Condec UPS3000 Instruction manual

63253
UPS3000, UPS3110 &
UPS3210
Portable & Rack-mountable Pneumatic Pressure Standards
CONDEC Sales Phone No.: (888) 295-8475
CONDEC Web Site: WWW.4CONDEC.COM
Operation & Maintenance
Manual

2UPS3000/UPS3110/UPS3210 Operation and Maintenance Manual

i
Contents
1.0 Introduction.................................................................................................................................. 2
2.0 Operation...................................................................................................................................... 4
2.1 Unpacking and Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.2 Standard Display/Keyboard Front Panel Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.3 Rear Panel Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.4 Configuration Switch Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.4.1 APC4000/APC4001 Interface Option Enable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.4.2 Peak Hold or MAX/MIN Option Enable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.4.3 Conversion Enable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.4.4 Digital Averaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.4.5 AUTO-ZERO Maintenance (AZM) Enable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.4.6 Automatic Span Maintenance (ASM) Enable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.5 Freeze Mode Option Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.6 Engineering Conversion with PEAK HOLD or MAX/MIN Option. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.7 UPS3000 or UPS3210 Initial Setup Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.7.1 UPS3000 or UPS3210 Display of Pressure Sequence GAGE Mode . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.7.2 UPS3000 or UPS3210 Display of Pressure Sequence Absolute (ABS) Mode. . . . . . . . . . . . . . . . . . . . 13
2.8 UPS3110 Initial Setup Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.8.1 UPS3110 Pressure Measurement Sequence GAGE Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.8.2 UPS3110 Pressure Measurement Sequence Absolute (ABS) Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.9 GAGE Mode Self-Check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.10 Battery Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.11 Serial Output: 20 mA Loop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.11.1 Hardware Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.11.2 Serial Output Software Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.11.3 DATA FORMAT TABLE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.11.4 SERIAL OUTPUT CONFIGURATION TABLE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.0 Calibration and Adjustment Procedure ..................................................................................... 19
3.1 Pneumatic Calibration Set-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.2 Instrument Calibration Set-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.3 Zero/Span Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.4 Linearity and Hysteresis Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.5 Shunt Resistor Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.6 Permanent Data Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.7 Barometric Offset - Absolute/Gage Switch Selectable Units ONLY . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.7.1 For UPS3000/UPS3210 Absolute/Gage Switch Selectable Units ONLY . . . . . . . . . . . . . . . . . . . . . . . 24
3.7.2 For UPS3110 Absolute/Gage Switch Selectable Units ONLY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4.0 Maintenance & Service ............................................................................................................. 28
4.1 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4.2 Maintenance & Service Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4.2.1 Case Removal and Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4.2.2 ORION-2C (PN 55283) or ORION-3A (PN 55287) Manifold Removal. . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.2.3 ORION Manifold - Valve Seat Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
4.2.4 ORION Manifold - Vernier Control Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.2.5 ORION Manifold - Vernier Control Reassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.2.6 ORION Manifold - Valve Seat Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.2.7 ORION - Manifold, Panel Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4.2.8 ORION 2C Manifold - Valve Adjustment Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.2.9 ORION 3A Manifold - Valve Adjustment Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.2.10 Pressure Limit Control (Standard Pneumatic), Regulator Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.2.11 Pressure Limit Control (Standard Pneumatic), Regulator Installation . . . . . . . . . . . . . . . . . . . . . . . . . . 36

ii UPS3000/UPS3110/UPS3210 Operation & Maintenance Manual
4.2.12 Pressure Limit Control (Tescom), Regulator Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.2.13 Pressure Limit Control (Tescom), Regulator Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.2.14 Panel Gauge Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.2.15 Panel Gauge Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.2.16 Test Port Quick-Connect Fitting (PN 59762) Removal and Installation - 15, 50 and 100 Full Scale PSI 39
4.2.17 Test Port Quick-Connect Fitting (PN 55426), Removal and Installation - 500, 1000 and 2000 Full Scale PSI
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.2.18 Test Port Filter (PN 54188), Removal and Installation - 500, 1000 and 2000 Full Scale PSI . . . . . . . . . 40
4.2.19 Test Port Filter (PN 54188), Removal and Installation - 15, 50 and 100 Full Scale PSI . . . . . . . . . . . . . 40
4.2.20 Test Port Quick-Connect Fitting (PN 59004) and Filter (PN 54188) Removal and Installation - 5000 and
10000 Full Scale PSI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
4.2.21 AC Fuse (PN 58076), Removal and Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
4.2.22 AC Power/EMI Line Filter (PN 58870), Removal and Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
4.2.23 Power Switch Cable (PN 55351), Removal and Installation (Battery Units). . . . . . . . . . . . . . . . . . . . . . 42
4.2.24 Power Switch (PN 58878), Removal and Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
4.2.25 Range Select Switch Cable (PN 56014), Removal and Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
4.2.26 Power Supply Assembly, Removal and Installation (Battery Units) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
4.2.27 BATTERY (55851), Removal, Installation and Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
4.3 ORION 2C Valve Assembly Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
4.4 ORION 3A Valve Assembly Parts List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
4.5 UPS3000 Assembly Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5.0 Model Number System .............................................................................................................. 50
6.0 Available Ranges, Multi-Conversions and Resolutions ............................................................ 51
7.0 Options, Replacement Kits ........................................................................................................ 54
7.1 Freeze Mode Option - PN 57778. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
7.2 Peak Hold Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
7.3 Min and Max Mode Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
7.4 Analog Output Option (0 VDC to +10 VDC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
7.5 Analog Output Option (+4 mA DC to +20 mA DC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
7.6 RS232 Simplex Output Option Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
7.7 Replacement Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
8.0 Specifications............................................................................................................................ 59

