRoll ...................................................................................................... 4
PoweringOn/O................................................................................................................. 5
AccessingtheMainMenu .................................................................................................. 6
ClarityPlanumUrineAnalyzerUserInterface .................................................................. 7
2. OPERATIONS ....................................................................................................... 8
2.1 PERFORMING A URINALYSIS TEST .................................................................. 9
PreparingaUrinalysisTest.................................................................................................. 9
PerformingaUrinalysisTest ............................................................................................. 10
ViewingtheUrinalysisTestResult .................................................................................... 10
PrinngtheUrinalysisTestResult..................................................................................... 11
ComplengtheUrinalysisTest ......................................................................................... 11
3. TEST RESULTS MANAGEMENT ............................................................................ 12
RecallAllResults ............................................................................................................... 12
SearchResultsbyID.......................................................................................................... 12
SearchResultsbyDate ..................................................................................................... 13
DeleteResults................................................................................................................... 13
4. SYSTEM CONFIGURATION .................................................................................. 14
4.1 BASIC SETTINGS ............................................................................................. 14
StripType.......................................................................................................................... 14
Language .......................................................................................................................... 14
Date&Time...................................................................................................................... 14
UnitSeng....................................................................................................................... 15
PrinterSeng................................................................................................................... 15
SystemInformaon .......................................................................................................... 15
RestoreDefaultSeng..................................................................................................... 15
(Connued on next page.)
WeSupportBeerOutcomes™
Table of Contents

4.2 ADVANCED SETTINGS .................................................................................... 16
Operator ........................................................................................................................... 16
TestMode......................................................................................................................... 16
QualityControl ................................................................................................................. 16
PerformaQualityControlTest ......................................................................................... 18
4.3 DEVICE DIAGNOSTICS .................................................................................... 20
5. TROUBLESHOOTING ........................................................................................ 21
FailureModeErrorMessageandCorrecveAcon......................................................... 22
6. MAINTENANCE ................................................................................................ 28
RemovingandReplacingtheStripBed............................................................................. 28
Daily/RouneCleaningoftheStripBed ........................................................................... 29
WeeklyCleaningoftheStripBed ..................................................................................... 29
CleaningtheAnalyzerCase............................................................................................... 29
CleaningtheTouchScreen................................................................................................ 29
APPENDIX A. Clarity Planum Urine Analyzer Specicaons .................................. 30
APPENDIX B. Clarity Planum Urine Analyzer Default Sengs ............................... 30
APPENDIX C. Clarity Planum Urine Analyzer Unit Convenon Table ..................... 31
APPENDIX D. Symbols ............................................................................................ 33
WeSupportBeerOutcomes™
Table of Contents

WeSupportBeerOutcomes™
Clarity Planum Urine Analyzer
OPERATOR’S MANUAL
Page1
1. Introducon
Theintroduconexplainshowtounpack,install,andbeginusingtheClarity
PlanumUrineAnalyzer.Theintroduconalsoincludesanoverviewofthe
analyzeranditskeyfeatures.
TheClarityPlanumUrineAnalyzerTestSystemisforCLIAWaivedUseand
onlytobeusedwithClarityDiagnosc’sUrineReagent(CLA-10P)Strips.This
testiswaivedundertheCLIA’88regulaons.Thistestisonlywaivedforurine
specimens,andfailuretoadheretotheinstruconsforusewillresultinthe
testbeingconsideredhighcomplexityandsubjecttoallCLIArequirements.
ACLIACercateofWaiverisrequiredtoperformthistest.ACercateof
WaivercanbeobtainedfromtheCentersforMedicare&MedicaidServices
(CMS).Visitwww.cms.govtoobtainanapplicaon(FormCMS-116).Youmust
followthemanufacturer’sinstruconstoperformtests.Youshouldreadthe
completetestprocedurebeforeperformingthetest.
Intended Use
TheClarityPlanumanalyzerisabenchtopanalyzerdesignedtoreadClarity’s
proprietaryurinalysisstrips(CLA-10P).Theinstrumentisintendedtobe
usedtogetherwithClarityDiagnoscs’UrineReagentStripsasasystemfor
semi-quantavedeteconofGlucose,Bilirubin,Ketone,SpecicGravity,
Blood,pH,Protein,Urobilinogen,NitriteandLeukocytesinurine.These
measurementsareusedtoaidinthediagnosisofmetabolicdisorders,
kidneyfunconanomalies,urinarytractinfecons,andliverfuncon.
TestsperformedbytheClarityPlanumanalyzerareintendedforin vitro
diagnoscuseonly.

