DOC-037 Iss 1 3
4) Illumination is not flat and even at the sample
(homogenous).
Has the optical setup procedure been carried out (Section 9 of the User Manual)?
Is the system being used on the microscope that it was ordered for?
If the system has been moved to an alternative microscope, please confirm with
CoolLED that the system is compatible in its current form. (Some microscopes
may require a different internal optical setting, even if the physical microscope
adaptor is the same).
5) Illumination is not centralized over the field of view.
Rotate the Light Source in the epi-port whilst observing the field of view.
If the offset moves with the rotation of the Light Source then the system has lost
its alignment. This will need to be returned to CoolLED to be reset.
If the offset does not move with the rotation of the system then the issue is with
the optical path of the microscope. A person that is competent in microscope
servicing will need to investigate this cause.
6) No light is illuminating the sample.
Ensure that a wavelength is selected and in the ON state on the Control Pod. The
intensity bar on the Control Pod will be filled if so.
Confirm whether the LEDs are illuminating. This can be done by removing the
system from the scope and shining at a wall or desk (take care not to directly view
the light output or shine at others). Please note that most of the light output from
the UV channel is not visible to the eye, it will however cause white paper to
fluoresce.
If the LEDs are illuminating then check the optical path of the microscope.
Confirm that any shutters are open and that the filter sets are compatible with
the excitation wavelengths being used.
If the LEDs are not illuminating then the system will need to be returned to
CoolLED for investigation.