1
About This Manual
The bench-top UPS3000 and the rack-mountable UPS3110 and UPS3210 pneumatic pressure standards are
rugged, compact instruments manufactured by CONDEC. They are designed to provide superior accuracy, range
of calibration and ease of operation when used to assist in the manufacture, test and/or calibration of a wide
variety of pressure sensing and measuring devices.
These instruments utilize an extremely repeatable sensor coupled to microprocessor-based electronic circuitry
and a selectable units display system. This provides easily readable and accurate digital representation of the
measured pressure. If required for portable use, the optional battery models, are able to provide up to six hours of
use.
The UPS3110 all electro-mechanical device has our precision Orion vernier and one test port. Some models
contain a front panel gage, to show the operator the input pressure, as well as, a pressure regulator that will act as
a pressure limiter so that the operator can not over pressure a unit under test. A test hose is supplied for the
customer. Standard front panel buttons and switches provide selection of the desired pressure range, push-button
zeroing and the unique internal self-check feature.
This manual has been written to give the user a simple and clear explanation of how to operate, calibrate and
maintain these instruments.
Before attempting to use either style, Pressure Standard, the following instructions
must be carefully read and understood by personnel utilizing the equipment. This is a
high-pressure system. While a substantial effort has been expended to make this
equipment safe, simple and fool-proof to operate, it is strongly recommended that only
personnel formally trained in the use of pneumatic pressure equipment be permitted to
operate it. Potentially dangerous conditions could be produced through negligent
handling or operation of the console due to the high pressure contained within the unit.
UPS3110 units are strictly for use with pneumatic pressures. Erroneous readings and
potential damage could result from the introduction of hydraulic fluids into the internal
tubing lines.
Authorized distributors and their
employees can view or download this
manual from the Condec distributor
site at www.4condec.com.
#AUTION

2UPS3000/UPS3110/UPS3210 Operation & Maintenance Manual
1.0 Introduction
Utilizing microprocessor technology, the UPS3000, rack-mountable UPS3110 and UPS3210 instruments offer a
combination of features, performance, versatility and reliability not previously available in a single,
self-contained pressure calibration instrument. The following list notes the features and benefits of each unit:
• Three independent switch-selectable pressure ranges per instrument.
• Nine button-selectable units (PSI, kPa, Bar, mBar, kg/cm2, mmHg, inHg, cmH2O, inH2O) plus percent
of full scale may be displayed on instrument.
• Accuracy of each range equal to or better than ±0.05% Full Scale.
• Both “Gage” and “Absolute” pressure calibrations available via front panel switch selection.
• Automatic self-check: Computer-controlled internal circuitry provides automatic maintenance of both
zero and span calibration data to insure long-term stability and accuracy.
• Digital Display: Eliminates parallax, interpolation and operator judgement errors. Large, bright, red LED
digits provide excellent readability under all lighting conditions.
• UPS3110 only - Using a manually adjustable regulator, the maximum system input pressure is adjusted
to any desired value higher (typically 20 to 50%) than the full scale range of the device being tested. By
virtue of this technique, the unit under test is fully protected from being inadvertently over-pressurized.
• Fast Response: Pressure data refreshed and updated at the rate of 12 times per second.
• Data Output: Serial Data, 20 mA loop, ASCII code format with start, stop and parity bits supplied as
standard practice. RS232 Simplex output available as an option.
• Calibration Integrity: Tamper-proof design. Once calibrated, numerous safeguards guarantee the
integrity of pressure readings obtained. Display prompting provides the operator with functional status
information during both operation and calibration.
• Pressure Media: UPS3000 and UPS3210, any gas or fluid compatible with 17-4 PH stainless steel alloy.
UPS3110, only gas compatible with 17-4 PH stainless steel alloy.
• Simple Operation: All controls and indicators are accessible from the front panel. Accompanying
operator's manual provides clear, concise instructions for system operation
• Safe, Clean Operation: All pressure components are made of brass, copper, aluminum or stainless steel
and proof-tested to at least 150% of maximum operating pressure. In addition, the UPS3110 system
contains a high-pressure relief valve to protect both the operator and system components from harm in
the event of inadvertent over-pressurization.
The heart of this indicator/calibration system is a highly stable and repeatable pressure transducer. These sensors
produce an electrical output signal which is linearly proportional to the applied pressure. This transduction
technique has been employed by CONDEC for many years and has realized hundreds of thousands of hours of
reliable operation.
By combining these sensors with sophisticated microprocessor-based circuitry, an even higher degree of
operational accuracy and precision has been accomplished. For example, computer-generated correction curves
for both the non-linearity and the hysteresis of the sensors improve these characteristics by an order of magnitude
or more. In addition, a “self-calibration” feature insures long-term accuracy by utilizing the computer to generate
and control an internal “shunt calibration mode” of operation. The indicators full-scale reading is compared
against, and if necessary, corrected to the digitally-stored value for full scale obtained at the time of initial
pressure calibration.
The UPS3000 and UPS3210 indicators, as well as, the UPS3110 calibrator has capability for an internal,
rechargeable 12 volt lead acid battery, which provides a minimum of 6 hours of complete usage, when battery is
fully charged. An ON/OFF switch is provided to conserve energy when the instrument is not in use. It also has a
LO BATT indicator. After receiving and prior to operating new battery models, reinstall (+) cable, red wire, on
battery.