Clarity Planum Urine Analyzer
OPERATOR’S MANUAL
WeSupportBeerOutcomes™Page2
Geng Started
Unpacking the Clarity Planum
TheClarityPlanumanalyzerisdeliveredinonebox.Tounpacktheanalyzer,
carefullyremoveallcomponentsfromtheboxandinspectforanysignsof
damage.Iftheanalyzerappearstobedamaged,immediatelycontactyour
shippingcarrierandClarityDiagnoscs.Eveniftheboxappearstobedamaged,
thedevicecanbepreservedduetotheinternalpackaging.Verifythatyouhave
eachofthefollowingcomponents:
• ClarityPlanumAnalyzer
•ThermalPrinterPaper(preloadedinanalyzer)
•PowerSupplyAdaptorwithACPowerCord
•ClarityPlanumManual
• QuickStartOperator’sManual
• BarcodeReader
Ifyouaremissinganycomponents,pleasecontactClarityDiagnoscs
immediatelyat(877)485-7877,Monday-Friday,8am-6pmEST.
Clarity Planum Urine Analyzer Hardware Overview
TheClarityPlanumconsistsofthefollowinghardwarecomponents:
•TouchScreenDisplay
• TestStripBed
•Printer
•PowerCord
• ConneconPorts

WeSupportBeerOutcomes™
Clarity Planum Urine Analyzer
OPERATOR’S MANUAL
Page3
Figure 1. Clarity Planum
1. Touch Screen Display
2. Test Strip Bed
3. USB-A Port, Barcode Reader
4. USB-C Port, Data Transfer
5. Power Cord (DC IN)
6. Power Switch
7. Printer Cover
Assembling the Clarity Planum
1. PlacetheClarityPlanumanalyzeronalevelsurfaceawayfrom
sourcesofheatandliquids.TheClarityPlanumanalyzerperforms
opmallybetween59°-86°F(18–30°C).
2. Removethebarcodereaderfromtheboxandconnectittothe
USB-APortonthebackofyouranalyzerwhereitislabeled
‘BarcodeReader’.(Figure1,number3)
3. Plugthepowersupplyadaptorcordintothebackoftheanalyzer
whereitislabeled‘DCIN’.(Figure1,number5)
4. Connectthepowersupplyadaptorcordintoa120VACwalloutlet.
1
23 4 5 6 7

Clarity Planum Urine Analyzer
OPERATOR’S MANUAL
WeSupportBeerOutcomes™
Figure 2. Printer
Loading the Printer Roll
1. Opentheprintercoverbygentlyliingtheexternaltablocatedatthe
backoftheanalyzerupwardandliingthefrontpinsoutoftheirslots
inthefront.Setasidethecover.
2. MakesuretheClarityPlanumispluggedinfortheautomacpaper
feedtofuncon.
3. Placetherollofthermalpaperinthecradleatthebackofthemachine,
withtheshinysideofthepaperfacingthefrontofthemachine.
4. Feedtheendofthethermalpaperrollintotheboomoftheblack
rollerunlitiscaughtbytheauto-feedandpulledintotheroller.
5. Gentlyreplacetheprintercoverbyhookingtheplascfrontpinsrst,
withtheClarityDiagnoscslogofacingthefrontofthemachine.
Feedthepaperfromtheblackrollerthroughtheslotwiththe
perforatededgeonthecover,andthenpushtheexternaltabdown
thebackofthemachineunlthecoverisbackinplace.
Page4

