manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Codex
  6. •
  7. Laboratory Equipment
  8. •
  9. Codex BioXp 3200 User manual

Codex BioXp 3200 User manual

Popular Laboratory Equipment manuals by other brands

Radleys Carousel 6 Plus instructions

Radleys

Radleys Carousel 6 Plus instructions

Hemotherm 400MR Operation & technical manual

Hemotherm

Hemotherm 400MR Operation & technical manual

mrclab DVO Series Operation manual

mrclab

mrclab DVO Series Operation manual

CoolLED pE-800 quick start guide

CoolLED

CoolLED pE-800 quick start guide

Luminex Guava easyCyte Maintenance Guide

Luminex

Luminex Guava easyCyte Maintenance Guide

IKA A10 basic operating instructions

IKA

IKA A10 basic operating instructions

Belden HIRSCHMANN RPI-P1-4PoE installation manual

Belden

Belden HIRSCHMANN RPI-P1-4PoE installation manual

Koehler K1223 Series Operation and instruction manual

Koehler

Koehler K1223 Series Operation and instruction manual

Globe Scientific GCM-12 quick start guide

Globe Scientific

Globe Scientific GCM-12 quick start guide

Getinge 86 SERIES Technical manual

Getinge

Getinge 86 SERIES Technical manual

CORNING Everon 6000 user manual

CORNING

CORNING Everon 6000 user manual

Biocomp GRADIENT MASTER 108 operating manual

Biocomp

Biocomp GRADIENT MASTER 108 operating manual

Cole Parmer Stuart SB3/1 instruction manual

Cole Parmer

Cole Parmer Stuart SB3/1 instruction manual

BEVS 2501/2A user manual

BEVS

BEVS 2501/2A user manual

Qiagen DML 3000 user manual

Qiagen

Qiagen DML 3000 user manual

Queensgate NANOSCAN OP400 Quick start instructions

Queensgate

Queensgate NANOSCAN OP400 Quick start instructions

Parr Instrument 4560 Operating instructions manual

Parr Instrument

Parr Instrument 4560 Operating instructions manual

Integra DOSE IT operating instructions

Integra

Integra DOSE IT operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Page 1 of 3
Part number 45027 | Revision 1.1 |Eective 17APR2020
© 2020 Codex DNA, Inc. All rights reserved.
BioXp™ 3200 system
Uncrating and crating
Page 2 of 3
Part number 45027 | Revision 1.1 |Eective 17APR2020
© 2020 Codex DNA, Inc. All rights reserved.
WARNINGS
The BioXp™ 3200 system is heavy. Do not attempt to lift
or move the instrument without the assistance and/or
supervision of colleagues. Be certain to use appropriate
moving equipment and proper lifting techniques (including but
not limited to making certain that you have a secure grip on
the instrument before lifting, keeping your spine in a neutral
position while lifting with your legs, and coordinating lifting and
moving movements with all appropriate personnel). Improper
lifting can cause permanent back injury.
Uncrating instructions
1. Remove the quarter turn latches from the crate and lift o
the sides and top.
Figure 1
2. Remove the plastic bag from the top of the unit and pull
down to expose the lower portion of the unit and remove
the accessory box and upper foam support.
3. If possible, lift the pallet approximately 24” o the ground
with a forklift and set it on a sturdy base.
4. Have a movable cart ready which is capable of holding
greater than 200 lbs.
5. Using two people, lift the unit clear of the base and set on
the movable cart.
6. Move the system using the cart to the prepared site.
7. Codex recommends keeping the shipping materials until
installation and acceptance testing have been completed.
8. Remove tape holding lid and open lid.
9. Remove any tape internal to the system holding the
thermal cycler door closed.
10. Remove the two white covers for the chillers from the
accessory box and place them in the storage locations
(shown in figure 2).
Figure 2: Thermal cover placement
11. Once in the final location, the shipping screw needs to
be removed which is located under the pipette head.
Caution: Use care when removing this screw as sharp
edges exist under the head (shown in figure 3).
Figure 3: Shipping screw location (shown with head cover removed)
12. After removing the power cord from packaging, connect it
to the power inlet on the system cart located on the right
rear side panel.
13. Check that the power switch located above the power
cord inlet is in the “o” position (0).
Page 3 of 3
Part number 45027 | Revision 1.1 |Eective 17APR2020
© 2020 Codex DNA, Inc. All rights reserved.
14. Connect the system to a grounded power outlet.
15. Switch the power on at the power inlet.
16. The system will boot up, perform a self check, and present
the BioXp™ 3200 home screen.
17. The system is now ready to use.
18. It is recommended that you run a QC job to ensure the
system is operating properly. Contact Codex customer
service if one is required.
Re-crating a BioXp™ 3200 system for shipping
Note that the lower crate is not symmetrical and the BioXp™
system must be placed with the display in the corner with no
foam as (shown in figure 4).
Figure 4: Correct orientation
1. Place the plastic bag into the crate facing up with the
bottom pushed into the four corners of the foam supports.
2. Using two people, place the BioXp™ system into the
shipping crate as shown in Figure 5.
Figure 5: Lowering the BioXp™ system into the crate
3. Wrap the bag over the instrument and tape the open end
down over the unit such that water cannot enter the bag
from the top.
4. Place the upper foam support on the unit.
5. Place any accessories needed into the accessory box and
place the accessory box on the side of the unit.
Place the sides back onto the crate, latch all lower latches,
and remove the top portion.
Figure 6: Placing the crate lid onto the base
6. If removed, place the lid back on the crate and lock all
of the latches. Use small wire ties through each latch to
secure them.
7. Contact Codex customer service for shipping information:
[email protected] |855.474.4362
8. Once bill of lading has been sent please attach it to the
top of the unit and bring to your shipping dock for pickup.