Introduction 3
With all its sophistication however, a great deal of effort has been expended to make the instruments simple to
operate and easy to calibrate.
Two micro-metering valves and vernier are provided to control the UPS3110 pressure source while the digital
display indicates precisely the magnitude of the applied test pressure. Also, a pair of simple push-button switches
provide both “zeroing” of the pressure display and the selection of either the “Gage” or “Absolute” calibration
mode. In like manner, the “Range” selection is accomplished via a clearly marked rotary switch. Over-pressure
protection is provided on lower pressure UPS3110 models via a fully adjustable pressure regulator which is
manually set for each new device being tested.
With respect to calibration, the instruments have been designed and programmed to be very user friendly in that
they provide the operator with various prompting symbols and legends during each phase of the calibration
sequence. Also, to prevent unauthorized tampering or calibrations, numerous features have been incorporated
which greatly minimize this potential danger. Also, the electronic circuitry has been designed without any
potentiometers or adjustments which eliminates the possibility of unwanted changes. Finally, the computer has
been programmed with a series of internal self-diagnostic routines which continually monitor and check every bit
of data stored and processed by this system, and immediately either notes or shuts down operation in the event of
an out-of-tolerance reading or outright failures.
The following schematic provides an overview of the UPS3110’s function.
Figure 1-1. UPS3110 Flow Diagram

4UPS3000/UPS3110/UPS3210 Operation & Maintenance Manual
2.0 Operation
2.1 Unpacking and Inspection
When received, carefully remove the instrument from its shipping container. A visual inspection of the
instrument’s external surfaces should be performed immediately after unpacking. If obvious damage has been
incurred during transit, the shipping agency and the distributor should be notified as soon as possible.
Instructions as to how to proceed after assessment of the damage will then be determined. If the instrument
shows no signs of damage, check to be sure all the required equipment and accessories have been received. See
“Options, Replacement Kits” on page 54, for all options.
NOTE: Prior to operating battery models, reinstall (+) cable, red wire, on battery.
2.2 Standard Display/Keyboard Front Panel Functions
UPS3000 See Figure 2-1 on page 5, UPS3110 See Figure 2-2 on page 5, UPS3210 See Figure 2-3 on page 6.
1. Six Digit LED display of the measured pressure. This display also prompts if a problem occurred in
performing an operation.
2. Three Annuciators (4). Two of the LEDs are used for indicating whether the instrument is operating as a
Gage pressure instrument or as an absolute (ABS) pressure instrument. The third LED indicates the
conversion units that the pressure data is currently being displayed in.
3. ZERO function key (1). In GAGE mode and ABS only instruments this key allows the indicator to be
zeroed.
NOTE: If attempting to perform a push to ZERO while the pressure data is in motion, the instrument will not go to ZERO.
NOTE: The four, or five if options have been installed, push button switches located across the lower face of the front
panel are individually actuated by momentarily pressing with a light fingertip motion anywhere within the black outlines
of the button. When actuated, each switch produces a visual feedback through the digital display.
4. ABS/GAGE function key (2), toggles instrument to be a Gage pressure instrument or an absolute (ABS)
pressure instrument.
NOTE: This key is blank and non-functional if unit is an Absolute (Abs) only pressure unit or a Gage only pressure unit.
5. CONV conversion key (3), is used to select the conversion units that the pressure data is currently being
displayed in.
6. CAL function key (5), activates the internal Self-Check feature. See “GAGE Mode Self-Check” on
page 15.
NOTE: CAL key is non-functional in absolute mode.
7. RANGE SELECT switch (6), used to select the desired pressure range.
NOTE: Do not turn switch between ranges during a cycle.
8. UPS3110 instrument with maximum range of 2K PSI or less: PRESSURE LIMIT CONTROL (7), for
regulating the input pressure to the unit.
9. UPS3110 instrument with maximum range of 2K PSI or less: PRESSURE LIMIT MONITOR (8), indicates
input pressure going to the unit.
10. UPS3110 instrument: TEST PORT (9), for connecting to the Device-Under-Test (DUT).
11. UPS3110 instrument: INPUT Valve (10), used for nearing target pressure of the DUT.
12. UPS3110 instrument: VERNIER Valve (11), used for precisely setting target pressure of the DUT.
13. UPS3110 instrument: VENT Valve (12), used for venting pressure of the DUT.