WeSupportBeerOutcomes™
Clarity Planum Urine Analyzer
OPERATOR’S MANUAL
Page5
Powering On/O
To Power On the Clarity Planum Analyzer:
1. Pressthepowerswitchlocatedonthebackoftheunit.
(Figure1,number6)
2. TheClarityPlanumwillperformsystemdiagnoscseachmethe
unitispoweredon,whichchecksthattheopcal,electronic,and
mechanicalsystemsareperformingasexpected.Ifthesystem
diagnoscsaresuccessful,theprinterwillautomacallyprint:
‘SystemDiagnoscsOK.ReadytoUse’,andthemainmenu
screenwillappear.
3. Ifanerroroccursduringinializaon,theanalyzerwilldisplaya
‘SystemCalibraonFailed’errormessageandtwooponswill
appearonscreen,‘Recalibrate’or‘TechnicalSupport.’
4. Presstherecalibratebuontorerunsystemdiagnoscs.Iftheerror
persists,pleasecontactClarityDiagnoscsTechnicalSupportat
(877)485-7877,Monday-Friday,8am-6pmEST.
To Power O the Clarity Planum Analyzer:
1. Ensurethatthestripbedisemptyandthatnotestsareinprogress.
2. Usewarmwaterandacoonswabtocleanthestripbed.
3. Pressthepowerswitchlocatedonthebackoftheunittoturnothe
analyzer.Pressingthepowerswitchwillalsoautomacallyretractthe
stripbedintothemachine.(Figure1,number6)

Clarity Planum Urine Analyzer
OPERATOR’S MANUAL
WeSupportBeerOutcomes™Page6
Figure 3. Clarity Planum Main Menu
From the main menu, you may select:
1. Test Strip
Thiswillbringyoutotheurinesampletest.
2. Results
FromtheResultspageyou,canrecallpreviousresultsandsearchstored
testresultsbasedonDate,PaentIDandTestID.
3. Sengs
Inthissecon,youcanaccesssengsforstriptype,displaylanguage,
date&me,andenable/disabletheautomacprinngfeatureandformat.
Youcanalsoviewsysteminformaon.
TheRestore Default SengfunconisforClarityDiagnoscs
administratorsonlyandrequiresavalidadministratorpasswordforaccess.
4. Advanced
TheAdvanced sengsmenuisaccessibletoalloperators.TheCLIA
Waivedoperatorcanperformfunconslikequalitycontroltesng,system
diagnoscs,andstripbedmaintenance.Moreadvancedfeatures,like
addingauthorizedoperators,modifyingoperatorpermissions,modifying
qualitycontrolsengs,selecngtestmode,andenablingordisabling
barcodescanning,areintendedforClarityDiagnoscsadministratorsonly
andrequireavalidadministratorpasswordforaccess.
Note: The strip bed can be retracted from the Strip Bed Device Maintenance
secon of the Advanced sengs menu. Retracng the strip bed can prevent
damage to the Planum while not in use. Full instrucons can be found in
Secon 6: Removing/Replacing Strip Bed.
Accessing the Main Menu
AerturningontheClarityPlanum
analyzer,thesystemdiagnoscswill
runautomacally,andthemachine
willbringyoutothemainmenu.

WeSupportBeerOutcomes™
Clarity Planum Urine Analyzer
OPERATOR’S MANUAL
Page7
Clarity Planum User Interface
TheClarityPlanumanalyzerusesatouchscreentodisplayuseropons,
results,sengs,anddiagnoscs.Thetouchscreenisusedforalluserinput.
Thebarcodereaderaachmentcanalsobeusedforinpungbarcodesfrom
theteststriporcontrolboles,aswellaspaentIDs.(SeeSecon 4.2:
Advanced Sengs)
Do not use any sharp or pointed objects on the screen, as they may damage
the display.
To Operate the Clarity Planum Analyzer:
• PresstheTest Stripicondisplayedonthescreento
runaUrineTestStrip.
• Scanorenterthebarcodefromthestripboleandpressenter.
• ScanorenterthePaentID,orskipitbypressingenter.
• FollowtheinstruconsintheQuickStartGuidetoperformatest.
• Presstheback/returnbuonontheboomleofthescreen
toreturntothemainmenu.

