CORNING 4930 User manual

Corning®
Aspirator
Instruction Manual
Catalog Number:
4930

Table of Contents
1.0 Product Description ........................ 1
2.0 Packing.................................... 1
3.0 Set Up..................................... 2
4.0 Aspirator Operation........................ 3
5.0 Cleaning and Service....................... 3
6.0 Aspirator Attachment Options.............. 4
7.0 Chemical Compatibility .................... 5
8.0 Spare Parts ................................ 6
9.0 Limited Warranty .......................... 7

Instruction Manual | 1
1.0 Product Description
The Corning® Aspirator is an aspiration device
used for the removal of liquids from a variety of
lab vessels. It is designed for use with an in-house
vacuum system. The aspirator allows precise
fingertip control of aspiration and presents a
sensible alternative to the use of glass pipets for
supernatant disposal.
Designed for a comfortable fit in either hand,
the aspirator can be used with a variety of
collection devices. These include some glass
pipets, 2 mL plastic pipets (Cat. No. 4020, 4021,
4486) or 2 mL aspiration pipets (Cat. No. 9186).
Included with the aspirator is a single-channel
adapter (Cat. No. 155850) for use with standard
1 to 200 µL disposable pipet tips (Cat. No. 4804,
4860, 4864), as well as a rubber adapter for glass
pipets (Cat. No. 155505).
An 8-channel adapter for standard 1-200 µL
disposable pipet tips is available for use with
96-well microplates.
2.0 Packing
Description Qty/Pk
Aspirator 1
Single-channel adapter 1
Rubber adapter for glass pipets 1

2 | CORNING ASPIRATOR
3.0 Set Up
Collection
Bottle
Safety
Trap
Vacuum
Source
The aspirator is designed for use with a vacuum
pump or an in-house vacuum system and a
suitable collection vessel.
Connect the hose barb end of the aspirator
handpiece to flexible 4 mm I.D. thick wall (to
prevent collapsing under vacuum) silicone
rubber tubing, and then connect the other end
to a suitable collection vessel. These heavy wall
filter flasks have a protective PVC coating that
helps prevent the glass from shattering. They are
autoclavable (121°C) and resistant to thermal
shock. The collection vessel is then connected
by tubing to the vacuum source. NOTE: When
working with infectious materials, connecting a
second vessel or 50 mm hydrophobic membrane
disc filter (Cat. No. 431227) as an in-line safety
trap between the collection vessel and the vacuum
source is strongly advised.
The aspirator is supplied with a soft rubber
grommet which holds both glass and plastic
2 mL pipets. To use the single-channel pipet tip
adapter, simply remove the orange fastening
sleeve from the aspirator handpiece by turning
one-quarter turn counterclockwise, take out
the rubber grommet, and replace with the tip
adapter, which has its own fastening sleeve.
Tighten with a clockwise one-quarter turn.
To connect the 8-channel adapter for disposable
tips, remove the fastening sleeve and rubber
grommet, and replace with the 8-channel adapter
which has its own fastening sleeve.

Instruction Manual | 3
4.0 Aspirator Operation
Vacuum pressure through the aspirator is
controlled by the pressure-sensitive orange control
button on the top of the handpiece.
Using the plastic screw opposite the control
button, you can set a residual vacuum level to
prevent waste liquids dripping from the tip or
the pipet. The aspirator is factory preset to allow
a low level of constant suction; by turning the
screw one full turn clockwise, the valve is shut
and residual vacuum is stopped.
5.0 Cleaning and Service
The aspirator is entirely autoclavable. When
working with biohazardous material, sterilize the
unit before disassembling for cleaning. Refer to
the diagram (below) for disassembly.
To clean, immerse the pieces in mild laboratory
detergent, but do not use abrasives, as a rough-
ened surface may affect the smooth operation of
the control valve. NOTE: When working with salt-
rich solutions, rinse out the aspirator handpiece
with distilled water after use to avoid build-up of
dried deposits.
Apply a thin coat of Dupont Molykote® BG-87
lubricant as needed to the control valve to keep
the action smooth.

4 | CORNING ASPIRATOR
6.0 Aspirator Attachment Options
Single-channel Adapter
The single-channel adapter is included with the
Corning® Aspirator and is for use with standard
1 to 200 µL disposable pipet tips.
8-channel Adapter
The 8-channel adapter fits standard 1 to 200 µL
disposable pipet tips, and is available for use with
96-well microplates (sold separately).

Instruction Manual | 5
7.0 Chemical Compatibility
Avoid prolonged contact (days or weeks) with:
Concentrated inorganic acids
Sodium hypochlorite
Methylene chloride
Concentrated aldehydes
Other halogenated hydrocarbons
Resistant to:
DMSO
Ethanol/methanol
Dimethylformamide
Acetone
NOTES:
Flushing the aspirator with ethanol
immediately after contact with organic
solvents is the best safeguard in all cases.
For O-ring replacement, use fluorocarbon
O-rings only.