Operation 5
Figure 2-1. UPS3000 Desktop/Panel-mountable Pressure Indicator Display/Keyboard Functions
Figure 2-2. UPS3110 Rack-mountable Front Pressure Calibrator Panel Functions

6UPS3000/UPS3110/UPS3210 Operation & Maintenance Manual
Figure 2-3. UPS3210 Rack-mountable Pressure Indicator Front Panel Functions
2.3 Rear Panel Configuration
UPS3000 Series, See Figure 2-4 below, contains the following:
1. AC power cord, and input receptacle (17).
2. INPUT PRESSURE port J1 (16), 7/16-20, 37o-4 AN male fitting.
Application of pressures greater than 1.5 times the highest pressure range of the indicator may cause
calibration errors or even permanent damage to the pressure transducer.
3. The unit’s identification plate (15).
4. Optional if required items:
• Connector J2 (14), 5 pin round MS style connector, for Serial or Analog Output communication board.
• Connector J3 (20), 15 pin D connector, for the PCM1000 Controller Interface.
• Connector (18), 5 pin round connector, for the Freeze Mode Cable.
NOTE: For further information, see “Options, Replacement Kits” on page 54.
Figure 2-4. UPS3000 Desktop/Panel-mountable Pressure Indicator Rear View
#AUTION

Operation 7
UPS3110 Rack-mountable Series, See Figure 2-5 below, contains the following:
1. AC power cord, fuse holder and input receptacle (17).
2. INPUT PRESSURE port (16), 7/16-20, 37o-4 AN male fitting. Location of port for AC only units.
NOTE: The maximum input pressure, supplied by user, is noted below port.
3. The unit’s identification plate (15).
4. VENT/VACUUM PORT (19), 7/16-20, 37o-4 AN male fitting.
5. Optional if required items:
Connector J1 (14), 5 pin round MS style connector, for Serial or Analog Output communication board or
location of Input Pressure port for models with battery.
Connector J2 (18), 5 pin round connector, for the Freeze Mode Cable.
NOTE: For further information, see “Options, Replacement Kits” on page 54.
Figure 2-5. UPS3110 Rack-mountable Pressure Calibrator Rear View
UPS3210 Rack-mountable Series, See Figure 2-6 on page 8, contains the following:
1. AC power cord, fuse holder and input receptacle (17).
2. INPUT PRESSURE port J3 (16), 7/16-20, 37o-4 AN male fitting.
NOTE: The maximum input pressure, supplied by user, is noted above port.
3. The unit’s identification plate (15).
4. Optional if required items:
Connector J1 (14), 5 pin round MS style connector, for Serial or Analog Output communication board.
Connector J2 (18), 5 pin round connector, for the Freeze Mode Cable.
Connector J4, located below identification plate, required for APC4000/APC4001 interface cable.
NOTE: For further information, see “Options, Replacement Kits” on page 54.

8UPS3000/UPS3110/UPS3210 Operation & Maintenance Manual
Figure 2-6. UPS3210 Rack-mountable Pressure Indicator Rear View
2.4 Configuration Switch Settings
As normally supplied, the UPS3000, UPS3110 and UPS3210 will be fully calibrated and configured to the
requirements specified by the customer purchase order. However, there are several functions or operational
features that may be selected or altered by the operator during usage. These are controlled by the eight-position
DIP switch, S1, located on the CPU board as shown in Figure 2-7.
NOTE: To gain access to the CPU board see “Case Removal and Installation” on page 28.
Figure 2-7. CPU Configuration Switch Locations