Clarity Planum Urine Analyzer
OPERATOR’S MANUAL
WeSupportBeerOutcomes™Page8
2. Operaons
By pressing the Test Strip buon on the main menu screen, you can perform
a urinalysis test.
Note: The Barcode Reader feature has been enabled (default conguraon)
and you will be prompted to scan or enter the barcode on the test strip bole
prior to reaching the Test Preparaon screen.
Entering Strip Barcode Informaon
ToentertheStripBarcodeinformaon,performthefollowingsteps:
1. Aerpressingthe Test Stripbuon,usethekeypadonthescreenor
theClarityPlanumbarcodescannertoenterthebarcodefromthe
CLA-10Pbole.
2. Aerenteringthebarcode,pressEnter toproceedtothenextstep.
Note: Test known posive and negave specimens/controls at each of the
following events in accordance with local, state, and/or federal regulaons or
accreditaon requirements. For CLIA Waived sengs, Clarity Urinalysis Quality
Controls must be tested under the following condions:
• When a new canister of strips is opened OR
• Test results appear to be inaccurate OR
• A new operator uses the analyzer OR
• Each new day of tesng OR
• Aer performing maintenance or service on the analyzer.
(See Secon 4.2: Advanced Sengs, Quality Control)

WeSupportBeerOutcomes™
Clarity Planum Urine Analyzer
OPERATOR’S MANUAL
Page9
2.1 Performing a Urinalysis Test
Pressing the Test Strip buon on the main menu screen will bring you directly
to the main tesng screen.
Note: You will be prompted to scan or enter the barcode on the test strip bole
prior to reaching the test preparaon screen.
Preparing a Urinalysis Test
1. Collectaurinesample,aCLA-10Purineteststrip,apapertowel,and
havetheanalyzerready.
2. AerpressingtheTest Stripbuon,usetheClarityPlanumscanneror
thekeypadonthescreentoenterthebarcodefromtheCLA-10Purine
reagentstripbole.
3. Next,theanalyzerwillpromptyouforaPaentID.Youcanusethe
scannerorkeypadtoinputthatinformaon,followedbyEnter,or
selectEntertoskip.
4. The Preparing Test screenwillappearandpromptyoutoperformthe
followingaconsbeforepressingtheStartbuon:
• Checktheexpiraondateontheteststripbole.Discardifit
isexpired.
•Removetheteststripfromtheboleandreplacethecap.
• Dipthestripintotheurinesample.
• PressStarttobeginthetest.
5.PressHelptoaccessmoredetailedinstrucons.Atanyme,youmay
returntothemainmenubypressingthebuon.
Biohazard:
Wearpersonalprotecveequipmentwhenhandlingthe
paentspecimenandperformingthetest.Useuniversal
precauonsforanybiohazardmaterials.

Clarity Planum Urine Analyzer
OPERATOR’S MANUAL
WeSupportBeerOutcomes™Page10
Performing a Urinalysis Test
1. Dipthestripintotheurinesampleandremoveitimmediately.
Ensurethatallpadsonthestriparesaturatedwiththeurinesample.
Whileyouaredoingthis,presstheStartbuon.
2. AerpressingtheStartbuon,theanalyzerbegins10secondsof
self-calibraon.Duringthisme,touchthelongedgeofthestriponto
apapertoweltoremovetheexcessurine,andplacethestriponthe
stripbedwiththetestpadsfacingup.Ensurethattheteststripis
insertedfullyintothebackofthestripbed.Thedippingstepshould
notexceed10seconds.
3. Whencalibraoniscomplete,theanalyzerretractsthestripbed
withthestrip,anditwillbeginanalysis.
4. Ameronscreenwillcountdownthemeremaininginthestrip
analysisprocess.
Note: You cannot cancel a test before the analyzer nishes the test.
Viewing the Urinalysis Test Result
1. Aerthecountdownends,theanalyzerdisplaysthetestresulton
thescreen.
2. PresstheResultsbuontoaccessthestoredresults,orpressthe
Next Testbuontoperformanothertest.
Cauon:
DONOTpushorpullthestripbedormovetheanalyzer
duringthetest.Itmightcausetheanalyzertofailcalibraon.

