
15
This procedure controls the high voltage level
used in the dielectric withstand test. It is
specified by safety agency personnel. Most
safety agencies allow a shorter duration test
(usually 1 sec. vs. 1 min.) if the voltage is
increased by 20%. The HT-3000 is factory set
for 1200 volts AC, a voltage level used for the
one second duration test for many types of
equipment. Consult the safety agencies for the
required voltage level for the type of equipment
being tested. If a different voltage level is
required, use this procedure to set it.
1. Defeat the Ground Check by turning the
rear panel switch Off.
2. Adjust the Test Duration control to
approximately the middle of its travel.
3. Connect the HT-3000 to a correctly rated
source of supply and turn the power
switch to the I or ON position. Push the
RESET button. The yellow TEST light
should light, indicating that the HT-3000
is ready to test.
4. Make sure there are no test leads
connected to the Tester. Push the TEST
button.
5. After the TEST button is pressed, the
voltage will ramp and hold. Set the
Voltage Adjust control on the rear panel
to the desired voltage.
6. If tamperproof operation is desired, we
suggest placing a calibration-style label
over the voltage adjust hole.
7. Check the position of the Ground Check
switch, and reset the test time using the
procedure Adjustment of the High
Voltage Test Time.
Adjustment of the High Voltage
Test Duration
This procedure sets the amount of time the
Tester will conduct the high voltage test. The
test time is specified by the safety agencies and
is tied to the test voltage. Most safety agencies
allow a much shorter test (usually 1 second vs. 1
minute) if the voltage is increased by 20%. The
HT-3000 is factory set for 2 seconds. Consult
the safety agencies for the test time for the type
of equipment being tested. If a different test
time is required, use this procedure to set it.
1. Defeat the Ground Check by turning the
rear panel switch Off.
2. Connect the HT-3000 to a correctly rated
source of supply and turn the power
switch to the I or ON position. Push the
RESET button. The yellow TEST
indicator should light, indicating that the
HT-3000 is ready to test.
3. Make sure there are no test leads
connected to the Tester. Push the TEST
button.
4. After the voltage has ramped, the Full
Voltage LED will light. The test time is
measured from when the Full Voltage
LED lights to when the voltage begins to
ramp down. Time the test.
5. After the test is complete, adjust the Test
Time control on the rear panel. If the time
desired is shorter, turn the Control
counterclockwise, and vice versa.
6. Push the RESET Button, then the TEST
Button and time the new test. Adjust the
Test Time control until correct.
7. Check the position of the Ground Check
switch before conducting testing.
Setting the Ground Check Switch
This switch controls whether the
Ground Test is conducted between the chassis
and the power supply cord grounding pin of the
unit being tested. The safety agencies require
that this test be conducted on all equipment
using a three-wire power supply cord. For other
types of equipment, such as some portable lamps
and all double-insulated equipment, there are
only two wires in the power supply cord and the
ground continuity check is waived by the safety
agencies.
If the power supply cord of the
equipment being tested has only two pins, the
Ground Check switch should be set to the Off
position. When the Ground Check switch is
turned Off, both the Ground OK and Ground
Open LED's on the front panel are lit as a
reminder to the operator that the ground
continuity test is not being conducted.
If the power supply cord of the
equipment being tested has three pins, the
Ground Check switch should be set to the On
position.
When testing with the Ground Check
switch Off, the operator must ensure that the
Return lead is properly connected to exposed