Operation 9
2.4.1 APC4000/APC4001 Interface Option Enable
The APC4000 interface enable is activated only for models that are purchased and used with the PCM1000-1
controller or part of an APC4001 controller system. When UPS3000 or UPS3210 is used in conjunction with a
controller, Peak Hold, Freeze Mode or MIN/MAX Mode options are not available.
NOTE: UPS3110 is not used with a controller.
Approximately 1994 multi conversion software was added to UPS3000 models, therefore UPS3000 and PCM1000 units
made prior to this will not work with the ones manufactured after that date. Consult factory for information on upgrading
units.
2.4.2 Peak Hold or MAX/MIN Option Enable
See “Peak Hold Option” on page 54 or “Min and Max Mode Option” on page 55.
NOTE: * Absolute or Gage modes of operation are factory set, depending upon the style of transducer supplied and the
type of internal memory configuration utilized.
When the Absolute/Gage switching is not utilized; the front panel ABS/GAGE push-button switch will be
programmed to inactive and covered with a blank overlay.
When the PEAK HOLD Option or MAX/MIN option is supplied switching from Absolute to Gauge mode via
the front panel is possible and the Freeze Mode option, if supplied, is operable.
NOTE: When the PEAK HOLD or MAX/MIN option is supplied, the CONV button is no longer active or present on front
panel. Therefore, the required conversion must be set thru reconfiguration. See “Engineering Conversion with PEAK
HOLD or MAX/MIN Option” on page 12 for applicable method to change to required conversion.
Switch Settings S1
Position 1 Position 2 Position 3 Position 4 Position 5 Position 6 Position 7 Position 8 UPS 3000
UPS 3110
UPS 3210
00100011With Controller
00100010Without
Controller
Table 2-1. APC 4000/APC 4001 Interface Enable Switch Settings
Switch Settings S1 Peak Hold or MAX/MIN Mode *ABS/GAGE Mode
Position 1 Position 2
0 0 Disabled ABS & Gauge (Front Panel Switch Selectable)
0 1 Disabled Gauge or ABS only
1 0 Enabled ABS & Gauge (Front Panel Switch Selectable)
1 0 Enabled Gauge or ABS only
Table 2-2. Peak Hold or MAX/MIN Enable Switch Settings

10 UPS3000/UPS3110/UPS3210 Operation & Maintenance Manual
2.4.3 Conversion Enable
As standard practice, the instruments are supplied with display indication in PSI (either Absolute, Gage or both)
and capable of being converted to a range of nine units via the front panel CONV push-button switch.
NOTE: When the PEAK HOLD or MAX/MIN option is supplied with a model having the ABS or GAGE mode, switch
selectable from front panel, the “CONV” button will be disabled and will not be present on the front panel. Therefore, the
required conversion must be set thru reconfiguration. See “Engineering Conversion with PEAK HOLD or MAX/MIN
Option” on page 12 for applicable method to change to required conversion.
2.4.4 Digital Averaging
Digital Averaging is a technique whereby numerous update cycles are averaged together before the numerical
display data is changed. In essence, this feature acts as a variable rate electronic filter to provide a more stable
pressure indication reading.The “AUTO” mode of this filter allows the display to update rapidly (12/sec.) when
the input pressure is being quickly changed and yet provides extremely stable display operation (3/sec.) when the
desired pressure input value has been obtained.
2.4.5 AUTO-ZERO Maintenance (AZM) Enable
The Automatic Zero Maintenance (AZM) feature is used to “hold” the indicator reading to a “zero” value as long
as the actual pressure input is maintained at zero. If the input pressure changes by more than one half a least
significant display digit between two consecutive display update cycles, the “hold” feature is automatically
disabled and the exact magnitude of pressure being exerted will be displayed. In some applications, it may be
better to operate the instrument without the AZM circuit enabled. If so, pressing the ZERO push-button switch,
with zero pressure applied to the instrument, will guarantee that each new pressure cycle begins at zero.
2.4.6 Automatic Span Maintenance (ASM) Enable
The Automatic Span Maintenance (ASM) circuit operates in conjunction with the front panel CAL push-button to
provide a Self-Check feature that insures long-term accuracy by utilizing the computer to generate and control an
internal “shunt calibration mode” of operation wherein the indicator reading obtained is compared against, and if
necessary, corrected to the digitally-stored value for the same shunt calibration reading obtained at the time of
initial pressure calibration.
Switch Setting S1 Position 3 Front Panel Conv Key Mode
0Disabled
1Enabled
Table 2-3. Conversion Enable Switch Setting
Switch Setting S1 Average Select Approximate Update Rate
Position 4 Position 5
0 0 1 (Off) 12/sec.
1 0 2 6/sec.
0 1 4 3/sec.
1 1 Auto 3-12/sec.
Table 2-4. Digital Averaging Switch Setting
Switch Setting S1 Position 6 AZM Mode
0Disabled AZM
1Enabled AZM
Table 2-5. AZM Enable Switch Setting
Switch Setting S1 Position 7 ASM Mode
0Disabled ASM
1Enabled ASM
Table 2-6. ASM Enable Switch Setting

Operation 11
2.5 Freeze Mode Option Wiring
See Figure 2-7 on page 8 for TB location on CPU.
Connections may be made using the Freeze Mode Kit, see “Freeze Mode Option - PN 57778” on page 54. The
small five pin connector (pins A & B) which, in turn, are connected to the main CPU board via terminal block
TB2-7 (+) (Pin A) and TB2-8 (Gnd Ret) (Pin B). The current through these wires is approximately 0.5 mA as a
non-inductive load. The voltage between these two wires is 5 VDC.
NOTE: Option may not be used with the APC4000/APC4001 Interface Option.
Figure 2-8. Freeze Mode Option Wiring