WeSupportBeerOutcomes™
Clarity Planum Urine Analyzer
OPERATOR’S MANUAL
Page11
Prinng the Urinalysis Test Result
Oncethetestiscompleted,thetestresultisprintedautomacallyifthe
machineisinthedefaultseng.Thetestresultsprintoutcouldincludeany
ofthefollowinginformaon,dependingonthesengs:
• OperatorID
• PaentID
• TestID
Youcanchangetheprintersengtomanualortheprintoponsinthe
Printer Sengmenu,asdescribedinSecon 4.1: Basic Sengs.
Compleng the Urinalysis Test
1. Removetheteststripfromthestripbedanddisposethestrip
accordingtoyourlocal,state,andfederalguidelines.
2. Cleanthestripbedusingasoluonofsodiumhypochlorite
525-615ppm(diluon1:100of5.25-6.15%sodiumhypochlorite)
or70%isopropylalcoholwithacoonswab.
3. SelectResultstoviewthestoredresults,orselectNext Test to
performthenexttest.
4.Youmayreturntothemainmenubypressingthebuon.

Clarity Planum Urine Analyzer
OPERATOR’S MANUAL
WeSupportBeerOutcomes™Page12
3. Test Result Management
PresstheResults buononthemainmenuscreentoaccessupto2,000
previouslystoredresults.YoumaysearchresultsbyIDortestdate,ordelete
allresults.
Note: When the results list reaches 2,000 tests, the analyzer deletes the oldest
test from the list. You cannot recall the deleted test.
Recall All Results
1. Ontheresultscreen,selectRecall Result toreviewallthestoredtest
resultsarrangedinchronologicalorder.
2. Thescreenwilldisplaythemostrecenttestresult,includingthestrip
type,testdate&me,paentIDandtestresult.
3. SelectPrinttoprintthetestresultusingtheanalyzer’sbuilt-in
thermalprinter.
4. Ontheresultscreen,selectSend tosendthetestresulttoaPC,
hostcomputer,orLIS,ifconguredforthisopon.
5. Ontheresultscreen,presstheoriconnexttothetestID
toviewthepreviousornextresult.
6. Pressattheboomofthescreentoreturntotheresults
mainmenu.
Search Results by ID
1. Ontheresultscreen,selectSearch by IDtosearchthestoredtest
resultsbasedonIDinformaon.
2. IfyouselectPaent ID,enterthepaentIDusingthekeypadonthe
screenandpressENTER.Allthematchingresultswillbelisted.Select
thedesiredtestresulttoreviewonthescreen.
3. IfyouselectTest ID,enterthetestIDusingthekeypadonthescreen
andpressENTER.Thematchingresultswillbedisplayedonthescreen.

WeSupportBeerOutcomes™
Clarity Planum Urine Analyzer
OPERATOR’S MANUAL
Page13
Search Results by Date
1. Ontheresultscreen,selectSearch by Datetosearchthestoredtest
resultsbasedonthetestdate.
2. EnterthedateusingthekeypadonthescreenandpressENTER.
3. Allthematchingresultswillbelisted.Selectthedesiredtestresult
toreviewonthescreen.
Delete Results
Youmaywanttodeleteallthetestresultsunderthefollowingcircumstances:
•Acompletedownloadofallthestoredresultstoahostcomputer
hasoccurred.
•ToprotectpaentcondenalityandcomplywithHIPAAregulaons
whentheanalyzerismovedtoanotherclinicalsite.
•Torepairordiscardtheanalyzer.
1. Ontheresultscreen,selectDeletetodeleteallthestoredtestresults.
2. Apromptwillsay“DeleteAllResults?”Presstodeleteallresults.
3. Youcanselecttoreturntotheresultmainmenu.
Cauon:
Selecngthedeletefunconwilldeleteallthestoredtest
results.Thesystemdoesnotallowthedeleonofaspecic
testresultfordatasecurityreasons.Onceresultsaredeleted,
youcannolongerrecallthemfromthedatabase.Ifyouwanttokeepall
thedatafromyouranalyzer,pleasetransferthedatatoahostcomputer
beforeyouproceedwiththisfuncon.

