
4.3 Low Temp Cycle
This section applies to machines equipped with LOW-TEMP STERILIZATION CYCLE
ONLY.
If your machine is equipped with Low-Temp cycle capability, to start a cycle, you should
make sure that the Low-Temp selector switch (Figure 4.2), the Chamber Steam Low-
Temp Control switch (Figure 4.3), and the Jacket Steam Low-Temp Control switch
(Figure 4.4) are all in Low-Temp mode. Make sure that the sterilizer door is sealed and
tightened. Set the chart recorder (Figure 5.1) to a temperature in the range of 172 F to
212 F. Press the red ON/OFF switch (Figure 3.1) to the ON position. You do not need to
depress any of the cycle selector switches. The Low-Temp cycle will commence. After
the end-of-cycle buzzer sounds, make sure the Chamber pressure gauge reads ZERO
before opening the door.
When running a LOW-TEMP cycle, you should pay attention to the following:
1) Make sure you set the chart recorder (see section 5) to a temperature in the range
of 172 F – 212 F.
2) Make sure that you leave sufficient space between the articles in the chamber to
allow circulation of the air-steam mixture throughout the load.
3) During a LOW-TEMP cycle, the chamber is NOT pressurized. You should
always check the chamber pressure gauge before opening the door of the sterilizer
to make sure it registers zero.
4) It is highly recommended that you schedule all LOW-TEMP sterilization at the
beginning of the day when the sterilizer is cold and the jacket has not been
pressurized. This will prevent moisture from entering the chamber, which may
de-stabilize the temperature in the chamber.
WARNING: BURN HAZARD
The operator MUST wear protective clothing including face shield, thermal gloves,
and proper lab attire, when attempting to load and unload the chamber materials.
Steam released from the sterilizer chamber can cause serious burns. Stand away
while opening the door.
If water leaks from the front of the sterilizer, DO NOT open the door. Contact
service personnel.
To run a LOW-TEMP cycle, make sure you toggle the selector switch on the MKII
controller to LOW-TEMP. See Figure 4.2 below. The LOW-TEMP light will be on
during the cycle.
The MKII Controller Operator’s Guide
16