12 UPS3000/UPS3110/UPS3210 Operation & Maintenance Manual
2.6 Engineering Conversion with PEAK HOLD or MAX/MIN Option
See Figure 2-7 on page 8 for location of “Data Enter” and rocker switches.
NOTE: When the PEAK HOLD or MAX/MIN option is supplied with a model having the ABS/GAGE mode, switch
selectable from front panel, the CONV button will be disabled and will not be present on the front panel.
Figure 2-9. Peak Hold Option Front Panel Switches
Figure 2-10. MIN/MAX Option Front Panel Switches
In order to step the light bar on conversion display to required conversion the following steps must be taken:
1. Open UPS3000/UPS3110/UPS3210 per Section 4.2.1 on page 28.
2. At this time, record the settings of rocker switches S1 and S3. These settings must be changed in order to
program the new conversion configuration.
3. Close the following rocker switches S1 position 3 and S3 positions 1 and 3.
4. Verify Numeric display has “1 XXX”, where X’s are some random number.
NOTE: If 1 is not on display open rocker switch S1 position 3 and rocker switch S3 positions 1 and 3, then restart at step
3 above.
5. Push the ABS/GAGE switch located on front panel as needed to move light bar on conversion display to
required conversion.
6. Momentarily push the “Data Enter” switch S2 to enter new conversion.
7. Open the following rocker switches S1 position 3 and S3 positions 1 and 3.
NOTE: Verify rocker switches S1 and S3 conform to original settings which had been recorded earlier.
8. To verify proper conversion is active remove and restore power to the UPS3000/UPS3110/UPS3210.

Operation 13
2.7 UPS3000 or UPS3210 Initial Setup Sequence
For UPS3000, refer to Figure 2-1 on page 5 and Figure 2-4 on page 6.
For UPS3210, refer to Figure 2-3 on page 6 and Figure 2-6 on page 8.
1. Connect the pressure source to the instrument via the INPUT PRESSURE port (16), 7/16-20, 37o-4 AN male
fitting provided on the rear panel. It is suggested that a Cheat Seal, PN 54854, be used between INPUT
PRESSURE port and pressure source fitting.
NOTE: Valves for venting and pressure input should be installed in line with the pressure source.
2. Apply power to the instrument using AC Cord (17) or, if applicable, by pressing POWER switch (13) on
front panel and allow it to stabilize for at least 5 to 10 minutes.
3. Select the desired full scale pressure range via the three-position rotary RANGE SELECT switch (6). For
the best accuracy, the selected range must be greater than, but close as possible to, the full scale range of
the Device-Under-Test.
NOTE: Do not switch pressure ranges during a pressure cycle.
4. If applicable or required, select the mode of operation by momentarily depressing the ABS/GAGE
push-button switch (2). The applicable “ABS” or “GAGE” light bar (4) will be lit to provide mode of
operation.
5. To select the desired measurement display units, depress CONV push-button switch (3).
NOTE: When the PEAK HOLD or MAX/MIN option is supplied with a model having the ABS or GAGE mode, switch
selectable from front panel, the “CONV” button will be disabled and will not be present on the front panel. Refer to
Section 2.6 on page 12.
6. If operating the unit in GAGE mode go to Section 2.7.1 or if operating the unit in ABS mode go to
Section 2.7.2.
Application of pressures greater than 1.5 times the highest pressure range of the indicator may cause
calibration errors or even permanent damage to the pressure transducer.
2.7.1 UPS3000 or UPS3210 Display of Pressure Sequence GAGE Mode
1. Vent the Input Pressure port (16) to atmosphere.
2. Momentarily depress the ZERO push-button switch (1). The display will indicate 0.00.
3. The instrument is now fully configured and ready to display the applied pressure.
2.7.2 UPS3000 or UPS3210 Display of Pressure Sequence Absolute (ABS) Mode
NOTE: If local barometric pressure is not 14.7, ABS/GAGE switch selectable units only, may need barometric offset. See
“Barometric Offset - Absolute/Gage Switch Selectable Units ONLY” on page 23.
1. If only pressure measurements greater than barometric are required, continue to Step 1.1. If pressure
measurements above and below atmospheric pressure are required go to step 2.
1.1.To apply pressure, close the customer supplied vent valve attached in line with the INPUT PRESSURE port
(16). Unit will display applied pressure.
2. If pressure measurements above and below atmospheric pressure are required, connect a VACUUM
PUMP in line with the INPUT PRESSURE port (16).
3. Close the customer supplied vent and input valve attached in line with the INPUT PRESSURE port (16).
4. Apply power to the vacuum pump and allow it to evacuate the system for several minutes or until the
digital display reading reaches equilibrium near Zero psia. Momentarily press the ZERO push-button
switch (1) to establish a zero reference on the display.
5. The instrument is now fully configured and ready to display the applied pressure.
#AUTION