Clarity Planum Urine Analyzer
OPERATOR’S MANUAL
WeSupportBeerOutcomes™Page14
4. System Conguraon
Youcanconguretheanalyzer’ssengsandreviewthesysteminformaon
withthesengsandadvancedmenus.
4.1 Sengs
PresstheSengsbuononthemainmenuscreentoaccessthesengs
menu.Fromthismenu,youmayreviewthecurrentsengsandselectthe
typeofstriptobeused,thedisplaylanguage,date&me,enableordisable
theautomacprinngfuncon,andviewsysteminformaonfortheClarity
Planum.TheRestore Default SengandUnits Displayedoponsare
intendedforClarityDiagnoscsadministratorsonlyandrequireavalid
administratorpasswordtologin.
Strip Type
Note: The device will set by default for tesng Clarity Diagnoscs’
CLA-10P strips.
1. Onthesengsmenu,presstheStrip Typebuontoreviewthe
currentseng.TheClarityPlanumisonlycompablewithClarity
Diagnoscs’proprietarystrips.
2. Presstoreturntothesengsmenu.
Language
1. Onthesengsmenu,presstheLanguagebuontoreviewthe
currentseng.TheClarityPlanumusesEnglishasthedefault
language.TheusermaychooseSpanishfortheoperaonlanguage.
2. Selectthedesiredlanguagetochangethedisplaylanguage.
3. Presstoreturntothesengsmenu.
Date & Time
1. Reviewthedateandme,locatedonthetoprightofthescreen.
Fromthesengsmenu,presstheDate & Timebuontoreviewthe
currentseng.
2. Usethekeypadtoprogramthedate&meexactlyasshowninthe
followingformat:MM/DD/YYYYHH:MMamorpm(Pleasenote:
am/pmneedtobelowercase).PressEnter tosave.
3. Presstoreturntothesengsmenu.

WeSupportBeerOutcomes™
Clarity Planum Urine Analyzer
OPERATOR’S MANUAL
Page15
Unit Seng
1. TheClarityPlanumurineanalyzerusesC1(convenonalunits)asthe
defaultunitseng.
2. The Unit Seng funconisintendedforClarityDiagnoscs
administratorsonlyandrequiresavalidadministratorpasswordtologin.
Printer Seng
1. Onthesengsscreen,pressthePrinter Sengbuontoreviewthe
currentseng.
2. SelectAutomacorManualfortheanalyzertoprinttheresult
automacallyormanuallyaereachtest.
3. Modifytheprinngformatbyselecngthefollowinginformaonto
beincludedintheresultprintout:
•OperatorID•PaentID•TestID
4. Presstoreturntothesengsmenu.
System Informaon
1. Onthesengsscreen,presstheSystem Informaonbuontoview
theserialnumberoftheanalyzer,thecurrentClarityPlanumsoware
versioninstalled,andtechnicalsupportcontactinformaon.
2. Presstoreturntothesengsmenu.
Restore Default Seng
1. The Restore Default SengfunconisintendedforClarityDiagnoscs
administratorsonlyandrequiresavalidadministratorpasswordtologin.