6 | CORNING ASPIRATOR
8.0 Spare Parts
B
D
E
F
G
CA
Item Cat. No. Description Qty/Pk
– 4931 8-channel adapter 1
– 155850 Single-channel adapter 1
A 155855 Quick-locking screw,
orange
1
B 155505 Rubber adapter for glass
pipets
1
C 155830 O-ring, 6.75 x 1.78 mm
FKM for end piece
1
D 155805 O-ring, 2.57 x 1.78 mm
FKM 60 shore A for valve
plunger
1
E 155605 Tappet grey with O-ring,
valve plunger
1
F 151230 Plunger spring 1
G 155800 O-ring, 9.25 x 1.78 mm
FKM for base screw
1

Instruction Manual | 7
9.0 Limited Warranty
Corning Incorporated (Corning) warrants that
this product will be free from defects in material
and workmanship for a period of one (1) year
from date of purchase. CORNING DISCLAIMS ALL
OTHER WARRANTIES WHETHER EXPRESSED OR
IMPLIED, INCLUDING ANY IMPLIED WARRANTIES
OF MERCHANTABILITY OR OF FITNESS FOR A
PARTICULAR PURPOSE. Corning’s sole obligation
shall be to repair or replace, at its option, any
product or part thereof that proves defective in
material or workmanship within the warranty
period, provided the purchaser notifies Corning
of any such defect. Corning is not liable for any
incidental or consequential damages, commercial
loss or any other damages from the use of this
product.
This warranty is valid only if the product is
used for its intended purpose and within the
guidelines specified in the supplied instruction
manual. This warranty does not cover damage
caused by accident, neglect, misuse, improper
service, natural forces or other causes not arising
from defects in original material or workmanship.
This warranty does not cover O-rings, rubber
adapter or damage to paint or finish. Claims
for transit damage should be filed with the
transportation carrier.
In the event this product fails within the specified
period of time because of a defect in material or
workmanship, contact Corning Customer Service
at: USA/Canada 1.800.492.1110, outside the
U.S. +1.978.442.2200, visit www.corning.com/
lifesciences, or contact your local support office.
Corning’s Customer Service team will help
arrange local service where available or
coordinate a return authorization number and
shipping instructions. Products received without
proper authorization will be returned. All items
returned for service should be sent postage
prepaid in the original packaging or other
suitable carton, padded to avoid damage. Corning
will not be responsible for damage incurred by
improper packaging. Corning may elect for onsite
service for larger equipment.

8 | CORNING ASPIRATOR
Some states do not allow limitation on the
length of implied warranties or the exclusion
or limitation of incidental or consequential
damages. This warranty gives you specific legal
rights. You may have other rights which vary
from state to state.
No individual may accept for, or on behalf of
Corning, any other obligation of liability, or
extend the period of this warranty.
For your reference, make a note of the serial and
model number, date of purchase, and supplier
here.
Serial No. __________________________________________
Date Purchased _____________________________________
Model No. _________________________________________
Supplier ___________________________________________
Warranty/Disclaimer: Unless otherwise specified, all
products are for research use or general laboratory
use only.* Not intended for use in diagnostic or
therapeutic procedures. Not for use in humans. These
products are not intended to mitigate the presence
of microorganisms on surfaces or in the environment,
where such organisms can be deleterious to humans or
the environment. Corning Life Sciences makes no claims
regarding the performance of these products for clinical
or diagnostic applications. *For a listing of US medical
devices, regulatory classifications or specific information
on claims, visit www.corning.com/resources.
Corning’s products are not specifically designed and
tested for diagnostic testing. Many Corning products,
though not specific for diagnostic testing, can be used
in the workflow and preparation of the test at the
customers discretion. Customers may use these products
to support their claims. We cannot make any claims or
statements that our products are approved for diagnostic
testing either directly or indirectly. The customer is
responsible for any testing, validation, and/or regulatory
submissions that may be required to support the safety
and efficacy of their intended application.

Corning Incorporated
Life Sciences
www.corning.com/lifesciences
NORTH AMERICA
t 800.492.1110
t 978.442.2200
ASIA/PACIFIC
Australia/New Zealand
t 61 427286832
Chinese Mainland
t 86 21 3338 4338
India
t 91 124 4604000
Japan
t 81 3-3586 1996
Korea
t 82 2-796-9500
Singapore
t 65 6572-9740
Taiwan
t 886 2-2716-0338
EUROPE
CSEur[email protected]
France
t 0800 916 882
Germany
t 0800 101 1153
The Netherlands
t 20 655 79 28
United Kingdom
t 0800 376 8660
All Other European
Countries
t +31 (0) 206 59 60 51
LATIN AMERICA
Brasil
t 55 (11) 3089-7400
Mexico
t (52-81) 8158-8400
For a listing of trademarks, visit www.corning.com/clstrademarks.
Allother trademarks are the property oftheir respective owners.
© 2008, 2023 Corning Incorporated. All rights reserved.
1/23 CLS-AN-725DOC REV1
Table of contents
Other CORNING Medical Equipment manuals
Popular Medical Equipment manuals by other brands

ARJO HUNTLEIGH
ARJO HUNTLEIGH Alpha Active 3 Instructions for use

MTR+
MTR+ Myolito Operator's manual

ARJO HUNTLEIGH
ARJO HUNTLEIGH Maxi Air Instructions for use

Contec
Contec CMS50DL user manual

B. Braun
B. Braun Perfusor Space Service manual

Contec Medical Systems Co.
Contec Medical Systems Co. OXY-2 user manual

Smartoptics
Smartoptics Activity 850 operating manual

Hillrom
Hillrom Centrella Smart+ Bed Service manual

Bionet
Bionet Cardio P1 Service manual

ARJO HUNTLEIGH
ARJO HUNTLEIGH Malibu Quick reference guide

V-tech
V-tech EzRay Air Portable user manual

ARJO HUNTLEIGH
ARJO HUNTLEIGH Nimbus 3 Mattress Replacement System manual