14 UPS3000/UPS3110/UPS3210 Operation & Maintenance Manual
2.8 UPS3110 Initial Setup Sequence
Refer to Figure 2-2 on page 5 and Figure 2-5 on page 7.
1. Connect the pressure source to the instrument via the INPUT PRESSURE port (16), 7/16-20, 37o-4 AN male
fitting provided on the rear panel. It is suggested that a Cheat Seal, PN 54854, be used between INPUT
PRESSURE port and pressure source fitting.
2. Check that the INPUT valve (10) is closed (rotate clockwise until it stops) and that the VENT valve (12) is
open (two turns counter-clockwise from its stop).
3. Plug in the power cord (17) and energize the unit by pushing the POWER switch located on front panel.
Allow at least 10 minutes warm-up time.
4. Select the desired full scale pressure range via the three-position rotary RANGE SELECT switch (6). For
the best accuracy, the selected range must be greater than, but close as possible to, the full scale range of
the Device-Under-Test (DUT).
NOTE: Do not switch pressure ranges during a calibration cycle.
5. Using the PRESSURE LIMIT CONTROL regulator (7), adjust the maximum system input pressure, as read by
the PRESSURE LIMIT MONITOR (8), to any desired value higher (typically 20–50% higher) than the
full-scale range of the DUT. Using this technique, the DUT is fully protected from being accidentally
over-pressurized.
NOTE: UPS3110 []A[] and UPS3110 []G[] units do not have PRESSURE LIMIT CONTROL or MONITOR.
6. Connect the male end of the test hose to the TEST PORT (9) fitting.
7. Connect the swivel fitting end (7/16-20) of the Test (output) hose to the DUT using adapters if required.
Tighten all connections properly.
8. If applicable or required, select the mode of operation by momentarily depressing the ABS/GAGE (2)
switch. The applicable “ABS” or “GAGE” light bar (4) will be lit to provide mode of operation.
9. To select the desired measurement display units, depress CONV (3) switch.
NOTE: When the PEAK HOLD or MAX/MIN option is supplied with a model having the ABS or GAGE mode, switch
selectable from front panel, the “CONV” button will be disabled and will not be present on the front panel. Refer to
Section 2.6 on page 12.
10. If operating the unit in GAGE mode go to Section 2.8.1 or if operating the unit in ABS mode go to
Section 2.8.2.
2.8.1 UPS3110 Pressure Measurement Sequence GAGE Mode
1. Press the ZERO push-button switch (1), display will return to a normal “Zero” reading. The instrument
may be zeroed at anytime, as long as the VENT valve (12) is open, by momentarily depressing the ZERO
push-button switch (1).
2. To apply pressure, close the VENT valve (12), approximately two turns, until it stops, then open the INPUT
valve (10) approximately 1/2 turn counter-clockwise until the numerical display begins to move. The
pressure may change rapidly until reaching approximately 90% of the desired final value.
3. Use either the INPUT (10) or VENT valve (12) to obtain a specific pressure reading. Both provide precise
control. As the pressure approaches the desired value, the valve being used for control should be rotated
slowly clockwise to its closed position. With a little experience, pressure values very close to the desired
final value may be quickly achieved.
4. To obtain exact pressure readings, slowly rotate the VERNIER control (11) knob in the direction required
(clockwise to increase pressure) as indicated by the electronic numerical display.
Application of pressures greater than 1.5 times the highest pressure range of the indicator may cause
calibration errors or even permanent damage to the pressure transducer.
#AUTION

Operation 15
2.8.2 UPS3110 Pressure Measurement Sequence Absolute (ABS) Mode
NOTE: If local barometric pressure is not 14.7, ABS/GAGE switch selectable units only, may need barometric offset. See
“Barometric Offset - Absolute/Gage Switch Selectable Units ONLY” on page 23.
1. If only pressure measurements greater than barometric are required, continue to step 1.1. If pressure
measurements above and below atmospheric pressure are required go to step 2.
To apply pressure, close the VENT valve (12), approximately two turns, until it stops, then open the INPUT
valve (10) approximately 1/2 turn counter-clockwise until the numerical display begins to move. The
pressure may change rapidly until reaching approximately 90% of the desired final value.
1.1. Use either the INPUT (10) or VENT valve (12) to obtain a specific pressure reading. Both provide
precise control. As the pressure approaches the desired value, the valve being used for control should
be rotated slowly clockwise to its closed position. With a little experience, pressure values very close
to the desired final value may be quickly achieved.
1.2. To obtain exact pressure readings, slowly rotate the VERNIER control (11) knob in the direction
required (clockwise to increase pressure) as indicated by the electronic numerical display
Application of pressures greater than 1.5 times the highest pressure range of the indicator may cause
calibration errors or even permanent damage to the pressure transducer.
2. If pressure measurements above and below atmospheric pressure are required, connect a VACUUM
PUMP to the VACUUM/VENT port (19).
3. Open the VENT valve (12), close the INPUT valve (10) and apply power to the vacuum pump and allow it
to evacuate the system for several minutes or until the digital display reading reaches equilibrium near
Zero psia. Press the ZERO push-button switch (1) to establish a zero reference on the display.
4. With the vacuum pump still running, close the VENT valve (12) (approximately two turns to its stop) and
check for system leaks. If there are none, continue to step 4.1.
4.1. To apply pressure, open the INPUT valve (10) (approximately 1/2 turn counter-clockwise until the
numerical display begins to move). In general, the pressure may be changed rapidly until reaching
approximately 90% of it desired final value.
4.2. Use either the INPUT (10) or VENT (12) valve to obtain a specific pressure reading. Both provide
precise control. As the pressure approaches the desired value, the valve being used for control should
be rotated slowly clockwise to its closed position. With a little experience, pressure values very close
to the desired final value may be quickly achieved.
4.3. To obtain exact pressure readings, slowly rotate the VERNIER control (11) knob in the direction
required (clockwise to increase pressure) as indicated by the electronic numerical display.
2.9 GAGE Mode Self-Check
For UPS3000, refer to Figure 2-1 on page 5 and Figure 2-4 on page 6.
For UPS3110, refer to Figure 2-2 on page 5 and Figure 2-5 on page 7.
For UPS3210, refer to Figure 2-3 on page 6 and Figure 2-6 on page 8.
NOTE: Use of this Self -Check is not required for the proper operation of unit. CAL key is non-functional in absolute
mode due to the inability to simulate a perfect vacuum reference.
1. Vent the Input Pressure port (16) to atmosphere.
2. Momentarily depress the ZERO push-button switch (1). The display will indicate 0.00.
3. Momentarily depress the CAL push-button switch (5). The display will immediately blank except for two
“- -” which indicate that the unit is performing the self-check. If the self-check is successful, the display
will flash “100.00” and revert to its normal zero indication.
#AUTION