Clarity Planum Urine Analyzer
OPERATOR’S MANUAL
WeSupportBeerOutcomes™Page16
4.2 Advanced Sengs
PresstheAdvanced buononthemainmenuscreentoaccesstheadvanced
sengsmenu.Fromthismenu,itispossibletoperformqualitycontroltesng,
stripbedmaintenance,andsystemdiagnoscsonanalyzerhardware.
Operator
TheoperatormenuisonlyaccessibleforClarityDiagnoscsadministrators.
Test Mode
ThetestmodemenuisonlyaccessibleforClarityDiagnoscsadministrators.
Quality Control
OntheAdvancedsengsmenu,presstheQuality Control buontoaccess
thequalitycontrolmenu.Fromhere,operatorscanperformQualityControl
tesng.
TheQCsengsoponisonlyaccessibleforClarityDiagnoscsadministrators.
ThefrequencyofQCpromptsispresetto30daysor100testsbetweenQC,
whichevercomesrst.
Note: The Clarity Planum analyzer will automacally lock the test strip
funcon aer 30 days or 100 tests have been performed to ensure the
integrity of the urinalysis strips. In order to use the Clarity Planum analyzer,
CLIA Waived users are required to perform a successful QC test.
Signicance of QC tests and Consequences of Not Performing
QC Procedures
Qualitycontroltesngisaprocessofdetecngerrorswithinthetesngsiteto
ensureboththereliabilityandaccuracyoftestresultsinordertoprovidethe
bestpossiblepaentcare.Therefore,toensuretheaccuracyandperformance
oftheClarityPlanumTestSystem,qualitycontrolproceduresareincorporated
intotheClarityPlanumanalyzer.Thepurposeofincludingqualitycontrol
proceduresistoevaluatethereliabilityofthetestsystembyassayingstable
qualitycontrolmaterialthatresemblespaentsamples.

WeSupportBeerOutcomes™
Clarity Planum Urine Analyzer
OPERATOR’S MANUAL
Page17
Thequalitycontrolfunconisdesignedtodetect,reduce,andcorrect
decienciesinalaboratory’sinternalprocesspriortothereleaseofpaent
results.FailuretoconductqualitycontroltesngorcarryoutQCprocedures
couldleadtounreliableperformanceofthetestsystem,whichcouldresult
inmisdiagnosis,delayedtreatment,andincreasedcostsduetoretesng.Itis
ofgreatimportancetoensureallresultsarebothaccurateandreliable.
ClarityDiagnoscsproducesQualityControlSoluonLevelsIandII,which
areusedtoperformqualitycontroltesng.ForCLIAWaivedsengs,Clarity
Urinalysisqualitycontrolsmustbetestedunderthefollowingcondions:
• WhenanewcanisterofstripsisopenedOR
• 100testshavebeenperformedOR
• TestresultsappeartobeinaccurateOR
• AnewoperatorusestheanalyzerOR
• EachnewdayoftesngOR
• 30dayshavepassedsincethepreviousQCPassOR
• Aerperformingmaintenanceorserviceontheanalyzer.
IftheQCtestsdonotprovidetheexpectedresults,performthe
followingchecks:
• Ensurethestripsusedarenotpasttheirexpiraondate.
• Ensurethestripsarefreshfromanewcanister.
• Ensurethecontrolsarenotpasttheirexpiraondate.
• Repeatthetesttoensurenoerrorsweremadeduringthetest.
• IfQCtesngslldoesnotprovidetheexpectedresults,
callClarityDiagnoscsTechnicalSupport:(877)485-7877,
Monday-Friday,8am-6pmEST.
Note: For CLIA Waived sengs, only the QC controls produced by Clarity
Diagnoscs (CLA-UHCRL25) should be used, as the QC pass/fail criteria have
been opmized for the Clarity Planum analyzer.
Table of contents
Other Clarity Medical Equipment manuals
Popular Medical Equipment manuals by other brands

FIOR & GENTZ
FIOR & GENTZ NEURO HiSWING Assembly instructions

GlobalMIND
GlobalMIND VeinWalker pro quick start guide

Boston Scientific
Boston Scientific precision spectra Information for Prescribers

Altimate Medical
Altimate Medical EasyStand Evolv owner's manual

Boston Scientific
Boston Scientific Bionic Navigator 3d NM-7153-11A System programming manual

Accutome
Accutome ACCUPEN user guide