16 UPS3000/UPS3110/UPS3210 Operation & Maintenance Manual
2.10 Battery Operation
For replacement, see “Replacement Kits” on page 57.
When supplied with the battery, the UPS3000/UPS3110/UPS3210 has an internal, rechargeable 12 volt, lead
acid battery which provides a minimum of 6 hours of completely portable usage before having to be re-charged.
The UPS3000/UPS3110/UPS3210 may be operated and/or recharged by simply connecting to a standard AC
outlet via the line cord supplied. The battery re-charge cycle time is approximately 16 to 20 hours with the ON/
OFF switch in the OFF position. The charging circuit is designed to be left on indefinitely without adversely
affecting battery life.
The battery voltage reading will typically be between 11.5 and 12.6 volts. When the battery voltage reads 11.5
volts there are approximately 1 to 2 hours of useful operation left and a LOW BATT indicator will be illuminated. A
red LED in the lower left corner of unit near the POWER switch will turn on. The instrument will cease to
function when the battery voltage is 11.0 volts or less.
2.11 Serial Output: 20 mA Loop
NOTE: This may not be used with the following options in “DEMAND FORMAT”, APC4000 or APC4001 interface, Peak
hold, Min/Max or Battery. This may not be used with the following options in “CONTINUOUS FORMAT”, APC 4000 or
APC4001 interface or Battery.
Their are two modes of operation, Continuous or Demand.
Continuous Mode: Model is continuously sending data.
Demand Mode: Must be done from the front panel of the UPS3000/UPS3110/UPS3210 using a button. The
button is the hidden one between the CONV and CAL button’s This is not a two-way mode and cannot be done
externally by PC.
2.11.1 Hardware Configuration
NOTE: See “Case Removal and Installation” on page 28.
UPS3000 Required Parts: PN 60607 Quantity 1, PN 58707 Quantity 1, PN 14839 Quantity 2, PN 58302,
Quantity 1.
UPS3110 and UPS3210 Required Parts: Contact factory for requirements.
NOTE: Because of soldering it may be beneficial to have this installed at factory.
Solder Brown Wire to E13 and Red Wire to E14
The serial output is accessed at J3, a 15-pin D-subminiature female connector at the rear of the unit. Pin
designations are as follows:
J3-8 +Tx (Brown wire)
J3-9 -Tx (Red Wire)
2.11.2 Serial Output Software Configuration
NOTES:
1. See “Case Removal and Installation” on page 28.
2. See Figure 2-7 on page 8 to locate DIP switches S1 and S3,as well as, switch S2 on CPU board.
3. See Figure 3-2 on page 20 for “STEPPER” and “ENTER” switch locations.
1. Open up the unit and power it up.
2. On the main circuit board, close S3-1. The unit will display the following:
6 0 _ X X X where “X” can be any number
3. Momentarily hit the STEPPER push-button switch until the number 64 is on the left side of the display.
The unit will display the following:
6 4 _ X X X where “X” can be any number
4. Refer to the “DATA FORMAT TABLE” in this section to determine the appropriate output format.
This manual suits for next models
2
Table of contents
Popular Touch Panel manuals by other brands

ICP DAS USA
ICP DAS USA SmartView TPM-4100 user manual

3M
3M M1500SS MicroTouch quick start guide

Interwrite Learning
Interwrite Learning Interwrite Board Getting started

Avalue Technology
Avalue Technology LPC SERIES Quick reference guide

Waveshare
Waveshare 7inch HDMI LCD user manual

Moxa Technologies
Moxa Technologies MPC-2000 Series Quick installation guide